← Product Code [MUJ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ) · K041971

# WIMRT MODULE FOR ARTP PLANNING SYSTEM (K041971)

_Prodigm Group, Inc. · MUJ · May 20, 2005 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K041971

## Device Facts

- **Applicant:** Prodigm Group, Inc.
- **Product Code:** [MUJ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ.md)
- **Decision Date:** May 20, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The TOPSLANE product ARTP with WIMRT is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ARTP with WIMRT is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

## Device Story

WIMRT is a radiation therapy treatment planning system used in oncology clinics to plan external beam radiation treatments. It accepts CT, MR, and PET-Spec imaging data as input. The system utilizes a super-position convolution dose calculation algorithm and an optimization engine based on simulated annealing and descending gradient methods to determine optimal radiation delivery. It supports IMRT modalities including sliding window and step-and-shoot techniques. The software provides automated image registration and planning capabilities, including inhomogeneity corrections and forward IMRT via field-within-a-field or MLC-based compensators. Clinicians use the output to design radiation treatment plans for linear accelerators and teletherapy devices. By optimizing dose distribution, the device assists in delivering precise radiation therapy to patients, potentially improving treatment efficacy while sparing healthy tissue.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Radiation therapy treatment planning system; Windows-based software. Supports CT, MR, PET-Spec imaging. Dose calculation: super-position convolution. Optimization: simulated annealing and descending gradient. Supports IMRT (sliding window, step-and-shoot). Features automated image registration, inhomogeneity corrections, and forward IMRT (field-within-a-field, MLC-based).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- NOMOS Corvus Inverse Planning System ([K940663](/device/K940663.md))
- Varian Helios Inverse Planning Module ([K984532](/device/K984532.md))
- ADAC Laboratories P3IMRT Module ([K002237](/device/K002237.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K041971

# MAY 2 0 2005

# Summary of Safety and Effectiveness Compliance with 513 (i) of the Federal Food, Drug and Cosmetic Act

October 30, 2003

### General Provisions 1.

RADIATION TREATMENT PLANNING SYSTEM Common/Usual Name:

Proprietary Name: WIMRT

Applicant Name and Address:

Prodigm Group, Inc. 19078 Chaparral Drive Penn Valley, CA 95946 Telephone: 530-432-5846 Fax: 530-432-2882

### Name of Predicate Devices: 2.

- (1)
l

| Manufacturer                          | K Number |
|---------------------------------------|----------|
| NOMOS Corvus Inverse Planning System  | K940663  |
| Varian Helios Inverse Planning Module | K984532  |
| ADAC Laboratories P3IMRT Module       | K002237  |

Any statement made in conjunction with this submission regarding substantial equivalence to any other product only relates to whether the product can be lawfully marketed without pre-market approval or reclassification and is not to be interpreted as an admission or used as evidence in patent infringement litigation. As the Commissioner of the FDA has indicated, "... a determination of substantial equivalence under the Federal Food, Drug, and Cosmetic Act relates to the fact that the product can be lawfully marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits." 42 Fed. Reg. 42,520 et seg. (1977).

{1}------------------------------------------------

### Classification 3.

This device is classified as a class II device according to 21 CFR 892.5050 .

### 4. Performance Standards

Performance standards for Brachytherapy applicators have not been established by the FDA under Section 514 of the Food, Drug and Cosmetic Act.

### Intended Use and Device Description ડ.

The TOPSLANE product ARTP with WIMRT is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ARTP with WIMRT is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

### 6. Biocompatibility

No new issues of biocompataibility are raised with regard to this device.

### Summary of Substantial Equivalence 7.

This device is similar in design and construction, and has the same intended use and performance characteristics to the predicate device. It utilizes materials that are already in use in other medical devices. No new issues of safety or effectiveness are introduced by using this device.

