← Product Code [JAD](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAD) · K241174

# INTRABEAM (700); And accessories (INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator, INTRABEAM Spherical Sizer Set, INTRABEAM Needle Applicator) (K241174)

_Carl Zeiss Meditec, AG · JAD · Jan 10, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAD/K241174

## Device Facts

- **Applicant:** Carl Zeiss Meditec, AG
- **Product Code:** [JAD](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAD.md)
- **Decision Date:** Jan 10, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5900
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The INTRABEAM is intended for use in radiotherapy treatments. The INTRABEAM SMART Spherical Applicator is used with the INTRABEAM to deliver a prescribed dose of intraoperative radiation to the treatment margin or tumor bed during intracavity and intraoperative radiotherapy treatments. The safety and effectiveness of the INTRABEAM as a replacement for whole breast irradiation in the treast cancer has not been established. The INTRABEM Needle Applicator (comprising the Needle Applicator and guide shafts) is intended for use in combination with the INTRABEAM to intraoperatively administer radiation to tissue including irradiation of intracranial tumors. The INTRABEAM SMART Stand is designed as an instrument support and positioning unit for the INTRABEAM. The INTRABEAM Spherical Sizer Set shall support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.

## Device Story

INTRABEAM 700 is a mobile X-ray radiation therapy system for intraoperative, interstitial, and intracavity tumor treatment. System comprises a mobile workstation (Control Console 700, computer, touchscreen, UNIDOS Romeo electrometer), XRS 4 X-ray source, and accessories (SMART Stand, SMART Spherical Applicators, Needle Applicator, Spherical Sizer Set). X-ray source emits electrons from cathode, accelerated along drift tube to gold target, generating low-energy X-rays (max 50keV). SMART Stand provides positioning, auto-balancing, and vibration damping. Spherical Sizer Set assists clinicians in selecting appropriate applicator size. System includes quality assurance tools (PAICH, PDA, Ionization Chamber). Used in OR or practitioner's office by physicians. Output is controlled radiation dose to target volume. Connectivity includes PACS, ZEISS SMART Service, and cloud-based Health Data Platform. Benefits include targeted, minimally invasive radiation delivery to tumor beds.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included EMC, electrical, mechanical, and thermal safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-4-2), radiation safety (IEC 60601-2-8), software verification/validation, and usability/human factors (IEC 60601-1-6). Biocompatibility and sterility/shelf-life assessments performed per FDA guidance.

## Technological Characteristics

X-ray radiation therapy system (21 CFR 892.5900). XRS 4 X-ray source (max 50keV). Materials: Polyetherimide (spheres/shafts). Connectivity: Wireless/LAN, PACS, Cloud. Sterilization: EO (applicators), Gamma (sizers). Software: Version 5.0. System includes mobile workstation, electrometer, and QA tools. Complies with IEC 60601 series standards.

## Regulatory Identification

An x-ray radiation therapy system is a device intended to produce and control x-rays used for radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- INTRABEAM 600 ([K162568](/device/K162568.md))

## Reference Devices

- INTRABEAM Needle Applicator ([K110590](/device/K110590.md))
- UNIDOS Romeo Electrometer ([K951764](/device/K951764.md))
- Radiance 4 ([K171885](/device/K171885.md))
- Radiance 5 ([K233236](/device/K233236.md))
- INTRABEAM Spherical Applicator ([K121653](/device/K121653.md))
- INTRABEAM Flat Applicator Set ([K130549](/device/K130549.md))
- INTRABEAM Surface Applicator Set ([K130549](/device/K130549.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 10, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of an eagle and a human figure. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text to the right of the square.

Carl Zeiss Meditec AG Chaitali Gawde Senior Regulatory Affairs Specialist Carl Zeiss Meditec, Inc. 5300 Central Parkway Dublin, California 94588

### Re: K241174

Trade/Device Name: INTRABEAM (700); And accessories (INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator, INTRABEAM Spherical Sizer Set, INTRABEAM Needle Applicator) Regulation Number: 21 CFR 892.5900 Regulation Name: X-Ray Radiation Therapy System Regulatory Class: Class II Product Code: JAD Dated: April 26, 2024 Received: December 13, 2024

Dear Chaitali Gawde:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See 

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Locon Weidner

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241174

Device Name

INTRABEAM (700); And accessories (INTRABEAM SMART Spherical Applicator, INTRABEAM Spherical Sizer Set, INTRABEAM Needle Applicator)

Indications for Use (Describe) The INTRABEAM is intended for use in radiotherapy treatments.

