K232799 · Siemens Medical Solutions USA, Inc. · MUJ · Apr 26, 2024 · Radiology
Device Facts
Record ID
K232799
Device Name
syngo.via RT Image Suite
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MUJ · Radiology
Decision Date
Apr 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Autocontouring of anatomical structures
DL-based autocontouring algorithm
Dice coefficient > defined acceptance criteria
All tests passed defined acceptance criteria
—
—
Independent validation set: 413 subjects
>1 (expert team)
Indications for Use
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to view existing contours, create, edit, modify, copy contours of regions of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create simple geometric treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System. The software combines the following digital image processing and visualization tools: • Multi-modality viewing and contouring of anatomical, functional, and multiparametric images such as but not limited to CT, PET, PET/CT, MRI, Linac CBCT images · Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT) - · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique - · Automated Contouring on CT images - · Creation of contours on images supported by the application without prior assignment of a planning CT - · Manual and semi-automatic registration using rigid and deformable registration · Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points - · Supports multi-modality image fusion - · Visualization and contouring of moving tumors and organs - · Management of points of interest including but not limited to the isocenter - · Creation of simple geometric treatment plans - · Generation of a synthetic CT based on multiple pre-define MR acquisitions
Device Story
syngo.via RT Image Suite is image analysis software for radiation therapy planning. It ingests multi-modality medical images (CT, PET, PET/CT, MRI, Linac CBCT); performs 3D/4D visualization, multiplanar reconstruction, and image fusion; and provides manual, semi-automatic, and automated contouring of tumors and organs-at-risk. Operated by medical professionals in clinical settings, it supports isocenter management and simple geometric treatment plan creation. Outputs are exported to external Treatment Planning Systems. The device aids clinicians in contouring and treatment preparation, potentially improving workflow efficiency and accuracy in radiation therapy planning.
Clinical Evidence
Performance evaluation of the AI-based autocontouring feature conducted on 413 independent validation datasets (Head&Neck, Thorax&Abdomen, Pelvis). Data spanned diverse demographics, scanner manufacturers (Siemens, GE, Philips), and slice thicknesses. Performance measured via Dice coefficient comparing automated segmentations to expert-annotated ground truth. All tests met acceptance criteria for geometric overlap. No clinical data involving patient outcomes provided; evidence is based on technical verification and validation of software performance.
Technological Characteristics
Software-based image analysis tool. Features DL-based autocontouring, rigid/deformable registration, and multi-modality visualization. Connectivity via DICOM (NEMA PS 3.1-3.20 2022d). Complies with ANSI AAMI IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366-1 (usability).
Indications for Use
Indicated for 3D/4D image visualization, multimodality manipulation, and contouring to assist in radiation therapy treatment preparation (e.g., Brachytherapy, Particle Therapy, External Beam Radiation Therapy). Used by medical professionals to view, create, edit, and modify contours of body regions, targets, and organs-at-risk, and to create simple geometric treatment plans.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Siemens Medical Solutions USA, Inc. % Monsuru Bello Official Correspondent 810 Innovation Drive KNOXVILLE, TN 37932
### Re: K232799
Trade/Device Name: syngo.via RT Image Suite Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: March 27, 2024 Received: March 27, 2024
Dear Monsuru Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
April 26, 2024
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K232799
Device Name
syngo.via RT Image Suite
#### Indications for Use (Describe)
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to view existing contours, create, edit, modify, copy contours of regions of the body, such as but not limited to, skin outline, targets and organs-at-risk.
It also provides functionalities to create simple geometric treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System.
