K040474 · Direx Systems Corp. · MUJ · Jun 23, 2004 · Radiology
Device Facts
Record ID
K040474
Device Name
ACCUSOFT
Applicant
Direx Systems Corp.
Product Code
MUJ · Radiology
Decision Date
Jun 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
AccuSoft is intended to be used for the computation, display, evaluation, and output documentation of radiation dose estimations including those for Intensity Modulated Radiation Treatment (IMRT), to be submitted for independent clinical review and judgment by a physicist or physician prior to use. The application provides output data in the form of displays or hardcopy printouts to guide a physician in selecting the optimum patient treatment plan. It is intended to provide a report to be used by a qualified professional such as a radiation oncologist, medical physicist, radiation therapist or dosimetrist.
Device Story
AccuSoft is a radiation treatment planning system (RTPS) software package; used by radiation oncologists, medical physicists, and dosimetrists in clinical settings. Input includes patient imaging data and delineation of anatomical structures/lesions. Device utilizes algorithmic approaches to calculate radiation dose distributions; enables beam planning techniques for irregularly shaped lesions. Output consists of displays and hardcopy printouts documenting dose estimations. Healthcare providers use these outputs to select optimal patient treatment plans, maximizing dose to lesions while minimizing dose to surrounding normal tissue and critical structures. System requires independent clinical review and verification by a physicist or physician prior to use.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, software validation, and performance testing results.
Technological Characteristics
Radiation treatment planning software. Complies with IEC 60601-1-4 (Programmable electrical medical systems), IEEE 829 (Software Test Documentation), IEEE 830 (Software Requirements Specifications), and ISO 14971 (Risk Management). Features include image processing, anatomical delineation, and beam planning algorithms for dose calculation.
Indications for Use
Indicated for radiation dose computation, display, and evaluation for patients undergoing radiation therapy, including Intensity Modulated Radiation Treatment (IMRT). Intended for use by qualified professionals (radiation oncologists, medical physicists, radiation therapists, dosimetrists) to guide treatment planning.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
JUN 2 3 2004
# 510(K) Summary
# AccuSoft 510(k) Number K (40474
# Applicant's Name:
Direx Systems Corp. 11 Mercer Road, Natick Business park Natick, MA 01760 United States of America
# Contact Person:
Larisa Gershtein Direx Systems Corp. 11 Mercer Road Natick, MA 01760 Tel: (888) 874 7837 Fax: (508) 651-8125 E-mail: lgershtein@direxusa.com
# Trade Name:
AccuSoft
# Model:
AccuSoft
# Classification Name:
System, Planning, Radiation Therapy Treatment
# Classification:
The FDA has classified this type of devices as class II (product code 90 IYE, Regulation No. 892.5050); they are reviewed by the Radiology Panel.
# Applicable Standards:
No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
However, AccuSoft complies with the following consensus standards:
- > IEC 60812 (1985) Failure Mode and Effects Analysis.
> IEC 60601-1-4 - Consol. Ed. 1.1 (2000-04) Medical electrical equipment - Part 1-4: General requirements for safety -- Collateral Standard: Programmable electrical medical systems.
- IEEE Std 829 (1998) Standard for Software Test Documentation A
- IEEE Std 830 (1998) Recommended Practice for Software Requirements Specifications A
- ISO 14971 (2000) Medical devices Application of risk management to medical devices A
- Guidance on Medical Device Patient Labeling; Final Guidance for Industry and
{2}------------------------------------------------
FDA Reviewers, April 19, 2001
- > An Introduction to Human Factors in Medical Devices, December 1996
- > An ma oduction is = " = " = " = " = " = " = " = " = " = " = " = " = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = > Guidance for FDA Keviewers and Industry, Cartical Devices May 29, 1998
- Chiarter Buonnestone Software Validation; Final Guidance for Industry and FDA A Staff January 11, 2002
- Start July 21, 2017 11:41
> Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices September 9, 1999
#### Intended Use:
AccuSoft is intended to be used for the computation, display, evaluation, Accusoft is intended to be accurion dose estimations including and output documentation of radiation Treatment (IMRT) that thise for Thrensity arounded and clinical review and verification by a physicist or physician prior to use.
