K041448 · Medical Intelligence Medizintechnik GmbH · JAI · Jun 16, 2004 · Radiology
Device Facts
Record ID
K041448
Device Name
HEXAPOD RT COUCHTOP
Applicant
Medical Intelligence Medizintechnik GmbH
Product Code
JAI · Radiology
Decision Date
Jun 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5770
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The intended use of the device is to support and aid in positioning a patient during radiation therapy.
Device Story
HexaPOD™ RT CouchTop is a powered radiation therapy patient support assembly. It consists of two platforms connected by six powered, length-adjustable linear elements. By coordinatively adjusting these elements, the system enables movement of the upper platform relative to the lower one in all three spatial dimensions, including rotation around three axes, providing six degrees of freedom (6DOF) for accurate patient positioning. The device is operated via a cable-connected teach pad or an external GUI on a PC. It is used in clinical radiation therapy settings to support and position patients. A sheet is placed between the patient and the support surface during use. The device aids clinicians in achieving precise patient alignment for radiation delivery.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Powered radiation therapy patient support assembly; 6DOF movement via six adjustable linear elements; controller unit with cable-connected teach pad; external PC-based GUI; non-patient-contact design (requires sheet).
Indications for Use
Indicated for patients undergoing radiation therapy requiring precise positioning support.
Regulatory Classification
Identification
A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy.
Predicate Devices
Med-Tec 6 Degree Axis Couch
Elekta Oncology Systems Precise™ Treatment Table
Related Devices
K071870 — HEXAPOD RT COUCH TOP · Medical Intelligence Medizintechnik GmbH · Jul 8, 2007
K062639 — MODIFICATION TO HEXAPOD RT COUCH TOP · Medical Intelligence Medizintechnik-GmbH-An Elekta · Sep 21, 2006
K072898 — HEXAPOD EVO RT COUCH TOP · Medical Intelligence Medizintechnik GmbH · Oct 25, 2007
K042146 — ROBOCOUCH PATIENT SUPPORT SYSTEM · Accuray Incorporated · Sep 17, 2004
K031866 — MED-TEC 6 DEGREE AXIS COUCH, MODEL MT-6DAC · Med-Tec,Inc. · Aug 14, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with outstretched wings. The bird is composed of three thick, curved lines. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medical Intelligence Medizintechnik GmbH % Mr. Stefan Preiss 510(k) TPR Project Manager TÜV Product Service, Inc. 1775 Old Highway 8 NW, Suite # 104 NEW BRIGHTON MN 55112
Re: K041448
Trade/Device Name: HexPod™ RT Couch Top Regulation Number: 21 CFR 892.5770 Regulation Name: Powered radiation therapy patient support assembly Regulatory Class: II Product Code: 90 JAI Dated: May 28, 2004 Received: June 1, 2004
OCT
5 2004
. Dear Mr. Preiss:
This letter corrects our substantially equivalent letter of June 16, 2004 regarding the HexPod™ RT Couch Top and the incorrect regulation number, name and product code.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
HexaPOD™ RT CouchTop
Product documentalion
Medical Intelligence Medizintechnik GmbH
Image /page/2/Picture/3 description: The image shows a logo with the words "Medical Intelligence" to the right of a symbol. The symbol is a circle with a dollar sign inside of it. The dollar sign has a snake wrapped around it.
FDA 510k submission
## Ko41449
**JUN 16 2004**
## 510(k) SUMMARY
Applicant:
1.
