K182888 · Philips Medical Systems MR Finland · MUJ · Apr 30, 2019 · Radiology
Device Facts
Record ID
K182888
Device Name
MRCAT Pelvis
Applicant
Philips Medical Systems MR Finland
Product Code
MUJ · Radiology
Decision Date
Apr 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.
Device Story
MRCAT Pelvis is a software add-on for Philips Ingenia 1.5T and 3.0T MR systems; operates as an automated post-processing tool at the MR console. Input consists of mDixon MR scan data; algorithm performs model-based segmentation of five tissue types (air, fat, water-rich tissue, spongy bone, compact bone) and assigns density values to generate CT-like images. Output is a 3D MRCAT image stored in the MR database and exportable via DICOM for use in radiotherapy treatment planning systems. Used by clinicians in radiotherapy departments to estimate radiation attenuation for photon external beam therapy. Benefits include enabling MR-only radiotherapy planning by providing spatially accurate tissue density information, potentially reducing the need for separate CT scans.
Clinical Evidence
Clinical evaluation involved assessing the robustness of the MRCAT Pelvis algorithm by comparing dose plans generated from MRCAT images against CT-based plans. Analysis used gamma analysis with a 3%/3mm criterion. Results demonstrated that MRCAT Pelvis images provide spatially accurate radiation attenuation estimates suitable for EBRT planning of pelvic soft tissue cancers.
Technological Characteristics
Software plug-in for Ingenia 1.5T/3.0T MR systems; utilizes mDixon imaging protocol. Implements model-based segmentation and density assignment. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-6, IEC 60601-2-33, IEC 62304, and ISO 14971.
Indications for Use
Indicated for radiotherapy treatment planning of soft tissue cancers in the pelvic region.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Philips Medical Systems MR Finland % Janne Marvola, Ph.D. Regulatory Engineer Ayritie 4 01510 Vantaa FINLAND
April 30, 2019
## Re: K182888
Trade/Device Name: MRCAT Pelvis Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: March 22, 2019 Received: April 1, 2019
Dear Dr. Marvola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182888
Device Name MRCAT Pelvis
Indications for Use (Describe)
MRCAT Pelvis is a software add-on for Ingenia 1.5T and 3.0T MR systems.
Intended Use:
MRCAT imaging is intended to provide the operator with information of tissue properties for radiation estimation purposes in photon external beam radiotherapy treatment planning.
Indications for use:
MRCAT Pelvis is indicated for radiotherapy treatment planning of soft tissue cancers in the pelvic region.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Philips Medical Systems MR Finland Äyritie 4 01510 Vantaa Finland T: +358-9-615-800
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Owner/Contact/Date (807.92(a)(1):
Date:
March 22, 2019
| Owner/Submitter: | Philips Medical Systems MR Finland<br>Äyritie 4<br>01510 Vantaa<br>FINLAND<br>Phone: +358-9-615 800<br>Fax: +358-9-3487 2406 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Janne Marvola<br>Regulatory Engineer |
Regulatory Enginee Philips Medical Systems MR Finland Phone: +358-40-126 1214 Fax: +358-9-3487 2406 E-mail: regulatory.mr.therapy@philips.com
| Secondary Contact Person: | Osku Ilvonen<br>Senior Manager Q&R<br>Philips Medical Systems MR Finland<br>Phone: +358-40-552-6197<br>Fax: +358-9-3487-2406<br>E-mail: osku.ilvonen@philips.com |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Image /page/3/Picture/9 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines that resemble sound waves or water. Above and below the wavy lines are four-pointed stars. The word "PHILIPS" is written in blue at the top of the shield.
{4}------------------------------------------------
#### Device names (807.92(a)(2)):
Trade Name: MRCAT Pelvis Classification: Class II Regulatory Section: 21 CFR 892.5050 Product Code: MUJ
Predicate Device(s) K151435 MRCAT (primary) (807.92(a)(3):: K013644 AcQPlan 5.0 (reference device)
Device Description MRCAT Pelvis is a software application to Ingenia 1.5T and 3T MR (807.92(a)(4)):: systems. MRCAT Pelvis is available to the customer as an option to Ingenia MR-RT package, which is a set of accessories for Ingenia systems.
> Automated generation of MRCAT images takes place at the MR console of Ingenia. The embedded image post-processing runs in the background parallel to image acquisition. MRCAT algorithms enable automatic tissue characterization of five tissue types; air, fat, waterrich tissue, spongy bone and compact bone. Subsequent density assignment provides MRCAT images with CT-like density information.
Intended Use: (807.92(a)(5): Intended Use:
> MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.
Indications for use:
MRCAT Pelvis is indicated for radiotherapy treatment planning of soft tissue cancers in the pelvic region.
