K180478 · Qlrad International , Ltd. · PCT · Nov 13, 2018 · Radiology
Device Facts
Record ID
K180478
Device Name
RectalPro 75 Endorectal Balloon
Applicant
Qlrad International , Ltd.
Product Code
PCT · Radiology
Decision Date
Nov 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5720
Device Class
Class 2
Indications for Use
The RectalPro™75 Endorectal Balloon device is a single-use disposable, inflatable, non-powered positioning device intended for use in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam, and X-Ray when these imaging techniques are used for Radiation Therapy (RT) planning.
Device Story
RectalPro™ 75 is a single-use, non-powered, inflatable rectal balloon; used to immobilize the prostate and displace normal tissue during radiation therapy planning (CT/X-Ray). Device consists of a curved shaft with an inflatable balloon, a locking stopper for depth control, and a syringe for inflation with air or water. Operated by clinicians in a clinical setting; inserted into the rectum and inflated to stabilize the prostate; deflated and removed post-procedure. Provides a predictable, reproducible prostate position; benefits patient by improving radiation targeting accuracy and sparing adjacent healthy tissue.
Clinical Evidence
Bench testing only. Assessments included biocompatibility (ISO 10993-1), bioburden (ISO 11737-1), leak/burst testing, transportation testing (ASTM D4169:2016), and risk assessment (ISO 14971). Results met all acceptance criteria and demonstrated performance comparable to the predicate device.
Technological Characteristics
Materials: PVC with blue colorant, silicone, polypropylene. Principle: Inflatable balloon for mechanical immobilization. Energy: Non-powered (manual syringe inflation). Form factor: Curved shaft with balloon, locking stopper, and luer-lock tubing. Connectivity: None. Sterilization: Non-sterile, not intended for user sterilization. Standards: ISO 10993-1 (biocompatibility), ISO 11737-1 (bioburden), ASTM D4169:2016 (transportation), ISO 14971 (risk).
Indications for Use
Indicated for male patients requiring prostate stabilization during CT or X-Ray imaging for radiation therapy planning.
Regulatory Classification
Identification
A rectal balloon for prostate immobilization is a single use, inflatable, non-powered positioning device placed in the rectum to immobilize the prostate in patients undergoing radiation therapy. The device is intended to be used during all the phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery.
Special Controls
In addition to the general controls of the FD&C Act, the Prostate Immobilization Rectal Balloon is subject to the following special controls:
- 1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing:
- Biocompatibility testing of the final finished device; i)
- ii) If provided sterile, sterilization validation;
- iii) If not provided sterile, bioburden testing of the final finished device;
- iv) Shelf life and expiration date validation; and
- v) Performance testing, including, but not limited to:
- A) Venting mechanism (if device has a vent mechanism);
- B) Safety mechanism(s) to prevent advancement beyond its intended safe placement; and
- C) Structural integrity testing (e.g., tensile strength, balloon leakage and burst strength).
- 2) Labeling that includes:
- i) Appropriate warnings and contraindications, including, but not limited to the following statements:
- A ) "Do not transport the patient with the rectal balloon inserted. The balloon should be removed prior to transport.";
- B) "Failure to perform the standard imaging position verification protocol may cause the device to not perform as intended." ;
- C) "Reduce the rectal balloon fill volume if the patient experiences discomfort due to the rectal balloon inflation."; and
- D) "Do not apply excessive pressure/force on the shaft or tubing of the rectal balloon.";
- ii) Adequate instructions for use on the proper insertion procedure, positioning, and inflation of the rectal balloon;
- iii) Whether the device is sterile or non-sterile; and,
- iv) An expiration date.
*Classification.* Class II (special controls). The special controls for this device are:(1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing:
(i) Biocompatibility testing of the final finished device;
(ii) If provided sterile, sterilization validation;
(iii) If not provided sterile, bioburden testing of the final finished device;
(iv) Shelf life and expiration date validation; and
(v) Performance testing including but not limited to:
(A) Venting mechanism (if device has a vent mechanism);
(B) Safety mechanism(s) to prevent advancement beyond its intended safe placement; and
(C) Structural integrity testing (
*e.g.,* tensile strength, balloon leakage and burst strength).(2) Labeling that includes:
(i) Appropriate warnings and contraindications, including, but not limited to the following statements:
(A) “Do not transport the patient with the rectal balloon inserted. The balloon should be removed prior to transport.”;
(B) “Failure to perform the standard imaging position verification protocol may cause the device to not perform as intended.”;
(C) “Reduce the rectal balloon fill volume if the patient experiences discomfort due to the rectal balloon inflation.”; and
(D) “Do not apply excessive pressure/force on the shaft or tubing of the rectal balloon.”
