← Product Code [LHN](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/LHN) · K201042

# PROBEAT-CR (K201042)

_Hitachi , Ltd. · LHN · Jul 13, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/LHN/K201042

## Device Facts

- **Applicant:** Hitachi , Ltd.
- **Product Code:** [LHN](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/LHN.md)
- **Decision Date:** Jul 13, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

## Device Story

Proton beam therapy system; generates/delivers proton beams for tumor treatment. Components: synchrotron accelerator, beam transport system, treatment rooms (rotating gantry/fixed beam). Inputs: patient positioning data (CBCT, orthogonal X-ray), fiducial tracking data (via integrated Real Time Image Gating System). Operation: beam delivery system scans proton beam (discrete spot scanning) to match prescribed dose/distribution. Used in clinical oncology settings by trained medical staff. Integrates Real Time Image Gating System (K171049) to track implanted fiducials and gate beam delivery based on patient motion. Optional patient couch top extension allows expanded positioning angles (±90 degrees). Output: controlled proton beam irradiation. Benefits: precise dose delivery to localized tumors while sparing surrounding healthy tissue.

## Clinical Evidence

Bench testing only. No clinical data provided. Validation included design verification and validation testing for the optional couch top extension and software verification and validation for the updated Real Time Image Gating System (RGPT).

## Technological Characteristics

Proton beam therapy system; synchrotron accelerator (70-230 MeV); discrete spot scanning nozzle. Subsystems: beam production, beam transport, beam delivery (rotating gantry/fixed beam). Patient positioning: robotic 6-DOF couch, CBCT, orthogonal X-ray, laser alignment. Connectivity: integrated RGS for gating. Software: RGPT for fiducial tracking/gating. Sterilization: N/A (capital equipment).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Hitachi Ltd., Healthcare Hitachi Works, PROBEAT-CR Proton Beam Therapy Device ([K191801](/device/K191801.md))

## Reference Devices

- Hitachi Ltd., Healthcare Hitachi Works, Real Time Image Gating System for Proton Beam Therapy Systems ([K171049](/device/K171049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 13, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.

Hitachi Ltd. % Mr. Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004

Re: K201042

Trade/Device Name: PROBEAT-CR Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: April 20, 2020 Received: April 20, 2020

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting

{1}------------------------------------------------

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K201042

## Device Name

### PROBEAT-CR

Indications for Use (Describe)

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

区 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY

### Hitachi Ltd.'s PROBEAT-CR Proton Beam Therapy Device

### Submitter

K201042

Hitachi Ltd., Healthcare Radiation Oncology Systems, Kashiwa 2-1 Shintoyofuta, Kashiwa-shi Tiba-ken, 277-0804, Japan Telephone: +81 (4) 7131-4280 Facsimile: +81 (4) 7132-6837

Contact Person: Tomoyuki Seino

Date Prepared: April 20, 2020

### Name of Device:

PROBEAT-CR Proton Beam Therapy System

#### Common or Usual Name:

Proton Beam Therapy Device

### Classification Name:

System, Radiation Therapy, Charged-Particle, Medical

### Requlatory Class:

Class II

#### Product Code:

LHN

#### Predicate Devices

Hitachi Ltd., Healthcare Hitachi Works, PROBEAT-CR Proton Beam Therapy Device (K191801)

#### Reference Devices

Hitachi Ltd., Healthcare Hitachi Works, Real Time Image Gating System for Proton Beam Therapy Systems (K171049)

{4}------------------------------------------------

## Device Description

The PROBEAT-CR is a proton beam irradiation system. which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose. dose distribution and directed to the prescribed patient treatment site.

### Intended Use / Indications for Use

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

### Summary of Technological Characteristics

The PROBEAT-CR has two main subsystems: (1) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment, and (2) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered. The PROBEAT-CR comprises the following components and subsystems:

- . Beam production system
	- Accelerator system (LINAC. Synchrotron). o
	- Beam transport system (Low/High Energy Beam Transport systems). o
- Beam delivery system in 4 separate treatment rooms. Each of 3 rooms will have a ● rotating gantry and 1 room will have a fixed beam.
	- o Gantry Room
		- . Scanning Nozzle
		- . Rotating Gantry
		- l Patient Positioning System
		- l Orthogonal X-ray system
		- . Cone Beam CT
	- Fixed Beam Room o
		- . Patient Positioning System
		- . Orthogonal X-ray system
		- . Treatment Control and Safety System

The subject PROBEAT-CR is a modification to the cleared PROBEAT-CR to include the incorporation of the previously cleared Real Time Image Gating System for Proton Beam Therapy Systems ("RGS" or "RGPT") (K171049) for tracking implanted fiducials to qate the delivery of the proton beam, and the addition of an optional patient couch top extension as an accessory to allow for different patient positioning configurations.

