K020027 · Computerized Medical Systems, Inc. · MUJ · Oct 8, 2002 · Radiology
Device Facts
Record ID
K020027
Device Name
FOCUS RADIATION TREATMENT PLANNING SYSTEM
Applicant
Computerized Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Oct 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of FOCUS RTP System is to provide radiation treatment planning capability, for both external beam and brachytherapy sources, to satisfy the prescription of a Radiation Oncologist. The resultant treatment plan is to be evaluated, modified as necessary, approved and delivered by qualified medical personnel. Operation of the system is identical to FOCUS systems cleared under previous Premarket Notifications with the exception the user can now select a third type of external beam particle for therapy (protons) in addition to the earlier two particles (electrons and photons).
Device Story
FOCUS Radiation Treatment Planning System (RTP) assists clinicians in planning radiation therapy. Inputs: patient diagnostic images (CT, MR, film) and dosimetry data from linear accelerators. Operation: Dosimetrist/Medical Physicist defines target volumes and critical structures; creates treatment scenarios (beam type, number, position, energy, and treatment aids like wedges/blocks). System calculates and displays 2D/3D radiation dose distributions. Output: hard-copy plan for patient file. Benefit: enables selection of optimal plan to maximize target dose while minimizing healthy tissue exposure. Proton planning capability added via algorithm developed with Massachusetts General Hospital (MGH) based on Harvard Cyclotron Lab experience.
Clinical Evidence
No clinical testing performed. Safety and effectiveness demonstrated via non-clinical algorithm validation using test cases and comparison to measured dose data at the Harvard Cyclotron Laboratory.
Technological Characteristics
Software-based radiation treatment planning system; runs on UNIX-based OS with off-the-shelf hardware (computers, digitizing tablets, printers/plotters). Adds proton beam dose calculation module to existing electron/photon planning system. Algorithm developed in collaboration with MGH.
Indications for Use
Indicated for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
## K 020027 FOCUS Radiation Treatment Planning System Proton Planning 510(k) Summary of Safety and Effectiveness
| Submitter Name: | Computerized Medical Systems, Inc. |
|--------------------------|-----------------------------------------------------------------------------|
| Submitter Address: | 1145 Corporate Lake Drive<br>St. Louis, MO 63132-1716 |
| Submitter Phone: | 314 993 0003 |
| Submitter Fax: | 314 993-0075 |
| Contact Person: | Michael A. Parsons - Director - Quality Assurance<br>and Regulatory Affairs |
| Date Summary Prepared: | December, 2001 |
| Device Trade Name: | FOCUS Radiation Treatment Planning System |
| Device Common Name: | Radiation Treatment Planning System |
| Device Classification: | System, Simulator, Radiation Therapy per<br>21CFR892.5840 |
| Substantial Equivalence: | Varian ProtonVision (K000922 & K002312) |
Device Description: The FOCUS Radiation Treatment Planning System accepts a) patient diagnostic imaging data from CT and MR scans, or from films, and b) "source" dosimetry data, typically from a linear accelerator. The system then permits the user to display and define (contour) a) the target volume to be treated and b) critical structures which must not receive above a certain level of radiation, on these diagnostic images.
{1}------------------------------------------------
## Kozoozz
### FOCUS Radiation Treatment Planning System Proton Planning 510(k) Summary of Safety and Effectiveness Page 2 of 3
Based on the prescribed dose, the user, typically a Dosimetrist or Medical Physicist, can then create multiple treatment scenarios involving the type, number, position(s) and energy of radiation beams and the use of treatment aids between the source of radiation and the patient (wedges, blocks, ports, etc.). The FOCUS system then produces a display of radiation dose distribution within the patient, indicating not only doses to the target volume but to surrounding tissue and structures. The "best" plan satisfying the prescription is then selected, one which maximizes dose to the target volume while minimizing dose to surrounding healthy volume. The parameters of the plan are output in hard-copy format for later reference and for placement in the patient file.
Previously, for situations where external beam therapy was to be used, either Electron and/or Photon radiation beams could be selected.. These were delivered by a linear accelerator whose output characteristics are input to the treatment planning system prior to beginning planning.
This Premarket Notification addresses the addition of a third type of radiation beam -Proton. The algorithm for calculating dose was provided by the Massachusetts General Hospital (MGH), based on their years of experience at the Harvard Cyclotron Lab. In addition to providing the algorithm, MGH also worked with CMS in its implementation. Software developers at MGH were trained on the CMS software development process to permit them to create code directly for use in FOCUS. As the final step, MGH provided the verification testing to assure the algorithm had been implemented correctly, measuring calculated dose against measures. A FOCUS RTP System with proton planning capability is now in clinical use at the Northeast Proton Therapy Center.
Device Intended Use: The FOCUS RTP System with will continue to be used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan sct-up.
Summary of Technological Characteristics Compared to Predicate Devices: The Proton Planning algorithm described in this 510(k) consists of modules "added-on" to the existing FOCUS Radiation Treatment Planning System previously cleared under K915691 in February, 1995, K973936 in June, 1998 and K002147 in October, 2000.
{2}------------------------------------------------
Kozouzą FOCUS Radiation Treatment Planning System Proton Planning 510(k) Summary of Safety and Effectiveness Page 3 of 3
Summary of Clinical Testing: Actual testing in a clinic was not performed as part of the development of this feature. Clinical testing is not required to demonstrate substantial equivalence or safety and effectiveness of the device.
Summary of Non-Clinical Testing: Algorithm test cases were written and executed to assure the system is calculating dose correctly for proton treatment plans. The results of testing on the Proton Algorithm feature can be found in the Validation of the CMS Proton Treatment Planning System for Treatments In Large Field Beam Line at the Harvard Cyclotron Laboratory (HCL) authored by Skip Rosenthal of MGH. This document is included in Tab 15 of this 510(k).
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a caduceus, composed of three curved lines.
#### Public Health Service
OCT 0 8 2002
Food and Drug Administrat 9200 Corporate Boulevard Rockville MD 20850
Mr. Michael A. Parsons Director, Quality Assurance and Regulatory Affairs Computerized Medical Systems, Inc.™ 1145 Corporate Lake Drive ST LOUIS MO 63132
- Re: K020027 Trade/Device Name: FOCUS Radiation Treatment Planning System with Proton Planning Capability Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: 90 MUJ Dated: August 22, 2002 Received: August 23, 2002
Dear Mr. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Statement of Indication for Use
# 510(k) Number: Ko20027
Device Name: FOCUS Radiation Treatment Planning System with Proton Planning Canability
Indication for use: The FOCUS RTP System consists of CMS-developed software running on UNIX-based operating systems and off the shelf hardware including computers and various computer peripherals (digitizing tablet, printer/plotter, etc.).
The intended use of FOCUS RTP System is to provide radiation treatment planning capability, for both external beam and brachytherapy sources, to satisfy the prescription of a Radiation Oncologist. The resultant treatment plan is to be evaluated, modified as necessary, approved and delivered by qualified medical personnel. Operation of the system is identical to FOCUS systems cleared under previous Premarket Notifications with the exception the user can now select a third type of external beam particle for therapy (protons) in addition to the earlier two particles (electrons and photons)..
> Concurrence of the Center for Devices and Radiological Health, Office of Device Evaluation (ODE)
Prescription Use
OR per 21 CFR 801.109 Over the Counter Use
David A. Heymann
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
Computerized Medical Systems Inc. - FOCUS RTP System with Proton Planning - Indications For Use Statement
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.