← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K150706

# VitalBeam (K150706)

_Varian Medical Systems, Inc. · IYE · May 8, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K150706

## Device Facts

- **Applicant:** Varian Medical Systems, Inc.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** May 8, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The VitalBeam™ system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

## Device Story

VitalBeam Radiotherapy Delivery System is a medical linear accelerator for stereotactic radiosurgery and precision radiotherapy; derived from TrueBeam system. System comprises beam-producing component in radiation-shielded vault and control console outside vault. Operated by clinicians to deliver radiation to tumors/lesions throughout body. Device enables precise radiation delivery; output used by healthcare providers to treat malignancies and benign conditions. Benefits include targeted therapy for wide range of oncological and non-oncological indications.

## Clinical Evidence

Bench testing only. Verification and validation testing demonstrated conformance to requirements specifications and proper functioning of hazard safeguards. No clinical data presented.

## Technological Characteristics

Medical linear accelerator; radiation-shielded vault installation. Conforms to IEC 60601-1, 60601-2-1, 60601-2-29, 60601-2-44, 60825, 61217, 62274, 62366, 62304, and 60976 standards. Software development follows IEC 62304.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- TrueBeam Radiotherapy System and Accessories ([K143224](/device/K143224.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2015

Varian Medical Systems, Inc. % Mr. Peter J. Coronado Director, Varian Oncology Systems Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304

Re: K150706

Trade/Device Name: VitalBeam Radiotherapy Delivery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: April 7, 2015 Received: April 9, 2015

Dear Mr. Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

for

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150706

#### Device Name

VitalBeam Radiotherapy Delivery System

#### Indications for Use (Describe)

The VitalBeam System is intended to provide stereotactic radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

The VitalBeam System may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial), gastrontestinal tumors (such as gastric, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid turnors (such as Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavemous malformation, trigeninal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pitutary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Premarket Notification [510(k)] Summary VitalBeam Radiotherapy Treatment System

The following information is provided following the format of 21 CFR 807.92.

| I. Submitter's Name:             | Varian Medical Systems, Inc.<br>3100 Hansen Way C-260<br>Palo Alto, CA 94304<br>Contact Name: Peter J. Coronado<br>Phone: 650.424.5731<br>Fax: 650.842.9200<br>Date: March 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| II. Trade Name:                  | VitalBeam™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common Name:                     | Linear accelerator radiation therapy system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name:             | Medical charged-particle radiation therapy system<br>21 CFR 892.5050, Class II<br>Product Code: 90 IYE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| III. Predicate Device:           | TrueBeam Radiotherapy System and Accessories: K143224                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| IV. Device Description:          | The VitalBeam™ Radiotherapy Delivery System is a medical linear<br>accelerator based on the previously cleared TrueBeam Radiotherapy<br>system and associated accessories.<br><br>The system consists of two major components: a beam-producing<br>component that is installed in a radiation-shielded vault (hereafter<br>referred to as the treatment room) and a control console located outside<br>the treatment room.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| V. Intended Use<br>Statement     | The VitalBeam™ system is intended to provide stereotactic radiosurgery<br>and precision radiotherapy for lesions, tumors, and conditions anywhere in<br>the body where radiation treatment is indicated.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indications for Use<br>Statement | The VitalBeam System is intended to provide stereotactic radiosurgery and<br>precision radiotherapy for lesions, tumors, and conditions anywhere in the<br>body where radiation therapy is indicated for adults and pediatric patients.<br><br>The VitalBeam System may be used in the delivery of radiation for<br>treatment that includes: brain and spine tumors (such as glioma,<br>meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors,<br>hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and<br>skull based tumors), head and neck tumors (such as unknown primary of<br>the head and neck, oral cavity, hypopharynx, larynx, oropharynx,<br>nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic<br>tumors (such as lung cancer, esophageal cancer, thymic tumors, and<br>mesothelioma), gynecologic tumors (such as ovarian, cervical,<br>endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric,<br>pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary<br>tumors (such as prostate, bladder, testicular, and kidney), breast tumors,<br>sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's |

{5}------------------------------------------------

lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).

#### VI. Technological Comparisons to the predicate device are listed below: Characteristics:

| Feature                                         | Cleared device | Device with change |
|-------------------------------------------------|----------------|--------------------|
| Bar-code Conical Collimator Verification (BCCV) | Yes            | No                 |
| Patient, Accessory Verification System (PAVS)   | Yes            | No                 |
| High Definition 120-leaf collimator             | Yes            | No                 |

# VII. Summary of performance testing:

Results of verification and validation testing showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly.

The Varian VitalBeam™ medical linear accelerator conforms in whole Standards conformance: or in part with the following FDA recognized consensus standards:

| AAMI/ANSI/IEC 60601-1:2005 | IEC 60601-2-1: 2009  |
|----------------------------|----------------------|
| IEC 60601-1-2:2007         | IEC 60601-2-29: 2008 |
| IEC 60601-1-3: 2008        | IEC 60601-2-44: 2009 |
| IEC 60601-1-6:2010         | IEC 60976: 2007      |
| IEC 60825: 2007            | IEC 62274: 2005      |
| IEC 61217: 2011            | IEC 62366:2007       |
| IEC 62304: 2006            |                      |

Conclusion: Based on the verification and validation non-clinical testing, the Varian VitalBeam™ medical linear accelerator is as safe, effective and perform as well as or better than the legally marketed device identified in section III above.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K150706](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K150706)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