The INTRABEAM SMART Spherical Applicator is used with the INTRABEAM to deliver a prescribed dose of intraoperative radiation to the treatment margin or tumor bed during intracavity and intraoperative radiotherapy treatments.

The safety and effectiveness of the INTRABEAM as a replacement for whole breast irradiation in the treast cancer has not been established.

The INTRABEM Needle Applicator (comprising the Needle Applicator and guide shafts) is intended for use in combination with the INTRABEAM to intraoperatively administer radiation to tissue including irradiation of intracranial tumors.

The INTRABEAM SMART Stand is designed as an instrument support and positioning unit for the INTRABEAM.

The INTRABEAM Spherical Sizer Set shall support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:12px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

In accordance with 21 CFR 807.92 the K241174 510(k) Summary for the INTRABEAM (Model No: 700), and accessories (INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator, INTRABEAM Spherical Sizer Set, and INTRABEAM Needle Applicator) is provided below.

#### 1. SUBMITTER

| Applicant:              | Carl Zeiss Meditec AG<br>Rudolf-Eber-Strasse 11<br>73447 Oberkochen<br>Germany                                                                                                                             |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Contact Name  | Anke Seitz<br>Head of Regulatory and Clinical Affairs Innovation<br>MCS<br>Rudolf- Eber- Str. 11<br>Oberkochen, Germany 73447<br>+49 7364 20 8613 Phone<br>E-mail: anke.seitz@zeiss.com (preferred)        |
| Primary Correspondent   | Chaitali Gawde<br>Senior Regulatory Affairs Specialist<br>Carl Zeiss Meditec, Inc.<br>5300 Central Parkway   Dublin, CA 94568<br>(224) 300-3992 Phone<br>E-mail: chaitali.gawde@zeiss.com (preferred)      |
| Secondary Correspondent | Paul Swift<br>Head of Regulatory and Clinical Affairs, Americas<br>Carl Zeiss Meditec, Inc.<br>5300 Central Parkway   Dublin, CA 94568<br>(817) 925.8507 Phone<br>E-mail: paul.swift@zeiss.com (preferred) |

Date Prepared:

January 09, 2025

{5}------------------------------------------------

#### 2. DEVICE

| Device Trade Name: | INTRABEAM (Model No: 700), and accessories (INTRABEAM<br>SMART Stand, INTRABEAM SMART Spherical Applicator,<br>INTRABEAM Spherical Sizer Set, and INTRABEAM Needle<br>Applicator) |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification:    | 21 CFR 892.5900 X-ray radiation therapy system                                                                                                                                    |
| Regulatory Class:  | II                                                                                                                                                                                |
| Product Code:      | JAD                                                                                                                                                                               |

#### LEGAL MARKETED PREDICATE DEVICE 3.

| Predicate Device Name: | INTRABEAM 600 (K162568)                        |
|------------------------|------------------------------------------------|
| Classification:        | 21 CFR 892.5900 X-ray radiation therapy system |
| Regulatory Class:      | II                                             |
| Product Code:          | JAD                                            |

#### 4. DEVICE DESCRIPTION SUMMARY

The INTRABEAM 700 is a radiation therapy device intended for targeted treatments of selected lessions for minimally invasive, intraoperative, interstital, intracavity and contact radiation therapy of tumors or tumor beds within the body of cancer patients. By applying the radiation source in conjunction with various applicators, a prescribed dose of low energy radiation can be delivered to the target volume. The delivery of the radiation dose is controlled via the integrated control unit and software.

The INTRABEAM 700 is provided as a mobile workstation. Like the previously cleared versions of the INTRABEAM system, the INTRABEAM 700 provides several tools for Quality Assurance of radiation delivery, which are intended to verify the proper functioning of the radiotherapy treatment system.

