ASSISTO Universal Applicator Clamping Device

K201279 · Geomed Medizin-Technik GmbH & Co. KG · JAQ · Jul 10, 2020 · Radiology

Device Facts

Record IDK201279
Device NameASSISTO Universal Applicator Clamping Device
ApplicantGeomed Medizin-Technik GmbH & Co. KG
Product CodeJAQ · Radiology
Decision DateJul 10, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2

Indications for Use

The ASSISTO® Universal Applicator Clamping Device is indicated for use as an accessory for HDR or PDR brachytherapy treatment. Specifically, it is intended to hold dedicated applicator holding devices by Varian Medical Systems in place during cancer treatment of the vaginal stump, rectum, cervix, uterus and endometrium using HDR or PDR brachytherapy.

Device Story

ASSISTO® Universal Applicator Clamping Device is a self-retaining, table-mounted mechanical support system. It consists of a stainless steel tubular, articulated arm connected to a vertical stand, featuring a screw clamp at the distal end. The device is used in clinical settings to hold dedicated Varian Medical Systems applicator holding devices in a fixed position during HDR or PDR brachytherapy for gynecological and rectal cancers. The arm is manually adjustable within its articulating radius. The device is reusable, provided non-sterile, and requires user cleaning and sterilization before use. By securely positioning the applicator, the device assists healthcare providers in maintaining precise treatment geometry, potentially improving the accuracy of radiation delivery to the target tissue.

Clinical Evidence

Bench testing only. Verification testing confirmed the device's capacity to securely hold Varian Medical Systems applicator holding devices. Validation testing confirmed the adequacy of recommended end-user reprocessing procedures. All acceptance criteria were met.

Technological Characteristics

Materials: Stainless steel and thermoplastic polyamide. Design: Tubular, articulated arm with screw-clamp, table-mounted via base plate and clamp socket. Dimensions: 310 mm working length. Energy source: None (mechanical). Connectivity: None. Sterilization: Reusable, non-sterile (user-reprocessed).

