← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K170345

# SOFTDISO (K170345)

_Best Nomos · IYE · Oct 27, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K170345

## Device Facts

- **Applicant:** Best Nomos
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Oct 27, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The SOFTDISO system is a standalone software solution designed to be used by medical physicists, radiation oncologists, and dosimetrists to have an overview of the treatment plans delivered from Treatment Systems (LINAC Systems using high energy x-rays) to the patient. This solution is intended to be used as a quality control check purposes only. SOFTDISO checks the correctness of the treatment by comparing the planned dose (from a treatment planning system) with the applied dose measured from the Electronic Portal Imaging Device (EPID) during treatment delivery and presents the results. SOFTDISO does not provide any conclusions or policy for interpretation of the results based on the comparison results data. The output from SOFTDISO cannot be and must not be used for changing the planning or treatment strategy or as a means of proving the effectiveness of the quality control process/chain during treatment. It is the physician's responsibility to verify the working of the quality control process/chain and the correctness of the dose delivered.

## Device Story

SOFTDISO is a standalone software for in-vivo dosimetric analysis (IVD) of 3DCRT, IMRT, and VMAT radiation therapy. It ingests patient treatment plans (via DICOM-RT) and transit signals from Si-EPID portal systems during treatment delivery. The system uses generalized correlation factors—derived from measurements of 70 Varian, Elekta, and Siemens linac beams—to reconstruct dose at the isocenter and perform gamma analysis between reference and current EPID images. Operated by medical physicists, oncologists, or dosimetrists in clinical settings, the software outputs dose comparison ratios (R ratio) and gamma analysis indices. These results are stored in an internal database and can be printed for the patient's clinical folder. The device assists clinicians in identifying discrepancies caused by patient setup, laser misalignment, or morphological changes (e.g., tumor shrinkage, weight loss), enabling proactive quality control. It does not provide clinical interpretations or treatment strategy recommendations; the physician remains responsible for verifying dose delivery accuracy.

## Clinical Evidence

Validation included non-clinical functional/performance testing and clinical validation. External validation across multiple sites for 3DCRT (192 tests), IMRT (48 tests), and VMAT (60 tests) showed R ratios within ±5% tolerance and gamma indices (γ% ≥ 95%, γmean ≤ 0.3). Clinical validation involved 1,287 treatments, confirming ease of use (avg 2.5 mins/patient). A study of 823 patients (340 3DCRT, 483 VMAT) demonstrated the system's ability to identify errors related to patient setup, laser misalignment, and morphological changes.

## Technological Characteristics

Standalone software for IVD. Uses generalized correlation factors between EPID transit signals and water-equivalent phantom measurements. Connectivity via DICOM and DICOM-RT protocols. Conforms to IEC 62304 (software life cycle) and BS EN 62366 (usability).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Dosimetry Check with Exit Dose ([K101503](/device/K101503.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 27, 2017

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Best NOMOS % Vineet Gupta, Ph.D. Director-R&D One Best Drive PITTSBURGH PA 15202

Re: K170345

Trade/Device Name: SOFTDISO Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 5, 2017 Received: October 10, 2017

Dear Dr. Gupta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Robert Oaks

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170345

Device Name

SOFTDISO

Indications for Use (Describe)

The SOFTDISO system is a standalone software solution designed to be used by medical physicists, radiation oncologists, and dosimetrists to have an overview of the treatment plans delivered from Treatment Systems using high energy x-rays) to the patient. This solution is to be used as a quality control check purposes only.

SOFTDISO does not provide any conclusions or policy for interpretation of the results based on the comparison results data. The output from SOFTDISO cannot be and must not be used for changing the planning or treatment strategy or as a means of proving the effectiveness of the quality control process/chain during treatment. It is the physician's responsibility to verify the working of the quality control process/chain and the correctness of the dose delivered.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

x | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Traditional 510(k) Summary

## A. SUBMITTERS NAME

Best NOMOS

## B. ADDRESS

One Best Drive Pittsburgh, PA 15202

#### C. CONTACT

| Name:  | Vineet Gupta, Ph.D.    |
|--------|------------------------|
| Phone: | (412) 312-6819         |
| Fax:   | (412) 312-6701         |
| Email: | vineet.gupta@nomos.com |

#### D. DATE PREPARED

Date: 5-Oct-2017

#### E. ESTABLISHMENT REGISTRATION NUMBER

Corporate Office Registration Number: 2434141 Manufacturing and Packaging Registration Number: 2434141

