← Product Code [MUJ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ) · K022036

# SIEMENS VIRTUAL SIMULATION (VSIM) (K022036)

_Siemens Medical Solutions USA, Inc. · MUJ · Nov 5, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K022036

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [MUJ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ.md)
- **Decision Date:** Nov 5, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The SYNGO workstation, K010938, encompasses a number of software applications for viewing, processing, filming, and archiving of medical images. VSim is one of the software applications that are offered on the SYNGO workstation. VSim permits CT Simulation to be performed on the SYNGO workstation. The CT scans are first loaded into the VSim software. On VSim the user is able to create 3D models of targets and organs. On VSim the user is able to identify the patient isocenter, place treatment beams and identify beam modifiers (blocks, apertures, and Multi-Leaf Collimators (MLCs)). This information is then sent to a radiation treatment planning system for dose calculation. The plans are then reviewed and approved by the clinician prior to transfer to the delivery system for the actual treatment.

## Device Story

VSim is a 3D post-processing software application running on the Siemens Syngo workstation. It ingests CT scans (with optional PET/MR overlays) to facilitate virtual radiation therapy simulation. Clinicians use the software to delineate anatomical targets and critical structures, define patient isocenters, and design beam geometry, including placement of jaws, blocks, and Multi-Leaf Collimators (MLCs). The system outputs DICOM-RT Structure-Sets, Plans, and Digitally Reconstructed Radiographs (DRRs). These outputs are transferred to a separate radiation treatment planning station for dose calculation. By enabling virtual setup and beam optimization, the device allows for precise treatment planning without requiring the patient to be present during the simulation process, ultimately supporting clinical decision-making for radiation delivery.

## Clinical Evidence

No clinical data provided; bench testing only. The device is a software application for treatment planning, and substantial equivalence is based on technological characteristics and functional comparison to predicate simulation systems.

## Technological Characteristics

Software application for Syngo workstation (K010938). Functions as a 3D post-processing tool for CT-based radiation therapy simulation. Supports DICOM-RT standards for data exchange (Structure-Sets, Plans, Images). Connectivity includes integration with CT scanners and linear accelerators (Siemens, Varian, Philips/Elekta).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- FocalSim ([K013112](/device/K013112.md))
- ACQSIM ([K923851](/device/K923851.md))

## Reference Devices

- SYNGO workstation ([K010938](/device/K010938.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

5 2002 NOV

Image /page/0/Picture/1 description: The image shows the word "SIEMENS" in bold, black letters. The letters are large and evenly spaced, filling most of the frame. The font appears to be sans-serif, and the overall impression is one of clarity and prominence.

K022036

510(k) Summary

| Submitter:           | Siemens Medical Solutions USA, Inc.<br>Oncology Care Systems Group<br>4040 Nelson Avenue<br>Concord, CA 94520 |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Contact person:      | Sean M. Curry<br>16787 Bernardo Center Drive, Suite A<br>San Diego, CA 92128                                  |
| Phone:               | (858) 675-8200                                                                                                |
| FAX:                 | (858) 675-8201                                                                                                |
| Proprietary name:    | Siemens Virtual Simulation (VSIM)                                                                             |
| Common name:         | Treatment Planning System                                                                                     |
| Classification:      | 892.5050                                                                                                      |
| Product Code:        | MUJ                                                                                                           |
| Classification name: | System Planning Radiation Therapy Treatmen                                                                    |

#### Substantial equivalence claimed to:

K013112, FocalSim, Computerized Medical Systems, Inc. K923851 , ACQSIM, Philips Medical Systems(Cleveland), Inc.

#### Description:

Siemens Virtual Simulation (VSim) is a software application that runs on Siemens Medical Workstation, Syngo (K010938). It is intended to give the user general Viewing & Examination tools for viewing medical diagnostic images. Computed Tomography (CT) scans are the centerpiece of the diagnostic images used by the VSim. It will be possible to load other modality images, Positron Emission Tomography (PET) and Magnetic Resonance (MR), in conjunction with the CT images for treatment planning.

VSim is intended to provide tools for delineating and representing targets and critical Structures. It takes specifications and dimension for the dose delivery system (Siemens and other vendors). It will then enable the user to design complex beam profiles and place them for optimum treatment of the discase. 3D Graphical representation and visualization of all the relevant objects allow for a virtual setup and treatment of the patient without involving the patient.

