Browse hierarchy Radiology (RA) Subpart F — Therapeutic Devices 21 CFR 892.5050 Product Code IYE K151426 — ScandiDos Delta4 Discover, Alternative Trade Name: ScandiDos Delta4 Discover+
ScandiDos Delta4 Discover, Alternative Trade Name: ScandiDos Delta4 Discover+
K151426 · Scandidos AB · IYE · Jan 29, 2016 · Radiology
Device Facts
Record ID K151426
Device Name ScandiDos Delta4 Discover, Alternative Trade Name: ScandiDos Delta4 Discover+
Applicant Scandidos AB
Product Code IYE · Radiology
Decision Date Jan 29, 2016
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.5050
Device Class Class 2
Indications for Use
quality assurance of patient specific treatment delivery during and before external radiotherapy treatment, including IMRT, VMAT and 4DRT (respiratory gating and tumour tracking) quality assurance of the radiation delivery system.
Device Story
Device is a 2D semiconductor detector matrix for radiotherapy quality assurance (QA). Detector matrix is placed in the radiation field of a medical linear accelerator; radiation exposure generates signals; signals are transferred to a computer for analysis. Software compares measured dose distributions against planned treatment distributions. Used by medical physicists/technicians in clinical radiotherapy settings to verify treatment plan delivery and accelerator performance. Provides quantitative and subjective assessments of dose distribution; enables verification of respiratory gating and tumor tracking. Benefits include improved accuracy of radiation therapy delivery and validation of treatment planning systems.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison and bench testing of performance characteristics.
Technological Characteristics
2D semiconductor detector matrix; acquisition, display, and analysis software. Designed to IEC 60601-1 and IEC 60601-1-2 standards. Operates by detecting radiation-induced signals in the linear accelerator field. Connectivity via computer interface for data processing.
Indications for Use
Indicated for quality assurance of patient-specific external radiotherapy treatment delivery (IMRT, VMAT, 4DRT) and radiation delivery system performance verification.
Regulatory Classification
Identification A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Related Devices
K183606 — Delta4 Phantom+ MR (PMMA), Delta4 Phantom+ MR (PW) · Scandidos AB · Apr 8, 2019
K162629 — IQM Integral Quality Monitor (also IQM) · Irt Systems · Oct 20, 2016
K151180 — ScandiDos Delta4 Phantom+ · Scandidos AB · Aug 19, 2015
K030175 — 2 BASIC · Capintec, Inc. · Apr 17, 2003
K062817 — PTW DAVID IMRT QC VERIFICATION SYSTEM · Ptw-New York Corp. · Nov 22, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2016
ScandiDos AB % Mr. Thomas Matzen Product Manager Dag Hammarskjolds vag 52A Uppsala, SE-75237 Uppsala Lan SWEDEN
Re: K151426
Trade/Device Name: "Delta Discover, and "Delta" Discover+" Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: December 28, 2015 Received: December 28, 2015
Dear Mr. Matzen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a sequence of handwritten numbers and letters. The sequence starts with "D012", followed by "04", then "045", and ends with "01". All the characters are written in blue ink and appear to be on a white background.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K151426
Device Name "Delta4 Discover" and "Delta4 Discover+"
Indications for Use (Describe)
The intended use of the device is
· quality assurance of patient specific treatment delivery during and before external radiotherapy treatment, including IMRT, VMAT and 4DRT (respiratory gating and tumour tracking)
· quality assurance of the radiation delivery system.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 5px;">☒</span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span style="padding-left: 20px;"> <span style="padding-right: 5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
ScandiDos
© ScandiDos AB
This document may not be copied without the written consent of ScandiDos AB.
Confidential
| Document Title: | FDA 510(k) - Summary - Delta4 Discover | Document ID: | Page |
|-----------------|----------------------------------------|----------------|---------------------|
| Subject: | Delta4 Discover | D012 04 016 01 | 002_06_1 (002_06_2) |
| Author: | Signature | Date | Thomas Matzen |
D012 04 016 01 FDA 510(k) - Summary - Delta4 Discover.docx
| Version | Comment | Author | Description |
|---------|----------|---------------|---------------|
| 01 | Released | Thomas Matzen | First release |
| | | | |
## 1 APPLICANT
| 510(k) owner's Name | ScandiDos AB |
|-----------------------------------------------|---------------------------|
| Street address | Dag Hammarskjölds väg 52A |
| Postal Code, City | SE-75237 Uppsala |
| Country | Sweden |
| Phone Number | +46-18-472 30 30 |
| Fax Number | +46-18-10 74 02 |
| Name of official correspondent | Thomas Matzen |
| Date the summary<br>was prepared (yyyy-mm-dd) | 2015-05-20 |
### 2 DEVICE
| Trade Name | "Delta4 Discover" and "Delta4 Discover+" |
|----------------------------|------------------------------------------------------------------------------|
| Common/Usual Name | QA device for verification of radiotherapy treatments and<br>treatment plans |
| Device Classification Name | Medical charged-particle radiation therapy system<br>(§892.5050) |
| Product code | IYE |
## 3 PREDICATE DEVICES
| Predicate Devices | Registered<br>Establishment Number | 510(k) Premarket submission<br>number |
|------------------------------------------|------------------------------------|---------------------------------------|
| PTW DAVID IMRT QC<br>verification system | 9681124 | K062817 |
| IBA Compass | 9614952 | K072374 |
{4}------------------------------------------------
| | ScandiDos | © ScandiDos AB | |
|-----------------|----------------------------------------|----------------|------------------------|
| Document Title: | FDA 510(k) - Summary - Delta4 Discover | Document ID: | Page |
| | | D012 04 016 01 | 002_06_2<br>(002 06 2) |
#### 4 DESCRIPTION OF THE DEVICE
The device is an instrument specifically designed to check delivered treatment plans by medical accelerator systems used for radiation therapy applications for quality assurance (QA) purposes.
The device consists of a 2-dimensional matrix of semiconductors and associated acquisition, display and analysis computer programs. It is a quality assurance (QA) device enabling detailed mapping of therapeutic radiation dose distributions. This dose information is used in both quantitative and subjective assessments of the performance of radiation therapy treatment planning systems and therapeutic radiation delivery systems.
The detector matrix is inserted into the radiation field of a medical linear accelerator. If radiation (from a radiotherapy treatment field) hits the detectors a signal is created and transferred to a computer where it is evaluated (e.g. by comparing planned with measured dose distributions).
#### 5 INTENDED USE OF THE DEVICE
The intended use of the device is
- o quality assurance of patient specific treatment delivery during and before external radiotherapy treatment, including IMRT, VMAT and 4DRT (respiratory gating and tumour tracking)
- quality assurance of the radiation delivery system. .
## 6
# SAFETY AND ESSENTIAL PERFORMANCE
The new device is designed in full accordance with the applicable sections of the following standards:
- IEC 60601-1
- . IEC 60601-1-2
## 7
# TECHNOLOGICAL COMPARISON WITH THE PREDICATE DEVICES
All devices are intended for quality assurance of patient specific treatment delivery and quality assurance of the radiation delivery system.
Based on the technological characteristics, the intended use, safety and effectiveness, essential performance, performance the new device is substantially equivalent to the predicate devices.
The documentation submitted for review supports this claim.
### 8 SUMMARY
The new device is superior or at least equivalent with the predicate devices regarding safety, design and performance.