K210645 · RaySearch Laboratories AB (PUBL) · MUJ · Jun 29, 2021 · Radiology
Device Facts
Record ID
K210645
Device Name
RayStation 10.1
Applicant
RaySearch Laboratories AB (PUBL)
Product Code
MUJ · Radiology
Decision Date
Jun 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Therapeutic
Indications for Use
RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.
Device Story
RayStation 10.1 is a software-based treatment planning system (TPS) for radiation therapy and medical oncology. It processes patient data (e.g., CT/CBCT/MR images) to generate treatment plans via dose and optimization engines. New features include medical oncology planning (medication formulary/dosage calculation), brachytherapy planning (TG43 dose calculation), and proton ocular treatment planning. It also supports robust planning by accounting for organ motion through deformed image sets. The system comprises modules: RayStation (planning), RayPhysics (commissioning), RayTreat (plan delivery), and RayCommand (session management). Operated by clinicians in clinical settings, it assists in planning and administering treatments. It does not perform real-time delivery interaction or emergency shutdowns. Output is reviewed/approved by authorized users to guide clinical decision-making, aiming to improve workflow efficiency and reduce medication/planning errors.
Clinical Evidence
Bench testing only. No clinical data. Validation included system-level testing, risk analysis-based verification, unit/subsystem testing, and dose engine validation. Performance data confirmed accuracy for brachytherapy TG43 dose calculations, medical oncology dosage calculations, proton ocular treatment dose modeling, and robust planning for organ motion. Validation was conducted in cooperation with cancer clinics.
Technological Characteristics
Software-based treatment planning system. Operates on a common software platform shared with the primary predicate. Features include dose/optimization engines, brachytherapy TG43 formalism, and medical oncology dosage calculation modules. Supports DICOM-based imaging inputs. Software is classified as major level of concern. No changes to machine learning components compared to the primary predicate.
Indications for Use
Indicated for radiation therapy and medical oncology treatment planning, analysis, and administration. Intended for use by authorized clinicians to propose, review, and approve treatment plans and manage treatment delivery.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
June 29, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
RaySearch Laboratories AB (publ) % Mr. David Hedfors Quality and Regulatory Affairs Director Sveavägen 44 Stockholm, 111 34 SWEDEN
Re: K210645
Trade/Device Name: RayStation 10.1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: May 25, 2021 Received: May 28, 2021
Dear Mr. Hedfors:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210645
Device Name RayStation 10.1
### Indications for Use (Describe)
RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.
The system functionality can be configured based on user needs.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Document ID and Title | Version: |
|----------------------------------------------|----------|
| RSL-D-RS-10.1 510(k) Summary RayStation 10.1 | 1.0 |
### 510(k) Summary 1.
#### 1.1 510(k) owner
RaySearch Laboratories AB (publ) Sveavägen 44 111 34 Stockholm Sweden
Tel: +46 8 510 530 00
#### 1.2 Contact person
David Hedfors Quality and Regulatory Affairs Director RaySearch Laboratories AB (publ) Email: quality(@raysearchlabs.com +46 722 366 110 Tel:
#### 1.3 Preparation date
February 26th, 2021
#### 1.4 Trade name
The trade name is RayStation.
The trade name and version number are written together, i.e. "RayStation 10.1" to easily distinguish the submitted device from the primary predicate device RayStation 9.1.
The marketing name is RayStation 10B.
#### 1.5 Common name
Radiation therapy treatment planning system
#### 1.6 Classification name
Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUI)
#### 1.7 Predicate devices
- 1.7.1 Primary predicate device
K200569 RayStation 9.1
#### 1.7.2 Secondary predicate devices
K203172 MOSAIQ Oncology Information System K181145 Eclipse Treatment Planning System
#### 1.8 Device description
RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.
RayStation consists of multiple applications:
- The main RayStation application is used for treatment planning.
- . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
- . The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.
- o The RayCommand application is used for treatment session management including treatment preparation and sending instructions to the treatment delivery devices.
{4}------------------------------------------------
| Document ID and Title | Version: |
|----------------------------------------------|----------|
| RSL-D-RS-10.1 510(k) Summary RayStation 10.1 | 1.0 |
The device to be marketed, RayStation version 10.1, marketing name "RayStation 10B", contains modified features compared to version 9.1. The main changes are:
- . the addition of medical oncology and corresponding update of the intended use. This feature is included in the intended use of the MOSAIQ OIS secondary predicate device.
- . the addition of brachy planning. This feature is included in the intended use of the Eclipse TPS secondary predicate device.
- RayStation 10.1 is compatible with the Bebig SagiNova system afterloader. ●
- the addition of eye treatment planning (ocular planning) with protons. This feature is included in the ● intended use of the Eclipse TPS secondary predicate device.
- the extension of the robust planning to take organ motion into account and
- the extension of the treatment session management, from the RayTreat module in RayStation 9.1 to a ● new module called RayCommand. Like RayTreat, RayCommand is not intended for real-time interaction with the delivery device and does not contain any emergency shutdown functionality.
#### 1.9 Intended use
RavStation is a software system for radiation therapy and medical oncology. Based on user inout. RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.