## F. DEVICE NAME

Device Trade Name: SOFTDISO Standalone Software Quality Control System Common Name: Medical Charged Particle Radiation Therapy System (21 CFR Classification Name: 892.5050)

#### G. DEVICE CLASS

Class II Panel: Radiology Product Code: IYE Regulation Number: 21CFR 892.5050

{4}------------------------------------------------

#### H. STATEMENT OF INDICATIONS FOR USE

The SOFTDISO system is a standalone software solution designed to be used by medical physicists, radiation oncologists, and dosimetrists to have an overview of the treatment plans delivered from Treatment Systems (LINAC Systems using high energy x-rays) to the patient. This solution is intended to be used as a quality control check purposes only.

SOFTDISO checks the correctness of the treatment by comparing the planned dose (from a treatment planning system) with the applied dose measured from the Electronic Portal Imaging Device (EPID) during treatment delivery and presents the results. SOFTDISO does not provide any conclusions or policy for interpretation of the results based on the comparison results data. The output from SOFTDISO cannot be and must not be used for changing the planning or treatment strategy or as a means of proving the effectiveness of the quality control process/chain during treatment. It is the physician's responsibility to verify the working of the quality control process/chain and the correctness of the dose delivered.

#### DEVICE DESCRIPTION I.

SOFTDISO software permit an IN-VIVO dosimetric analysis (IVD) in patient for 3DCRT, IMRT and VMAT beams using a Si-EPID portal systems.

The dosimetric analysis consists of two tests elaboration, dose reconstruction in the isocenter point and EPID images y-analysis. In case of out of tolerance levels, the system is capable of underline the right type of quality control useful to remove the causes of discrepancy.

SOFTDISO software is based on the usage of generalized correlation factors between transit signals measured by EPID and doses measured in water equivalent solid phantom, along the beam central axis.

SOFTDISO software uses generalized functions taken from measurements of 70 beams of Varian, Elekta and Siemens linacs. So only a small set of dosimetric measurements must be performed by the user for the software commissioning, and some of those measurements are already performed during the linac's beams calibration. In particular some of the measurements that the user performs include calculating the beam dose in cGy/UM in reference conditions (field 10.10 cm2 at drif =10 cm depth), the beam quality indicator (TPR20, 10) and the attenuation factor WF for wedged beams. The user also performs a measure of the EPID signal in reference condition (field 10.10 cm2 at SED distance) for EPID calibration.

SOFTDISO software can use information coming from record and verify systems through DICOM and DICOM-RT protocols. DICOM protocol is used for transferring images from CT and from EPID and the DICOM-RT protocol is used for transferring information related to the TPS system.

{5}------------------------------------------------

As soon as the treatment plan and the EPID images (fraction images for each day) are imported into the system, the user is able to visualize the comparison results.

At the end of fraction SOFTDISO provides result for two tests:

- Calculation of R ratio between in patient reconstructed dose and the one planned by the TPS
- Gamma analysis between a reference EPID image and a current one with Py<1 € Ymean . indexes

SOFTDISO can print the tests results and once signed by a Medical Physicist it can become part of the patient's clinical folder. The patient images and the comparison results are stored in the system's internal database for easy access at any time.

## J. PREDICATE DEVICE INFORMATION

The predicate devices identified within this submission are as follows:

- a. Math Resolution, LL, Dosimetry Check with Exit Dose (510(K) No. K101503 (Decision Date: 04-August-2010)).
## K. COMPARISON TO THE PREDICATE DEVICE

This section provides the summary of comparison of SOFTDISO to the predicate device.

|                        | Proposed Device<br>SOFTDISO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>Dosimetry Check with Exit Dose<br>(K101503)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The SOFTDISO system is a<br>standalone software solution designed<br>to be used by medical physicists,<br>radiation oncologists, and dosimetrists<br>to have an overview of the treatment<br>plans delivered from Treatment<br>Systems (LINAC Systems using high<br>energy x-rays) to the patient. This<br>solution is to be used as a quality<br>control check purposes only.<br><br>SOFTDISO does not provide any<br>conclusions or policy for<br>interpretation of the results based on<br>the comparison results data. The<br>output from SOFTDISO cannot be | The product is to be used by radiation<br>oncologist, dosimetrist, and radiation<br>therapy physicist to check the<br>correctness of the x-ray treatment<br>fields from high energy treatment<br>machines that are planned to be or<br>have been applied to a patient. This<br>product is to be used in addition to<br>the treatment planning system to<br>provide a means for additional and<br>redundant verification that the plan is<br>in fact successfully accomplished.<br><br>This product is not a treatment<br>planning system and is not to be used<br>as one. This product only checks the<br>applied dose based on the |