The main components of the INTRABEAM 700 system are:

- INTRABEAM Workplace - mobile cart containing the following:
	- Control Console 700 (CC700) O
	- Computer with Software Version 5.0 O
	- Touchscreen monitor and mouse O
	- UNIDOS Romeo Electrometer o
	- V-guide O
- XRS 4 X-ray Source ●
- Quality Assurance Tools: PAICH, PDA, and Ionization Chamber with Ionization Chamber Holder ●
- radiance Third party treatment planning simulation software ●

The INTRABEAM SMART Stand is connected to the INTRABEAM 700 and is used to mount the X-ray generator (XRS 4) and the necessary applicator, in order to deliver the prescribed radiation dose at the target site.

{6}------------------------------------------------

The INTRABEAM Spherical Applicator is a sterile disposable product that shall be placed in contact with the tumor mass and/or tumor resection cavity to deliver a prescribed dose of intraoperative radiation.

The INTRABEAM Spherical Sizer Set is a sterile disposable product that shall be placed in contact with body part and/or tumor mass to help support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.

The INTRABEAM Needle Applicator has not been updated since the last clearance, K162568.

#### INTENDED USE/INDICATIONS FOR USE ર.

The INTRABEAM is intended for use in radiotherapy treatments.

The INTRABEAM SMART Spherical Applicator is used with the INTRABEAM to deliver a prescribed dose of intraoperative radiation to the treatment margin or tumor bed during intracavity and intraoperative radiotherapy treatments.

The safety and effectiveness of the INTRABEAM as a replacement for whole breast irradiation in the treatment of breast cancer has not been established.

The INTRABEM Needle Applicator (comprising the Needle Applicator and guide shafts) is intended for use in combination with the INTRABEAM to intraoperatively administer radiation to tissue including irradiation of intracranial tumors.

The INTRABEAM SMART Stand is designed as an instrument support and positioning unit for the INTRABEAM.

The INTRABEAM Spherical Sizer Set shall support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.

#### 6. SUBSTANTIAL EQUIVALENCE

Table 1. Subject to Predicate Device Comparison Table - Indications for Use

| Item                   | Subject Device<br>INTRABEAM<br>(K241174)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantial<br>Equivalence<br>Assessment      |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Manufacturer           | Carl Zeiss Meditec AG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Carl Zeiss Meditec AG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | N/A                                           |
| Regulation Name        | X-ray radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | X-ray radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical                                     |
| Regulation             | 21 CFR 892.5900                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 21 CFR 892.5900                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                     |
| Device Class           | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                     |
| Product Code           | JAD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | JAD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                     |
| Classification<br>Name | System, Therapeutic, X-Ray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | System, Therapeutic, X-Ray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                     |
| Use Environment        | Operating room or practitioner's<br>office                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Operating room or practitioner's<br>office                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                     |
| Indications for Use    | The INTRABEAM is intended<br>for use in radiotherapy<br>treatments.<br>The INTRABEAM SMART<br>Spherical Applicator is used with<br>the INTRABEAM to deliver a<br>prescribed dose of intraoperative                                                                                                                                                                                                                                                                                                                                                                             | The INTRABEAM 600 is<br>indicated for radiation therapy<br>treatments.<br>The INTRABEAM Spherical<br>Applicators are indicated for use<br>with the INTRABEAM 600 to<br>deliver a prescribed dose of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent –<br>new accessories<br>were added |
| Item                   | Subject Device<br>INTRABEAM<br>(K241174)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantial<br>Equivalence<br>Assessment      |
|                        | radiation to the treatment margin<br>or tumor bed during intracavity<br>and intraoperative radiotherapy<br>treatments.<br><br>The safety and effectiveness of<br>the INTRABEAM as a<br>replacement for whole breast<br>irradiation in the treatment of<br>breast cancer has not been<br>established.<br><br>The INTRABEAM Needle<br>Applicator (comprising the<br>Needle Applicator and guide<br>shafts) is intended for use in<br>combination with the<br>INTRABEAM to<br>intraoperatively administer<br>radiation to tissue including<br>irradiation of intracranial tumors. | radiation to the treatment margin<br>or tumor bed during intracavity<br>and intraoperative radiotherapy<br>treatments.<br><br>The INTRABEAM Spherical<br>Applicators used with the<br>INTRABEAM 600 are able to<br>deliver a prescribed dose of<br>intraoperative radiation in<br>conjunction with whole breast<br>irradiation, based upon the<br>medical judgment of the<br>physician.<br><br>The safety and effectiveness of<br>the INTRABEAM 600 as a<br>replacement for whole breast<br>irradiation in the treatment of<br>breast cancer has not been<br>established.<br><br>The Needle Applicator set<br>(comprising the Needle<br>Applicator and guide shafts) is<br>intended for use in combination<br>with the INTRABEAM 600 to<br>intraoperatively administer<br>radiation to tissue including<br>irradiation of intracranial tumors.<br><br>The INTRABEAM Flat<br>Applicator is intended to supply<br>a specified radiation dose during<br>applications in combination with<br>the INTRABEAM 600,<br>• during intraoperative<br>radiotherapy, on a surgically<br>exposed surface or in a tumor<br>bed.<br>• during treatment of tumors on<br>the body surface.<br><br>The INTRABEAM Flat<br>Applicator is designed to deliver<br>a flat radiation field at a distance<br>of 5mm from its circular<br>application surface in water.<br>The INTRABEAM Surface<br>Applicator is intended to supply |                                               |
| Item                   | Subject Device<br>INTRABEAM<br>(K241174)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantial<br>Equivalence<br>Assessment      |
|                        | The INTRABEAM SMART<br>Stand is designed as an<br>instrument support and<br>positioning unit for the<br>INTRABEAM.<br>The INTRABEAM Spherical<br>Sizer Set shall support the doctor<br>(surgeon and/or radiation<br>oncologist) in assessing which<br>spherical-shaped applicator shall<br>be used for the radiation therapy<br>procedure, involving<br>INTRABEAM.                                                                                                                                                                                                             | a specified radiation dose during<br>applications in combination<br>with the INTRABEAM 600.<br>• during intraoperative<br>radiotherapy, on a surgically<br>exposed surface or in a tumor<br>bed.<br>• during treatment of tumors on<br>the body surface.<br>The INTRABEAM Surface<br>Applicator is designed to deliver<br>a flat radiation field directly at<br>the applicator's surface.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                               |

