← Product Code [JAQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAQ) · K183332

# BRAXX Esophageal Brachytherapy Applicator (K183332)

_Braxx Biotech Co., Ltd. · JAQ · Apr 16, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAQ/K183332

## Device Facts

- **Applicant:** Braxx Biotech Co., Ltd.
- **Product Code:** [JAQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAQ.md)
- **Decision Date:** Apr 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5700
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Esophageal Brachytherapy Applicator is intended for use with commercially available afterloader during brachytherapy. The purpose of the device is for delivery of radioactive source to the esophagus. This device is sterile, disposable and single-use.

## Device Story

Device is a sterile, single-use esophageal brachytherapy applicator; consists of a main catheter with inflatable balloons. Used by radiation oncologists in hospital OR settings. Operates by inserting the catheter into the esophagus; balloons are inflated to center the treatment catheter within the lumen. Facilitates precise delivery of radioactive sources from a commercially available afterloader to the target site. Benefits patient by ensuring accurate radiation dose delivery to esophageal tumors while minimizing exposure to surrounding healthy tissue.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10, -11, USP 151) and sterilization validation (ISO 11135:2014) to SAL 10^-6. Mechanical testing confirmed design specifications and structural integrity.

## Technological Characteristics

Main catheter with inflatable balloons for centering. Sterilization via Ethylene Oxide (ISO 11135:2014). Biocompatibility per ISO 10993. Single-use, disposable. No electronic or software components.

## Regulatory Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

## Predicate Devices

- E-App ([K142719](/device/K142719.md))
- Intraluminal Balloon Applicator for Brachytherapy ([K162499](/device/K162499.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Section 4 Indications for Use Statement

FORM FDA 3881

{1}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

### K183332

Device Name Esophageal Brachytherapy Applicator

### Indications for Use (Describe)

The Esophageal Brachytherapy Applicator is intended for use with commercially available afterloader during brachytherapy. The purpose of the device is for delivery of radioactive source to the esophagus. This device is sterile, disposable and single-use.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{2}------------------------------------------------

# Section 5 510(k) Summary

### 510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

The assigned 510(k) Number: K183332

- Submitter 1.

| Mailing Address | BRAXX Biotech Co. Ltd<br>6F., No.10, Songde Rd., Xinyi Dist<br>Taipei City, 11076, Taiwan R.O.C<br>Establishment Registration No.: NA (1st submission) |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person  | Judy Kao                                                                                                                                               |
| Phone:          | 886-921988206                                                                                                                                          |
| E-mail:         | judy.kao@braxx.com.tw                                                                                                                                  |
| Date Prepared   | 2018/11/27                                                                                                                                             |

#### 2 Device Name

Proprietary Name Common name Product Code Device CFR Classification Device Class Classification Panel

"BRAXX" Esophageal Brachytherapy Applicator Esophageal Brachytherapy Applicator JAQ System, Applicator, Radionuclide, Remote-Controlled 21 CFR 892.5700 II Remote controlled radionuclide applicator system

#### 3 Predicate Device Name

510(k) number: Trade or proprietary name: Manufacturer:

510(k) number: Trade or proprietary name: Manufacturer:

#### 5. Intended Use:

K142719 E-App (Esophageal Applicator) BIONIX RADIATION THERAPY, LLC

## K162499

Intraluminal Balloon Applicator for Brachytherapy Best Medical International, Inc

The Esophageal Brachytherapy Applicator is intended for use with commercially available afterloader during brachytherapy. The purpose of the device is for delivery

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BRAXX. The logo consists of the word "BRAXX" in a bold, sans-serif font, with a stylized "B" that incorporates a small orange circle. Below the word "BRAXX" are three Chinese characters. 

of radioactive source to the esophagus. This device is sterile, disposable and single-use. For prescription use only

Special Conditions for Use Statement(s):

Characteristics and

Substantial Equivalence

Comparison with Predicate:

Technological

A comparison of the device features, intended use, and other information demonstrates that the Esophageal Brachytherapy Applicator is substantially equivalent to the predicate device as summarized in Table 1.

| Device name                  | Subject device:<br>Esophageal<br>Brachytherapy<br>Applicator                                                                                                                                                                                                                                 | Predicate device:<br>E-App                                                                                                                                                                                                    | Predicate device:<br>Best® Intraluminal<br>Balloon Applicator for<br>Brachytherapy                                                                                                                                                                                                                                                                                                    |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                 | The Esophageal<br>Brachytherapy Applicator<br>is intended for use with<br>commercially available<br>afterloader during<br>brachytherapy. The<br>purpose of the device is for<br>delivery of radioactive<br>source to the esophagus.<br>This device is sterile,<br>disposable and single-use. | The GC Medtech<br>Esophageal<br>applicator(E-App) is an<br>applicator used to facilitate<br>delivery of a prescription of<br>radiation to the esophagus<br>when used in conjunction<br>with a high dose rate<br>after-loader. | Best® Intraluminal<br>Balloon Applicator for<br>Brachytherapy is intended<br>to provide brachytherapy<br>when the physician<br>chooses to deliver radiation<br>to an existing body site<br>such as the esophagus or<br>bronchus. It is supplied as a<br>single-use sterile device.<br>The device is indicated for<br>use under the direct<br>supervision of a qualified<br>physician. |
| Target patient<br>Population | Esophageal cancer patients                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Target User<br>Population    | Radiation oncologists                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Anatomical<br>Site           | Esophagus                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Where Used                   | Hospital O.R. room                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Design                       | A main catheter with<br>inflatable balloons                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Performance                  | For centering the treatment<br>catheter (radioactive<br>source) within the lumen<br>by inflated balloons.                                                                                                                                                                                    | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Biocompatible                | Fully biocompatible                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterilization<br>Method      | Ethylene Oxide                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |
| Single Use                   | Single use                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                  |

Table 1

6.

{4}------------------------------------------------

# 7. Performance Testing

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device.

# Biocompatibility testing

The biocompatibility evaluation and testing of the Esophageal Brachytherapy Applicator was conducted in accordance with the following standards and guidance, as recognized by the FDA:

- FDA Draft Guidance Use of International Standard ISO- 10993, "Biological Evaluation O of Medical Devices, Part 1: Evaluation and Testing", dated 04-23-2013
- 0 ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- 10993-10, Biological evaluation of medical devices- Part 10: Tests for irritation and skin 0 sensitization
- 0 ISO 10993-11, Biological evaluation of medical devices- Part 11: Tests for systemic toxicity.
- O USP 151, Pyrogen Test in White Rabbits

# Sterilization Validation

Sterilization validation was conducted in accordance with ISO 11135:2014 "Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices". Sterilization has been validated to achieve a Sterility Assurance Level (SAL) of 10-6.

## Mechanical testing

Esophageal Brachytherapy Applicator's mechanical function and structure integrity were tested and demonstrated that the design specification from design input are fulfilled and that the device raises no safety concern.

## 8. Conclusion

Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Esophageal Brachytherapy Applicator is substantially equivalent to the predicate device and raises no new questions of safety or effectiveness.

{5}------------------------------------------------

BRAXX Biotech Co., Ltd. % Ms. Judy Kao Regulatory Affairs Representative 6F., No.10, Songde Rd., Xinyi Dist Taipei City, 11076 Taiwan REPUBLIC OF CHINA

April 16, 2019

### Re: K183332

Trade/Device Name: "BRAXX" Esophageal Brachytherapy Applicator Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ Dated: February 27, 2019 Received: March 8, 2019

Dear Ms. Kao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{6}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAQ/K183332](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAQ/K183332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
