C-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner

K211001 · Medtec, Inc. Dba Civco Medical Solutions and Civco · IYE · May 20, 2021 · Radiology

Device Facts

Record IDK211001
Device NameC-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner
ApplicantMedtec, Inc. Dba Civco Medical Solutions and Civco
Product CodeIYE · Radiology
Decision DateMay 20, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography (CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age.

Device Story

Patient positioning system for radiation therapy and imaging; consists of base, back support, bottom stop, and overhead arm positioner. Supports patient in supine position; allows angular adjustments (5° to 25°) for back support; provides arm support and head cushioning. Used in healthcare facilities by clinical staff. Facilitates patient immobilization during CT and MR imaging for treatment planning. Benefits include stable, reproducible patient positioning for radiation delivery and image acquisition. Device is reusable and non-sterile.

Clinical Evidence

No clinical data. Bench testing only. MR safety and compatibility testing performed per ASTM F2052-15, F2213-06, and F2119-07; device demonstrated safety at 1.5T and 3.0T. Biocompatibility testing performed per ISO 10993-5:2009 and ISO 10993-10:2010 for limited contact duration (<24 hours).

Technological Characteristics

Mechanical patient positioning assembly; includes base, back support, bottom stop, and overhead arm positioner. Materials evaluated for biocompatibility (ISO 10993). MR safety tested per ASTM F2052-15, F2213-06, F2119-07. Non-sterile, reusable. No electronic or software components.