{7}------------------------------------------------

{8}------------------------------------------------

Table 2. Subject to Predicate Device Comparison Table – INTRABEAM main device

| Item                                          | Subject Device<br>INTRABEAM (Model: 700)<br>(K241174)                                                                                                                                               | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                     | Substantial<br>Equivalence<br>Assessment                                                                                                       |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                   | INTRABEAM 700                                                                                                                                                                                       | INTRABEAM 600                                                                                                                                                                      | N/A                                                                                                                                            |
| System<br>Component<br>Storage Cart           | INTRABEAM Workplace – a<br>fully enclosed cart that<br>provides dedicated storage<br>space for each component<br>secured inside the cart                                                            | INTRABEAM Workplace – a<br>fully enclosed cart that<br>provides dedicated storage<br>space for each component<br>secured inside the cart                                           | Identical                                                                                                                                      |
| System<br>Components on<br>Cart               | Control Console 700<br>(CC700)                                                                                                                                                                      | Control Console 600 (CC600)                                                                                                                                                        | Equivalent- subject<br>device makes use of a<br>different power<br>supply concept.<br><br>Safety is proven by<br>established methods           |
| Item                                          | Subject Device<br>INTRABEAM (Model: 700)<br>(K241174)                                                                                                                                               | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                     | Substantial<br>Equivalence<br>Assessment                                                                                                       |
|                                               | Computer, Touchscreen<br>Monitor, Mouse, (Keyboard<br>functionality is available via<br>the touchpad)                                                                                               | Computer, Touchscreen<br>Monitor, Keyboard, Mouse                                                                                                                                  | IEC 60601-1 and IEC<br>60601-2-8                                                                                                               |
|                                               | UNIDOS Romeo<br>Electrometer                                                                                                                                                                        | UNIDOS E (Dosimeter)                                                                                                                                                               | Equivalent -<br>UNIDOS Romeo<br>Electrometer is the<br>successor model and<br>cleared under<br>(K951764)                                       |
|                                               | Ionization Chamber<br>V-guide                                                                                                                                                                       | Ionization Chamber<br>V-guide                                                                                                                                                      | Identical                                                                                                                                      |
| Radio Modules                                 | RFID Reader                                                                                                                                                                                         | N/A                                                                                                                                                                                | Equivalent - RFID<br>reader will not have<br>an impact on the use<br>of the device.                                                            |
| Network                                       | Wireless or LAN                                                                                                                                                                                     | LAN                                                                                                                                                                                | Equivalent - used for<br>connection of PC with<br>hospital network                                                                             |
| Connectivity                                  | - Picture Archiving and<br>Communication System<br>(PACS)<br>- ZEISS SMART Service<br>- ZEISS Server / Cloud -<br>Health Data Platform<br>(HDP)                                                     | - Picture Archiving and<br>Communication System<br>(PACS)                                                                                                                          | Equivalent -<br>additional<br>connectivity features<br>in the subject device,<br>no impact on the<br>intended use of the<br>device.            |
| Applicators<br>required for<br>Treatment      | Yes                                                                                                                                                                                                 | Yes                                                                                                                                                                                | Identical                                                                                                                                      |
| Method of<br>treatment /<br>application       | Intraoperative<br>Intracavitary<br>Interstitial<br>Post-Operative                                                                                                                                   | Intraoperative<br>Intracavitary<br>Interstitial<br>Post-Operative                                                                                                                  | Identical                                                                                                                                      |
| Radiation Source                              | XRS 4                                                                                                                                                                                               | XRS 4                                                                                                                                                                              | Identical                                                                                                                                      |