Indications for Use

Indicated for use as an accessory for HDR or PDR brachytherapy treatment to hold Varian Medical Systems applicator holding devices in place during cancer treatment of the vaginal stump, rectum, cervix, uterus, and endometrium.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Geomed Medizin-Technik GmbH & Co. KG % Ms. Angelika Scherp Regulatory Affairs Consultant Business Support International Aalsmeerweg 123-3 Amsterdam, Noord-Holland 1059AH THE NETHERLANDS July 10, 2020 ## Re: K201279 Trade/Device Name: ASSISTO® Universal Applicator Clamping Device Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ, GWG, and GCJ Dated: May 12, 2020 Received: May 13, 2020 Dear Ms. Scherp: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201279 Device Name ASSISTO® Universal Applicator Clamping Device Indications for Use (Describe) The ASSISTO® Universal Applicator Clamping Device is indicated for use as an accessory for HDR or PDR brachytherapy treatment. Specifically, it is intended to hold dedicated applicator holding devices by Varian Medical Systems in place during cancer treatment of the vaginal stump, rectum, cervix, uterus and endometrium using HDR or PDR brachytherapy. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------| | <span> <b> \[X] Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> | | <span> \[] Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary of Safety and Effectiveness | Date: | June 15, 2020 | | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Name:<br>Address: | Geomed Medizin-Technik GmbH & Co. KG<br>Ludwigstaler Strasse 27<br>D-78532 Tuttlingen<br>Germany | | | Contact Person:<br>Telephone: | Hanno Haug<br>+49.7461.93550 | | Product: | Name of Device:<br>Common Name:<br>Classification Name:<br>Regulatory Class:<br>21 CFR Section: | ASSISTO® Universal Applicator Clamping Device<br>Holding Accessory for Applicator System<br>Remote Controlled Radionuclide Applicator System<br>II<br>892.5700 | | Predicate Device: | K170203 – Kobold® Prostate HDR Template(s) and Kobold® HDR Stepper Template(s) | | | Reference Device: | K052745 - ASSISTO® Arm Systems | | | Device Description: | The ASSISTO® Universal Applicator Clamping Device is a self-retaining device that consists of a stainless steel tubular, articulated arm connected to a vertical stand. The arm is freely adjustable within the articulating radius. A screw clamp mounted on the distal end of the device serves to hold dedicated applicator holding devices by Varian Medical Systems in place during cancer treatment of the vagina, vaginal stump, rectum, cervix, uterus and endometrium using HDR or PDR brachytherapy. Two accessories, a stainless steel clamp socket and a base plate manufactured of thermoplastic and stainless steel, are provided to table-mount the device during clinical use.<br><br>The ASSISTO® Universal Applicator Clamping Device and the clamp socket are reusable and provided non-sterile for cleaning and sterilization by the user before use. The non-sterile, reusable base plate is intended for cleaning and disinfection by the user before use. | | | Indications for Use: | The ASSISTO® Universal Applicator Clamping Device is indicated for use as an accessory for HDR or PDR brachytherapy treatment. | | | | Subject Device | Predicate Device | Reference Device | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | ASSISTO® Universal<br>Applicator Clamping<br>Device | Kobold® Prostate<br>HDR Template(s) and<br>Kobold® HDR<br>Stepper Template(s) | ASSISTO® Arm Systems<br>Model No. CA 235-30-01 | | 510(k)<br>Number | K201279 | K170203 | K052745 | | Manufacturer | Geomed Medizin-<br>Technik | Kobold, LLC | Geomed Medizin-Technik<br>GmbH & Co. | | Indications<br>for Use | The ASSISTO®<br>Universal Applicator<br>Clamping Device is<br>indicated for use as an<br>accessory for HDR or<br>PDR brachytherapy<br>treatment.<br>Specifically, it is<br>intended to hold<br>dedicated applicator<br>holding devices by<br>Varian Medical Systems<br>in place during cancer<br>treatment of the vagina,<br>vaginal stump, rectum,<br>cervix, uterus and<br>endometrium using HDR<br>or PDR brachytherapy. | The Kobold®<br>Prostate HDR<br>Template(s) and<br>Kobold® HDR<br>Stepper Template(s)<br>are indicated for use<br>as an accessory for<br>high dose rate<br>brachytherapy<br>treatment of the<br>prostate. | ASSISTO® Arm Systems<br>consist of a table-mounted<br>endoscope holder system<br>intended for use by surgeons<br>to hold endoscopes and<br>arthroscopes with a diameter<br>of 2.7mm to 10mm during<br>general diagnostic and<br>therapeutic procedures. The<br>device is also intended for<br>use by qualified surgeons for<br>holding endoscopes during<br>diagnostic and therapeutic<br>neurologic procedures. | | Design | Tubular, articulated arm<br>connected to a vertical,<br>table-mounted stand and<br>provided with a screw-<br>clamp holding<br>component. The position<br>of the articulated arm is<br>freely adjustable within<br>the articulating radius. | Template provided<br>with holes in a grid<br>array and a locking<br>mechanism to lock<br>needles in place. The<br>template is equipped<br>with four suture holes<br>to secure it to the<br>perineum during use. | Tubular, articulated arm<br>connected to a vertical,<br>table-mounted stand and<br>provided with a selection of<br>screw clamp holding<br>components. The position of<br>the articulated arms is freely<br>adjustable within the<br>articulating radius. | | Working<br>Length | 310 mm | N/A | 310 mm | | Materials | Stainless steel and<br>thermoplast polyamid | N/A | Stainless steel and<br>thermoplast polyamid | | Patient<br>Contact | No patient contact | Intact skin | No patient contact | | Sterility | Non-sterile | Non-sterile | Non-sterile | | Reusable | Yes | No | Yes | ### Comparison Table: {4}------------------------------------------------ Performance Design verification and validation testing to support determination of substantial Testing: equivalence consisted of the following tests: - Verification of the device's capacity to securely hold the Varian ● Medical Systems applicator holding devices. - Validation testing of the recommended end user devicereprocessing ● procedures Acceptance criteria were met for all tests performed. The information provided in this 510(k) submission, including results of Conclusion: nonclinical testing, indicates that the ASSISTO® Universal Applicator Clamping Device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...