A table comparing the key features of the subject and predicate devices is provided below.

{5}------------------------------------------------

# Performance Data

The following testing was performed to validate the modifications to the device:

- Design verification and validation testing for the addition of the optional top couch . extension
- Software verification and validation for the updated RGS (RGPT) software .

# Conclusions

The company's subject PROBEAT-CR is as safe and effective as the predicate PROBEAT-The subject PROBEAT-CR has the same intended use and indications as the CR. previously cleared PROBEAT-CR. In addition, the subject PROBEAT-CR has very similar technological characteristics and principles of operation as its predicate. Although there are minor differences between the PROBEAT-CR and the predicate, namely the incorporation of the cleared RGS (RGPT) software system and the addition of an optional patient couch top extension, those differences do not raise new questions of safety or efficacy. The performance testing demonstrates that the device is as safe and effective as the predicate. Thus, the PROBEAT-CR is substantially equivalent.

# Substantial Equivalence Table

|                                    | Hitachi PROBEAT-CR (Subject<br>Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Hitachi PROBEAT-CR (K191801)                                                                                                                                                                                                                         |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                | The PROBEAT-CR is a medical<br>device designed to produce and<br>deliver a proton beam for the<br>treatment of patients with localized<br>tumors and other conditions<br>susceptible to treatment by<br>radiation.                                                                                                                                                                                                                                                                                                                 | The PROBEAT-CR is a medical<br>device designed to produce and<br>deliver a proton beam for the<br>treatment of patients with localized<br>tumors and other conditions<br>susceptible to treatment by<br>radiation.                                   |
| Accelerator                        | Synchrotron                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Synchrotron                                                                                                                                                                                                                                          |
| Particle                           | Protons                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Protons                                                                                                                                                                                                                                              |
| Variable Energy                    | 70-230 MeV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 70-230 MeV                                                                                                                                                                                                                                           |
| Nozzles                            | Discrete Spot Scanning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Discrete Spot Scanning                                                                                                                                                                                                                               |
| Support for Patient<br>Positioning | Gantry built-in type Cone Beam<br>CT is used to verify patient<br>positioning.<br>A conventional x-ray system can<br>be also equipped in the same<br>treatment room.<br>A computer assisted patient<br>position system, (PIAS) is<br>available for use.                                                                                                                                                                                                                                                                            | Gantry built-in type Cone Beam CT<br>is used to verify patient positioning.<br>A conventional x-ray system can<br>be also equipped in the same<br>treatment room.<br>A computer assisted patient<br>position system, (PIAS) is available<br>for use. |
|                                    | Hitachi PROBEAT-CR (Subject<br>Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Hitachi PROBEAT-CR (K191801)                                                                                                                                                                                                                         |
| Treatment Room                     | Typically 3 rotating gantry rooms<br>and 1 fixed beam room (Maximum<br>5 total).<br><Note>All rooms are dedicated<br>to the DSSS                                                                                                                                                                                                                                                                                                                                                                                                   | Typically 3 rotating gantry rooms<br>and 1 fixed beam room (Maximum<br>5 total).<br><Note>All rooms are dedicated to<br>the DSSS                                                                                                                     |
| Gantry rotating angle              | 360 degrees<br>(-180-180degrees)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 360 degrees<br>(-180-180degrees)                                                                                                                                                                                                                     |
| Patient Positioner                 | Rotating gantry room and fixed<br>beam room,<br>(a) Patient Couch<br>Swing type Robotic Patient<br>Positioning System with 6 Degrees<br>of freedom<br>Isocentric rotating angle of the<br>couch is ±45 degrees for standard<br>base and extension configuration.<br>With the optional couch top<br>extension, isocentric rotating angle<br>can be extended to ±90 degrees<br>for short base and overlay<br>configuration, with a ±5 degree<br>movement range for the rolling and<br>pitching angles.<br>(b) Laser Alignment System | Rotating gantry room and fixed<br>beam room,<br>(a) Patient Couch<br>Isocentric rotating angle of the<br>couch is ±45 degrees for standard<br>base and extension configuration.<br>(b) Laser Alignment System                                        |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                      |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                      |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                      |
| Patient Positions<br>Supported     | Head first-supine ("HFS"), head<br>first-prone ("HFP"), feet first-<br>supine ("FFS"), and feet first-prone<br>(                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                      |

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/LHN/K201042](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/LHN/K201042)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
