ADVANTAGE-STRAND/ ADVANTAGE-LOAD BRACHYTHERAPY KIT, MODEL IASLBKA

K091117 · Isoaid, LLC · KXK · Sep 15, 2009 · Radiology

Device Facts

Record IDK091117
Device NameADVANTAGE-STRAND/ ADVANTAGE-LOAD BRACHYTHERAPY KIT, MODEL IASLBKA
ApplicantIsoaid, LLC
Product CodeKXK · Radiology
Decision DateSep 15, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The IsoAid Brachytherapy Kit is intended for the treatment of selected localized tumors. The devices are implanted as a source of nuclear radiation for therapy.

Device Story

Pre-sterilized brachytherapy kit containing 18-gauge stainless steel needle and custom-loaded radioactive seeds (I-125 or Pd-103). Seeds/spacers loaded into absorbable suture strand or directly into needle; maximum 20 components per needle. Bone wax used at needle tip to retain implants. Used by clinicians for implantation into tissue/body cavities to treat localized tumors via nuclear radiation. Provides localized radiation therapy; benefits patients with unresectable or radiosensitive tumors.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

18-gauge stainless steel needle; absorbable suture material; I-125 or Pd-103 radioactive sources; bone wax sealant. Pre-sterilized kit. No software or electronic components.

Indications for Use

Indicated for localized, unresectable tumors or tumors with low to moderate radiosensitivity.

Regulatory Classification

Identification

A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.

Special Controls

*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 691111 SEP & 5 2009 ## 510 (k) SUMMAR Y (as required by 807.92(c)) Submitter of 510(k): IsoAid, LLC 7824 Clark Moody Blvd., Port Richey, FL 34668 Phone: 727-815-3262 Fax: 727-815-1973 Benjamin Roedell April 16, 2009 Advantage-Strand™ / Advantage-Load™ Brachytherapy Kit Brachytherapy Seed Strand in Needle Class II (21 CFR 892.5730, Product Code KXK) Radionuclide Brachytherapy Source Device 510(k) # Brachytherapy Strand Device K040339 The IsoAid Brachytherapy Kit is a pre-sterilized kit containing brachytherapy needle and a custom-loaded strand (K013975) of seeds spaced at a precise distance within absorbable suture. The strand is optional, when no strand is requested the seeds and spacers are custom-loaded directly into the needle. A maximum total of 20 seeds and spacers can be loaded into a needle. The spacers (K010621) are made from the same material as the sutures. The stranded Pd-103 (K033770) and I-125 (K011205) implants are placed inside the needle. Bone Wax (K024372) is used at the tip of the need to keep the implants from falling out. The needle is made from 18-gauge stainless steel. The IsoAid Brachytherapy Kit is intended for the treatment of selected localized tumors. The devices are implanted as a source of nuclear radiation for therapy. The IsoAid Brachytherapy Kit is indicated for tumors that are localized, unresectable, or have low to moderate radiosensitivity. l Contact Person: Date of Summary: Trade Name: Common Name: Classification: Classification Name: Predicate Devices: Device Description: Intended Use: Indications for Use: Comparison to Predicate Device: {1}------------------------------------------------ | | IsoAid Brachytherapy Kit | Bebig Brachytherapy Strand Device | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K091117 | K040339 | | Indications for Use | Same | | | Description | The IsoAid Brachytherapy Kit is a pre-sterilized kit containing brachytherapy needle and a custom-loaded strand (K013975) of seeds spaced at a precise distance within absorbable suture. The strand is optional, when no strand is requested the seeds and spacers are custom-loaded directly into the needle. A maximum total of 20 seeds and spacers can be loaded into a needle. The spacers (K010621) are made from the same material as the sutures. The stranded Pd-103 (K033770) and I-125 (K011205) implants are placed inside the needle. Bone Wax (K024372) is used at the tip of the need to keep the implants from falling out. The needle is made from 18-gauge stainless steel. | The Brachytherapy Strand Device is used for the treatment of localized tumors and is placed into a body cavity or tissue. It consists of a pre-sterilized kit containing a prostate seeding needle and a custom-loaded strand of seeds spaced at a precise distance within absorbable suture. The spacers are made from the same material as the sutures. The customized strand can contain a variable number (1-12) of seeds and/or seeding spacers (maximum 12 components per strand). The stranded Pd-103 and I-125 implants are placed inside the needle. The needle is made from 18 gauge stainless steel. | | Radioactive<br>Isotope(s) | Iodine-125 and/or Palladium-103 | Same | | Application Method | Through an 18 gauge needle | Same | : {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of a bird-like symbol, possibly representing an eagle or other bird of freedom. The bird is depicted with three curved lines forming its body and wings. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service SEP 1 5 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002 Mr. Benjamin Roedell Quality Assurance Manager IsoAid LLC 7824 Clark Moody Blvd. PORT RICHEY FL 34668 Re: K091117 Trade/Device Name: Advantage-Strand™ / Advantage-Load™ Brachytherany Kit Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: KXK Dated: August 14, 2009 Received: August 17, 2009 Dear Mr. Roedell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K091117 Advantage-Strand™ / Advantage-Load™ Brachytherapy Kit Device Name: Indications for Use: The IsoAid Brachytherapy Kit is indicated for tumors that are localized, unresectable, or have low to moderate radiosensitivity. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 1 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) hoanm Wha (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...