K133247 · Varian Medical Systems, Inc. · MUJ · Feb 7, 2014 · Radiology
Device Facts
Record ID
K133247
Device Name
ECLIPSE TREATMENT PLANNING SYSTEM
Applicant
Varian Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Feb 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Device Story
Eclipse TPS is a software-based treatment planning system for radiation therapy. It ingests patient imaging data (CT, MR, PET) to design and simulate radiation treatments. Clinicians use the software to perform image segmentation, contouring, and dose calculation for external beam (photon, electron, proton) and internal (brachytherapy) radiation. The system utilizes various algorithms—including AAA, AXB, and AcurosBV—to calculate dose distributions, which are visualized as isodose levels, surfaces, and dose-volume histograms (DVH). The output assists physicians in optimizing treatment plans to maximize tumor coverage while sparing healthy tissue. Used in clinical settings by trained medical professionals, the system integrates with hospital infrastructure (ARIA) to manage patient data and treatment parameters. It supports advanced features like proton layer-by-layer MLC, deformable registration, and biological optimization to improve treatment precision and patient outcomes.
Clinical Evidence
Bench testing only. Verification and validation were performed for all new features, with regression testing against existing features. System requirements were traced to test outcomes, confirming the product conforms to defined user needs and intended uses. No clinical data was required or provided.
Technological Characteristics
Software-based treatment planning system; 64-bit architecture. Supports photon, electron, and proton beam planning. Algorithms include AAA, AXB, AcurosXB, and AcurosBV. Connectivity via DICOM RT, ARIA integration, and ESAPI. Infrastructure includes SQL Server, Unicode support, and MD5 checksums for dose calculation integrity. Operates on standard clinical workstations.
Indications for Use
Indicated for patients with malignant or benign diseases requiring radiotherapy. Used for planning external beam irradiation (photon, electron, proton) and internal irradiation (brachytherapy). Proton Eye algorithm indicated for proton treatment planning of ocular neoplasms.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K133247
Page 1 of 6
# · PREMARKET NOTIFICATION
## 510(k) Summary
# 7 2014
Eclipse Treatment Planning System
#### As required by 21 CFR 807.92
Submitter's Name:
Varian Medical Systems 3100 Hansen Way, m/s E110 Palo Alto CA94304
Contact Name: Peter J. Coronado-Director Regulatory Affairs Phone: 650/424.6230 Fax:650/842.5040 E-mail: submissions.support@varian.com Date: 30d October 2013
Eclipse Treatment Planning System
system,planning,radiation therapy treatment 21CFR892.5050, MUJ, Class II
Eclipse TPS, Eclipse, Treatment Planning System.
Eclipse Treatment Planning System 12 (K131891)
The Varian Eclipse™ Treatment Planning System (Eclipse TPS) provides software tools for planning the treatment of malignant or benign diseases with radiation. Eclipse TPS is a computer-based software device used by trained medical professionals to design and simulate radiation therapy treatments. Eclipse TPS is capable of planning treatments for external beam irradiation with photon, electron, and proton beams, as well as for internal irradiation, (brachytherapy) treatments.
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Proprietary Name:
Classification Name:
Common/Usual Name:
Predicate Devices:
Device Description:
Indications for Use:
{1}------------------------------------------------
# Changes in Technological Characteristics:
The significant changes in this device are those associated with the change to the Proton optimizer, modifications to the Proton Dose Calculation algorithm, Proton Layer by Layer MLC support and the Improved second source modeling in AAA and AXB.
# Device Comparison Table ·
(Changed Features are in bold).
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | PREDICATE DEVICE | DEVICE WITH CHANGE |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K131891 | ECLIPSE 13 |
| | ECLIPSE 12 | |
| General usage | External beam PHOTON planning External beam PHOTON inverse planning External beam ELECTRON planning External beam PROTON planning External beam OCULAR PROTON planning Internal BRACHYTHERAPY planning Stereotactic Frame Localization | External beam PHOTON planning External beam PHOTON inverse planning External beam ELECTRON planning External beam PROTON planning External beam OCULAR PROTON planning Internal BRACHYTHERAPY planning Stereotactic Frame Localization |
| Supported External<br>Beams &<br>Accessories | Photon beams Electron beams Proton beams Coplanar fields Non-coplanar fields Multi-leaf Collimators Asymmetric collimators Stereotactic Cone collimators Arc fields Poured Blocks Compensators Physical wedges Dynamic wedges Flattening filter free support (FFF) Rotating treatment couch TrueBeam 1.5 support TrueBeam 1.6 support. | Photon beams Electron beams Proton beams Coplanar fields Non-coplanar fields Multi-leaf Collimators Asymmetric collimators Stereotactic Cone collimators Arc fields Poured Blocks Compensators Physical wedges Dynamic wedges Flattening filter free support (FFF) Rotating treatment couch TrueBeam 1.5 support TrueBeam 1.6 and 2.0 support. Support for<br>LE imaging (2.5X for imaging only) and<br>support for new portal imager. |
| Supported<br>Brachytherapy<br>Sources &<br>Accessories | Plan for high dose rate afterloader Manual low dose rate brachytherapy: seeds,<br>line sources, wire Applicator library Needle templates | Plan for high dose rate afterloader Manual low dose rate brachytherapy: seeds,<br>line sources, wire Applicator library Needle templates |
| Graphical User<br>Interface | Multiple-instance application Multiple-workspace layout Graphical display/editing of field parameters Beam's-Eye-View display 3D patient image display Model for human Eye SRS Localization application SRS Planning application Biological Optimization application Biological Evaluation application 3D Conformal Optimization application Nexus Phase 0 - Home screen integration and<br>navigation | Multiple-instance application Multiple-workspace layout Graphical display/editing of field parameters Beam's-Eye-View display 3D patient image display Model for human Eye SRS Localization application SRS Planning application Biological Optimization application Biological Evaluation application 3D Conformal Optimization application Nexus Phase 0 - Home screen integration and<br>navigation Limited RT Prescription information available<br>in Eclipse |
| Image Processing | Orthogonal image displays (3) Oblique image display Edge enhancement filters Image smoothing filters CT/MR/PET Image Registration | Orthogonal image displays (3) Oblique image display Edge enhancement filters Image smoothing filters CT/MR/PET Image Registration |
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | PREDICATE DEVICE | DEVICE WITH CHANGE |
| | K131891 | |
| | ECLIPSE 12 | ECLIPSE 13 |
| Image<br>Segmentation | Image blending utility 4D image display (registration of time series of<br>3D images) Digitally reconstructed radiographs Enclosed Volume measurement Stereotactic Frame Coordinate transformation Geometrical shapes Manual editing and manipulation tools Automatic/semi-automatic tools Automatic/semi-automatic on-demand and<br>post-processing tools for individual<br>organs/structures No Automatic on-demand and pre-processing<br>tools for multiple organs/structures (SmartAdapt) toolset utilizing changed CT-MR<br>and MR-MR deformable registration 3D Automargin Logical operators | Image blending utility 4D image display (registration of time series of<br>3D images) Digitally reconstructed radiographs Enclosed Volume measurement Stereotactic Frame Coordinate transformation Geometrical shapes Manual editing and manipulation tools Automatic/semi-automatic tools Automatic/semi-automatic on-demand and<br>post-processing tools for individual<br>organs/structures No Automatic on-demand and pre-processing<br>tools for multiple organs/structures (SmartAdapt) toolset utilizing changed CT-MR<br>and MR-MR deformable registration Separate Selection Workspace and enhanced<br>Contouring 3D Automargin Logical operators |