The system functionality can be configured based on user needs.
#### 1.10 Technological characteristics summary
The technological characteristics are the same for RayStation 10.1 as for the primary predicate device RavStation 9.1. Both versions are built on the same software platform and share design to a high degree. Both versions have been developed under the same quality system, by the same development teams, meeting the same requirements for safety and effectiveness.
Comparing RayStation 10.1 with RayStation 9.1, the newer version includes usability improvements as well as new features.
The most significant new feature is the addition of medical oncology and the corresponding update of the intended use. For this feature the K203172 MOSAIQ OIS is used as a secondary predicate device. Its 510(k) Summary describes this feature as "Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology", and this is the same as in RayStation 10.1.
Other main changes are the additions of brachytherapy planning, aka ocular planning. For these features the K181145 Eclipse Treatment Planning System is used as a secondary predicate device. Its 510(k) Summary describes this feature as "Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irrachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye." This is the same features as in RayStation 10.1.
Other significant changes are:
- the extension of the robust planning to take organ motion into account and ●
- the extension of the treatment session management, from the RayTreat module in RayStation 9.1 to a ● new module called RayCommand. Like RayTreat, RayCommand is not intended for real-time interaction with the delivery device and does not contain any emergency shutdown functionality.
Related to machine learning, there is no change compared to the primary predicate device.
#### 1.11 Assessment of non-clinical performance data
The test specification of RayStation 10.1 is a further developed version of the test specification of RayStation 9.1. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 10.1 therefore support the substantial equivalence of the above RayStation versions.
{5}------------------------------------------------
| Document ID and Title | Version: |
|----------------------------------------------|----------|
| RSL-D-RS-10.1 510(k) Summary RayStation 10.1 | 1.0 |
#### 1.11.1 Brachytherapy TG43 dose calculation performance data
RayStation 10.1 accurately models the output from single sources and from sources combined into brachytherapy treatment setups in clinical plans. Since the TG43 dose computation formalism treats the full patient geometry as water, no dedicated test cases for validation of heterogenous geometries are included in the specification. This also means that all doses are reported as dose-to-water with radiation transport in water, DWw.
The brachy TG43 dose computation algorithm in RayStation10B has been successfully validated for accuracy in clinically relevant settings according to specification.
#### 1.11.2 Medical oncology dose calculation performance data
Medical oncology planning in RayStation supports dose calculation based on patient weight, patient body surface area, and flat-rate dose specification. These units of calculation are used in the majority of medical oncology regimens and are the available options in other systems.
Numerous publications describe features of computerized treatment planning for medical oncology that reduce medication errors and improve workflows. The RayStation medical oncology functions have been validated to be appropriate, when used by an intended qualified user according to the Instructions for Use, for supporting medical oncology planning workflows.
#### 1.11.3 Proton ocular treatment dose calculation performance data
The validation covers ocular treatments using the single scattering (SS) delivery technique. Since ocular treatments require small field sizes, achieved by using a field-specific block aperture, the generated Spread Out Bragg Peaks (SOBPs) are approximately laterally uniform. Hence, the SS delivery technique is modelled in RayStation as the double scattering (DS) treatment technique. A range compensator is not used.
The proton dose computation for proton ocular treatments in RayStation 10.1 has been successfully validated for accuracy in clinically relevant settings according to specification.
#### 1.11.4 Robust planning of organ motion performance data
The robustness of a plan can be evaluated by taking uncertainties such as CT inaccuracy and setup errors into account. Robustness handling in RayStation 10.1 has been expanded to also take the uncertainty of the patient's internal organ motion into account. Based on user defined criteria the system computes a set of scenario doses for evaluation.
Deformed image sets can be generated to simulate organ motion without taking multiple CT/CBCT/MR scans of the patient. Structures that are not considered to move can also be used as input to the deformation by being specified as fixed. When the new anatomy has been defined, all other structures will be mapped deformably to the generated deformed image sets.
The generated motion image sets may be used both for evaluation and as input when planning robustly against intra-fractional or inter-fractional organ motion.
The robust planning of organ motion in RayStation 10.1 has been successfully validated according to specification.
#### 1.12 Test conclusion
The determination of substantial equivalence compared to the primary predicate device is not based on an assessment of non-clinical performance data. However, the entire system verification and validation specifications and reports are included in the submission as required by a software device of major concern.
A number of different types of verification activities have been performed:
- . System Tests of RayStation
- . Risk analysis-based tests for use error mitigation verification
- Unit and subsystem testing for low-level testing
- Dose engine validation including internal testing ●
{6}------------------------------------------------
- User validation in cooperation with cancer clinics ●
- Reviews of design, code and Master Labeling .
The data obtained from the verification show that system tests, unit and subsystem tests have passed, and the validations been completed successfully. The reviews of design, code and labeling are also passed.
From the successful verification and validation activities, including the brachytherapy TG43, the medical oncology and the proton ocular treatment dose calculations and the robust planning of organ motion performance data, the conclusion can be drawn that RayStation 10.1 have met specifications and is as safe, as effective and performs as well as or better than the legally marketed primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.