RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.
Device Story
RayStation 11B is a radiation therapy treatment planning system (TPS) used by clinicians to plan, analyze, and administer radiation therapy and medical oncology treatments. Input data includes patient imaging (CT, CBCT) and treatment machine parameters. The system utilizes dose and optimization engines to propose treatment plans based on user input. New features include EQD2 dose computation using biological linear quadratic models; synthetic CT generation from CBCT by combining patient-specific CT data to improve photon dose calculation accuracy; and computation of dose-weighted Linear Energy Transfer (LET) for proton/light ion plans to assess radiobiological effects. The system comprises three applications: RayStation (planning), RayPhysics (commissioning), and RayTreat (plan delivery/record keeping). A simplified version, RayPlan 11B, offers a limited license set. Clinicians review and approve plans generated by the system before treatment administration. The device benefits patients by providing accurate, optimized radiation dose delivery and advanced radiobiological evaluation tools.
Clinical Evidence
Bench testing only. Verification and validation included system tests, risk analysis-based testing for use error mitigation, unit/subsystem testing, and dose engine validation. Clinical evaluation and user site validation were performed for new features (EQD2, synthetic CT, and LET/RBE components) to confirm adequacy for clinical use. No clinical trial data provided.
Technological Characteristics
Software-based radiation therapy treatment planning system. Operates on a shared software platform with the predicate. Features include dose optimization engines, EQD2 biological modeling, synthetic CT generation algorithms, and LET/RBE calculation modules. Software is classified as a device of major concern. Connectivity includes plan transfer to treatment delivery devices.
Indications for Use
Indicated for radiation therapy and medical oncology treatment planning, analysis, and administration. Intended for use by authorized clinicians to propose, review, and approve treatment plans for patients requiring radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
RaySearch Laboratories AB (publ) % David Hedfors Quality and Regulatory Affairs Director Eugeniavägen 18 Stockholm. 113 68 Sweden
Re: K220141
Trade/Device Name: RayStation 11B Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: January 10, 2022 Received: January 18, 2022
Dear David Hedfors:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Sullivan, Ph.D. Assistant Director Nuclear Medicine and Radiation Therapy Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220141
Device Name RayStation 11B
Indications for Use (Describe)
RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.
The system functionality can be configured based on user needs.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K220141
Image /page/3/Picture/1 description: The image contains the logo for RaySearch Laboratories. The logo consists of the text "RaySearch" in blue, with the word "Laboratories" underneath in gray. To the right of the text is a graphic of several gray lines intersecting at a central blue point, resembling rays emanating from a source.
# 510(k) Summary
RayStation 11B
{4}------------------------------------------------
| Document ID and Title | Version: |
|-------------------------------|----------|
| 510(k) Summary RayStation 11B | 1.0 |
## Table of Contents
| 1. | 510(k) Summary | 3 |
|----|--------------------------------------------------|---|
| | 1.1 510(k) owner | 3 |
| | 1.2 Contact person | 3 |
| | 1.3 Preparation date | 3 |
| | 1.4 Trade name | 3 |
| | 1.5 Common name | 3 |
| | 1.6 Classification name | 3 |
| | 1.7 Predicate devices | 3 |
| | 1.8 Device description | 3 |
| | 1.9 Intended use | 4 |
| | 1.10 Technological characteristics summary | 4 |
| | 1.11 Assessment of non-clinical performance data | 4 |
| | 1.12 Test conclusion | 4 |
{5}------------------------------------------------
Document ID and Title
510(k) Summary RayStation 11B
Version:
1.0
## 510(k) Summary
#### 1.1 510(k) owner
1.
RaySearch Laboratories AB (publ) Eugeniavägen 18 113 68 Stockholm Sweden
Tel: +46 8 510 530 00
#### 1.2 Contact person
David Hedfors Quality and Regulatory Affairs Director RaySearch Laboratories AB (publ) quality@raysearchlabs.com Email: +46 722 366 110 Tel:
#### Preparation date 1.3
January 10th, 2022
#### 1.4 Trade name
The trade name is RayStation.
The marketing name is RayStation 11B and RayPlan 11B.