| 2. | Address: | Feyerabendstrasse 13 – 15<br>86830 Schwabmünchen<br>Germany |
|----|-------------------|-------------------------------------------------------------|
| 3. | Contact Person: | Christian Hieronimi<br>Tel. +49 (0) 8232 9692-0 |
| 4. | Preparation Date: | February 10, 2004 |
| 5. | Device Submitted: | HexaPOD™ RT Couch Top |
6. Proprietary Name: HexaPOD™ RT CouchTop
7. Common Name: Hexapod
8. Classification Name: Couch, Radiation Therapy, Powered Powered radiation therapy patient support assembly Product Code JAI
். Substantial Equivalence: The HexaPOD is substantially equivalent to the following legally marketed devices: Med-Tec Inc.'s "Med-Tec 6 Degree Axis Couch", and Elekta Instrument Inc.'s "Elekta Oncology Systems Precise™ Treatment Table". The characteristics of this device are similar to those of the predicate devices identified on the comparison chart, which is provided with the premarket notification submission. It is our opinion that the HexaPOD does not have technological characteristics that raise additional types of questions related to terms of safety and effectiveness. 10. Device Description: The HexaPOD consists of two platforms, which are connected by six linear, rigid but length adjustable elements which are powered. By appropriate coordinative adjustment of these elements, the system is able to move the upper platform relative to the lower one. The movement can occur in all three dimensions in space. Additionally the upper platform can rotate around these three axes which results in a tilt or a rotation of the upper platform relative to the lower one. Finally an accurate positioning within all six degrees of freedom (6DOF) can be provided. The HexaPOD consists of a controller unit which is directed by a
cable connected teach pad. Additionally it can be directed via an external graphics user interface (GUI) which is installed on a PC.
## Intended Use: 11. The intended use of the device is to support and aid in positioning a patient during radiation therapy.
| JOC # | ారు గ్రామం నుండి 10 కి.మీ. దూర<br>109 11<br>C<br>· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>" 2004-01-<br>(<br>version | |
|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1.<br>Super States and Canadian Career States of Children | and and any and any and the county of the county of<br>ﻟ<br>Addressed Holm | BE IT US AND MANAGER AND ARREAL<br>AND IN INC. IN<br>445 .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>wounder to F I f A free more and a l'e a l'e manus<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>"We'll "Party and state the country with the would be the county of the county of the county of the county of the former of the former of the former of the former of the form | And answer announce and a some more a more a more a commend thereason works and |
{3}------------------------------------------------
| HexaPOD™ RT CouchTop | Part H 02 | Image: Medical Intelligence logo |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Product documentation | | |
| FDA 510k submission | K041448 | |
| 12. Biocompatibility: | The HexaPOD is not in contact with patient. At any<br>time when in use a sheet is to be placed between the<br>patient's skin surface and the treatment support when<br>in use. Additionally there are no new materials<br>introduced in the manufacture of the HexaPOD.<br>Therefore, no biocompatibility studies were<br>undertaken for this device. | |
| 13. Performance Data: | No performance data is required for this Class II<br>device nor requested by the Food and Drug<br>Administration (Office of Device Evaluation). | |
| CA PI P<br>.<br>Box Angelession A.A., LA LA COLLECT PERSONAL CALL PROPERTY | CARLES COLLECT<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A Book of Children Company of Children Company of Children Company of | TT'LA TH' LIFELALA<br>ירושים בין המועד המועד המועד המועד המועד המועד המועד המועד המועד המילה המועד המועד המילה המועד המועד המועד המועד המועד המילה המועד המועד המועד המילה המועד המועד המועד המועד<br>Jarsin<br>tant T | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <br>Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Chi | <br>i | Canadian Cases of Children of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>And Change<br>NAME OF COLUMNAL PROPERTY OF CLAIM OF CLEAR<br>1 State a de 1<br>and the country of the county of the county of<br>.<br>11 All 11 - 1 - 1 Microsons debon "Anno "Anno a for A<br>11 11119/11/11 11:11:14 11:00 11:00 11:00 11:00 11:00 11:00 11:00 11:00 11:00 11:00 11:00 11:00<br>Comments a . Man former to the Picture Phone | |
{4}------------------------------------------------
## Indications for Use
1041448
510(k) Number (if known):
Device Name:
HexaPOD™ RT CouchTop
Indications For Use:
The intended use of the device is to support and aid in Intended Use: positioning a patient during radiation therapy.
Prescription Use _YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Legron
(Division Sign-Off) Division of Reproductive, Abd and Radiological Devices 510(k) Number .
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.