K141283 · Varian Medical Systems, Inc. · MUJ · Aug 7, 2014 · Radiology
Device Facts
Record ID
K141283
Device Name
ECLIPSE TREATMENT PLANNING SYSTEM
Applicant
Varian Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Aug 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Device Story
Eclipse TPS is a computer-based software tool for medical professionals to design and simulate radiation therapy. Inputs include patient medical images (CT, MR, PET) and clinical data. Software transforms inputs into treatment plans via dose calculation algorithms (e.g., AcurosXB, AcurosBV, Gaussian Pencil Beam, Monte Carlo) and optimization modules. Outputs include dose distributions, field parameters, and hardcopy reports for clinical review. Used in clinical settings by trained professionals to guide radiation delivery. Benefits include precise treatment planning, optimized dose delivery to target volumes, and sparing of healthy tissue. Supports integration with ARIA RadOnc and DICOM RT standards.
Clinical Evidence
Bench testing only. Verification and validation performed for new features and regression testing for existing features. System requirements traced to test outcomes. No safety-intolerable discrepancy reports remained.
Technological Characteristics
Software-based treatment planning system. Features include dose calculation algorithms (AcurosXB, AcurosBV, Monte Carlo, Gaussian Pencil Beam), image processing (segmentation, registration, blending), and optimization modules. Connectivity via DICOM RT and ARIA integration. Operates on standard clinical computing hardware. No specific materials or energy sources as it is a software device.
Indications for Use
Indicated for patients with malignant or benign diseases requiring radiotherapy. Includes external beam irradiation (photon, electron, proton) and internal irradiation (brachytherapy). Eclipse Proton Eye algorithm indicated for proton treatment of ocular neoplasms.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, forming a trefoil shape. The profiles are rendered in a dark color, contrasting with the white background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2014
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K141283
Trade/Device Name: Eclipse Treatment Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: April 30, 2014 Received: May 16, 2014
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141283
Device Name Eclipse Treatment Planning System
#### Indications for Use (Describe)
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Michael D'Hara
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# PREMARKET NOTIFICATION
# 510(k) Summary
# Eclipse Treatment Planning System
# As required by 21 CFR 807.92
| Submitter's Name: | Varian Medical Systems<br>3100 Hansen Way, m/s E110<br>Palo Alto CA94304 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Name: Peter J. Coronado-Director Regulatory Affairs<br>Phone: 650/424.6230<br>Fax: 650/646.9200<br>E-mail: submissions.support@varian.com<br>Date: 30th April 2014 |
| Proprietary Name: | Eclipse Treatment Planning System |
| Classification Name: | system,planning,radiation therapy treatment<br>21CFR892.5050, MUJ, Class II |
| Common/Usual Name: | Eclipse TPS, Eclipse, Treatment Planning System. |
| Predicate Devices: | Eclipse Treatment Planning System 13 (K133247)<br>Eclipse Treatment Planning System 12 (K131891) |
| Device Description: | The Varian Eclipse™ Treatment Planning System (Eclipse TPS)<br>provides software tools for planning the treatment of malignant<br>or benign diseases with radiation. Eclipse TPS is a computer-based<br>software device used by trained medical professionals to<br>design and simulate radiation therapy treatments. Eclipse TPS is<br>capable of planning treatments for external beam irradiation with<br>photon, electron, and proton beams, as well as for internal<br>irradiation, (brachytherapy) treatments. |
| Indications for Use: | The Eclipse Treatment Planning System (Eclipse TPS) is used to<br>plan radiotherapy treatments for patients with malignant or<br>benign diseases. Eclipse TPS is used to plan external beam<br>irradiation with photon, electron and proton beams, as well as for<br>internal irradiation (brachytherapy) treatments. In addition, the<br>Eclipse Proton Eye algorithm is specifically indicated for<br>planning proton treatment of neoplasms of the eye. |
{5}------------------------------------------------
#### Changes in Technological Characteristics:
The significant changes compared with the predicate are changes associated with RapidPlan (previously known as Dose Volume Histogram Estimation) and also changes to the algorithm related to support for mARC treatment planning by Siemens treatment machines.
The complete list of changes and their related requirements can be found in the document Tracing Changed/New Features to System Requirements in Section 18 of this submission.