(ii) Adequate instructions for use on the proper insertion procedure, positioning, and inflation of the rectal balloon;
(iii) Whether the device is sterile or non-sterile; and
(iv) An expiration date.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 13, 2018
QLRAD International Ltd. % Audrey Swearingen Regulatory Affairs Specialist Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 AUSTIN, TX 78746
Re: K180478
Trade/Device Name: RectalPro™ 75 Endorectal Balloon Regulation Number: 21 CFR 892.5720 Regulation Name: Prostate Immobilizer Rectal Balloon Regulatory Class: Class II Product Code: PCT Dated: October 11, 2018 Received: October 12, 2018
Dear Mr. Swearingen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hse 2. Nils
for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180478
Device Name RectalPro™75 Endorectal Balloon
#### Indications for Use (Describe)
The RectalPro™15 Endorectal Balloon device is a single-use disposable, inflatable, non-powered positioning device intended for us in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam, and X-Ray when these imaging techniques are used for Radiation Therapy (RT) planning.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# RectalPro™75 Endorectal Balloon
## K180478
### 1. Submission Sponsor
QLRAD International Ltd. (QLRAD)
Wormerweg 6
1311 XB Almere
The Netherlands
Contact: Arjen Winkel
Title: President and CEO
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Audrey Swearingen Title: Regulatory Affairs Manager
## 3. Date Prepared
February 21, 2018
### 4. Device Identification
| | Trade/Proprietary Name: | RectalPro™ Endo Rectal Balloon (ERB) (“RectalPro 75”) |
|--|-------------------------|-------------------------------------------------------|
| | Common/Usual Name: | Rectal Balloon for Prostate Immobilization |
| | Classification Name: | Prostate immobilizer rectal balloon |
| | Regulation Number: | 21 CFR 892.5720 |
| | Product Code: | PCT |
| | Device Class: | Class II |
| | Classification Panel: | Radiology |
{4}------------------------------------------------
### 5. Legally Marketed Predicate Device(s)
K150234, Myriad Prostate Caddy™ Immobilization Rectal Balloon Item, Myriad Medical, LLC
### 6. Indication for Use Statement
The RectalPro™75 Endorectal Balloon device is a single-use disposable, inflatable, non-powered positioning device intended for use in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam, and X-Ray when these imaging techniques are used for Radiation Therapy (RT) planning.
#### 7. Device Description
The RectalPro™ 75 Endorectal Balloon (ERB) is single use, disposable, inflatable, non-powered rectal device placed in the rectum to immobilize the prostate in patients undergoing radiation therapy. The device is intended to be used during all the phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery. The RectalPro75 is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The ERB is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The ERB is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session.
The RectalPro ERB is designed with a slim curved, extended shaft and balloon. The device is provided non-sterile to the end user, is not intended to be sterilized by the end user. It is packaged in a kit configuration, consisting of three primary components:
- 1. ERB shaft This is the shaft with attached balloon and attached clear plastic tubing for filling the balloon with air / water. The end of the tubing has a female luer lock. There are white markings on the side of the ERB shaft as insertion depth markings.
- 2. Stopper This is the circular plastic piece with one flat side, and one side containing four prongs. There is a hole in the middle of the stopper for placement onto the ERB shaft.
- 3. Air / water valve This is the small plastic valve on the air tube connected to the ERB shaft. The clamp on the air / water tube stops the air / water flow by squeezing the tube closed. This component stops air / water from releasing out of the balloon after it has been filled.
The kit also includes a standard 100 cc syringe with male luer lock connector, for connecting to the female luer connection of the ERB tubing, and is used to deliver the water or air to the ERB.
#### 8. Substantial Equivalence Discussion
The following table compares the RectalPro™75 Endorectal Balloon to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance.
{5}------------------------------------------------
The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | QLRAD International Ltd. | Myriad Medical, LLC | Device Comparison |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | RectalPro™75 Endorectal<br>Balloon | Myriad Prostate Caddy™<br>Immobilization Rectal<br>Balloon | |
| 510(k) Number | To be determined | K150234 | Not applicable |
| Product Code | PCT | PCT | Same |
| Regulation<br>Number | 21 CFR 892.5720 | 21 CFR 892.5720 | Same |
| Regulation Name | Rectal balloon for<br>prostate immobilization | Rectal balloon for prostate<br>immobilization | Same |