| Software on<br>Computer                       | Version 5.0                                                                                                                                                                                         | Version 4.0                                                                                                                                                                        | Equivalent - Updated<br>Software Version                                                                                                       |
| Controller Set                                | 40keV<br>50keV                                                                                                                                                                                      | 40keV<br>50keV                                                                                                                                                                     | Identical                                                                                                                                      |
| Energy range used                             | 40keV<br>50keV                                                                                                                                                                                      | 40keV<br>50keV                                                                                                                                                                     | Identical                                                                                                                                      |
| Energy of<br>radiation emitted                | Max 50keV                                                                                                                                                                                           | Max 50keV                                                                                                                                                                          | Identical                                                                                                                                      |
| Item                                          | Subject Device<br>INTRABEAM (Model: 700)<br>(K241174)                                                                                                                                               | Predicate Device<br>INTRABEAM 600<br>(K162568)                                                                                                                                     | Substantial<br>Equivalence<br>Assessment                                                                                                       |
| General mode of operation of the X-Ray source | Electrons are emitted by cathode, accelerated by an electrical field along a drift tube inside the X-Ray source and hit a gold target resulting in the generation of X-rays                         | Electrons are emitted by cathode, accelerated by an electrical field along a drift tube inside the X-Ray source and hit a gold target resulting in the generation of X-rays        | Identical                                                                                                                                      |
| Maximum radiation output                      | 0.6Gy/min (at 2cm from isocenter)                                                                                                                                                                   | 0.6Gy/min (at 2cm from isocenter)                                                                                                                                                  | Identical                                                                                                                                      |
| Maximum photon energy                         | 50keV                                                                                                                                                                                               | 50keV                                                                                                                                                                              | Identical                                                                                                                                      |
| Geometry of dose emitted (without applicator) | Mostly spherical                                                                                                                                                                                    | Mostly spherical                                                                                                                                                                   | Identical                                                                                                                                      |
| Does fall-off (in water)                      | ~ 1/r3                                                                                                                                                                                              | ~ 1/r3                                                                                                                                                                             | Identical                                                                                                                                      |
| Maximum Power Range                           | 2W (50kV x 40 $\mu$ A)                                                                                                                                                                              | 2W (50kV x 40 $\mu$ A)                                                                                                                                                             | Identical                                                                                                                                      |
| Maximum Beam Current                          | 40 $\mu$ A                                                                                                                                                                                          | 40 $\mu$ A                                                                                                                                                                         | Identical                                                                                                                                      |
| System Quality Assurance Tools                | Probe Adjuster Ionization Chamber Holder (PAICH)<br>Photo Diode Array (PDA)<br>Ionization Chamber (IC)                                                                                              | Probe Adjuster Ionization Chamber Holder (PAICH)<br>Photo Diode Array (PDA)<br>Ionization Chamber (IC)                                                                             | Identical                                                                                                                                      |
| Radiation Treatment Planning Software         | Radiance V5                                                                                                                                                                                         | Radiance V4                                                                                                                                                                        | Equivalent -<br>Radiance 4<br>(K171885)<br>Radiance 5<br>(K233236)                                                                             |
| Compatible Applicators                        | – INTRABEAM SMART Spherical Applicator (K241174)<br>– INTRABEAM Needle Applicator (K110590)                                                                                                         | – INTRABEAM Spherical Applicator (K121653)<br>– INTRABEAM Flat Applicator Set (K130549)<br>– INTRABEAM Surface Applicator Set (K130549)<br>– INTRABEAM Needle Applicator (K110590) | Equivalent – new compatible applicator is introduced for INTRABEAM 700                                                                         |