#### Device Comparison Table
| | ECLIPSE TPS V12.0<br>(510(k) ID # K131891) | ECLIPSE TPS V13.0<br>(510(k) ID #K133247) | ECLIPSE TPS V13.5 |
|------------------------------------------------|--------------------------------------------|-------------------------------------------|-------------------|
| 2. General Usage | | | |
| External beam PHOTON<br>planning | yes | yes | yes |
| External beam PHOTON<br>inverse planning | yes | yes | yes |
| External beam ELECTRON<br>planning | yes | yes | yes |
| External beam PROTON<br>planning | yes | yes | yes |
| External beam OCULAR<br>PROTON planning (EOPP) | no | yes | yes |
| Internal BRACHYTHERAPY<br>planning | yes | yes | yes |
| Stereotactic Frame Localization | yes | yes | yes |
| 3. Supported External Beams<br>& Accessories | | | |
| Photon beams | yes | yes | yes |
| Electron beams | yes | yes | yes |
| Proton beams | yes | yes | yes |
| Coplanar fields | yes | yes | yes |
| Non-coplanar fields | yes | yes | yes |
| Multi-leaf Collimators | yes | yes | yes |
| Asymmetric collimators | yes | yes | yes |
| Stereotactic Cone collimators | yes | yes | yes |
| Arc fields | yes | yes | yes |
| Poured Blocks | yes | yes | yes |
| Compensators | yes | yes | yes |
| Physical wedges | yes | yes | yes |
{6}------------------------------------------------
| Dynamic wedges | yes | yes | yes |
|---------------------------------------------------------------------|-----|-----|-----|
| Flattening filter free support<br>(FFF) | yes | yes | yes |
| Rotating treatment couch | yes | yes | yes |
| 4. Supported Brachytherapy<br>Sources & Accessories | | | |
| Plan for high dose rate<br>afterloader | yes | yes | yes |
| Manual low dose rate<br>brachytherapy: seeds, line<br>sources, wire | yes | yes | yes |
| Applicator library | yes | yes | yes |
| Needle templates | yes | yes | yes |
| Seed templates | no | no | yes |
| 5. Graphical User Interface | | | |
| Multiple-instance application | yes | yes | yes |
| Multiple-workspace layout | yes | yes | yes |
| Graphical display/editing of<br>field parameters | yes | yes | yes |
| Beam's-Eye-View display | yes | yes | yes |
| 3D patient image display | yes | yes | yes |
| Model for human Eye | yes | yes | yes |
| SRS Localization application | yes | yes | yes |
| SRS Planning application | yes | yes | yes |
| Biological Optimization<br>application | yes | yes | yes |
| Biological Evaluation<br>application | yes | yes | yes |
| 3D Conformal Optimization<br>application | yes | yes | yes |
| 6. Image Processing | | | |
| Orthogonal image displays (3) | yes | yes | yes |
| Oblique image display | yes | yes | yes |
| Edge enhancement filters | yes | yes | yes |
| Image smoothing filters | yes | yes | yes |
| CT/MR/PET Image Registration | yes | yes | yes |
| Image blending utility | yes | yes | yes |
| 4D image display (registration<br>of time series of 3D images) | yes | yes | yes |
| Digitally reconstructed<br>radiographs | yes | yes | yes |
{7}------------------------------------------------
| Enclosed Volume<br>measurement | yes | yes | yes |
|-----------------------------------------------------------------------------------------------------------------------|--------------|--------------|--------------|
| Stereotactic Frame Coordinate<br>transformation | yes | yes | yes |
| 7. Image Segmentation | | | |
| Geometrical shapes | yes | yes | yes |
| Manual editing and<br>manipulation tools | yes | yes | yes |
| Automatic /semi-automatic<br>tools | yes | yes | yes |
| Automatic/semi-automatic on-<br>demand and post-processing<br>tools for individual<br>organs/structures | yes | yes | yes |
| 3D Automargin | yes | yes | yes |
| Logical operators | yes | yes | yes |
| Enhanced 2D and 3D<br>contouring tools | no | yes | yes |
| Enhanced 4D functionality,<br>including structure propagation<br>and display of respiratory<br>amplitude distribution | no | yes | yes |
| 8. Dose Calculation | | | |
| Distributed Calculation<br>Framework | yes | yes | yes |
| Photon calculation | yes | yes | yes |
| - Energy Range | 1 MV – 50 MV | 1 MV – 50 MV | 1 MV – 50 MV |
| - CT-based volumetric<br>calculation | yes | yes | yes |
| - Non-CT based IRREG<br>calculation | yes | yes | yes |
| - Convolution method | yes | yes | yes |
| - Combined electron/photon<br>scatter | yes | yes | yes |
| - Directional heterogeneity<br>correction | yes | yes | yes |
| - Treatment head modelling | yes | yes | yes |
| - Photon Monitor Unit<br>calculation | yes | yes | yes |
| - Compensator monitor unit<br>calculation | yes | yes | yes |
| - Beam Angle Optimization<br>(GEOS) | yes | yes | yes |
{8}------------------------------------------------
| - Dose Dynamic Arc planning | yes | yes | yes |
|------------------------------------------------------|------------------|-----------------------------------------------------------|------------------|
| - Cone Dose Calculation | yes | yes | yes |