Technology (807.92(a)(6)):: MRCAT Pelvis functionality is implemented as a software plug-in for the MR main software and it contains the following main features:
- Automatic post-processing tool delivering MRCAT images 1)
- 2) Examcard with mDixon imaging protocol
Image /page/4/Picture/12 description: The image features the Philips logo, which is a shield-shaped emblem in a shade of blue. The word "PHILIPS" is displayed in white, bold, sans-serif letters at the top of the shield. Below the text, there are two wavy lines that resemble sound waves or water, and two four-pointed stars are positioned above and below the wavy lines, adding a celestial or dynamic element to the logo.
{5}------------------------------------------------
3) DICOM export of MRCAT image.
## MRCAT Image Generation
MRCAT images are generated with an ExamCard post-processing step, which uses the images from the previous mDixon scan.
The post-processing logic takes care of launching MRCAT algorithm executable calculating a new 3D MRCAT image. The algorithm enables model based segmentation of bones and soft tissues. Subsequent density assignment provides MRCAT images with CT-like density information. Once the process is running, post-processing logic exchanges information with the algorithm:
- Image source data to algorithm, and image output data back to the post-processing step
- . Progress notifications
- . Error and warning notifications
The 3D MRCAT image from the post-processing step is stored into the MR image database.
### mDIXON scan
An mDixon imaging protocol, with imaging parameters optimized for MRCAT image post-processing and for geometric accuracy, is delivered as a part of MRCAT option. MRCAT uses fixed parameters for mDixon scan, only the image stack location is configurable.
### DICOM Export
The MRCAT post-processing step stores the image data returned by the MRCAT algorithm into MR database.
MRCAT images can be exported in DICOM format enabling the use as primary images in the treatment planning systems.
Image /page/5/Picture/13 description: The image shows the Philips logo, which consists of the word "PHILIPS" in blue, positioned above a blue shield-like shape. Inside the shield, there are two stylized waves and two four-pointed stars. The overall design is simple and clean, with a focus on the brand name and a symbolic representation of waves and stars.
{6}------------------------------------------------
### Hardware platform description
MRCAT Pelvis does not require any change of the hardware platform of Ingenia scanners. The new software extensions introduced by MRCAT Pelvis run on the MR console of Ingenia.
Summary of Non-Clinical Tests:
Determination of Substantial Equivalence, non-clinical (807.92(b)(1))::
The MRCAT software complies with voluntary standards as detailed below. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Testing on unit level (Subsystem verification)
- Integration testing (System verification)
- Final acceptance testing (Validation) ●
- Performance testing (Verification)
- Safety testing (Verification)
The MRCAT software was designed and tested for compliance to the following standards:
- 1. ANSI/AAMI ES60601-1: 2012, Medical electrical equipment -Part 1: General requirements for basic safety and essential performance – Third Edition
- 2. IEC 60601-1-6:2006, Medical electrical equipment Part 1-6: General requirements for safety - Collateral standard: Usability.
- 3. IEC 60601-2-33:2015, Medical electrical equipment Part 2-33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis.
- 4. IEC 62304:2006, Medical device software Software life-cycle processes
- 5. ISO 14971:2007, Medical devices. Application of risk management to medical devices
All requirements are met for the MRCAT Pelvis application.
Image /page/6/Picture/18 description: The image shows the Philips logo, which consists of the word "PHILIPS" in a sans-serif font above a shield-like emblem. The emblem is blue and features two wavy lines and four stars. The wavy lines are positioned horizontally in the center of the shield, and the stars are placed above and below the wavy lines.
{7}------------------------------------------------
The conclusion from these reports is:
All the tests performed for MRCAT Pelvis software were successful. All defects have been analyzed and are confirmed that they are not safety defects and will not cause any hazardous situation on using this application.
- Summary of Clinical Tests: Clinical (807.92(b)(2)):
The robustness of the MRCAT Pelvis algorithm for producing equivalent dose plans to CT using gamma analysis with criterion of 3%/3mm is shown by post-processing MRCAT images from patients, and calculating dose using the MRCAT images.
In summary, the MRCAT Pelvis images are spatially accurate radiation attenuation estimates that can aid in the EBRT planning of pelvic soft tissue cancers.
Conclusion (807.92(b)(3)):: MRCAT Pelvis software is a tool to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning for pelvic cancer patients.
> lt is the opinion of Philips Medical Systems MR Finland that the MRCAT Pelvis software option for Philips Ingenia MR scanners is substantially equivalent and raises no new issues of safety or effectiveness compared to the primary predicate Philips MRCAT, K151435 cleared by FDA on February 25, 2016 and the reference device AcQPlan 5.0, K013644, cleared by FDA on September 12, 2002.
Image /page/7/Picture/7 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and two stars. The word "PHILIPS" is written in white at the top of the shield.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.