The software combines the following digital image processing and visualization tools:
• Multi-modality viewing and contouring of anatomical, functional, and multiparametric images such as but not limited to CT, PET, PET/CT, MRI, Linac CBCT images
· Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
- · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- · Automated Contouring on CT images
- · Creation of contours on images supported by the application without prior
- assignment of a planning CT
- · Manual and semi-automatic registration using rigid and
deformable registration
· Supports the user in comparing, contouring, and adapting contours based
- on datasets acquired with different imaging modalities and at different time points
- · Supports multi-modality image fusion
- · Visualization and contouring of moving tumors and organs
- · Management of points of interest including but not limited to the isocenter
- · Creation of simple geometric treatment plans
- · Generation of a synthetic CT based on multiple pre-define MR acquisitions
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
# K232799 - 510(k) Summary
### 1. Identification of the Submitter
| Submitter / Primary Contact Person | Monsuru K Bello |
|------------------------------------|-----------------------------------------------|
| | Regulatory Affairs |
| | monsuru.bello@siemens-healthineers.com |
| | +1(202) 856-6099 |
| Secondary Contact Person | Clayton Ginn |
| | Regulatory Affairs |
| | clayton.ginn@siemens-healthineers.com |
| | +1 (865) 898-2692 |
| Submitter Address | Siemens Medical Solutions, Inc. USA |
| | Molecular Imaging |
| | 810 Innovation Drive |
| | Knoxville, TN 37932 |
| | Establishment Registration Number: 1034973 |
| Legal Manufacturer | Siemens Healthcare GmbH |
| | Siemensstr 1 |
| | D-91301 Forchheim, Germany |
| | Establishment Registration Number: 3004977335 |
| Importer/Distributor | Siemens Medical Solutions USA, Inc. |
| | 40 Liberty Boulevard |
| | Malvern, PA 19355 |
| | Establishment Registration Number: 2240869 |
# 2. Device Name and Classification
| Product Name: | syngo.via RT Image Suite |
|-----------------------|---------------------------------------------------|
| Propriety Trade Name: | syngo.via RT Image Suite |
| Classification Name: | Medical Charged-Particle Radiation Therapy System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.5050 |
| Device Class: | Class II |
| Product Code: | MUJ |
# 3. Predicate Devices
### Predicate Device:
| Trade Name: | syngo.via RT Image Suite |
|-----------------------|---------------------------------------------------|
| Classification Name: | Medical Charged-Particle Radiation Therapy System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.5050 |
| Device Class: | Class II |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
Product Code: MUJ 510(k) Number: K220783
#### 4. Device Description
The subject device with the current software version SOMARIS/8 VB80 is an image analysis software for viewing, manipulation, 3D and 4D visualization, comparison of medical images from multiple imaging modalities and for the segmentation of tumors and organs-at-risk, prior to dosimetric planning in radiation therapy. syngo.via RT Image Suite combines routine and advanced digital image processing and visualization tools for manual and software assisted contouring of volumes of interest, identification of points of interest, sending isocenter points to an external laser system, registering images and exporting final results. syngo.via RT Image Suite supports the medical professional with tools to use during different steps in radiation therapy case preparation. The current software version SOMARIS/8 VB80 are as follows:
- Modifications in Advanced Contouring: Data for training and validation of Advanced Contouring was obtained through clinical collaborations from Asia, Australia, Europe, and America to provide variability in age, gender, geographic origin, etc. Both native and contrasted CT images are included.
- . Revised User Interface
#### 5. Indications for Use
syngo.via RT Image Suite is a 3D and 4D image visualization, multi-modality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).
lt provides tools to view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create simple geometric treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System.
The software combines the following digital image processing and visualization tools:
- Multi-modality viewing and contouring of anatomical, functional, and multi-parametric images such as but not limited to CT, PET, PET/CT, MRI, Linac CBCT images
- Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
- Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- · Automated Contouring on CT images
- Creation of contours on images supported by the application without prior assignment of a planning CT
- Manual and semi-automatic registration using rigid and deformable registration
- Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points
- Supports multi-modality image fusion
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots in a circular pattern.
