KOBOLD FLETCHER TYPE APPLICATOR SET, KOBOLD HENSCHKE TYPE APPLICATOR SET

K123912 · Spencer Fillmore · JAQ · May 14, 2013 · Radiology

Device Facts

Record IDK123912
Device NameKOBOLD FLETCHER TYPE APPLICATOR SET, KOBOLD HENSCHKE TYPE APPLICATOR SET
ApplicantSpencer Fillmore
Product CodeJAQ · Radiology
Decision DateMay 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines. The CT/MRI Compatible HDR Henschke Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines.

Device Story

Kobold Fletcher and Henschke Type GYN T&O Applicator Sets are medical devices used for high dose rate (HDR) brachytherapy. The sets consist of intrauterine tandems and interlocking colpostats (Fletcher or Henschke style) with adjustable ovoids. These devices are used in clinical settings by physicians to deliver radiation boosts to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The applicators are CT/MRI compatible, allowing for visualization and localization within the treatment volume. They feature a click-fit connector interface compatible with GammaMed and VariSource afterloaders. By providing a stable delivery system for radioactive sources, the applicators enable precise dosimetry, potentially improving treatment outcomes for patients with cervical or uterine conditions.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing and biocompatibility documentation were provided to support substantial equivalence.

Technological Characteristics

Materials: Titanium, Ultem®, Stainless Steel, Silicone, High Temperature Vinyl. Design: Intrauterine tandems (straight, 15°, 30°, 45° curvature) and interlocking colpostats with adjustable ovoids (1cm-3cm). Compatibility: CT/MRI, GammaMed/VariSource afterloaders. Sterilization: Steam (134°C, 44psi, 18 min).