{2}------------------------------------------------

# 

| Specification                                            | WIMRT Module                                         | ADAC P3IMRT Module                      | Varian Helios Inverse<br>Planning Module | NOMOS Corvus Inverse<br>Planning System        |
|----------------------------------------------------------|------------------------------------------------------|-----------------------------------------|------------------------------------------|------------------------------------------------|
| K Number                                                 | This filing                                          | K002237                                 | K984532                                  | K940663                                        |
| Application<br>(Use)                                     | Inverse Planning                                     | Inverse Planning                        | Inverse Plannning                        | Inverse Planning                               |
| Imaging<br>Modalities<br>Supported                       | CT, MR, PET-Spec                                     | CT, MR, PET-Spec                        | CT, MR. PET-Spec                         | CT, MR, PET-Spec                               |
| Dose<br>Calculation<br>Algorithm                         | Super-position<br>Convolution                        | Super-position<br>Convolution           | Pencil Beam Convolution                  | Finite Sum Pencil Beam                         |
| Optimization<br>Engine                                   | Simulated<br>Annealing and<br>Descending<br>Gradient | Collapsing<br>Gradient                  | Cone Descending Gradient                 | Simulated Annealing and<br>Descending Gradient |
| IMRT<br>Modalities<br>Supported                          | Sliding Window<br>and Step and<br>Shoot              | Sliding Window<br>and Step and<br>Shoot | Sliding Window and Step<br>and Shoot     | Step and Shoot                                 |
| Operating<br>System                                      | Windows                                              | UNIX                                    | Windows 2000/Windows<br>NT               | NeXT Step and UNIX                             |
| Image<br>Registration                                    | Automatic                                            | Manual and Automatic                    | Manual and Automatic                     | Manual and Automatic                           |
| Automated<br>Planning                                    | Yes                                                  | Yes via Scripts                         | Yes via Templates                        | No                                             |
| Inhomogeneity<br>Corrections                             | Yes                                                  | Yes                                     | Yes                                      | Yes                                            |
| Forward IMRT<br>via Field<br>within a Field<br>technique | Yes                                                  | Yes                                     | Yes                                      | No                                             |
| Forward IMRT<br>via MLC Based<br>compensators            | Yes                                                  | No                                      | Yes                                      | No                                             |

Prodigm Group, Inc: WIMRT - 510(K)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 0 2005

Lee Potts President Prodigm Group, Inc. 19078 Chaparral Drive PENN VALLEY CA 95946

Re: K041971 Trade/Device Name: WIMRT Regulation Number: 21 CFR §892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: April 25 2005 Received: May 2, 2005

Dear Mr. Potts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced w chave and have determined the device is substantially equivalent (for the indications for use stated in above and nave determinerarketed predicate devices marketed in interstate commerce prior to the chorolore) to regarly man date of the Medical Device Amendments, or to devices that have been May 20, 1770, the charantinent of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require upproval or a provisions of the Act. The general controls provisions of the Act device, suojoer to the general registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket n your do roo is classion (t to such additional controls. Existing major regulations affecting your Apploval), it they of sabject to actively Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any I DA has made a deceminations administered by other Federal agencies. You must comply with all the I oderal building, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin maneting of substantial equivalence of your device to a legally premarket nothleation: "The PDA midning of backanding or your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your derice of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation charact, cornation on your responsibilities under the Act from the 807.97). You may obtain other general missance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number: To be assigned Ko41971

Device Name: WIMRT

## Indications for Use:

The TOPSLANE product ARTP with WIMRT is a Radiation Therapy Treatment in the The TOF SEANL produce Artifing of patients undergoing external beam treatment in the oncology clinic. ARTP with WIMRT is used to plan radiation treatments with linear accelerators oncology cinne. Arth mail relies with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use:

or

ﻢ

ﺎ ﺗﻌ

Over-The Counter Use: ____ (Per 21 CFR 801.109)

David A. Johnson

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number _

Prodigm Group, Inc: WIMRT - 510(K)

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K041971](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K041971)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