- Visualization and contouring of moving tumors and organs
- Management of points of interest including but not limited to the isocenter
- Creation of simple geometric treatment plans
- Generation of a synthetic CT based on multiple pre-define MR acquisitions
| Subject Device | Predicate Device |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| syngo.via RT Image Suite | syngo.via RT Image Suite |
| (software version: SOMARIS/8 VB80) | (software version SOMARIS/8 VB70) |
| (510(k) number: K232799) | (510(k) number: K220783) |
| syngo.via RT Image Suite is a 3D and 4D image visualization, | syngo.via RT Image Suite is a 3D and 4D image visualization, |
| multi-modality manipulation and contouring tool that helps | multi-modality manipulation and contouring tool that helps |
| the preparation of treatments such as, but not limited to | the preparation of treatments such as, but not limited to |
| those performed with radiation (for example, | those performed with radiation (for example, |
| Brachytherapy, Particle Therapy, External Beam Radiation | Brachytherapy, Particle Therapy, External Beam Radiation |
| Therapy). | Therapy). |
| It provides tools to view existing contours, create, edit, | It provides tools to view existing contours, create, edit, |
| modify, copy contours of regions of the body, such as but | modify, copy contours of regions of the body, such as but |
| not limited to, skin outline, targets and organs-at-risk. It also | not limited to, skin outline, targets and organs-at-risk. It also |
| provides functionalities to create simple geometric | provides functionalities to create simple geometric |
| treatment plans. Contours, images and treatment plans can | treatment plans. Contours, images and treatment plans can |
| subsequently be exported to a Treatment Planning System. | subsequently be exported to a Treatment Planning System. |
| The software combines the following digital image<br>processing and visualization tools:<br>• Multi-modality viewing and contouring of anatomical,<br>functional, and multi-parametric images such as but not<br>limited to CT, PET, PET/CT, MRI, Linac CBCT images<br>• Multiplanar reconstruction (MPR) thin/thick, minimum<br>intensity projection (MIP), volume rendering technique<br>(VRT)<br>• Freehand and semi-automatic contouring of regions-of-<br>interest on any orientation including oblique<br>• Automated Contouring on CT images<br>• Creation of contours on images supported by the<br>application without prior assignment of a planning CT<br>• Manual and semi-automatic registration using rigid and<br>deformable registration | The software combines the following digital image<br>processing and visualization tools:<br>• Multi-modality viewing and contouring of anatomical,<br>functional, and multi-parametric images such as but not<br>limited to CT, PET, PET/CT, MRI, Linac CBCT images<br>• Multiplanar reconstruction (MPR) thin/thick, minimum<br>intensity projection (MIP), volume rendering technique<br>(VRT)<br>• Freehand and semi-automatic contouring of regions-of-<br>interest on any orientation including oblique<br>• Automated Contouring on CT images<br>• Creation of contours on images supported by the<br>application without prior assignment of a planning CT<br>• Manual and semi-automatic registration using rigid and<br>deformable registration |
| Supports the user in comparing, contouring, and<br>adapting contours based on datasets acquired with<br>different imaging modalities and at different time points | Supports the user in comparing, contouring, and<br>adapting contours based on datasets acquired with<br>different imaging modalities and at different time points |
| Supports multi-modality image fusion | Supports multi-modality image fusion |
| • Visualization and contouring of moving tumors and<br>organs | • Visualization and contouring of moving tumors and<br>organs |
| Management of points of interest including but not<br>limited to the isocenter | Management of points of interest including but not<br>limited to the isocenter |
| Creation of simple geometric treatment plans | Creation of simple geometric treatment plans |
| Generation of a synthetic CT based on multiple pre-<br>define MR acquisitions | Generation of a synthetic CT based on multiple pre-<br>define MR acquisitions |
# 6. Indications for Use Comparison to the Predicate Device
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a cluster of orange dots.
# 7. Comparison of Technological Characteristics with the Predicate Device
The differences between the above referenced predicate device are listed at a high-level in the following table:
| | Subject Device | Predicate Device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Siemens | Siemens |
| | syngo.via RT Image Suite<br>(software version SOMARIS/8 VB80) | syngo.via RT Image Suite<br>(software version SOMARIS/8 VB70)<br>(510(k) number: K220783) |
| Advanced<br>Contouring | Advanced Contouring tools (automatic contouring of<br>structures, nudge 3D tool, etc.). Support of Rapid<br>Results Technology. Streamlined workflow to adapt<br>contours from a prior to a current planning CT<br>("adaptive contouring").<br>Modifications:<br>• The existing, cleared Advanced Contouring<br>feature has been extended by additional<br>anatomical structures. There are no changes<br>on the underlying DL-based autocontouring<br>algorithm. | Advanced Contouring tools (automatic contouring of<br>structures, nudge 3D tool, etc.). Support of Rapid<br>Results Technology. Streamlined workflow to adapt<br>contours from a prior to a current planning CT<br>("adaptive contouring"). |
| User<br>Interface | syngo.via based GUI<br>Modifications:<br>• Minor modifications on the UI. Restructured<br>and revised UI elements. | syngo.via based GUI |
The remaining functions in syngo.via RT Image Suite remain unchanged compared to the predicate version.
