← Product Code [JAI](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAI) · K033316

# TABLE TILT DEVICE (K033316)

_Brainlab AG · JAI · May 14, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAI/K033316

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [JAI](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAI.md)
- **Decision Date:** May 14, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5770
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Table Tilt Device is an additional device for treatment tables used for radiation treatment. It enables a motorized table tilt around the lateral and longitudinal axis to compensate for patient rotation. In this fashion misalignments and shift of the patient can be precisely compensated.

## Device Story

Powered patient support assembly for radiotherapy/radiosurgery; enables motorized tilt around lateral and longitudinal axes. Input: user-controlled adjustments to correct patient misalignment. Output: physical table tilt. Used in clinical radiation therapy environments; operated by trained personnel. Eliminates manual patient repositioning; improves precision of patient alignment; facilitates accurate radiation delivery.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Powered patient support assembly; motorized tilt mechanism for lateral and longitudinal axes. Intended for integration with radiation treatment tables.

## Regulatory Identification

A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy.

## Predicate Devices

- Med-Tec 6 degree axis couch, model MT-6DAC ([K031866](/device/K031866.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 1 4 2004

# 510 (k) Summary of Safety and Effectiveness for Table Tilt Device

Image /page/0/Picture/2 description: The image shows a handwritten text string. The string appears to be "K033316". The text is written in a dark ink on a white background. The handwriting is somewhat stylized, with some of the characters slightly slanted or distorted.

## Manufacturer:

| Address:        | BrainLAB AG<br>Ammerthalstrasse 8<br>85551 Heimstetten         |  |
|-----------------|----------------------------------------------------------------|--|
|                 | Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 |  |
| Contact Person: | Mr. Rainer Birkenbach                                          |  |
| Summary Date:   | January 8, 2004                                                |  |

| Device Name:                |                     |
|-----------------------------|---------------------|
| Trade name:                 | Robotic Tilt Module |
| Common/Classification Name: | Table Tilt Device   |

| Predicate Device:           | Med-Tec 6 degree axis couch, model MT-6DAC (K031866) |
|-----------------------------|------------------------------------------------------|
| Device Classification Name: | Powered radiation therapy patient support assembly   |
| Regulatory Class:           | Class II                                             |

### Intended Use:

The Table Tilt Device is an additional device for treatment tables used for radiation treatment. It enables a motorized table tilt around the lateral and longitudinal axis to compensate for patient rotation. In this fashion misalignments and shift of the patient can be precisely compensated.

### Device Description:

The Table Tilt Device is a powered patient support assembly for use in radiotherapy and radiosurgery treatments. It allows the user to undertake motorized tilt adjustment of a patient misalignment to eliminate the need to manually move the patient on the table.

### Substantial equivalence:

The Table Tilt Device will be verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device of Med-Tec: Med-Tec 6 degree axis couch, model MT-6DAC (K031866).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) of the United States. The logo features the department's name in a circular arrangement around a stylized eagle symbol. The eagle is depicted with three curved lines representing its body and wings, conveying a sense of motion and flight. The overall design is simple and recognizable, representing the HHS's role in protecting the health of all Americans.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 4 2004

Mr. Rainer Birkenbach Executive Vice President BrainLAB AG Ammerthalstraße 8 85551 Heimstetten GERMANY

Re: K033316 Trade/Device Name: Table Tilt Device Regulation Number: 21 CFR 892.5770 Regulation Name: Powered radiation therapy patient support assembly Regulatory Class: II Product Code: 90 JAI Dated: February 25, 2004 Received: March 9, 2004

Dear Mr. Birkenbach:

We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your Section > re(t) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreative to regally including the Medical Device American be lo commerce private to May 20, 1770, the claculance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). and Cosment Act (Act) that do not reques of the general controls provisions of the Act. The Y ou may, therefore, market the detros, books of the courements for annual registration, listing of general controls provisions of the rice labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classince (soo avoro) its . Existing major regulations affecting your device can may be subject to such additional controller. Title 21, Parts 800 to 898. In addition, FDA may be found in the Code of Poderal Cleg.
publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act mall FUA has made a decemmance. Inc. Jour stered by other Federal agencies. You must comply of any Pederal Statures and regalations and but not limited to: registration and listing (21 CFR Part with an the Act s requirements, menumacturing practice requirements as set forth in the 807), laoching (21 CFR Part 820); and if applicable, the electronic product quality Systems (QB) regions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I has lotter will and work your finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogden

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K033316

510(k) Number (if known):

Table Tilt Device Device Name:

Indications For Usc:

The Table Tilt Device is a device used to compensate rotational patient misalignment (roll and pitch) in a linear accelerator environment for stereotactic radiosurgery or radiotherapy procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRII, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K033316

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAI/K033316](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAI/K033316)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