Indications for Use

Indicated for patients requiring high dose rate (HDR) brachytherapy for the cervix and uterus. Contraindications: As per clinical guidelines and standard clinical practice.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123912 Page 1 of 5 б Kobold Applicator Kits # 510(k) Summary MAY 1 4 2013 This summary of 510(k) safety and effectiveness information is furnished in accordance with requirements detailed in 21 CFR 807.92. 1 . The assigned 510(k) number is K-TBD Submitter's Identification: Spencer Fillmore 2410 S Sumner Lane Greenacres, WA, 99016 Correspondence: Christina Bernstein BB Medical Surgical, Inc. 2670 Leavenworth Street San Francisco, CA 94133 Tel: 415-450-0515 Email tina@bbmedicalsurgical.com Date of submission 07 December 2012 2. Device name: 1. Fletcher type Applicator Set Proprietary name: Kobold Fletcher Type GYN T&O Applicator Set™ 2. Henschke type Applicator Set Proprietary name: Henschke Tvpe GYN T&O Applicator Set™ - Regulation Section 892.5700 A. - B. Classification: Class II - C. Product Code: JAQ - D. Panel: Radiology 3. #### Intended Use: 1. The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines. {1}------------------------------------------------ 2. The CT/MRI Compatible HDR Henschke Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines. ### 4. ## Device Description: 1. The CT/MRI Compatible HDR Fletcher Type Applicator Set™ design is based on the conventional modified Fletcher GYN applicator with a click fit connector interface. This applicator is particularly useful in providing an HDR radiation boost to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The set consists of precision manufactured intrauterine tandems and a pair of interlocking colpostats with Fletcher type unshielded ovoids ranging in size from 1cm to 3cm. All components are designed to allow for customization of the treated geometry. The distance between the ovoids is adjustable with a lock knob. Four different tandems are provided: one straight, one with a slight curvature of 15 degrees, one with medium curvature of 30 degrees, and one with a maximal curvature of 45 degrees. 2. The CT/MRI Compatible HDR Henschke Type Applicator Set™ design is based on the conventional modified Henschke GYN applicator with a click fit connector interface. This applicator is particularly useful in providing an HDR radiation boost to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The set consists of precision manufactured intrauterine tandems and a pair of interlocking colpostats with Henschke type unshielded ovoids ranging in size from 2cm to 3cm. All components are designed to allow for customization of the treated geometry. The distance between the ovoids is adjustable with a lock knob. Four different tandems are provided: one straight, one with a slight curvature of 15 degrees, one with medium curvature of 30 degrees, and one with a maximal curvature of 45 degrees. #### 5. ### Substantial Equivalence Information: Information presented supports substantial equivalence of the Kobold Fletcher and Henschke Type GYN T&O Applicator Set™ to the predicate device. Each of the two proposed devices has the same indications for use, has similar nature of body contact, is similar in shape and design, has the same fundamental technology and uses the same sterilization procedures. Please see predicate device brochure in Appendix C. 7 - Predicate device names: Mick® Ring Type GYN Applicator Set A. - Predicate device K numbers: 011657 B. - C. Comparison with predicate: {2}------------------------------------------------ . : | SUBSTANTIAL<br>EQUIVALENCE<br>TABLE | KOBOLD LLC | KOBOLD LLC | MICK RADIO-NUCLEAR<br>INSTRUMENTS, INC.® | |-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | K-Number | TBD | TBD | K-011657 | | Device Description | Applicator Set.<br>Fletcher Type | Applicator Set.<br>Henschke Type | Applicator Set, Ring Type | | Indications for Use | The CT/MRI<br>Compatible HDR<br>Fletcher Type<br>Applicator is<br>indicated for use in<br>any case where high<br>dose rate (HDR)<br>radiation treatment<br>of the cervix and<br>uterus is accepted by<br>up-to-date clinical<br>guidelines. | The CT/MRI<br>Compatible HDR<br>Henschke Type<br>Applicator is<br>indicated for use in<br>any case where high<br>dose rate (HDR)<br>radiation treatment<br>of the cervix and<br>uterus is accepted by<br>up-to-date clinical<br>guidelines. | The Mick Radio-Nuclear<br>Instruments, Inc. HDR<br>Tandem/Ring Applicator with<br>Rectal Retractor is intended for<br>use in Brachytherapy. The<br>delivery of intra-cavitary<br>radiation therapy requires not<br>only proper visualization and<br>localization of the applicator<br>within the treatment volume,<br>but precise dosimetry and a<br>stable delivery system from<br>which treatment can be<br>administered. The Mick Radio-<br>Nuclear HDR Tandem/Ring<br>Applicator with Rectal<br>Retractor meets these<br>requirements. | | Afterloader<br>Compatibility | GammaMed®<br>VariSource® with<br>click-fit connector. | GammaMed®.<br>VariSource® with<br>click-fit connector. | GammaMed®.<br>VariSource® with<br>click-fit connector. | | Design Geometry | Flexible | Flexible | Flexible | | Materials | Titanium. Ultem®<br>Stainless Steel,<br>Silicone, High<br>Temperature Vinyl | Titanium. Ultem®.<br>Stainless Steel,<br>Silicone, High<br>Temperature Vinyl | Titanium, Stainless Steel,<br>Polysulphone, Silicone | | Packing | Tray | Tray | Cassette | | Sterility | Provided Non-Sterile | Provided Non-Sterile | Provided Non-Sterile | | Sterilization<br>Methods,<br>Sterilization<br>Conditions. | Steam 18 min at 273°F<br>(134°C), 44psi (3 bar)<br>Dry 15 min; no<br>Sterrad® compatibility<br>documented | Steam 18 min at 273°F<br>(134°C), 44psi (3 bar)<br>Dry 15 min; no<br>Sterrad® compatibility<br>documented | Steam 20 min at 132°C<br>Dry 15 min; no Sterrad®<br>compatibility documented | | Biocompatibility | Documented | Documented | Not Documented | | Anatomical Sites | Uterus, Cervix | Uterus, Cervix | Cervix | | Environmental<br>Compatibility | CT/MRI Compatible | CT/MRI Compatible | CT/MRI Compatible | ・ : 8 · {3}------------------------------------------------ Kobold Applicator Kits 6. # Test Principle, Performance Characteristics: FDA has not established special controls or performance standards for this device. # 7. . # Bench Top Testing Extensive testing in accordance with known standards is documented by the manufacturer. The standard is a specification for the materials used in surgical implant applications, predicate brachytherapy devices and manufacturer's acceptance procedures are detailed in Appendix A. Mechanical testing of the finished devices is additionally described in Appendix A. ### 8. ### Conclusions Both versions of Kobold`s Applicator Set are similar in intended use and technological characteristics to predicate devices reviewed. Each of the two versions of the device is similar with respect to indications for use and physical characteristics to predicate devices in terms of section 510(k) substantial equivalency. Contraindications: As per clinical guidelines and standard clinical practice. Warnings and Precautions: The precautions are provided in the device labeling for the Kobold Fletcher Style GYN T&O Applicator Set™ and Kobold Henschke Style GYN T&O Applicator Set™. There is no warning associated with this type of device. ### ੇ. ### Summary | Description | Comparison with Predicate Device | |------------------|----------------------------------| | Biocompatibility | Safe as Predicate Device | 9 {4}------------------------------------------------ K123912 Page 5 of 5 1 0 Kobold Applicator Kits : . | Performance Characteristics | Substantially equivalent | |-----------------------------|--------------------------| | Intended Use | Substantially equivalent | | Performance Tests | Not Required | Each version of the device 1. Kobold Fletcher Type GYN T&O Applicator Set™ and 2. Kobold Henschke Style GYN T&O Applicator Set™, based on the information submitted in this 510(k) application has been demonstrated to be substantially equivalent to the predicate devices. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 14, 2013 Kobold, LLC % Ms. Christina Bernstein Regulatory Director 2670 Leavenworth Street SAN FRANCISCO CA 94133 Re: K123912 Trade/Device Name: Fletcher Type GYN T&O Applicator Set™ Regulation Number: 21 CFR 892.5700 Regulation Name: Remote Controlled Radionuclide Applicator System Regulatory Class: Class II Product Code: JAQ Dated: April 19, 2013 Received: May 2, 2013 Dear Ms. Bernstein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {6}------------------------------------------------ Page 2 - Ms. Christina Bernstein comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of nedical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesTorYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Michael D. DiHara for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): K123912 Device Name: Fletcher Type GYN T&O Applicator Set™ Indications for Use: The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines. Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Michal D. O-Ling (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K123912 510(k) Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...