K990085 · Varian Assoc., Inc. · IYE · Feb 8, 1999 · Radiology
Device Facts
Record ID
K990085
Device Name
MILLENNIUM MLC -120
Applicant
Varian Assoc., Inc.
Product Code
IYE · Radiology
Decision Date
Feb 8, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Varian Millennium Multileaf 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac® dose fraction or gantry angle. The intended use to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
Device Story
Millennium MLC-120 is an x-ray collimator accessory for Varian Clinac® C or EX Series linear accelerators. It mounts beneath standard field-defining jaws to shape radiation beams. Operates in static mode (replacing custom shadow blocks) or dynamic mode (leaf positions indexed to dose fraction or gantry angle). Used by radiation oncologists in clinical settings to deliver precise radiation to target volumes while sparing healthy tissue/organs. Enables 3D dose distribution through dynamic beam shaping during treatment.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
X-ray collimator accessory for Varian Clinac C or EX Series linear accelerators. Features 120-leaf design for field shaping. Operates via mechanical leaf positioning indexed to gantry angle or dose fraction. Connectivity is integrated with the host linear accelerator system.
Indications for Use
Indicated for patients requiring radiation therapy where the device assists radiation oncologists in delivering radiation to defined target volumes while sparing surrounding normal tissue and critical organs. Used in static mode as a replacement for customized shadow blocks or in dynamic mode to create changing beam shapes indexed to dose fraction or gantry angle for 3D dose distribution.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Varian Multileaf Collimator with Dynamic Arc Therapy (K973889)
Related Devices
K050442 — MILLENNIUM MULTI-LEAF COLLIMATOR · Varian Medical Systems · Mar 23, 2005
K973963 — 120 LEAF MULTILEAF COLLIMATOR · Mitsubishi Electronics America, Inc. · Jan 14, 1998
K973889 — VARIAN MULTILEAF COLLIMATOR WITH DYNAMIC ARC THERAPY FEATURE · Varian Medical Systems, Inc. · Nov 21, 1997
K133240 — VARIAN MULTILEAF COLLIMATOR · Varian Medical Systems, Inc. · Jan 31, 2014
K071992 — HIGH-DEFINITION 120 MULTILEAF COLLIMATOR (HD120 MLC) · Varian Medical Systems · Aug 15, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 8 1559 KG90085
### varian@ oncology systems
# Premarket Notification [510K] Summary as required by 21 CFR 807.92
Date Summary was prepared; January 8, 1999
Submitter's Name: Varian Oncology Systems 3045 Hanover Street Palo Alto, CA 94304
#### Contact Person:
Linda S. Nash Regulatory Compliance & Radiation Safety Manager Phone (650) 424-6990 (650) 424-4830 FAX E-Mail linda.nash@os.varian.com
Device Name: Millennium MLC - 120
Classification Name: Medical Linear Accelerator Accessory, 90 IYE (per 21 CFR section 892.5050)
Predicate Device: Varian Multileaf Collimator with Dynamic Arc Therapy, K973889
#### Product Description:
The Varian Millennium Multileaf Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator beneath the standard field defining the collimator jaw.
Image /page/0/Picture/13 description: The image shows a logo for "Partners for Life". The words "Partners" and "For" are stacked on top of each other in a bold, sans-serif font. The word "Life" is written in a cursive script and extends to the right of the other words. The logo is in black and white.
3 printed on recycled paper
{1}------------------------------------------------
### Intended Use:
The Varian Millennium Multileaf 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac® dose fraction or gantry angle. The intended use to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
### Technological Characteristics:
See the attached "Specification Comparison Chart" (Tab F).
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
FEB 8 1999
Linda Nash Regulatory Compliance And Radiation Safety Manager Varian Associates, Inc. 3045 Hanover Street Palo Alto, CA 94304-1129
Re:
K990085 Millenium MLC-120 (Multileaf Collimator) Dated: January 8, 1999 Received: January 11, 1999 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE
Dear Mr. Nash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have We neve reviewed your oodlor of the indications for use stated in the enclosure) to legally decemined the devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the manced production as noo manova lices that have been reclassified in accordance with the provisions of the Mcaral Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may ir your additional controls. Existing major regulations affecting your device can be found in the Code of De Subject to Julianal Controlly - Extentially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Mourden of the arreral voyalerily such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal rogatory dollor. In adallion, 1971 hay paur premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
~~~~
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Varian Oncology Systems. The word "varian" is in bold, black letters, with a circled "VA" symbol to the right of the word. Below the word "varian" is the phrase "oncology systems" in a smaller, non-bold font.
# Indications For Use
510(k) Number (if known): _K990085
Device Name: Millennium MLC - 120
Indications for Use: The Varian Millennium 120 Multileaf Collimator (MLC) is provided to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. In a static mode, the MLC performs the same function as the customized shadow blocks. In a dynamic mode, a series of MLC leaf positions can be indexed to either Clinac® dose fraction or gantry angle to create a changing beam shape while the radiation beam is on to create a three dimensional dose distribution.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH/Office of Device Evaluation (ODE) | |
|--------------------------------------------------------------------|--------------------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K990085 |
| Prescription Use (per 21 CFR 801.109) | OR Over-The-Counter Use ______ |
Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129 U.S.A. 650.493.4000 FAX: 650.424.4830
25 printed on recycled pa
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.