| Stand                                         | INTRABEAM SMART Stand:<br>- Wireless Module<br>- RFID Module with auto-draping functionality<br>- auto-balancing<br>- active vibration damping & application modes                                  | NC32 INTRABEAM Floor Stand                                                                                                                                                         | Equivalent - SMART Stand offers more functions for INTRABEAM 700                                                                               |
| Item                                          | Subject Device<br>INTRABEAM SMART<br>Spherical Applicator<br>(K241174)                                                                                                                              | Predicate Device<br>INTRABEAM Spherical<br>Applicator (K121653)                                                                                                                    | Substantial<br>Equivalence<br>Assessment                                                                                                       |
| Device Name                                   | INTRABEAM SMART<br>Spherical Applicator                                                                                                                                                             | INTRABEAM Spherical<br>Applicator                                                                                                                                                  | N/A                                                                                                                                            |
| Energy source                                 | X-ray source                                                                                                                                                                                        | X-ray source                                                                                                                                                                       | Identical                                                                                                                                      |
| Maximum Voltage<br>(kV)                       | 50 kV or 40 kV                                                                                                                                                                                      | 50 kV or 40 kV                                                                                                                                                                     | Identical                                                                                                                                      |
| Maximum Current<br>(μΑ)                       | 40 μΑ                                                                                                                                                                                               | 40 μΑ                                                                                                                                                                              | Identical                                                                                                                                      |
| Number of<br>applicator sizes                 | Eight (8)                                                                                                                                                                                           | Eight (8)                                                                                                                                                                          | Identical                                                                                                                                      |
| Applicator<br>Diameters                       | Spherical Diameter:<br>1.5 cm; 2 cm; 2.5 cm; 3 cm;<br>3.5 cm; 4 cm; 4.5 cm and 5 cm                                                                                                                 | Spherical Diameter:<br>1.5 cm; 2 cm; 2.5 cm; 3 cm;<br>3.5 cm; 4 cm; 4.5 cm and 5 cm                                                                                                | Identical                                                                                                                                      |
| Materials                                     | Sphere:<br>Polyetherimide<br><br>Shaft: Polyetherimide                                                                                                                                              | Sphere:<br>Polyetherimide (Ultem<br>1000)<br><br>Shaft: Ultem 1000 and<br>stainless steel                                                                                          | Equivalent - Stainless<br>steel no longer used,<br>does not affect the<br>intended use of the<br>device.                                       |
| Sterilization                                 | Supplied sterile;<br>EO Sterilization                                                                                                                                                               | Supplied non-sterile. User<br>performs steam sterilization.                                                                                                                        | Equivalent –<br>sterilization is carried<br>out in both.                                                                                       |
| Number of uses                                | Single-Use                                                                                                                                                                                          | Multiple uses. Can be<br>sterilized 100 times by the<br>user.                                                                                                                      | Equivalent - there is no<br>impact on the use of<br>the device.                                                                                |
| Primary Packaging                             | Lid: Tyvek®<br>Blister: PETG transparent                                                                                                                                                            | N/A                                                                                                                                                                                | Equivalent – Primary<br>packaging was not<br>required before, as<br>device was sterilized<br>by the user.                                      |
| Connectivity                                  | RFID tag                                                                                                                                                                                            | N/A                                                                                                                                                                                | Equivalent - RFID tag<br>will not have an impact<br>on the use of the<br>device.                                                               |
| Sizing support tool                           | INTRABEAM Spherical Sizer<br>Set<br>- Gamma sterilization<br>- Single-Use<br>- 8 different sizes<br>- Base Material:<br>polycarbonate<br>- Color Batch Material:<br>Acrylnitril-Butadien-<br>Styrol | N/A                                                                                                                                                                                | Equivalent –<br>INTRABEAM<br>Spherical Sizer Set will<br>support the user to pick<br>appropriate<br>INTRABEAM<br>SMART Spherical<br>Applicator |

