← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K131807

# HIGH ENERGY LINEAR ACCELERATOR (K131807)

_Varian Medical Systems, Inc. · IYE · Oct 11, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K131807

## Device Facts

- **Applicant:** Varian Medical Systems, Inc.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Oct 11, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

## Device Story

Radiotherapy treatment unit comprising gantry, couch, stand, and control console; permanently installed. Generates radiation beam via electron gun, waveguide, and collimator. Operated by clinical staff to deliver stereotactic radiosurgery and precision radiotherapy. New functionality introduces 'motion zone rules' to limit couch and gantry movement initiated from outside the treatment room, reducing collision risk between gantry and patient. System utilizes interlocks to prevent or terminate beam-on unless treatment parameters and operating conditions are safe. Healthcare providers use the system to target lesions/tumors; output is controlled radiation dose delivery. Benefits include precise treatment delivery with enhanced safety protocols regarding mechanical motion.

## Clinical Evidence

Bench testing only. Verification and validation activities were performed to confirm that the modified system, including the new motion zone rules and updated control software (C-Series 9.1), met all acceptance criteria and safety requirements.

## Technological Characteristics

Medical linear accelerator; permanently installed. Components: gantry, couch, stand, control console, electron gun, waveguide, collimator. Software: C-Series 9.1. Features motion zone rules for collision avoidance. Energy source: high-energy photon beams. Dose delivery rates up to 2400 MU/min (FFF).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Varian High Energy Linear Accelerator ([K112839](/device/K112839.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K131807
Page 1 of 4

ww.varian.com

June 17, 2013

## 510(k) Summary

CT 11 2013

The information below is provided for the Varian High Energy Linear Accelerator. following the format of 21 CFR 807.92.

- 1. 510(k) Owner: Varian Medical Systems 3100 Hansen Way, M/S C 260 Palo Alto, CA 94304 Contact Name: Peter J. Coronado - Director, Regulatory Affairs Phone: 650/424.6320 Fax: 650/842.5040 E-mail: submissions.support@varian.com

| 2. Name of the Device:   | Varian High Energy Linear Accelerator                                                                                                 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Proprietary Names: | Novalis Tx, Trilogy. Trilogy Tx<br>Clinac iX, Clinac Cx<br>Clinac 2100C, 2100 C/D, 2300 C/D<br>Clinac 21 EX, 23 EX<br>Clinac DHX. DMX |
| Common Name:             | Medical Linear Accelerator                                                                                                            |
| Classification Name:     | Medical Charged Particle Radiation Therapy System<br>21 CFR §892.5050<br>Class II<br>Product Code 90 IYE                              |

- 3. Predicate Device: Varian High Energy Linear Accelerator K112839
- 4. Description of the Device:

The Varian High Energy Linear Accelerator models provide various selections among the features, specifications, and accessories that have been most recently cleared as the Varian High Energy Linear Accelerator, K112839.

The High Energy Linear Accelerator is a radiotherapy treatment unit. The equipment consists of a gantry, couch, stand and control console. The device is permanently installed. The radiotherapy treatment beam is generated by a linear accelerator assembly consisting of an electron gun, waveguide and collimator.

{1}------------------------------------------------

An extensive system of interlocks is designed to prevent or terminate beam-on unless essential treatment parameters are in place and system operating conditions relevant to safe operation are correct.

The changes to the Varian High Energy Linear Accelerator establish motion zone rules, which limit the motions of couch and gantry that can be initiated or directed from outside the treatment room. The purpose of the zone rules is to increase patient safety by reducing the possibility of collision between the gantry and the patient.

All other features and technological characteristics of the Varian High Energy Linear Accelerator models remain as cleared by K112839.

- 5. Intended Use Statement
The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

- 6. Indications for Use Statement
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions. tumors and conditions anywhere in the body when radiation treatment is indicated.

- 7. Substantial Equivalence
The modified device, the Varian High Energy Linear Accelerator. is substantially equivalent to the predicate device, the Varian High Energy Linear Accelerator (K112839).