| Dose Calculation | Distributed Calculation Framework Photon calculation Energy Range: 1 MV - 50 MV CT-based volumetric calculation Non-CT based IRREG calculation Convolution method Combined electron/photon scatter Directional heterogeneity<br>correction Treatment head modeling Photon Monitor Unit calculation Compensator monitor unit<br>calculation Beam Angle Optimization (GEOS) Leaf Motion Sequencing Dose Dynamic Arc planning Cone Dose Calculation Biological optimization 3D Conformal Optimization IMRT optimization AcurosXB dose calculation<br>algorithm RapidArc: enhancements in<br>intermediate dose calculation IMRT: intermediate dose<br>calculation RapidArc Varian linac, and Elekta<br>VMAT support FFF: Support for C3 and TrueBeam Dosimetrically equivalent machines | Distributed Calculation Framework Photon calculation Energy Range: 1 MV - 50 MV CT-based volumetric calculation Non-CT based IRREG calculation Convolution method Combined electron/photon scatter Directional heterogeneity<br>correction Treatment head modeling Photon Monitor Unit calculation Compensator monitor unit<br>calculation Beam Angle Optimization (GEOS) Leaf Motion Sequencing Dose Dynamic Arc planning Cone Dose Calculation Biological optimization 3D Conformal Optimization IMRT optimization AcurosXB dose calculation<br>algorithm RapidArc: enhancements in<br>intermediate dose calculation IMRT: intermediate dose<br>calculation RapidArc Varian linac, and Elekta<br>VMAT support FFF: Support for C3 and TrueBeam Dosimetrically equivalent machines Plan uncertainty for photon plans Isocenter handling improvements Addition of Point Dose<br>functionality into AAA dose<br>calculation algorithm Improved second source modeling<br>in AAA and AXB |
| | SmartIMRTDose-Volume Histogram (DVH)<br>Estimation* | |
| | Electron calculation | Electron calculation |
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | PREDICATE DEVICE | DEVICE WITH CHANGE |
| | K131891<br><br>ECLIPSE 12 | ECLIPSE 13 |
| | Gaussian Pencil Beam Model Electron Monte Carlo algorithm Electron Monitor Unit calculation Dosimetrically equivalent machines Proton calculation Energy Range: 50 MeV - 300 MeV Single scattering technique, support block and MLC Double scattering technique, support block and MLC Uniform scanning technique, support block and MLC Modulated scanning technique (spot and line scanning), support block and MLC Single scattering technique Optimization for modulated scanning Monitor unit calculation for modulated scanning Range uncertainty feature Spot editor user interface improvements Dosimetrically equivalent treatment units (for different gantries) Block drill bit corrections for milling machines | Gaussian Pencil Beam Model Electron Monte Carlo algorithm Electron Monitor Unit calculation Dosimetrically equivalent machines Proton calculation Energy Range: 50 MeV - 300 MeV Single scattering technique, support block and MLC Double scattering technique, support block and MLC Uniform scanning technique, support block and MLC Modulated scanning technique (spot and line scanning), support block and MLC Single scattering technique Optimization for modulated scanning Monitor unit calculation for modulated scanning Range uncertainty feature Spot editor user interface improvements Dosimetrically equivalent treatment units (for different gantries) Support for Sumitomo tx machine with Layer-by-Layer MLC Non-linear universal proton optimizer PCS enhancements Block drill bit corrections for milling machines |
| | Brachytherapy calculation AAPM TG 43 compliant Point Dose calculation Optimization to point dose constraints Geometric optimization AcurosBV dose calculation algorithm in 64 bit environment AcurosBV calculates does to transport media Nexus phase 0 support ** | Brachytherapy calculation AAPM TG 43 compliant Point Dose calculation Optimization to point dose constraints Geometric optimization AcurosBV dose calculation algorithm in 64 bit environment AcurosBV calculates does to transport media and reports dose to transport media Addition of TG186 source model and standard applicator for AcurosBV Nexus phase 0 support Source model approval Enhanced Normal Tissue and Source Fitting Simple contouring tools in Planning workspace Nominal Time Constraints in Volumetric Optimizer New version of AcurosBV algorithm Dose tracking support Ability to calculate AcurosBV on non-CT image |
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | PREDICATE DEVICE | DEVICE WITH CHANGE |
| | K131891 | ECLIPSE 13 |
| | ECLIPSE 12 | |
| | Eclipse Algorithm Application Programming<br>Interface (EAAPI) 64 bit External Beam Planning, BrachyVision<br>and Proton, PRO, AcurosXB, AcurosBV and<br>BAO & DVO algorithms Unified fluence calculation in Eclipse & DCF by<br>the final 3D dose calculation algorithm | Eclipse Algorithm Application Programming<br>Interface (EAAPI) 64 bit External Beam Planning, BrachyVision<br>and Proton, PRO, AcurosXB, AcurosBV and<br>BAO & DVO algorithms Unified fluence calculation in Eclipse & DCF by<br>the final 3D dose calculation algorithm |