#### 1.5 Common name
Radiation therapy treatment planning system
#### 1.6 Classification name
Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ)
#### 1.7 Predicate devices
K211867 RayStation 11.0
#### 1.8 Device description
RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.
RayStation consists of multiple applications:
- . The main RayStation application is used for treatment planning.
- . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
- The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.
The device to be marketed, RayStation 11B, contains modified features compared to RayStation 11.0 as indicated below:
A simplified license configuration of RayStation is marketed as RayPlan. RayPlan has a limited set of purchasable licenses and some modules will not be accessible. RayPlan is marketed as RayPlan 11B.
EQD2 dose computation (new) - From photon and/or braction doses, it is possible to compute, deform and accumulate the two Gray equivalent (EQD2) dose. The computation of the EQD2 dose uses the biological linear quadratic model, which is also the basis for the already released biological optimization and evaluation functionality.
{6}------------------------------------------------
| Document ID and Title | Version: |
|-------------------------------|----------|
| 510(k) Summary RayStation 11B | 1.0 |
Generation of synthetic CT from CBCT (new) - Two new methods (algorithms) for synthetic CT generation will be included. The synthetic CT images are created by combining information in the CBCT image and a CT image for the specific patient to allow for dose computation using the HU values in the image, as for regular CT images. In RayStation 11.0 it is possible to compute dose on CBCT images for photons using bulk density assignments. The added functionality will improve the photon dose calculation accuracy on CBCT images. Handling of LET and other RBE components (new) - This functionality enables possibility to compute and evaluate the dose weighted LET (Linear Energy Transfer) for proton and light ion plans. LET is an additional dosimetric measure that can be used to assess the radiobiological effect of the proton and light ion radiation. Radiobiological equivalent (RBE) dose is a derived quantity with dependence on both the physical dose and the LET. In RayStation 11.0, it is possible to compute and evaluate RBE doses.
#### 1.9 Intended use
RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.
The system functionality can be configured based on user needs.
#### 1.10 Technological characteristics summary
The technological characteristics are the same for RayStation 11B as for the predicate device RayStation 11.0. Both versions are built on the same software platform and share design to a high degree. Both versions have been developed under the same quality system, by the same development teams, meeting the same requirements for safety and effectiveness.
The device to be marketed, RayStation 11B contains new features as indicated below:
- . EQD2 dose computation (new)
- . Generation of synthetic CT from CBCT (new)
- Handling of LET and other RBE components (new)
Related to machine learning, there is no change compared to the predicate device.
#### 1.11 Assessment of non-clinical performance data
The test specification of RayStation 11B is a further developed version of the test specification of RayStation 11.0. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 11B therefore support the substantial equivalence of the above RayStation versions.
EOD2 dose computation (new) – Validation for photon and/or braction doses were validated as part of the Clinical Evaluation for Brachy and User Site Validation demonstrates that the dose computation adequate for clinical use.
Generation of synthetic CT from CBCT (new) – Validation of improved photon dose calculation accuracy on CBCT images was performed in CBCT Conversion validation demonstrates that the dose computation adequate for clinical use.
Handling of LET and other RBE components (new) - Validation of dose weighted LET (Linear Energy Transfer) for proton and light ion plans were performed as part of the Proton PBS Monte Carlo validation. The validation demonstrates that the dose computation adequate for clinical use.
#### 1.12 Test conclusion
The determination of substantial equivalence is not based on an assessment of non-clinical performance data. However, the entire system verification and validation specifications and reports are included in the submission as required by a software device of major concern.
A number of different types of verification activities have been performed:
- . System Tests of RayStation
{7}------------------------------------------------
- Risk analysis-based tests for use error mitigation verification ●
- . Unit and subsystem testing for low-level testing
- . Dose engine validation including internal testing
- User validation in cooperation with cancer clinics
- . Reviews of design, code and Master Labeling
The data obtained from the verification show that system tests, unit and subsystem tests have passed, and the validations been completed successfully. The reviews of design, code and labeling are also passed.
From the successful verification and validation activities, the conclusion can be drawn that RayStation 11B has met specifications and is as safe, as effective and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.