| - Biological optimization | yes | yes | yes |
| - 3D Conformal Optimization | yes | yes | yes |
| - IMRT optimization | yes | yes | yes |
| - AcurosXB dose calculation<br>algorithm | yes | yes | yes |
| - Range Uncertainty feature<br>for photons | no | yes | yes |
| - Siemens mArc | no | no | yes |
| Electron calculation | yes | yes | yes |
| - Energy Range | 1 MeV - 50 MeV | 1 MeV - 50 MeV | 1 MeV - 50 MeV |
| - Gaussian Pencil Beam Model | yes | yes | yes |
| - Electron Monte Carlo<br>algorithm | yes | yes | yes |
| - Electron Monitor Unit<br>calculation | yes | yes | yes |
| Proton calculation | yes | yes | yes |
| - Energy Range | 50 MeV - 300 MeV | 50 MeV - 300 MeV | 50 MeV - 300 MeV |
| - Single scattering technique | yes | yes | yes |
| - Double scattering technique | yes | yes | yes |
| - Uniform scanning technique | yes | yes | yes |
| - Modulated scanning<br>technique | yes | yes | yes |
| - Optimization for modulated<br>scanning | yes | yes<br>(new implementation<br>for<br>Proton Optimizer) | yes |
| - Monitor unit calculation for<br>modulated scanning | yes | yes | yes |
| - Range uncertainty feature<br>for Protons | yes | yes | yes |
| Brachytherapy calculation | yes | yes | yes |
| - AAPM TG 43 compliant | yes | yes | yes |
| - Point Dose calculation | yes | yes | yes |
| - Optimization to point dose<br>constraints | yes | yes | yes |
| - Geometric optimization | yes | yes | yes |
| - AcurosBV dose calculation<br>algorithm | yes | yes<br>(with support for<br>report the dose to<br>medium) | yes |
{9}------------------------------------------------
| Eclipse Algorithm Application<br>Programming Interface (EAAPI) | yes | yes | yes |
|----------------------------------------------------------------|--------|----------------------------------------|----------------------------------------------------------------|
| RapidPlan - Dose Volume<br>Histogram (DVH) Estimation | yes | no | Yes<br>(extended Dose<br>Volume Histogram<br>(DVH) Estimation) |
| 9. Dose evaluation | | | |
| Dose color wash | 2D, 3D | 2D, 3D | 2D, 3D |
| Isodose levels | 2D, 3D | 2D, 3D | 2D, 3D |
| Isodose Surface | 3D | 3D | 3D |
| Reference point dose summary | yes | yes | yes |
| Dose Volume Histogram plot | yes | yes | yes |
| Plan summing tool | yes | yes | yes |
| Plan comparison tools | yes | yes | yes |
| Evaluation using biological<br>models | yes | yes | yes |
| 10. Plan Output - Hardcopy | | | |
| Graphics window screen dump | yes | yes | yes |
| Patient administration data | yes | yes | yes |
| Plan parameters | yes | yes | yes |
| Geometrical displays of plan<br>data | yes | yes | yes |
| Dose distribution | yes | yes | yes |
| DVH plot | yes | yes | yes |
| BEV display | yes | yes | yes |
| Patient orientation | yes | yes | yes |
| User Configurable hardcopy<br>layouts | yes | yes | yes |
| Import/Export<br>11.<br>Interfaces | | | |
| ARIA RadOnc integration | yes | yes | yes |
| DICOM RT | yes | yes | yes |
| Other image formats | yes | yes | yes |
| Electromagnetic Digitizer | import | import | import |
| Eclipse Scripting API (ESAPI)<br>read only access | yes | yes<br>(includes also<br>BrachyVision) | yes<br>(includes also<br>BrachyVision) |
| Eclipse Scripting API (ESAPI)<br>write access | no | yes<br>(in research database,<br>only) | yes<br>(in research database,<br>only) |
{10}------------------------------------------------
| Export field coordinates to<br>Laser System | yes | export | export |
|------------------------------------------------|-----|--------|--------|
| Basic RT Prescription<br>information available | no | yes | yes |
### Non-clinical Testing
Verification and Validation were performed for all the new features and regression testing was performed against the existing features of Eclipse. System requirements created or affected by the changes can be traced to the test outcomes.
### Conclusion of Non-Clinical testing
The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports) remaining which had a priority of Safety Intolerable or Customer Intolerable. Varian therefore considers Eclipse 13.5 to be safe and to perform at least as well as the predicate device.
### Argument for Substantial Equivalence to the Predicate Device
A subset of features of the current device is different to the predicate. Of these, the significant changes compared with the predicate are associated with RapidPlan (previously known as Dose Volume Histogram Estimation) and also changes to the algorithm related to support for mARC treatment planning by Siemens treatment machines. Other changes are the addition of new BrachyVision features.
These changes are all considered by Varian to be enhancements of the Indications for Use and the Intended Use remain unchanged. There are no changes in the principle of operation of the software. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes that Eclipse TPS is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.