| Dose evaluation | Dose color wash 2D, 3D Isodose levels 2D, 3D Isodose Surface 3D Reference point dose summary Dose Volume Histogram plot Plan summing tool Plan comparison tools Evaluation using biological models | Dose color wash 2D, 3D Isodose levels 2D, 3D Isodose Surface 3D Reference point dose summary Dose Volume Histogram plot Plan summing tool Plan comparison tools Evaluation using biological models |
| Plan Output -<br>Hardcopy | Graphics window screen dump Patient administration data Plan parameters Geometrical displays of plan data Dose distribution DVH plot BEV display Patient orientation User Configurable hardcopy layouts | Graphics window screen dump Patient administration data Plan parameters Geometrical displays of plan data Dose distribution DVH plot BEV display Patient orientation User Configurable hardcopy layouts |
| Import/Export<br>Interfaces | ARIA RadOnc integration DICOM RT Other image formats Electromagnetic Digitizer import No Film Scanner import Eclipse Scripting API (ESAPI) read only access Export field coordinates to Laser System | ARIA RadOnc integration DICOM RT Other image formats Electromagnetic Digitizer import No Film Scanner import Eclipse Scripting API (ESAPI) read only access<br>including brachytherapy plans, write access<br>in non-clinical system Export field coordinates to Laser System |
| Infrastructure | SQL Server migration Zero Clinical Downtime: Faster DB upgrades Zero Clinical Downtime: Remote deployment<br>of the clients Nexus Phase O RT Prescription integration Plan validation and status change<br>service Dosimetrically Equivalent machine<br>change service Approval modifications | SQL Server migration Zero Clinical Downtime: Faster DB upgrades Zero Clinical Downtime: Remote deployment<br>of the clients Nexus Phase O RT Prescription integration Plan validation and status change<br>service Dosimetrically Equivalent machine<br>change service Approval modifications Unicode support Electronic acceptance of Customer Approval<br>Tests MD5 checksums in dose calculation related |
{2}------------------------------------------------
.
:
:
.
:
:
.
:
{3}------------------------------------------------
・
1
·
:
{4}------------------------------------------------
*For business reasons this feature is not being commercialized at this time.
.
.
·
** Nexus is a project involving the Aria product. Aria is not part of this submission
{5}------------------------------------------------
K133247
Page 6 of 6
#### Non-clinical Testing
Verification and Validation were performed for all the new features and regression testing was performed against the existing features of Eclipse. System requirements created or affected by the changes can be traced to the test outcomes.
#### Conclusion of Non-Clinical testing
The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports) remaining which had a priority of Safety Intolerable or Customer Intolerable. Varian therefore considers Eclipse 13 to be safe and to perform at least as well as the predicate device.
#### Argument for Substantial Equivalence to the Predicate Device
A subset of features of the current device is different to the predicate. Of these, the significant changes compared with the predicate are changes to the Proton Optimizer, modifications to the Proton Dose Calculation algorithm, introduction of Proton Layer by Layer MLC support, and the improved second source modeling in AAA and AXB. Other changes are related to dose calculations and scripting for Brachytherapy, image segmentation, import and export interfaces and infrastructure.
These changes are all considered by Varian to be enhancements of the Indications for Use and the Intended Use remain unchanged. There are no changes in the principle of operation of the software. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes that Eclipse TPS is substantially equivalent to the predicate.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Silver Spring. MD 20993-0002
February 7, 2014
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K133247
Trade/Device Name: Eclipse Treatment Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: October 3, 2013 Received: November 4, 2013
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Mr. Coronado
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
For
Janine M Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K133247
Device Name: Eclipse Treatment Planning System
Indications For Use:
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
· (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.