K171536 · RaySearch Laboratories AB (PUBL) · MUJ · Jun 27, 2017 · Radiology
Device Facts
Record ID
K171536
Device Name
RayStation
Applicant
RaySearch Laboratories AB (PUBL)
Product Code
MUJ · Radiology
Decision Date
Jun 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Device Story
RayStation 6.1 is a radiation therapy treatment planning system (TPS) software. Inputs include patient image data (CT, PET, MRI) via DICOM. System enables contouring, multi-criteria optimization, dose tracking, treatment adaptation, and deformable registration. Workflow: user imports images; defines structures (ROIs); sets up treatment parameters (machine, energy, delivery type); optimizes plan; evaluates dose distributions (2D/3D displays, DVH curves); and exports approved plans to DICOM servers. Used in clinical settings by qualified radiation therapy staff. Output provides treatment unit set-up parameters and dose distribution estimates to guide clinical decision-making and treatment administration. Benefits include optimized, patient-specific radiation therapy planning.
Clinical Evidence
Bench testing only. Verification and validation were performed based on updated requirements and test specifications derived from the predicate device (RayStation 6.0). No clinical data provided.
Technological Characteristics
Software-based radiation therapy treatment planning system. Features include image fusion (CT, PET, MRI), contouring, collapsed cone convolution dose computation, 4D data compatibility, multi-criteria optimization, dose tracking, and deformable registration. Operates on a unified software platform. Connectivity via DICOM standards.
Indications for Use
Indicated for radiation therapy treatment planning and analysis in patients requiring radiation therapy, to be used by clinically qualified radiation therapy staff.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2017
RaySearch Laboratories AB (publ) % Mr. David Hedfors Ouality and Regulatory Affairs Director Sveavägen 44 Stockholm、111 34 Sweden
Re: K171536
Trade/Device Name: RayStation 6.1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: May 21, 2017 Received: May 25, 2017
Dear David Hedfors:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D.'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name RayStation 6.1
Indications for Use (Describe)
RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
The system functionality can be configured based on user needs.
The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Document ID and Title | Version: |
|-----------------------------------------------------------|----------|
| RSL-D-RS-6.1 Traditional 510(k) Submission RayStation 6.1 | 1.0 |
### 510(k) Summary 5.
#### 5.1 510(k) owner
RaySearch Laboratories AB (publ) Sveavägen 44 111 34 Stockholm Sweden
Tel: +46 8 510 530 00
#### 5.2 Contact person
David Hedfors Quality and Regulatory Affairs Director RaySearch Laboratories AB (publ) Email: david.hedfors@raysearchlabs.com Tel: +46 722 366 110
#### 5.3 Preparation date
May 18th, 2017
#### 5.4 Trade name
The trade name is RayStation.
The trade name and version number are written together, i.e. "RayStation 6.1" to easily distinguish the submitted device from the predicate device RayStation 6.0.
#### 5.5 Common name
Radiation therapy treatment planning system
#### 5.6 Classification name
Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ)
#### 5.7 Predicate device
RayStation 6.0 K170355
#### 5.8 Device description
RayStation 6.1 is a radiation therapy treatment planning system, i.e. a software program for planning and analysis of radiation therapy treatment plans. Functionality includes fusion capabilities (CT, PET and MRI), contouring, collapsed cone convolution dose computation and 4D data compatibility, as well as unique features such as multi-criteria optimization, dose tracking, treatment adaptation and deformable registration, all available on one platform.
The main workflow, creating a treatment plan from imported patient image data, is described below:
| Flow of Events | |
|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| User | System |
| 1. The user launches RayStation 6.1 | |
| 2. The user imports a patient and case with<br>CT images through DICOM | 3. The system imports the data and checks<br>consistency of in-data |
| 4. The user enters the Structure Definition<br>module and creates ROIs using the<br>contouring tools | 5. The system adds the ROIs to the patient<br>case |
| 6. The user enters the Plan Setup module and<br>creates a plan and a treatment setup with<br>specified machine, treatment energy and | |
{4}------------------------------------------------
| Document ID and Title<br>RSL-D-RS-6.1 Traditional 510(k) Submission RayStation 6.1 | Version:<br>1.0 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| delivery type<br>7. The user specifies beam configuration<br>including isocenter, dose grid and fluence<br>grid resolution<br><br>9. The user enters the Plan Optimization<br>module and optimizes the plan parameters<br><br>12. The user reviews the plan<br>13. The user enters the Plan Evaluation<br>module and evaluates the plan<br>14. The user approves and exports the plan<br>together with dose, structure sets and<br>images | 8. The system adds the plan and treatment<br>setup to the patient case<br><br>10. The system generates a deliverable plan<br>11. The system displays the plan as<br>- 2D and 3D dose and patient displays<br>- DVH curves<br>- Plan data (beams, segments etc.)<br><br>15. The system exports the plan and patient<br>data to a DICOM server |
#### 5.9 Intended use
RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
The system functionality can be configured based on user needs.
The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
The intended use for RayStation 6.1 is the same as for the predicate device RayStation 6.0.
#### 5.10 Technological characteristics summary
The technological characteristics are the same for RayStation 6.1 as for the predicate device RayStation 6.0.
Comparing RayStation 6.1 with RayStation 6.0, the newer version includes usability, computational speed and connectivity improvements. Both versions are built on the same software platform and share design to a high degree. Both versions have been developed under the same quality system meeting the same requirements for safety and effectiveness.
#### 5.11 Assessment of non-clinical performance data
The test specification of RayStation 6.1 is a further developed version of the test specification of RayStation 6.0. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 6.1 therefore support the substantial equivalence of the above RayStation versions.
#### 5.12 Test conclusion
The summary of the performed non-clinical tests shows that RayStation 6.1 is as safe and effective, and performs as well as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.