← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K172013

# VitalBeam (K172013)

_Varian Medical Systems, Inc. · IYE · Jul 28, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K172013

## Device Facts

- **Applicant:** Varian Medical Systems, Inc.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Jul 28, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

VitalBeam is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

## Device Story

VitalBeam is a medical linear accelerator (LINAC) delivering photon, electron, and kV X-ray radiation for stereotactic radiosurgery and precision radiotherapy. System comprises a radiation-shielded vault component and an external control console. Operated by clinicians in a clinical setting; device uses imaging inputs (CBCT) to guide radiation delivery. Features include HyperArc automated treatment, gated CBCT, 4D CBCT reconstruction, and Automatic Exposure Control (AEC) for kV imaging. Output is controlled radiation beams directed at lesions/tumors. Healthcare providers use the system to plan and execute precise radiation treatments, potentially improving tumor control while sparing healthy tissue.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation testing conducted per FDA Quality System Regulation (21 CFR 820), ISO 13485, and ISO 14971. Electrical safety and EMC testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2 standards.

## Technological Characteristics

Medical linear accelerator; photon, electron, and kV X-ray sources. Features: HyperArc, gated CBCT, 4D CBCT reconstruction, AEC. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-1, IEC 60825, IEC 61217, IEC 62274, IEC 62304, IEC 62366, ISO 10993-1. Software class: Major level of concern.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- TrueBeam Radiotherapy System and Accessories ([K162472](/device/K162472.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles facing right, suggesting health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2017

Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Varian Oncology Systems Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304

Re: K172013

Trade/Device Name: VitalBeam Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: June 30, 2017 Received: July 3, 2017

Dear Mr. Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172013

Device Name VitalBeam

#### Indications for Use (Describe)

VitalBeam is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, turnors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size:14px">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:14px">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Premarket Notification [510(k)] Summary

The following information follows the format of 21 CFR 807.92.

| I. Submitter's Name:             | Varian Medical Systems, Inc.<br>3120 Hansen Way C-260<br>Palo Alto, CA 94304<br><br>Contact Name: Peter J. Coronado<br>Phone: 650.424.5731<br>Fax: 650.842.5040<br>Date: July 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| II. Trade Name:                  | VitalBeam                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name:                     | Linear accelerator radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name:             | Medical charged-particle radiation therapy system<br>21 CFR 892.5050, Class II<br>Product Code: 90 IYE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| III. Predicate Device:           | TrueBeam Radiotherapy System and Accessories: K162472                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| IV. Device Description:          | VitalBeam is a medical linear accelerator.<br><br>The system consists of two major components, a photon, electron, and<br>diagnostic kV X-ray radiation beam-producing component that is<br>installed in a radiation-shielded vault and a control console area<br>located outside the treatment room.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| V. Intended Use<br>Statement     | VitalBeam is intended to provide stereotactic radiosurgery and precision<br>radiotherapy for lesions, tumors, and conditions anywhere in the body<br>where radiation treatment is indicated.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Indications for Use<br>Statement | VitalBeam may be used in the delivery of radiation for treatment that includes:<br>brain and spine tumors (such as glioma, meningioma, craniopharyngioma,<br>pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular<br>tumors, optic nerve tumors, and skull based tumors), head and neck tumors<br>(such as unknown primary of the head and neck, oral cavity, hypopharynx,<br>larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer),<br>thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and<br>mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial,<br>vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic,<br>hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as<br>prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid<br>tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as<br>squamous cell, basal cell, and melanoma), benign diseases (such as<br>schwannoma, arteriovenous malformation, cavernous malformation, trigeminal<br>neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including<br>all parts of the body such as brain, bone, liver, lung, kidney, and skin) and<br>pediatric tumors (such as glioma, ependymoma, pituitary tumors,<br>hemangioblastoma, craniopharyngioma, meningioma, metastasis,<br>medulloblastoma, nasopharyngeal tumors, arteriovenous malformation,<br>cavernous malformation and skull base tumors) |

{4}------------------------------------------------

## Significant changes to the predicate device are listed below.

# VI. Technological Characteristics:

| Feature                    | Cleared device | Device with change |
|----------------------------|----------------|--------------------|
| Enlarged Bounding Box      | No             | Yes                |
| Delta Couch shift          | No             | Yes                |
| Single Isocenter Multiplan | No             | Yes                |

Features cleared in TrueBeam™-TrueBeam STx™-Edge™ v2.7 (K162472) and included in VitalBeam v2.7:

- Hyperarc™ new automated treatment type ●
	- Cone Beam Computed Tomography (CBCT) changes:
	- Gated CBCT New CBCT acquisition technique O
	- Online 4D & Extended Length CBCT New advanced reconstructor -O imaging application integration
	- iCBCT New CBCT images reconstruction algorithms o
	- 4D CBCT Reconstruction New algorithms for binning CBCT projections o into different phases within existing 4D CBCT feature.
- Automatic Exposure Control (AEC) - New kV imaging system capability to control the kV exposure automatically
- Auxiliary Device Interface (ADI) 3.0 - New 3rd party device interface version
- Single Camera for Respiratory Gating with new algorithms ●

### VII. Summary of performance testing:

Hardware and software verification and validation bench testing was conducted according to the FDA Quality System Regulation (21 CFR §820). ISO 13485 QMS standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards listed below including electrical safety and EMC. Test results showed passing criteria was met for conformance to applicable requirements specifications and assured hazard safeguards functioned properly.

Software verification and validation testing were conducted and recorded as recommended by FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern. A failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

There was no change to the patient-contact materials biocompatibility in this medical device. Therefore, no change occurred in conformance to ANSI/AAMI/ ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1"

Electrical safety and electromagnetic compatibility (EMC) testing was conducted on this medical device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.

{5}------------------------------------------------

#### Varian VitalBeam™ medical linear accelerator model conforms in whole Standards or in part with the following FDA recognized consensus standards: conformance:

| AAMI/ANSI ES60601-1: 2005/<br>(R)2012 & A1:2012; C1:2009/<br>(R)2012 & A2:2010/ (R)2012<br>ANSI/AAMI/ISO 10993-1: 2009 | IEC 60601-1-2:2007<br>IEC 60601-1-3: 2013<br>IEC 60601-1-6: 2013<br>IEC 60601-2-1: 2014<br>IEC 60825: 2007 | IEC 61217: 2011<br>IEC 62274: 2005<br>IEC 62304: 2006<br>IEC 62366:2015<br>IEC 60976: 2007 |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|

#### Conclusion: The results of verification, validation and safety standards testing demonstrate that the VitalBeam medical linear accelerator is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K172013](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K172013)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
