← Product Code [KPQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ) · K132944

# ADVANTAGESIM MD MR PELVIC ORGAN AT RISK SEGMENTATION (K132944)

_Ge Hungary Kft · KPQ · Mar 14, 2014 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K132944

## Device Facts

- **Applicant:** Ge Hungary Kft
- **Product Code:** [KPQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ.md)
- **Decision Date:** Mar 14, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5840
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

AdvantageSim™ MD is used to prepare geometric and anatomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position. Definition of the anatomical volumes may be assisted by additional CT, MR or PET studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user to define the target or treatment volume over a defined range of the respiratory cycle. The geometric parameters of a proposed treatment field are selected to allow non-dosimetric, interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3 D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph.

## Device Story

AdvantageSim™ MD is an oncology software application for radiotherapy treatment planning. It processes CT, MR, and PET images to assist clinicians in defining anatomical volumes and organs-at-risk. The MR pelvic organ at risk segmentation option provides semi-automated contouring tools; users place seed points within an organ to initiate segmentation. The software overlays generated contours onto co-registered CT planning images. Clinicians use manual editing tools to modify, accept, or reject contours based on clinical judgment. The device is used in clinical settings by radiation oncologists and other specialists. It does not perform auto-detection or auto-saving. The output assists in geometric and anatomical data preparation for external beam radiotherapy, potentially reducing inter-operator variability and improving clinical efficiency. The clinician retains ultimate responsibility for all diagnoses and patient management decisions.

## Clinical Evidence

A usability study was conducted with three board-certified radiation oncologists using consented clinical images. The study evaluated measurement accuracy, precision, efficiency (time comparison), and qualitative user feedback. Results demonstrated that the software reduces inter-operator variability and provides statistically significant and practically meaningful improvements in clinical efficiency, confirming the device is easy to learn and useful for increasing throughput.

## Technological Characteristics

Software-based oncology application for radiotherapy planning. Employs semi-automated segmentation algorithms requiring user-placed seed points. Supports co-registration of CT, MR, and PET images. Provides manual contour editing tools. Operates on standard clinical workstations. Complies with voluntary standards for risk analysis, design, and verification testing.

## Regulatory Identification

A radiation therapy simulation system is a fluoroscopic or radiographic x-ray system intended for use in localizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irradiation field produced. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- AdvantageSim™ MD with CT Atlas-based Contouring and Replanning Options ([K132045](/device/K132045.md))
- Segasist Prostate Auto-Contouring Software ([K111311](/device/K111311.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a document page with handwritten text and a logo. The text at the top reads "K132 944 page 1 of 4". Below the text is the GE logo, which consists of the letters "GE" inside a circular design.