#### Siemens Medical Solutions USA, Inc.

{1}------------------------------------------------

# SIEMENS

The VSim software will be used in a typical scenario summarized below:

- Patient is registered, and the relevant data is entered into the system, ı
- Schedule is created for radiation therapy, I
- Patient is called in for a CT scan,
- CT scan is performed, 트
- VSim is used for initial viewing and definition of isocenter reference (for patient I marking)
- 트 Patient is marked prior to leaving the CT room,
- 트 VSim is used to delineate the Structures targeted for radiation and other critical Structures.
- . VSim is used to establish a reference point, known as the isocenter reference, which will be marked on the patient, and any other coordinates will be based on this reference coordinate,
- 트 VSim is used to design the beam geometry,
- . VSim is used to place the jaws and to define the blocks and/or MLC,
- 트 VSim is used to print charts used for flow of information,
- 트 VSim is used to print images on film,
- t The plans are transferred to radiation therapy planning station for dose calculation.
- l The isocenters may be changed by the dosimetrist,
- . The plans may be transferred back to the VSim workstation for verification.

VSim is a 3D post-processing software application that uses CT planning images as input and creates the following data objects as output:

- 1. Structure sets stored in the form of DICOM-RT Structure-Set,
- 2. Reference Points (including isocenters) stored in the form of DICOM-RT Structure-Set.
- 3. Plans, including beams stored in the form of DICOM-RT-Plan, and
- 4. Reference Images in the form of Digitally Reconstructed Radiographs (DRRs), (one DRR for each beam in the plan) stored in the form of DICOM-RT-Image.

#### Intended use:

The SYNGO workstation, K010938, encompasses a number of software applications for viewing, processing, filming, and archiving of medical images. VSim is one of the software applications that are offered on the SYNGO workstation.

VSim permits CT Simulation to be performed on the SYNGO workstation. The CT scans are first loaded into the VSim software. On VSim the user is able to create 3D models of targets and organs. On VSim the user is able to identify the patient isocenter, place treatment beams and identify beam modifiers (blocks, apertures, and Multi-Leaf Collimators (MLCs)). This information is then sent to a radiation treatment planning system for dose calculation. The plans are then reviewed and approved by the clinician prior to transfer to the delivery system for the actual treatment.

### Siemens Medical Solutions USA, Inc.

{2}------------------------------------------------

# SIEMENS

### Summary of technological characteristics Compared to Predicate Devices:

The VSim software described in this 510(k) will be an add-on to the SYNGO workstation, K010938. The VSim system incorporates no technological characteristics not currently in the predicate simulation devices. VSim supports the most popular CT Scanners as well as the three most common linear accelerators (Siemens, Varian, and Philips/Elekta)

## Siemens Medical Solutions USA, Inc.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with a stylized eagle or bird symbol on the right. To the left of the bird symbol, the text "DEPARTMENT OF HEALTH &" is arranged in a circular fashion. The bird symbol is composed of three curved lines that suggest the shape of a bird in flight.

5 2002 NOV

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions USA. Inc. % Mr. Sean M. Curry Chief Operating Officer Certified Software Solutions, Inc. 16787 Bernardo Center Drive Suite A SAN DIEGO CA 92128

Re: K022036

Trade/Device Name: Siemens Virtual Simulation (VSIM) Release 1.0 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: 90 MUJ Dated: October 18, 2002 Received: October 21. 2002

Dear Mr. Curry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Henry C. Snodgin

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):_

K022036

Device Name: Siemens Virtual Simulation (VSIM)

Indications for Use:

The SYNGO workstation encompasses a number of software applications for viewing, processing, filming, and archiving of medical images. VSim is one of the software applications that are offered on the SYNGO workstation, K010938.

VSim permits CT Simulation to be performed on the SYNGO workstation. The CT scans are first loaded into the VSim software. On VSim the user is able to create 3D models of targets and organs. On VSim the user is able to identify the patient isocenter, place treatment beams and identify beam modifiers (blocks, apertures, and MLCs). This information is then sent to a radiation treatment planning system for dose calculation. The plans are then reviewed and approved by the clinician prior to transfer to the delivery system for the actual treatment.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use (Per 21 CER 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K022036](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/MUJ/K022036)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
