The Flexitron Remotely Controlled Brachytherapy Afterloading System is intended to enable an operator to apply by remote control a radionuclide source (Flexisource) into the body or to the surface of the body for radiation therapy. Flexitron is intended to be used in combination with applicators which makes it suitable for intracavitary, interstitial, intralumenary, bronchial, endovascular, intra-operative and surface brachytherapy treatment. Flexitron is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and authorized medical professional.
Device Story
Flexitron is a remotely controlled brachytherapy afterloading system; used in clinical settings for radiation therapy. System components: Treatment Delivery Unit (TDU) housing radioactive source (Ir-192) in shielded safe; Treatment Communications Console (TCC) for data entry/storage; Treatment Control Panel (TCP) touchscreen for operation. Physician programs dwell positions (1mm intervals) and times (0.1s increments) to create patient-specific isodose shapes. TDU uses microprocessor-controlled drive mechanism to move source through transit tubes into applicators; check cable verifies channel patency before source delivery. System logs performance and treatment data. Benefits include precise radiation delivery to target sites while sparing healthy tissue; remote operation protects clinicians from radiation exposure.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
System includes TDU with shielded safe, microprocessor-controlled drive/channel selector mechanisms, TCC, and TCP. Radioactive source (Ir-192) in stainless steel capsule on stranded cable. Supports up to 40 treatment channels; 1mm dwell position resolution; 0.1s dwell time increments. Connectivity via serial data links. Powered by 28V DC via high-isolation transformer. Includes X-Ray markers and source position check ruler for QA.
Indications for Use
Indicated for patients requiring radiation therapy via intracavitary, interstitial, intralumenary, bronchial, endovascular, intra-operative, or surface brachytherapy. Use restricted to prescription by trained, authorized medical professionals.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
GammaMed Plus 3/24 High Dose Rate (HDR) Remote Afterloader (K031524)
Submission Summary (Full Text)
{0}------------------------------------------------
K070574
# 5. 510(k) Summary as required by 21 CFR 807.92
### 5.1. Submitter of 510(k)
| 510(k) owner's name | : | Isodose Control BV | |
|-------------------------------|---|--------------------------------------------------------|-------------|
| address | : | Landjuweel 11<br>3905 PE Veenendaal<br>The Netherlands | APR 13 2007 |
| phone | : | +31 318 502950 | |
| fax | : | +31 318 502528 | |
| name of contactperson | : | Hub van de Bergh | |
| e-mail | : | hvandebergh@isodosecontrol.com | |
| date the summary was prepared | : | February 20, 2006 | |
#### 5.2. Device
#### 5.2.1. Device: Flexitron
name of the device / trade or proprietary name: Flexitron common or usual name: Brachytherapy Afterloader Remote controlled radionuclide applicator system classification name: (per 21 CFR section 892.5700 Product code JAQ)
#### 5.2.2. Accessory: Flexisource
| | name of the device / trade or proprietary name: Flexisource |
|----------------------|-------------------------------------------------------------------------|
| | common or usual name: Radioactive Source for Brachytherapy Afterloading |
| classification name: | Radionuclide brachytherapy source |
| | (per 21 CFR section 892.5730 Product code JAQ) |
#### 5.2.3. Accessory: Cervix Rotterdam Applicator
name of the device / trade or proprietary name: Cervix Rotterdam Applicator common or usual name: Brachytherapy Applicator classification name: Remote controlled radionuclide applicator system (per 21 CFR section 892.5700 Product code JAQ)
#### 5.2.4. Accessory: Vaginal Cylinder Applicator
name of the device / trade or proprietary name: Vaginal Cylinder Applicator common or usual name: Brachytherapy Applicator classification name: Remote controlled radionuclide applicator system (per 21 CFR section 892.5700 Product code JAQ)
## 5.3. Legally Marketed Device(s)
The Flexitron device can be substantial equivalent to the legally marketed devices cited in the table below.
| Device | Manufacturer | 510(k) # |
|---------------------------------------------------------------|----------------------------|----------|
| microSelectron-HDR Version 2 | Nucletron bv | K953946 |
| GammaMed Plus 3/24 High Dose Rate (HDR)<br>Remote Afterloader | Varian Medical Systems Inc | K031524 |
lsodose Control B.V. ඹ PO Box 615, 3900 AP Veenendaal # Landjuweel 11, 3905 PE Veenendaall # The Netherlands 2 +31 318 502950 & B +31 318 502528 #@ info@isodosecontrol.com € KvK: 30191600 ww.isodosecontrol.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a close-up of text that appears to be inverted and slightly blurry. The text is in a bold, sans-serif font, and the letters are closely spaced together. There is a decorative element between the two words, possibly a logo or emblem. The overall impression is that the text is part of a title or heading, but the inversion and blurriness make it difficult to read.