| Indications for Use | The RectalPro™75<br>Endorectal Balloon device<br>is a single-use disposable,<br>inflatable, non-powered<br>positioning device<br>intended for use in the<br>temporary positioning of<br>the rectal wall and<br>adjacent structure in the<br>male human anatomies.<br>The purpose of the device<br>is to stabilize the prostate<br>during Computed<br>Tomography (CT) exam,<br>and X-Ray when these<br>imaging techniques are<br>used for Radiation<br>Therapy (RT) planning. | The Myriad Prostate<br>Caddy™ Immobilization<br>Rectal Balloon is a single-<br>use disposable, inflatable,<br>non-powered positioning<br>device intended for use in<br>the temporary positioning<br>of the rectal wall and<br>adjacent structure in the<br>male human anatomies.<br>The purpose of the device<br>is to stabilize the prostate<br>during Computed<br>Tomography (CT- exam,<br>and X-Ray when these<br>imaging techniques are<br>used for Radiation<br>Therapy (RT) planning.<br>The placement of the<br>balloon requires a<br>Physician, or a Physician<br>directed healthcare<br>professional and is<br>performed as a separate<br>procedure apart from the | Similar Indications<br>statements. The intended<br>use to immobilize /<br>stabilize the prostate<br>during radiation therapy is<br>the same for both. |
| Manufacturer | QLRAD International Ltd. | Myriad Medical, LLC | Device Comparison |
| Trade Name | RectalPro™75 Endorectal<br>Balloon | Myriad Prostate Caddy™<br>Immobilization Rectal<br>Balloon<br>standard CT exam and RT treatment. | |
| Mechanism of<br>Action | The balloon is inserted in<br>the patient's rectum and<br>inflated with air or water<br>using the syringe. The<br>stopper is used to lock the<br>device at the desired<br>depth. The balloon is<br>intended to stabilize the<br>prostate and displace<br>normal tissue away from<br>the radiation treatment<br>area. | The balloon is inserted in<br>the patient's rectum and<br>inflated with air or water<br>using the syringe. The<br>stopper is used to lock the<br>device at the desired<br>depth. The balloon is<br>intended to stabilize the<br>prostate and displace<br>normal tissue away from<br>the radiation treatment<br>area. | Same |
| Device<br>Components | Rectal balloon attached to<br>Curved Shaft, with<br>protective sleeve;<br>Locking Stopper;<br>Tubing,<br>Syringe | Rectal balloon attached to<br>Shaft, no protective<br>sleeve;<br>Locking Stopper;<br>Tubing;<br>Syringe | Same basic components<br>and configuration.<br>RectalPro 75 has a<br>protective sleeve on the<br>balloon to protect it<br>during handling and<br>transportation;<br>RectalPro shaft is curved<br>on the end to assist<br>insertion into the rectum.<br>These minor design<br>changes do not raise new<br>questions of safety and<br>effectiveness. |
| Inflated Diameter | 4.5 cm with 60cc air<br>4.6 cm with 80cc air<br>5.0 cm with 100cc air | 4.6 cm with 80cc air<br>5.0 cm with 100cc air | Same |
| Gas Venting Valve | No | None verified | Not known. The RectalPro<br>ERB is open-ended to<br>allow release of gas. |
| Manufacturer | QLRAD International Ltd. | Myriad Medical, LLC | Device Comparison |
| Trade Name | RectalPro™75 Endorectal<br>Balloon | Myriad Prostate Caddy™<br>Immobilization Rectal<br>Balloon | Device Comparison |
| Visualization<br>Method | CT / MRI | CT / MRI | Same |
| Sterilization | Not provided sterile nor<br>intended to be sterilized | Not provided sterile nor<br>intended to be sterilized | Same |
| Single Use | Yes | Yes | Same |
| Shelf Life | 2.5 years | 2 years | Similar |
| Patient Contacting<br>Materials | PVC with Blue colorant;<br>Silicone;<br>Polypropylene | PVC;<br>'polymer';<br>Other materials unknown | All materials used in the<br>manufacture of the<br>RectalPro75™ Endorectal<br>Balloon are suitable for<br>this use and have been<br>used in numerous<br>previously cleared<br>products. The materials<br>were tested per ISO 10993<br>and found to be<br>biocompatible. |
| Non-Clinical Bench<br>Testing | Bioburden - < 100CFU<br>Leakage (water) – no<br>leaking with 110cc water<br>after 60 min;<br>Tensile (Burst) (air) - did<br>not burst with 140cc air<br>under 7.7 lbs force for 12<br>min.;<br>Stopper Resistance -<br>stopper did not move<br>under 10 lbs of force | Bioburden - < 100CFU<br>Leakage (water) – no<br>leaking with 100cc water<br>after 10 min;<br>Tensile (Burst) (air) - did<br>not burst with 100cc air<br>under 7.1 lbs force;<br>Stopper Resistance - force<br>required to move was<br>4.91 lbs average | The performance testing<br>was consistent with the<br>predicate device. The<br>results showed that the<br>RectalPro™75™ Endorectal<br>Balloon performed as<br>intended and is at least as<br>safe and effective as the<br>predicate. |
## Table 5A – Comparison of Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
#### 9. Non-Clinical Performance Data
In demonstrating safety and effectiveness of RectalPro™75 Endorectal Balloon, and in showing substantial equivalence to the predicate device, QLRAD International Ltd. performed various assessments. The RectalPro™75 Endorectal Balloon meets the requirements for design and
{8}------------------------------------------------
functionality, with test results confirming that the design input meets the design inputs and specifications for the device, and performs comparably to the predicate device.
The RectalPro™75 Endorectal Balloon was assessed for the following in accordance with internal requirements and international standards, and met all requirements:
- Biocompatibility per ISO 10993-1. Device determined to be biocompatible.
- . Bioburden testing per ANSI/AAMI/ISO 11737-1. Device met acceptance criteria.
- Leak and Burst testing per internal protocol. Device met acceptance criteria.
- Transportation testing per ASTM D4169:2016. Device met requirements.
- Risk assessment per ISO 14971. Device has acceptable residual risks.
#### 10. Statement of Substantial Equivalence
The device has the same intended use and the same or similar technological characteristics to the Prostate Caddy Balloon predicate device. The differences do not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
Therefore, the RectalPro™75 Endorectal Balloon is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.