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

| Table 3. Subject to Predicate Device Comparison Table - INTRABEAM SMART Spherical Applicator |
|----------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|

{12}------------------------------------------------

| Item | Subject Device<br>INTRABEAM SMART<br>Spherical Applicator<br>(K241174)           | Predicate Device<br>INTRABEAM Spherical<br>Applicator (K121653) | Substantial<br>Equivalence<br>Assessment |
|------|----------------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------|
|      | - Primary Packaging:<br>Tyvek® and<br>laminate film transparent<br>peel PET-O/PP |                                                                 |                                          |

{13}------------------------------------------------

#### NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY OF STUDIES 7.

### Sterilization and Shelf Life

The INTRABEAM was assessed for Sterility and Shelf-iffe information per recommendations in FDA 's Sterility guidance document and were deemed to demonstrate substantial equivalence.

## Biocompatibility

The device consists of patient-contacting materials; therefore, a biocompatibility assessment was performed in accordance with FDA's biocompatibility guidance documents.

The biocompatibility evaluation demonstrated substantial equivalence.

## Performance Testing - Bench

Non-clinical system testing provided an evaluation of the system relevant to each of the system specifications. The functional and system level testing showed that the defined specifications. Software verification testing has been performed to demonstrate that software is performing as intended.

The following parameters/specifications were tested in conformity with the relevant requirements:

- . EMC, Wireless, Electrical, Mechanical, and Thermal Safety
The INTRABEAM (model no: 700) with INTRABEAM SMART Stand were assessed for conformity with the relevant requirements of IEC 60601-1-2 "Medical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests" and were found to comply.

The INTRABEAM (model no: 700) with INTRABEAM SMART Stand were assessed for conformity with the relevant requirements of IEC TR 60601-4-2 "Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems" and were found to comply.

The INTRABEAM (model no: 700) with INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator and INTRABEAM Needle Applicator (as detachable parts) were assessed for conformity with the relevant requirements of IEC 60601-1 "Medical electrical equipment – Part 1: General requirements for basic safety and essential performance" and were found to comply.

- Radiation Safety ●
The therapeutic X-ray equipment of INTRABEAM (model no: 700) with INTRABEAM SMART Stand were assessed for conformity with the relevant requirements of IEC 60601-2-8 "Medical electrical equipment - Part 2-8: Particular requirements for the safety of therapeutic X-ray equipment operating in the range 10 kV to 1 MV" and were found to comply.

- System Level Verification
The system level testing verified that the device performed according to requirements.

- Software Verification and Validation ●
Software documentation was provided in accordance with FDA's software guidance documents. Enhanced documentation was submitted for INTRABEAM (Model: 700) and its accessories. The results of verification and validation testing demonstrate that the software performs in accordance with its

{14}------------------------------------------------

established requirements and will therefore meet user needs and intended uses. Also, the FDA's cybersecurity recommendations were fulfilled.

- Usability / Human Factors
The INTRABEAM (model no: 700) with INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator and INTRABEAM Needle Applicator (as detachable parts) were assessed for conformity with the relevant requirements of IEC 60601-1-6 "Medical electrical equipment – Part1-6: General requirements for basic safety and essential performance - Collateral standard: Usability". And the FDA's recommendations for human factors validation are fulfilled.

All the performance testing evaluations demonstrated substantial equivalence.

#### 8. CONCLUSION

The indications for use are equivalent to the indications for use of the predicate devices; and therefore, are deemed to demonstrate substantial equivalence.

The technological characteristics and risk profile of the subject device are equivalent to the predicate devices; and therefore, are deemed to demonstrate substantial equivalence.

Testing methods are equivalent to those of the predicate devices and support substantial equivalence.

Accordingly, the INTRABEAM (Model: 700 and accessories) is substantially equivalent to the predicate device, the INTRABEAM 600 (K162568).

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAD/K241174](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/JAD/K241174)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
