K162476 · Varian Medical Systems, Inc. · IYE · Dec 12, 2016 · Radiology
Device Facts
Record ID
K162476
Device Name
Varian High Energy Linear Accelerator
Applicant
Varian Medical Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Dec 12, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated
Device Story
Radiotherapy treatment unit consisting of gantry, couch, stand, and control console; permanently installed. Generates radiation beam via electron gun, waveguide, and collimator. Modified to include yield monitoring capability: detects beam deviations via target hardware anomalies; triggers interlock to prevent treatment if 6X, 6FFF, or 6SRS beams fall outside output limits. Adds firmware check for Primary/Secondary Position Readout (PRO/SPRO) polarity mismatch across motion axes (jaws, gantry, couch, collimator). Operated by clinical staff in radiation oncology settings. Output used for precision dose delivery; yield monitoring and polarity checks enhance safety by preventing undetected dose decreases or positioning errors. Benefits patient through increased treatment accuracy and safety.
Clinical Evidence
No clinical data included in this submission. Safety and effectiveness supported by non-clinical bench testing and software verification/validation.
Technological Characteristics
Medical charged-particle radiation therapy system. Components: gantry, couch, stand, control console. Beam generation: electron gun, waveguide, collimator. Features: dual ion chamber dosimetry, yield monitoring interlock, PRO/SPRO firmware polarity checks. Connectivity: networked control console. Software: major level of concern.
Indications for Use
Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body requiring radiation treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K132705 — UNIQUE · Varian Medical Systems, Inc. · Oct 31, 2013
K131807 — HIGH ENERGY LINEAR ACCELERATOR · Varian Medical Systems, Inc. · Oct 11, 2013
K091132 — 4D INTEGRATED TREATMENT CONSOLE · Varian Medical Systems · Jun 17, 2009
K231317 — TrueBeam, TrueBeam STx, Edge, VitalBeam · Varian Medical Systems, Inc. · Sep 12, 2023
K112839 — VARIAN HIGH ENERGY LINEAR ACCELERATOR · Varian Medical Systems · Nov 14, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three wave-like lines below them. The profiles and waves are black, and the background is white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2016
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K162476
Trade/Device Name: Varian High Energy Linear Accelerator Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 10, 2016 Received: November 14, 2016
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162476
## Device Name
Varian High Energy Linear Accelerator
Indications for Use (Describe)
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 201 Subpart D) | |
|----------------------------------------------|--|
| Over-The-Counter Use (21 CFR 201 Subpart C) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 USA Tel +1 650 493 4000 www.varian.com
# Premarket Notification [510(k)] Summary
## Varian High Energy Linear Accelerator
The following information is provided following the format of 21 CFR 807.92(c).
| Submitter's Name: | Varian Medical Systems, Inc.<br>3100 Hansen Way E-110<br>Palo Alto, CA 94304<br><br>Contact Name: Peter J. Coronado<br>Phone: 650.424.6320<br>Fax: 650.646.9200 |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Date: September 2, 2016 |
| Name of the Device:<br>Trade/ Proprietary<br>Names: | Varian High Energy Linear Accelerator<br>Novalis Tx, Trilogy, Trilogy Tx<br>Clinac iX, Clinac Cx<br>Clinac 2100C, 2100 C/D, 2300 C/D<br>Clinac 21 EX, 23 EX<br>Clinac DHX, DMX |
| Classification Name: | Medical charged-particle radiation therapy system<br>21 CFR 892.5050, Class II<br>Product Code: IYE |
| Common/Usual<br>Name: | Medical Linear Accelerator |
| Predicate Device: | Varian High Energy Linear Accelerator K131807 |
| Device Description: | The Varian High Energy Linear Accelerator models provide various selections among the<br>features, specifications, and accessories that have been most recently cleared as the<br>Varian High Energy Linear Accelerator, K131807.<br><br>The High Energy Linear Accelerator is a radiotherapy treatment unit. The equipment<br>consists of a gantry, couch, stand and control console. The device is permanently<br>installed. The radiotherapy treatment beam is generated by a linear accelerator assembly<br>consisting of an electron gun, waveguide and collimator.<br><br>The High Energy (HE) Linear Accelerator device is modified to provide a yield monitoring<br>capability, which triggers an interlock in response to detected beam deviations when<br>specific target hardware anomalies are present. The yield monitoring capability provides<br>an additional check for unexpected changes in radiation beam output, beyond the<br>existing beam checks and dosimetry interlocks provided by the dual ion chamber design.<br>The purpose of the yield monitoring capability is to prevent an undetected decrease in<br>dose output, which can occur in the case of failures within the target hardware. If such<br>hardware failures are present, the yield monitoring capability triggers an interlock and<br>treatment is prevented if 6X, 6FFF, or 6SRS beams do not meet specified output limits.<br>For additional detail please see the Description of Modifications included in the |
| | submission. |
| | The Primary Position Readout (PRO) and Secondary Position Readout (SPRO) are<br>enhanced by the addition of a firmware check for polarity mismatch. The function of the<br>PRO and SPRO remains unchanged. The enhancement provides a secondary check of the<br>positions reported by the PRO and SPRO for all motion axes: jaws, gantry, couch and<br>collimator. The firmware enhancement is an improvement that verifies the polarity of<br>both the PRO and SPRO components, and notifies the user in the case of a polarity<br>mismatch. Notifications continue until the polarity has been corrected. |
| | All other features and technological characteristics of the Varian High Energy Linear<br>Accelerator models remain as cleared by K131807. |
| Intended Use<br>Statement | The Varian High Energy Linear Accelerator is intended to provide stereotactic<br>radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in<br>the body when radiation treatment is indicated |
| Indications for Use<br>Statement | The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and<br>precision radiotherapy for lesions, tumors and conditions anywhere in the body when<br>radiation treatment is indicated. |
{4}------------------------------------------------
## Technological Characteristics:
### Performance Data:
#### Software Verification and Validation Testing
Software verification and validation was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
Clinical Tests No clinical tests have been included in this pre-market submission
#### Conclusions
The non-clinical data support the safety of the device and the software verification and validation demonstrate that the High Energy Linear Accelerator device with the new yield interlock functionality and firmware check for polarity mismatch performs as intended. Varian therefore considers the High Energy Linear Accelerator to be safe and effective and to perform at least as well as the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.