CADPLAN, VERSION 2.62. RADIATION THERAPY TREATMENT PLANNING SYSTEM

K962950 · Varian Medical Systems, Inc. · IYE · Jul 22, 1997 · Radiology

Device Facts

Record IDK962950
Device NameCADPLAN, VERSION 2.62. RADIATION THERAPY TREATMENT PLANNING SYSTEM
ApplicantVarian Medical Systems, Inc.
Product CodeIYE · Radiology
Decision DateJul 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

CADPLAN is used to plan radiation therapy treatments employing linear accelerators and other similar teletherapy devices with x-ray energies from 4 to 50 MV, as well as Cobalt-60, and electron energies from 4 to 50 MeV, and to plan brachytherapy treatments. CADPLAN will plan the 3D radiotherapy treatment approaches to combined modality plans, asymmetric and non-coplanar field, total body irradiation, multileaf collimators, motorized and dynamic wedges, customized blocking, compensating filters, and bolus. CADPLAN includes export capabilities to verify beam and patient data, dose planning results, and provide on-line information to block-cutting devices.

Device Story

CADPLAN is a 3D radiation therapy treatment planning (RTP) system operating on UNIX technology. It processes patient anatomical data and beam parameters to model radiation dose distributions for photon, electron, and brachytherapy treatments. The system supports complex configurations including non-coplanar fields, multileaf collimators, and dynamic wedges. Clinicians use the system to design treatment plans, which are then exported to verify data and drive block-cutting devices. The device assists in optimizing radiation delivery, potentially improving treatment accuracy and patient outcomes. It is intended for clinical use by radiation oncology professionals.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

UNIX-based 3D radiation therapy planning system. Supports photon/electron energies 4-50 MV/MeV and Cobalt-60. Features include 3D dose modeling, support for multileaf collimators, dynamic wedges, and customized blocking. Connectivity includes export capabilities for beam/patient data and block-cutting device integration.

Indications for Use

Indicated for planning radiation therapy treatments in patients requiring photon or electron therapy (4-50 MV/MeV) or brachytherapy. Supports complex treatment approaches including 3D planning, combined modalities, asymmetric/non-coplanar fields, total body irradiation, and use of multileaf collimators, wedges, and customized blocking.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 6962950 # Summary of Safety and Effectiveness Information 1. Submitter: Varian Oncology Systems 3045 Hanover Street Palo Alto, CA 94304 JUL 22 1997 Contact: Charles H. Will, Manager Regulatory Compliance & Safety Phone (415) 424-5036 FAX (415) 424-4830 cwill@os.varian.com Prepared: July 26, 1996 2. Device Name: CADPLAN Radiation Therapy Treatment Planning (RTP) System 3. Predicate Device: FOCUS, manufactured by Computerized Medical Systems, Inc. 4. Description: Varian CADPLAN is a comprehensive 3D RTP system for radiation dose modeling of patients undergoing photon or electron therapy or brachytherapy, based on modern UNIX technology. External beam calculation is fully 3-dimensional supporting coplanar and non-coplanar fields. The beam model takes into account collimator rotation, off-center calculation, and irregular field shape. Different energies and treatment modalities (fixed, isocentric, and arc) can be combined in the same plan; and external beam, intracavitary, and interstitial plans can be combined. Wedges, both static and dynamic, blocks, compensators, and boluses can be used. All images, graphical data, and texts are displayed on one high resolution monitor. Easy to use operation is provided with mouse selectable menus. 5. Intended Use: CADPLAN is used to plan radiation therapy treatments employing linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 50 MV, as well as Cobalt-60, and electron energies from 1 to 50 MeV, and to plan brachytherapy treatments. CADPLAN will plan the 3D radiotherapy treatment approaches to combined modality plans, asymmetric and non-coplanar field, total body irradiation, multileaf collimators, motorized and dynamic wedges, customized blocking, compensating filters, and bolus. {1} Page 2 Varian CADPLAN 510(k) Summary of Safety and Effectiveness Information 5. Intended Use: CADPLAN includes export capabilities to verify beam and patient data, dose planning results, and provide on-line information to block-cutting devices. Cont'd. 6. Technological Considerations: CADPLAN has no significant differences in design, materials, energy source or other technological characteristics compared to the predicate device. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 22 Charles H. Will Manager Regulatory Compliance & Safety Varian Oncology Systems Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129 Re: K962950 Varian Cadplan ver. 2.62, Radiotherapy Treatment Planning System Dated: June 30, 1997 Received: July 2, 1997 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE Dear Mr. Will: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number: K962950 Device Name: Varian CadPlan, Version 2.62 ## Indications for use: I state in my capacity as Manager, Regulatory Compliance and Safety, of Varian Oncology Systems that the CADPLAN Radiation Therapy Treatment Planning System which is the subject of this premarket notification, is intended to be used for the following: CADPLAN is used to plan radiation therapy treatments employing linear accelerators and other similar teletherapy devices with x-ray energies from 4 to 50 MV, as well as Cobalt-60, and electron energies from 4 to 50 MeV, and to plan brachytherapy treatments. CADPLAN will plan the 3D radiotherapy treatment approaches to combined modality plans, asymmetric and non-coplanar field, total body irradiation, multileaf collimators, motorized and dynamic wedges, customized blocking, compensating filters, and bolus. CADPLAN includes export capabilities to verify beam and patient data, dose planning results, and provide on-line information to block-cutting devices. Charles H. Will, Manager Regulatory Compliance & Safety Date Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K962950 Prescription Use ☑ (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...