## 5.4. Description of the Device
#### 5.4.1. Flexitron
Flexitron is a computer controlled medical device for brachytherapy treatment.
The Flexitron Remotely Controlled Brachytherapy Afterloading System comprises subjoined listed subsystems:
- . TDU - Treatment Delivery Unit It houses the radioactive source in its built-in shielding safe, the microcomputer controlled driving mechanism and channel selector mechanism.
- . TCC - Treatment Communications Console For treatment program data entry, data storage and data displaying.
- TCP Treatment Control Panel . It comprises the basic operator elements for start and stop treatment and for displaying device status and used and remaining treatment time.
The TDU houses the shielding safe for storage of the radioactive source and the source drive mechanism for moving the radioactive source into treatment positions within an applicator and back into the safe. There is a channel selector mechanism for selecting one out of 40 treatment channels.
The TDU source drive mechanism and the channel selector mechanism are controlled by microprocessor based electronics. Sensors are applied and their signals are input for the embedded microprocessor for safe and reliable operation. The TDU is connected to mains via a built-in high isolation transformer and an AC/DC power supply converting the mains voltage level to 28 volt for internal machine operation.
The TDU is designed for positioning the source at 401 separate dwell positions in each of the 40 treatment channels and together with dwell time settings for each individual dwell position it enables the creation of isodose shapes targeted to the patients anatomy and geometry in order to irradiate the target patient body site and spare the healthy surrounding tissue.
Dwell positions can be programmed at 1 mm intervals. The maximum source outdrive distance is 1400 mm from the channel selector front plate.
Dwell times are programmable with increments of 0.1 second. Maximum dwell time is 999.9 seconds per position. Minimum dwell time is 0.1 second per position. A dwell position is skipped if its dwell time is 0 seconds.
Before the radioactive source is moved into a treatment channel a check cable is first moved into it. A check cable is of a similar construction of a source cable but without radioactive isotope. It is a measure to check the treatment channel for free passage of the source immediately after the check.
The TDU is connected to TCC via a serial data communications link to receive the patient treatment program and to return data of the performed patient treatment.
TCC allows for manually entering a patient treatment program or for electronically transferring a patient treatment program from Brachytherapy Treatment Planning Software (Treatment planning software is not subject of this 510(k) PreMarket Notification).
During patient treatment TCC displays treatment progress and logs patient treatment data and machine performance data in separate log files.
TDU is connected via another serial communications port to TCP.
TCP is a touchscreen device which is directly controlled from the TDU embedded microprocessor. TCP allows for starting the patient treatment when all conditions are fulfilled.
| Carlo Company Con- | पु<br>BARREAL |
|--------------------|---------------|
| | |
Section 5: 510(k) Summary - page 2 of 4 900-5008-20070220w_Flexitron510(k)AbbreviatedPmr
lsodose Control B.V. « PO Box 615, 3900 AP Veenendaal » Landjuweel 11, 3905 PE Veenendaall » The Netherlands 율 +31 318 502950 % 됨 +31 318 502528 *@ info@isodosecontrol.com « KvK: 30191600 www.lsodosecontrol.com
{2}------------------------------------------------
# I コリリリ
#### 5.4.2. Accessory: Flexisource
The Flexisource is the radioactive source for brachytherapy with Flexitron.
A stainless steel capsule accomodates the radioactive isotope Ir-192. This capsule is welded to the outer end of a steel stranded cable. The other end of the stranded cable is connected to the Flexitron built-in cable drive mechanism.
#### 5.4.3. Accessory: Transit tubes and applicators
Transit tubes are guiding the radioactive source to and from the applicator. Transit tubes are connected at one end to the Flexitron device at the channel selector mechanism and at the other end to the body site specific applicator. Applicators are positioned by the physician at the target patient body site.
#### 5.4.4. Other accessories
Other accessories to achieve the utmost performance of the Flexitron device are available to support Brachytherapy Treatment Planning and to support regular hospital QA procedures. Accessories in these categories included with this 510(k) submission are X-Ray Markers and Source Position Check Ruler.
- · X-Ray Markers are intended to visualize dwell positions in treatment channels and to distinguish the treatment channels on fluoroscopic images to be used for brachytherapy treatment planning.
- · Source Position Check Ruler is intended to be used for regular check of the source positioning accuracy as part of the hospital regular QA checks.
#### 5.5. Intended use of the Device
The Flexitron Remotely Controlled Brachytherapy Afterloading System is intended to enable an operator to apply by remote control a radionuclide source (Flexisource) into the body or to the surface of the body for radiation therapy.
Flexitron is intended to be used in combination with applicators which makes it suitable for intracavitary, interstitial, intralumenary, bronchial, endovascular, intra-operative and surface brachytherapy treatment.
Flexitron is intended to be used for medical procedures on patients to be prescribed and performerd by a suitably trained and authorized medical professional.