Indications for Use

Indicated for adult and adolescent patients (12+ years) requiring support and positioning during radiation therapy (electron, photon, proton) or image acquisition (CT, MR) for the breast and chest region.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 20, 2021 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. MEDTEC, Inc. dba CIVCO Medical Solutions and CIVCO Radiotherapy % Ms. Alena Newgren Regulatory Manager 1401 8th Street SE ORANGE CITY IA 51041 ### Re: K211001 Trade/Device Name: C-Qual M" Breastboard; Monarch™ Overhead Arm Positioner Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE, LNH Dated: March 31, 2021 Received: April 2, 2021 Dear Ms. Newgren: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {1}------------------------------------------------ 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K211001 #### Device Name C-Qual M Breastboard with Monarch Overhead Arm Positioner Indications for Use (Describe) The device is indicated to aid in supporting adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography (CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font. The logo is simple and modern, and the blue color gives it a professional and trustworthy look. # 510(k) Summary K211001 ## A. Submitter Information | | Submitter Name & Address: MEDTEC, Inc. d/b/a "CIVCO Medical Solutions" and "CIVCO<br>Radiotherapy"<br>1401 8th St. SE<br>Orange City, IA 51041 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Alena Krishna, Regulatory Affairs Manager<br>Telephone: 319-248-6650<br>Alena.Newgren@civcort.com | | Date Summary Prepared: | March 31, 2021 | | Trade Names: | C-Qual M™ Breastboard; Monarch™ Overhead Arm Positioner | | Common Names: | Breastboard; Overhead Arm Positioner | | Classification Names &<br>Numbers: | Medical charged-particle radiation therapy system (892.5050)<br>System, Nuclear Magnetic Resonance Imaging (892.1000) | | Device Class: | Class II | | Review Panels: | Radiology | | Product Codes: | IYE, LNH | ### B. Predicate Devices The Breastboard and Overhead Arm Positioner are substantially equivalent to the following predicate device: | Predicate Device | Manufacturer | |----------------------------------------------------------------------------|--------------------| | K180021: Proton Positioning and Immobilization Devices<br>(Breast Devices) | CIVCO Radiotherapy | The purpose of this 510(k) is to add MR to the indications for use for the MTM410. ## C. Device Descriptions The Breastboard and Overhead Arm Positioner are used to position the patient during external beam radiation therapy. The Breastboard is comprised of a Base, Back Support, and Bottom Stop. The Base serves as the foundation of the assembly and provides a mounting surface for the Back Support, Angle Supports, and the Bottom Stop. The inferior end of the Back Support ## www.civcort.com {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font. The logo is simple and modern, and the blue color gives it a professional and trustworthy feel. is mounted to the Base while the superior end of the Back Support is raised/lowered as needed for patient setup using Low Angle Supports. The Bottom Stop is located near the inferior end of the Base and prevents the patient from sliding down the Board. The Low Angle Support is used to position the Back Support at 5°, 7.5°, 10°, 12.5°, and 15° anqles. Features on the High Angle Support receive the cutouts on the Low Angle Support and hold the anqular position of the Support. The High Angle Support is used to position the Back Support at 17.5°, 20°, 22.5°, and 25° angles. Features on the Base receive the cutouts on the High Angle Support and hold the angular position of the Support. A Lock provides a method of locking the Back Support to the Base at 0° to facilitate easier handling and storage of the Breastboard. The Overhead Arm Positioner provides support for the patient's arms above the head and is attached to the Back Support via a locking assembly. The Overhead Arm Positioner adjusts longitudinally on the Back Support to accommodate a variety of patient sizes. A Hand Grip is mounted on the Overhead Arm Positioner and the location of the Hand Grip is adjustable longitudinally. A Head Support rests on a recessed area of the Overhead Arm Positioner and provides a cushion for positioning of the head of the patient. The Head Support can also be mounted directly on the Breastboard. A Thermoplastic Frame may also be mounted to the Overhead Arm Positioner to enable attachment of a thermoplastic mask for additional patient support. The Breastboard and Overhead Arm Positioner can also be used for supporting and elevating patients undergoing treatment who cannot lay down flat on the table due to their habitus. The Breastboard and Overhead Arm Positioner are reusable devices that are provided nonsterile. The devices are used in a healthcare facility/hospital. The following model is included in this submission: | Part No. | Device Name | |----------|---------------------------------------------------------------| | MTM410 | C-Qual MRTM Breastboard and MonarchTM Overhead Arm Positioner | ### D. Indications for Use/Intended Use Statements Indications for Use/Intended Use: The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. Device is also used to position the patient during image acquisition to support treatment planning including in Computed Tomography ### www.civcort.com {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font. A blue circle is placed above the V. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font. (CT) and Magnetic Resonance (MR) imaging systems. The device is not intended for use with patients under 12 years of age. ### E. Comparison of Technological Characteristics There are no technological differences between the predicate and proposed device. Both the proposed and predicate Breastboard are comprised of a Base, Back Support, and Bottom Stop. Both devices consist of High Angle and Low Angle Supports that allow elevation from 5 to 25° in 2.5 degree increments. The Overhead Arm Positioner adjusts longitudinally on the Back Support via a locking assembly. ### F. Non-Clinical Testing To add MR indications to the Breastboard and Overhead Arm Positioner (MTM410), the device was tested for MR safety and compatibility in accordance with ASTM Standard F2052-15. Testing was also completed using ASTM Standards F2213-06 and F2119-07 as guidance. The device passed the acceptance criteria for magnetically induced torque and magnetically induced displacement force and demonstrate that the device is safe for use in field strengths of 1.5 T and 3.0 T. Image artifact was observed, with the worst case artifact occurring near the location of the locking knob on the Breastboard. Information regarding size of this artifact has been included in the Instructions for Use. In addition to MR safety testing, other nonclinical testing demonstrated that the differences in technological characteristics did not raise any new issues of safety or effectiveness. Biocompatibility testing was completed for patient-contacting materials in accordance with ISO 10993-5:2009 and ISO 10993-10:2010. The Breastboard and Overhead Arm Positioner is intended for limited contact duration (<24 hours) for surface devices (skin). ### G. Conclusion This premarket submission for the Breastboard and Overhead Arm Positioner has demonstrated substantial equivalence as defined and understood in the Federal Food. Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. ### www.civcort.com
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...