- . Beam Placement
- Reference Point Management
- Patient Marking
- . Routine Contouring
- . Advanced Contouring
- Routine Structure Operations
- Duplication of Structures and POIs
- Structure Set Management
- Rigid Alignment
- Deformable Alignment
- Dose Evaluation
- Synthetic CT
- Contouring on 4D Image Data
- Lobe-Based Lung Ventilation
- Routine Reading Functionality
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
- Parallel Image Display
- . Routine Annotation Functionality
### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Non-Clinical Testing - Software Verification and Validation
Software Documentation for Enhanced documentation Level per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, is also included as part of this submission. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
#### Performance Evaluation of the Algorithm
The Al-based autocontouring feature of syngo.via RT Image Suite was tested on 413 subjects. The test data was generated from an independent set that was not seen by the model during training stage to cover a wide range of CT scanners and typical CT acquisition parameters. The general guideline was to reserve 20% of the available data for validation. The test data covers:
- . Regional distribution (Europe: IT, PT, CH, UK, NL, DE; North America: US, CA; South America: BR; Australia, Asia: JP, IN)
- Demographic distribution: Male/female
- Distribution by manufacturer of the scanner: GE, Siemens, Philips ●
- Distribution in protocols of different slice thicknesses: between < 1 mm till > 3 mm
- . Subgroup analysis regarding manufacturer, slice thickness and gender did not show any confounder.
| Subgroup | # Validation data sets |
|-------------------------|------------------------------------------------------------------------------------------|
| Data Source | Europe: 58, US: 165, Canada: 39, South America: 78, Australia: 28, Asia: 33, unknown: 12 |
| Body Region | Head&Neck: 113, Thorax&Abdomen: 216, Pelvis: 84 |
| Gender | Male: 188, female: 174. Unknown: 51 |
| Age | <=30: 1, [30-50]: 6, [50;70]: 46, >70: 20, unknown: 340 |
| Slice thickness (in mm) | <=1: 19, (1,2]: 207, (2,3]: 168, >3: 19 |
| Manufacturer | Siemens: 126, GE: 77, Philips: 140, unknown/others: 70 |
Table 1: Distribution of validation data across subgroups
Manual ground-truth segmentations were annotated by an expert team based on well accepted international contouring guidelines, followed by a rigorous independent quality assessment. The testing ensures the quantitative performance of the resulting segmentations by comparing them to the manually annotated ground truth based on the commonly used overlap metric Dice coefficient . All tests passed the defined acceptance criteria on the geometric overlap with the ground truth. The evaluation results confirm the clinical safety and performance of the autocontouring feature.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots that are arranged in a circular pattern.
#### Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Siemens hereby certifies that syngo.via RT Image Suite meets the following FDA Recognized Consensus standards listed below:
| Standard | Version | Content | FDA<br>Recognition<br>Number<br>(if applicable) |
|-----------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| ANSI AAMI IEC<br>62304 | :62304:2006/A<br>1:2016 | Medical device software - Software life cycle<br>processes [Including Amendment 1 (2016)] | 13-79 |
| NEMA PS 3.1 -<br>3.20 2022d | :2022 | Digital Imaging and Communications in<br>Medicine (DICOM) Set | 12-349 |
| ISO 14971 | :2019 | Application of Risk Management to Medical<br>Devices | 5-125 |
| IEC 62366-1 | Edition 1.1<br>2020-06<br>CONSOLIDATED<br>VERSION | Medical devices - Part 1: Application of<br>usability engineering to medical devices | 5-129 |
| ISO 15223-1 | Fourth edition<br>2021-07 | Medical devices - Symbols to be used with<br>information to be supplied by the<br>manufacturer - Part 1: General requirements | 5-134 |
| ISO 20417 | First edition<br>2021-04<br>Corrected<br>version 2021-<br>12 | Medical devices - Information to be supplied<br>by the manufacturer | 5-135 |
### 9. Conclusion
syngo.via RT Image Suite has the same intended use and indication for use as the predicate device. The result of all testing conducted was found acceptable to support the claim of substantial equivalence. The comparison of technological characteristics, clinical and non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate device that is currently marketed for the same intended use. Siemens used the same testing with the same workflows as used to clear the predicate device. Siemens considers syngo.via RT Image Suite to be as safe, as effective and with performance substantially equivalent to the commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.