Section 5: 510(k) Summary - page 3 of 4 900-5008-20070220w_Flexitron510(k)AbbreviatedPmnSub
Isodose Control B.V. & PO Box 615, 3900 AP Veenendaal & Landjuweel 11, 3905 PE Veenendaal & The Netherlands = +31 318 502950 * & +31 318 502528 = @ info@isodosecontrol.com * KvK: 30191600 www.lsodosecontrol.com
{3}------------------------------------------------
ISSOUEE RESOLUTION VI
#### 5.6. Technological characteristics of device compared to the predicate device
The Flexitron Remotely Controlled Brachytherapy Afterloading System has similar technological characteristics compared to the legally marketed predicate devices listed above.
All these devices have a subsystem comprising a shielding safe for storage of the radioactive source when not in use, drive mechanism and channel selection mechanism and controlling electronics which is connected to a personal computer for the entry of treatment program data.
All these devices are using a radioactive source sealed in a stainless steel capsule connected to a stranded cable.
The Flexitron device has built-in provisions for operation with two radioactive sources and one check cable. The channel selector mechanism enables the use of check cable and two radioactive sources simultaneously. The device can be configured with two drive mechanisms, one for driving a check cable and one for driving a source cable or it can be configured with three drive mechanisms, one for driving a check cable and two for driving a source cable. The Flexitron device contains an appropriate dimensioned shielding safe and appropriate shielding capability for housing two sources.
This 510(k) PreMarket Notification concerns the configuration with two drive mechanisms, one for driving the check cable and one for driving the source cable.
A variety of applicators dedicated to specific body sites are available with these devices. In all cases is the therapeutic effect based on the radiation property of the radioactive source.
#### 5.7. Substantial Equivalence
From the discussion in previous paragraphs it can be concluded that the Flexitron Remotely Controlled Brachytherapy Afterloading System has similar technological characteristics compared to the legally marketed predicate devices listed in paragraph 5.2.4.
The differences between Flexitron and predicate devices do not concern the basic principle of operation nor does it adversely effect the safety or effectiveness of the device.
The intended use of Flexitron and predicate devices are the same.
The conclusion is that Flexitron is substantial equivalent to the legally marketed predicate devices.
Section 5: 510(k) Summary - page 4 of 4 900-5008-20070220w_Flexitron510(k)Abb
Isodose Control B.V. & PO Box 615, 3900 AP Veenendaal & Landiuweel 11, 3905 PE Veenendaal @ The Netherlands ¥ +31 318 502950 ☎ 콤 +31 318 502528 좋@ info@isodosecontrol.com ▼ KvK: 30191600 www.isodosecontrol.com
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the address of the Food and Drug Administration. The address is 9200 Corporate Blvd., Rockville MD 20850. The text is in a simple, sans-serif font and is easy to read.
APR 1 3 2007
Mr. Hub van de Bergh Quality Assurance & Regulatory Affairs Officer Isodose Control B.V. Landjuweel 11 3905 PE Veenendaal THE NETHERLANDS
Re: K070574
Trade/Device Name: Flexitron Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radio-nuclide applicator system Regulatory Class: II Product Code: JAQ Dated: February 23, 2007 Received: February 28, 2007
### Dear Mr. van de Bergh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/8 description: The image is a black and white circular seal. The seal contains the text "1826-1926" at the top, followed by the letters "PA" in a bold, stylized font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are arranged in a horizontal line at the bottom of the seal. The seal appears to be a commemorative emblem for a centennial celebration.
tions and Promoting Publio Hoalth
{5}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Isodose Control. The logo consists of the word "isodose" in a bold, sans-serif font, followed by a circular symbol with concentric rings. The word "control" is written in the same font as "isodose" and is placed to the right of the circular symbol. A small registered trademark symbol is located in the upper right corner of the logo.
# Indications for Use
510(k) Number (if known): pending K070574
Device Name: Flexitron
Indications for Use:
The Flexitron Remotely Controlled Brachytherapy Afterloading System is intended to enable an operator to apply by remote control a radionuclide source (Flexisource) into the body or to the surface of the body for radiation therapy.
Flexitron is intended to be used in combination with applicators which makes it suitable for intracavitary, interstitial, intralumenary, bronchial, endovascular, intra-operative and surface brachytherapy treatment.
Flexitron is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and authorized medical professional.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lyman
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Company Confidential
Section 4: Indications for Use Statement - page 2 of 2 900-5008-20070220w_Flexitron510(k)AbbreviatedPmnSub
Isodose Control B.V. & PO Box 615, 3900 AP Veenendaal & Landjuweel 11, 3905 PE Veenendaal & The Netherlands 2 +31 318 502950 ¥ 黑 +31 318 502528 €@ info@isodosecontrol.com ¥ KvK: 30191600 www.isodosecontrol.com