# Table 1 Indications for Use & Intended Use Comparison

{6}------------------------------------------------

| and must not be used for changing<br>the planning or treatment strategy or<br>as a means of proving the<br>effectiveness of the quality control<br>process/chain during treatment. It is<br>the physician's responsibility to<br>verify the working of the quality<br>control process/chain and the<br>correctness of the dose delivered. | measurement of each x-ray field and<br>a theoretical calculation. This product<br>does not provide any quality<br>assurance that the fields are. in fact<br>correctly applied to and correctly<br>aligned with the patient anatomy as<br>planned. In addition, the product may<br>be used to display the above dose on<br>other fused image sets which could<br>provide additional clinical<br>information to the radiation<br>oncologist regarding the treatment. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Comparison Item     | Proposed Device<br>SOFTDISO                                                 | Predicate Device<br>Dosimetry Check with Exit<br>Dose (K101503)             |
|---------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Classification Name | Quality Control For Medical<br>Charged Particle Radiation<br>Therapy System | Quality Control For Medical<br>Charged Particle Radiation<br>Therapy System |
| Product Code        | IYE                                                                         | IYE                                                                         |
| Class               | II                                                                          | II                                                                          |
| Regulation Number   | 892.5050                                                                    | 892.5050                                                                    |

#### Table 2 General Comparison

The indications for use for both the proposed device (SOFTDISO) and the Predicate Device (Dosimetry Check with Exit Dose) are similar. Both the devices are intended to be used as QA check to determine the correctness of the dose delivery by using the planned dose and delivered dose as input to the software to generate the report of differences. Both the devices use the planning images and exit images, and are designed to work only on photons (x-rays). Since the target population, indications for use, intended use, and the technical features of the proposed device (SOFTDISO) are similar to that of the predicate device, any minor difference in features do not raise any concerns for safety, performance, or effectiveness of the device; thus indicating that the proposed device is substantially equivalent to its predicate device. The characteristics/features of SOFTDISO with respect to the predicate device is described in detail the comparison chart in the submission. The similarities and differences are discussed in detail as part of this submission.

{7}------------------------------------------------

#### L. SUMMARY OF TESTING

Design verification and validation testing was performed to ensure that the device functionality works as per its intended use, all risks are mitigated, and the product conforms to the required standards. The details of the design verification and validation activities are explained in the submission.

#### Non-Clinical Testing:

Non-clinical testing included functional testing, performance testing, verification of risk control measures implemented in the software, and installation testing.

#### Validation:

External validation to validate the accuracy of dosimetry procedure for 3DCRT was performed at three different clinical sites. The results of R ratios for all 192 tests performed were within the acceptable tolerance of ± 5%. The y% and ymean indexes for all tests were ≥ 95% and ≤ 0.3 respectively.

To validate the accuracy of dosimetry procedure for IMRT and VMAT, external validation was performed at three and two different clinical sites respectively. The results of the R ratios for 48 tests for IMRT was well within the acceptable tolerance of ± 5%. The results of the R ratios for 60 tests for VMAT were well within the acceptable tolerance of ± 5%. The y% and Ymean indexes for all tests for IMRT and VMAT were ≥ 95% and ≤ 0.3 respectively.

Clinical validation was performed by applying in-vivo dosimetry checks on 1287 treatments. The results of the in-vivo dosimetry check indicated that SOFTDISO was easy to setup and use. The average analysis time was about 2.5 mins per patient (2 hours for an average of 50 patients/day).

Clinical validation was performed on 823 patient data who underwent 3DCRT (340) and VMAT (483) treatments. The results of the study indicated that the SOFTDISO system was able to identify errors that occur during treatment that were a result of inadequate quality controls (for example: patient setup, laser misalignment. TPS beam-implementation) and patient morphological changes (for example: tumor shrinkage, gas pockets, loss of patient weight).

#### Conclusion:

Detailed results of these tests are included as part of this submission. The verification and validation results demonstrate that the SOFTDISO system met its design requirements and specifications, is safe and effective to use for its intended purpose, and conforms to the applicable sections of standards that includes IEC 63204:2006: Medical device software - Software life cycle processes, BS EN 62366:2008: Medical devices -Application of usability engineering to medical devices.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K170345](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K170345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
