K093250 · Varian Medical Systems · LHN · Dec 24, 2009 · Radiology
Device Facts
Record ID
K093250
Device Name
PROTON THERAPY MULTILEAF COLLIMATOR
Applicant
Varian Medical Systems
Product Code
LHN · Radiology
Decision Date
Dec 24, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Varian Proton Therapy Multileaf Collimator is an accessory radiation collimator intended to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
Device Story
The Proton Therapy Multileaf Collimator (MLC) is an accessory mounted on a proton radiation therapy system to shape the treatment field perimeter. It operates in two modes: Pencil Beam Scanning (PBS), where the MLC is parked with the carriage retracted to allow the nozzle magnets to direct the beam; and Beam Shaping, where the MLC collimates the beam for double/single scattering and uniform scanning modes, replacing patient-customized final collimators. Used in clinical radiation oncology settings, the device is operated by trained personnel. By replacing manual patient-specific collimators, it streamlines the treatment workflow and improves the precision of radiation delivery, potentially reducing radiation exposure to healthy tissues and critical organs.
Clinical Evidence
Bench testing only.
Technological Characteristics
Accessory radiation collimator; mechanical multileaf design; mounts to proton radiation therapy system; operates in two modes (Pencil Beam Scanning and Beam Shaping).
Indications for Use
Indicated for patients requiring proton radiation therapy, where the device assists radiation oncologists in shaping the proton beam to target defined volumes while sparing surrounding healthy tissue and critical organs.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Varian Medical Systems Millennium Multileaf Collimator (K050442)
Related Devices
K982549 — RADIONICS MINI MULTI-LEAF COLLIMATOR, MODEL RADIONICS MMLC · Radionics, Inc. · Dec 30, 1998
K121657 — PROTON RADIATION THERAPY BEAM-SHAPING APERTURE PROTON RADIAION THERAPY BEAM-SHPING RANGE COMPENSATOR · Proton Systems · Oct 23, 2012
K132236 — .DECIMAL PROTON APERTURE · .Decimal, Inc. · Oct 16, 2013
K083058 — IBA PROTON THERAPY SYSTEM - PROTEUS 235 · Ion Beam Applications S.A. · Apr 15, 2009
K050442 — MILLENNIUM MULTI-LEAF COLLIMATOR · Varian Medical Systems · Mar 23, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K093252
#### 510(k) Summary
The information below is provided for the Varian Medical Systems Proton Therapy Multileaf Collinator following the format of 21 CFR 807.92.
- 1. Submitter: Varian Medical Systems 3100 Hansen Wav. M/S e110 Palo Alto, CA 94304 Contact Name: Vy Tran Phone: 650/424.5731 Fax: 650/842.5040 E-mail: vy.tran@varian.com
# DEC 2 4 2009
.
| | 2. Name of the Device: | Proton Therapy Multileaf Collimator |
|--|----------------------------|------------------------------------------------------------------------------------------|
| | Trade / Proprietary Names: | iPTMLC ; Proton Therapy Multileaf Collimator; Proton Multileaf<br>Collimator; Proton MLC |
| | Common or Usual Name: | Accessory to Proton Therapy System |
| | Classification Name: | Medical Charged Particle Radiation Therapy System |
| | | 21 CFR §892.5050 |
| | | Class II |
| | Product Code: | LHN |
- 3. Predicate Device to claim substantial equivalence: Varian Medical Systems Millennjum Multileaf Collimator - K050442
- 4. Description of the Device:
The Varian Proton Therapy Multileaf Collimator is an accessory proton beam collimator designed to be mounted on a proton radiation therapy system and is designed to shape the treatment field perimeter.
The Proton Therapy Multileaf Collimator is designed to shape a proton beam for cancer treatment. There are two operating modes:
- Pencil Beam Scanning (PBS) where proton therapy equipment nozzle scanning magnets direct a. the proton beam to the voxel of treatment while the Proton Therapy Multileaf Collimator is parked with the carriage fully retracted for largest possible exposure field.
- b. Beam Shaping where the Proton Therapy Multileaf Collimator collimates the beam for treatment, replacing the current patient-customized final collimator. This Proton Therapy Multileaf Collimator mode of operating is used for double and single scattering and uniform scanning proton treatment modes.
- 5. Intended Use Statement
The Varian Proton Therapy Multileaf Collimator is an accessory radiation collimator intended to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
{1}------------------------------------------------
#### Indications for Use Statement ن
The Varian Proton Therapy Multileaf Collimator is an accessory radiation collimator intended to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
#### 7. Substantial Equivalence
The Varian Medical Systems Proton Therapy Multileaf Collimator submission illustrates substantial equivalence to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Vy Tran Vice President Corporate Regulatory Affairs Varian Medical Systems 3100 Hansen Way PALO ALTO CA 94304-1038
DEC 2 4 2009
Re: K093250
Trade/Device Name: Proton Therapy Multileaf Collimator Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: October 14, 2009 Received: October 16, 2009
Dear Ms. Tran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Enclosure
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lucia M. Morris
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Varian Medical Systems. The logo consists of the word "VARIAN" in a serif font, with a stylized "i" that has a curved line above it. Below the word "VARIAN" are the words "medical systems" in a smaller, sans-serif font. The logo is black and white.
Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 USA Tel +1 650 493 4000 www.varian.com
#### Indications for Use Statement
510(k) Number (if known):
K093256
Device Name:
Proton Therapy Multileaf Collimator
Indications for Use:
The Varian Proton Therapy Multileaf Collimator is an accessory radiation collimator intended to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR § 801.109)
Over-the-counter
Armu M. Mothy
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K093250
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches