← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K223839

# TrueBeam, TrueBeam STx, Edge, VitalBeam (K223839)

_Varian Medical Systems, Inc. · IYE · May 22, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K223839

## Device Facts

- **Applicant:** Varian Medical Systems, Inc.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** May 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carvinoma), gentourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, avernous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only). VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pitutary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal turnors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

## Device Story

Medical linear accelerator system; delivers photon, electron, and kV X-ray radiation for stereotactic radiosurgery and precision radiotherapy. System comprises radiation-producing component in shielded vault and external control console. Operated by clinical staff; delivers radiation based on physician prescription. Enables precise target localization, beam collimation, and patient immobilization. Used for wide range of oncological and benign conditions, including essential tremor. Benefits patients by providing targeted therapeutic radiation while sparing surrounding healthy tissue.

## Clinical Evidence

Literature review of over 1300 patients treated with SRS for tremor conditions. Reported outcomes and complication rates comparable to traditional surgical options and Gamma Knife-based SRS. Supports safety and efficacy of linear accelerator-based SRS for medically refractory essential tremor.

## Technological Characteristics

Medical linear accelerator; photon, electron, and kV X-ray radiation sources. Features advanced target localization, beam collimation, and patient immobilization. System architecture and performance specifications identical to K213977. No hardware modifications.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- TrueBeam Radiotherapy System and Accessories ([K213977](/device/K213977.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Varian Medical Systems, Inc. % Peter J. Coronado Sr. Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304

Re: K223839

Trade/Device Name: TrueBeam™, TrueBeam STxTM, Edge™, VitalBeam Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: April 20, 2023 Received: April 20, 2023

Dear Peter Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

May 22, 2023

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lora D.
Weidner -S

Digitally signed by
Lora D. Weidner -S
Date: 2023.05.22
17:25:26 -04'00'

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223839

Device Name

TrueBeam, TrueBeam STx and Edge Radiotherapy Delivery System; VitalBeam

Indications for Use (Describe) TrueBeam-TrueBeam STx-Edge:

The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere radiation therapy is indicated for adults and pediatric patients.

The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carvinoma), gentourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, avernous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only).

## VitalBeam:

VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pitutary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal turnors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

{3}------------------------------------------------

### Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### K223839 PREMARKET NOTIFICATION 510(k) Summary TrueBeam and VitalBeam Radiotherapy Treatment System

The following information follows the format of 21 CFR 807.92

| Submitter's Name:       | Varian Medical Systems<br>3100 Hansen Way, m/s E110<br>Palo Alto CA94304                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Peter J. Coronado<br>Sr. Director Regulatory Affairs<br>Phone: (650) 424.6320<br>Fax: (650) 646.9200<br>E-mail: submissions.support@varian.com                                                                                                                                                                                                                                                                                                                                                                                               |
| Date Prepared:          | 18 May 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Proprietary Name:       | TrueBeamTM /TrueBeam STxTM/EdgeTM/VitalBeam                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name:    | Medical charged-particle radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation:             | 21CFR892.5050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulatory Class        | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code:           | IYE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common/Usual Name:      | Linear accelerator radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Predicate Devices:      | TrueBeam Radiotherapy System and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Description:     | The TrueBeam and VitalBeam Radiotherapy System is a medical linear<br>accelerator that delivered therapeutic radiation to patient in accordance<br>with the physician's prescription.<br>The system consists of two major components – a photon, electron and<br>diagnostic kV X-ray radiation beam producing component that is installed in<br>a radiation-shielded vault and a control console area located outside the<br>treatment room.                                                                                                 |
| Intended Use            | The intended use is the same as the predicate.<br>TrueBeam-TrueBeam STx-Edge:<br>The TrueBeam TM radiotherapy delivery system is intended to provide<br>stereotactic radiosurgery and precision radiotherapy for lesions, tumors,<br>and conditions anywhere in the body where radiation treatment is<br>indicated.<br>VitalBeam:<br>The VitalBeam system is intended to provide stereotactic radiosurgery and<br>precision radiotherapy for lesions, tumors, and conditions anywhere in the<br>body where radiation treatment is indicated. |

{5}------------------------------------------------

### TrueBeam-TrueBeam STx-Edge:

The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only).

### VitalBeam:

VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, tumors, spinal coocord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors,

{6}------------------------------------------------

|                                 | arteriovenous malformation, cavernous malformation, and skull base<br>tumors).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics:  | The device has the same technological characteristics as the previously<br>cleared TrueBeam device K213977. The indications for use have expanded<br>to include "medically refractory essential tremor" in TrueBeam™,<br>TrueBeam STx and Edge™ System. This submission does not include any<br>other device modification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Summary of Performance Testing: | The predicate and subject device have equivalent performance<br>specifications and have the same principle of operation. The results of the<br>verification, validation and safety standards testing (Predicate device -<br>K213977) demonstrates that there are no changes to the safety profile of<br>the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Summary of Clinical Data:       | Stereotactic Radiosurgery (SRS) has been used for decades to treat tremor<br>conditions. In a literature review, over 1300 patients were treated with SRS<br>for tremor conditions with similar outcomes and complication rates as<br>traditional surgical options. The attributes of advanced target localization,<br>beam collimation, and patient immobilization are common to both linear<br>accelerators and Gamma Knife, allowing for both devices to deliver<br>comparable doses to precisely-defined targets. Linear accelerators are in<br>routine clinical use for SRS treatments  and  recent publications report<br>tremor improvement and complication rates for linear accelerator-based<br>SRS comparable to those achieved with the Gamma Knife. The majority of<br>studies reporting on the use of SRS for the treatment of tremor conditions<br>include a discussion on the importance of advanced imaging and a<br>dedicated/experienced treatment team in order to properly localize,<br>immobilize, and treat these patients. In conclusion, linear accelerator-based<br>SRS for medically refractory essential tremor is reported to be a safe and<br>efficacious treatment option in situations where other treatment options<br>are not available or are otherwise contraindicated. |

# Argument for Substantial Equivalence to the Predicate Device

The subject device and predicate device have the same intended use of stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The expansion of the indications for use to include "medically refractory essential tremor" does not affect the design/principle of operation and does not introduce changes in the risk profile.

The technological characteristics and features of the subject device remain unchanged from the previous submission (Predicate device - K213977). There are no other modifications to predicate and subject device have equivalent performance specifications and have the same principle of operation. The verification and safety testing standards that support HET v3.0 (TrueBeam STx-Edge and VitalBeam) have been performed for the predicate device.

Based on the methods of evaluation and the data from the clinical literature, Varian believes that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K223839](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K223839)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
