← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K963860

# THE NOMOS TALON POSITIONING AND IMMOBILIZATION SYSTEM (K963860)

_Nomos Corp. · IYE · Dec 23, 1996 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K963860

## Device Facts

- **Applicant:** Nomos Corp.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Dec 23, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The NOMOS TALON Positioning and Immobilization System is intended to provide an invasive means to repeatedly position and immobilize a patient during: a) diagnostic procedures in preparation for radiation therapy (RT) treatments of the head or head and neck, and b) during treatment procedures in the radiation therapy suite.

## Device Story

The NOMOS TALON Positioning and Immobilization System is an invasive stereotactic device used to secure a patient's head during diagnostic imaging (CT) and radiation therapy. The system consists of a skull mount, a rigid bracket (NOMOGrip), a targeting box for laser alignment, a CT reference pointer, and adapters for the treatment couch and CT head support. The skull mount is fixed to the patient's skull via screws. The device allows for precise, repeatable patient positioning relative to the linear accelerator's light field and lateral lasers. It is operated by clinical staff in radiation therapy suites and diagnostic imaging departments. By ensuring rigid immobilization, the device facilitates accurate delivery of radiation therapy, potentially improving treatment precision and patient outcomes.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Invasive skull-mounted positioning system. Components include skull screws, rigid brackets, targeting frames, and imaging/treatment couch adapters. Materials are consistent with previously cleared NOMOS stereotactic systems. Mechanical fixation principle; no energy source or software components.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- RSA XKnife Radiosurgical System/BRW Stereotactic System ([K811452](/device/K811452.md), [K912630](/device/K912630.md), [K923522](/device/K923522.md))
- NOMOS ReFix System (K930279B)
- gNomos Stereotactic System for neurosurgery (K901670B)
- CT & MR frame from BRW Stereotactic System ([K811452](/device/K811452.md))
- Vertical post of gNomos Stereotactic System (K901670B)
- ReFix Radiotherapy Table Adapter (K930279B)
- ReFix CT Head Support (K930279B)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K963860

NOMOS TALON 9/23/96 Page 31

DEC 23 1996

## Appendix 1 510(k) Summary (pursuant to 513(l))

## Summary of Safety and Effectiveness NOMOS TALON™ Positioning and Immobilization System

Pursuant to Section 513(i) of the Federal Food, Drug, and Cosmetic Act

## I. General Information:

Trade/Proprietary Name: The NOMOS TALON Positioning and Immobilization System

Common/Usual Name: Patient Positioning/Positioning and Immobilization System

Classification Name: Medical Charged Particle Radiation Therapy Systems and Accessories

## Applicant's Name and Address:

Marvin L. Sussman, Ph. D.
NOMOS Corporation (NOMOS)
2591 Wexford Bayne Road (Suite 400)
Sewickley, PA 15143

## II. Name of predicate device(s):

TALON Skull Mount:
RSA XKnife Radiosurgical System/BRW Stereotactic System (K811452, K912630 and K923522)

TALON Skull Mount and NOMOGrip:
NOMOS ReFix™ System (K930279B)
gNomos Stereotactic System for neurosurgery (K901670B)

Targeting Box:
CT &amp; MR frame from BRW Stereotactic System (K811452)

CT Reference Pointer:
Vertical post of gNomos Stereotactic System (K901670B)

Radiotherapy Table Adapter:
ReFix Radiotherapy Table Adapter (K930279B)

000029

{1}

NOMOS TALON 9/23/96 Page 32

CT Head Support:
ReFix CT Head Support (K930279B)

Each of these devices has been marketed prior to and/or following enactment of the Medical Device Amendments in May, 1976.

III. Classification:

Medical Charged Particle Radiation Therapy Systems and accessories are Class II (21 CFR 892.5050) - Radiological Devices Panel.

Radiographic Head Holders are Class I (21 CFR 892.1920) and are exempt from the premarket notification procedures in Subpart E and Part 807.

IV. Performance Standards:

No applicable performance standards have been established by FDA under section 514 of the Food, Drug and Cosmetic Act.

V. Intended Use and Device Description:

The NOMOS TALON Positioning and Immobilization System is intended to provide an invasive means to repeatedly position and immobilize a patient during:

a) diagnostic procedures in preparation for radiation therapy (RT) treatments of the head or head and neck, and

b) during treatment procedures in the radiation therapy suite.

A description of the function(s) of the major components follow:

The TALON Skull Mount is an invasive means of repeatedly positioning and immobilizing a patient during diagnostic imaging and/or therapeutic administrations of radiation therapy (RT).

The NOMOGrip is a rigid bracket that attaches the TALON Skull Mount to the Radiotherapy Table Adapter or CT head support.

The Target Box is an external, removable frame designed to permit positioning of the patient with respect to the lateral lasers and defining light field from the linear accelerator.

000030

{2}

NOMOS TALON 9/23/96 Page 33

CT Reference Pointer is a pointer which attaches with 2 screws to the NOMOGrip and produces a reference pointer on the CT scans.

Radiotherapy Table Adapter attaches to the treatment couch and provides an attachment point for the NOMOGrip during treatment.

CT Head Support attaches directly to the CT imaging couch and provides an attachment point for the NOMOGrip during imaging.

VI. Summary of Substantial Equivalence:

Indications: The indications for use are the same as the predicate NOMOS ReFix System (K930279B) and are similar to the other predicate devices.

Design: The design is similar to the components of the predicate NOMOS gNomos Stereotactic System (K901670B) and components of the predicate NOMOS ReFix System (K930279B).

Materials: The implanted material (Skull Screws) is similar to the material used in the predicate NOMOS gNomos Stereotactic System (K901670B). The non-implanted materials are the similar to those used in the NOMOS ReFix System (K930279B).

Manufacturing: The manufacturing processes are the same as those used in the NOMOS ReFix System (K930279B) and the gNomos Stereotactic System (K901670B).

Specifications: The specifications for repeat fixation are similar to those of the NOMOS gNomos Stereotactic System (K901670B).

Conclusions: The indications, design, materials, manufacturing, and specifications of the NOMOS TALON Positioning and Immobilization System do not raise any new issues relating to safety and effectiveness. NOMOS thus considers the NOMOS TALON Positioning and Immobilizing System equivalent to the predicate devices for positioning and immobilization of the head in therapeutic administrations of radiation.

'Any statement made in conjunction with this submission regarding substantial equivalence to any other product only relates to whether the product can be lawfully marketed without premarket approval or reclassification and is not to be interpreted as an admission or used as evidence in patent infringement litigation. As the Commissioner of the FDA has indicated, "... a determination of substantial equivalence under the Federal Food, Drug, and Cosmetic Act relates to the fact that the product can lawfully be marketed without premarket approval or reclassification. This determination is not intended to have any bearing whatever on the resolution of patent infringement suits." 42 Fed. Reg. 42,520 et seq. (1977).

000031

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K963860](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K963860)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