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

## 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | September 16, 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | GE Hungary Kft.<br>Akron utca 2<br>2040 Budörs, Hungary                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Contact<br>Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Angelique Cruz<br>Regulatory Affairs Leader<br>Tel: 0033 (0)1 30 70 47 55<br>Fax: (262)364-2506                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Secondary Contact<br>Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Huy Doan<br>Regulatory Affairs Director<br>Tel: 262-312-7751<br>Mobile: 414-5818553<br>Fax:: (262)364-2506<br><br>Helen Peng<br>Regulatory Affairs Manager<br>Tel: (262)5485091<br>Fax: (262)364-2506                                                                                                                                                                                                                                                                                                                        |
| Device Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | AdvantageSim™ MD MR pelvic organ at risk segmentation                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common/Usual Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | AdvantageSim™ MD MR pelvic organ at risk segmentation                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Names:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 21CFR 892.5840, Radiology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | KPQ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device(s):                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | K132045 - AdvantageSim™ MD with CT Atlas-based Contouring and<br>Replanning Options from GE Healthcare<br>K111311 Segasist Prostate Auto-Contouring Software from Segasist<br>Technologies                                                                                                                                                                                                                                                                                                                                   |
| Device Description /<br>Intended Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | AdvantageSim™ MD is a CT/MR/PET oncology application used by<br>clinicians (radiologist, radiation oncologist, medical oncologist                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| nuclear medicine physicians and trained healthcare professional)<br>to assist treatment planning.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| AdvantageSim MD with MR pelvic organ at risk segmentation<br>Option is used to provide MR based prostate and pelvic organs-at-<br>risk segmentation. A suite of semi-automated MR based organ<br>segmentation contouring allows generating complex structures<br>around organs at risk. These contours overlay on the co-registered<br>CT planning image.                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| The segmentation methods in the modified device are semi-<br>automatic. The user has to place seed points to identify an inner<br>point of the organ to contour.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| The software offers a suite of manual contour editing tools enabling<br>the user to edit, modify, or change contours generated from the MR<br>segmentation tools to their desired configuration based on their<br>medical and clinical knowledge and experience. The results<br>provided by the software needs to be approved by the experienced<br>clinician and can always be modified or corrected by him/her. It is<br>up to the expert user to accept the result without any change,<br>reject it completely and delineate manually, or modify the result<br>and then save it. The software does not provide any auto-detection<br>or auto-saving functionalities. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Same as the predicate devices, the clinician retains the ultimate<br>responsibility for making the pertinent diagnosis and patient<br>management decisions based on their standard practices and<br>visual comparison of the individual images, regardless of the<br>accuracy of the output generated by the software.                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indications for Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | AdvantageSim™ MD is used to prepare geometric and anatomical<br>data relating to a proposed external beam radiotherapy treatment<br>prior to dosimetry planning. Anatomical volumes can be defined<br>automatically or manually in three dimensions using a set of CT<br>images acquired with the patient in the proposed treatment<br>position.                                                                                                                                                                             |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Definition of the anatomical volumes may be assisted by additional<br>CT, MR or PET studies that have been co-registered with the<br>planning CT scan. Additionally, CT & PET data from a respiratory<br>tracked examination may be used to allow the user to define the<br>target or treatment volume over a defined range of the respiratory<br>cycle.                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The geometric parameters of a proposed treatment field are<br>selected to allow non-dosimetric, interactive optimization of field<br>coverage. Defined anatomical structures and geometric treatments<br>fields are displayed on transverse images, on reformatted sagittal,<br>coronal or oblique images, on 3 D views created from the images,<br>or on a beam eye's view display with or without the display of<br>defined structures with or without the display of digitally<br>reconstructed radiograph.               |
| Technology:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The AdvantageSim MD with MR pelvic organ at risk segmentation<br>Option software employs the same fundamental scientific<br>technology as that of its predicate devices.                                                                                                                                                                                                                                                                                                                                                     |
| Determination of<br>Substantial<br>Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Summary of Non-Clinical Tests:<br>The AdvantageSim MD MR pelvic organ at risk segmentation Option<br>software complies with voluntary standards as detailed in Section<br>9, 11 and 16 of this premarket submission. The following quality<br>assurance measures were applied to the development of the<br>system:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Integration testing (System verification)<br>• Performance testing (Bench testing, verification)<br>• Safety testing (Verification) |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Summary of Clinical Tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | A usability study using consented clinical images was conducted<br>by three board certified Radiation Oncologists who were<br>considered experts. The study was meant to assess the following:                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

y

Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and the border are black, contrasting with the white background.

{2}------------------------------------------------

GE Healthcare AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are black, contrasting with the white background.

:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K132944 Page 4 of 4" in a handwritten style. The text is arranged in two lines, with "K132944" on the top line and "Page 4 of 4" on the bottom line. Below the text, there is a logo of GE (General Electric) in a circular shape.

·

GE Healthcare

AdvantageSim™ MD with MR pelvic organ at risk segmentation Option 510(k) Premarket Notification Submission

|             | <ul><li>Accuracy of measurement</li><li>Precision of the measurement</li><li>Efficiency, as a comparison of amount of time</li><li>General user Qualitative feedback</li></ul> The study results demonstrated that when used by qualified  clinician the new software device has the potential to reduce inter-  operator variability, provides statistically significant and practically  meaningful clinical efficiency improvements and substantiates the  characteristics of this feature, among others, as easy to learn,  useful, efficient and providing increased throughput. |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | GE Healthcare considers the AdvantageSim MD with MR pelvic  organ at risk segmentation Option software application to be as  safe, as effective, and performance is substantially equivalent to  the predicate devices.                                                                                                                                                                                                                                                                                                                                                               |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a person with outstretched arms, with three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 14, 2014

GE Hungary Kft. % Mr. Stephen Slavens Regulatory Affairs Director GE Medical Systems, LLC dba GE Healthcare 3000 N Grandview WAUKESHA WI 53188

Re: K132944

Trade/Device Name: AdvantageSim" MD MR pelvic organ at risk segmentation Option Regulation Number: 21 CFR 892.5840 Regulation Name: Radiation therapy simulation system Regulatory Class: II Product Code: KPQ Dated: January 20, 2014 Received: January 22, 2014

Dear Mr. Slavens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Slavens

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known): K132944

## Device Name: AdvantageSim™ MD with MR pelvic organ at risk segmentation Option

Indications For Use:

AdvantageSim™ MD is used to prepare geometric and analomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position of the anatomical volumes may be assisted by additional CT, MR or PET studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user define the target or treatment volume over a defined range of the respiratory cycle.

The geometric parameters of a proposed treatment field are selected to allow non-dosimetric. interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Michael D. O'Hara

Page 1 of __ 1 _______________________________________________________________________________________________________________________________________________________________

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health K132944 510(k)

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K132944](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K132944)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