The application provides output data in the form of displays or The application provides output a physician in selecting the bound by natucopy printouts to garao a personal provide a report to be used by patient treathent professional such as a radiation oncologist, a comportist, radiation therapist or dosimetrist.
# Device Description:
AccuSoft is a radiation treatment planning system. It consists of a AccuSoft is a factation treatment proxected algorithmic approaches to Soltware package that aloous and of extensive verification and produce radiation dose coundations to enable the proper system and patient setup and adequate radiation delivery.
It includes the image, delineation and beam planning techniques. The it includes the miage, delineation beam enables the user to treat irregularly shaped lesions, maximizing the radiation dose to the lesion, while minimizing the radiation dose to the surrounding normal tissue and critical structures.
# Predicate Devices:
Based on validations and performance testing results, Direx Systems Corp. Dascu on valuations and pubstantially equivalent to the following belleves that hecabor is backening new safety and/or effectiveness issues.
- AccuSoft, v2.02., radiation treatment planning system cleared on . Accurson, V2.02., Facking number K032171), manufacturer Direx Systems Corp.
- DMLC IV ERGO cleared on October 23, 2000 (Tracking number . K001163). Manufacturer 3D Line
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle-like symbol with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 3 2004
Ms. Larisa Gershtein QA Manager Direx Systems Corp. 11 Mercer Road NATICK MA 01760 Re: K040474 Trade/Device Name: AccuSoft Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: 90 MUJ Dated: May 18, 2004 Received: May 21, 2004
Dear Ms. Gershtein:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Bection 310(t) promotion is substantially equivalent (for the indications felerenced above and nave coloring to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regally manated posice Americal Device Americal processor of Coast Drug commerce prior to May 20, 1970, the excordance with the provisions of the Federal Food, DNA devices that have been recuassincs in accordance wal of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval ap and Cosment Act (Act) that do not require apper o the general controls provisions of the Act. The You may, therefore, thanker the device, only of equirements for annual registration, listing of general Controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classince (see above) nits. Existing major regulations affecting your device can may be subject to sach additional ochilonial official of the 21, Parts 800 to 898. In addition, FDA may be found in the Ood of Pouchar Eng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
that FDA has made a determination that your device complies with one waters and that FDA has made a Gecermanation that Jointer Federal agencies. You must comply of CCP Post or any Federal statutes and regulations administry of the orders and listing (2) CFR Part with all the Act \$ requirements, moractice requirements as set forth in the 807), labeling (21 CFR Part 820), and if applicable, the electronic product quality systems (QB) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin hanceling your ever of your device of your device to a legally premarket notification. The PDA miding of bactiation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your dones on our of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Adultionally, for questions on the promotions. Also, please note the regulation entitled, "Misbranding Othece of Compliance at (301) 59 1-100 (21CFR Part 807.97) you may obtain. Other general by reletence to premarket notified.com (Dr of Act may be obtained from the Division of Strail information on your responsionalism Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
# 510(k) Number (if known): k040474 Device Name: AccuSoft Indications for Use:
AccuSoft is to be used for the computation, display, evaluation, and output Accumentation of radiation dose estimations including those for Intensity Modulated Radiation Treatment (IMRT), to be submitted for independent clinical review and judgment by a physicist or physician much to use. The device provides output data in the form of displays or prior to abores.
hardcopy printouts to guide a physician in selecting the optimum patient treatment plan. It is intended to provide a report to be used by a tratifies professional such as a radiation oncologist, medical physicist, radiation therapist or dosimetrist.
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
y Prescription Use _____________________________________________________________________________________________________________________________________________________________
Daniel A. Sprem
(Division Sign-Off) Division of Reproductive, Abdon ann Radiological Devices 51(Jik) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.