The functionality of the Varian High Energy Linear Accelerator is equivalent to the functionality of the Varian predicate device in safety and effectiveness. The design control procedures applied to the development of the Varian High Energy Linear Accelerator and its modifications include requirements reviews, risk analysis, and verification and validation testing. The results of verification and validation activities demonstrate that the acceptance criteria have been met.

Compared with the predicate device, the Varian High Energy Linear Accelerator (K112839), the basic operation and technological characteristics are the same. Operational differences are described in the Instructions for Use for the Varian High Energy Linear Accelerator. The intended use for the device is unchanged.

A comparison table illustrating the substantial equivalence of the modified device to the predicate device appears below.

{2}------------------------------------------------

| FEATURE AND/OR<br>SPECIFICATION                                                                                                                      | CLEARED DEVICE<br>(HIGH ENERGY LINEAR<br>ACCELERATOR K112839)                                                                                                                                                                                | DEVICE WITH CHANGE<br>(HIGH ENERGY LINEAR<br>ACCELERATOR)                                 |
|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Intended Use                                                                                                                                         | The Varian High Energy<br>Linear Accelerator is<br>intended to provide<br>stereotactic radiosurgery and<br>precision radiotherapy for<br>lesions, tumors and<br>conditions anywhere in the<br>body when radiation<br>treatment is indicated. | Unchanged                                                                                 |
| Indications for use                                                                                                                                  | The Varian High Energy<br>Linear Accelerator is<br>intended to provide<br>stereotactic radiosurgery and<br>precision radiotherapy for<br>lesions, tumors and<br>conditions anywhere in the<br>body when radiation<br>treatment is indicated. | Unchanged                                                                                 |
| Release version of control<br>software                                                                                                               | C-Series 9.0                                                                                                                                                                                                                                 | C-Series 9.1                                                                              |
| Gantry and couch rotation<br>from outside the treatment<br>room are prevented or<br>limited within Exclusion<br>Zone areas defined by Zone<br>Rules. | No                                                                                                                                                                                                                                           | Yes                                                                                       |
| Maximum rate at which dose<br>delivery occurs for 10x<br>photon energy                                                                               | 600 MU per minute                                                                                                                                                                                                                            | 600 MU per minute                                                                         |
| Maximum rate at which dose<br>delivery occurs for 10x FFF<br>photon energy                                                                           | 2400 MU per minute                                                                                                                                                                                                                           | 2400 MU per minute                                                                        |
| Maximum rate at which dose<br>delivery occurs for 6x SRS<br>photon energy                                                                            | 1000 MU per minute                                                                                                                                                                                                                           | 1000 MU per minute                                                                        |
| Maximum rate at which dose<br>delivery occurs for 6x FFF<br>photon energy                                                                            | 1400 MU per minute                                                                                                                                                                                                                           | 1400 MU per minute                                                                        |
| FEATURE AND/OR<br>SPECIFICATION                                                                                                                      | CLEARED DEVICE<br>(HIGH ENERGY LINEAR<br>ACCELERATOR K112839)                                                                                                                                                                                | DEVICE WITH CHANGE<br>(HIGH ENERGY LINEAR<br>ACCELERATOR)                                 |
| Maximum field size                                                                                                                                   | 3D Conformal Radiation<br>Therapy: 40cm x 40cm.<br>IMRT: 34cm x 40cm. SRS:<br>15cm x 15cm                                                                                                                                                    | 3D Conformal Radiation<br>Therapy: 40cm x 40cm.<br>IMRT: 34cm x 40cm. SRS:<br>15cm x 15cm |
| Maximum allowable dose<br>limit for fixed X treatment<br>type (for non-SRS and non-<br>FFF treatment types)                                          | 1999 MU                                                                                                                                                                                                                                      | 1999 MU                                                                                   |
| Maximum programmable<br>dose                                                                                                                         | 9999 MU                                                                                                                                                                                                                                      | 9999 MU                                                                                   |

{3}------------------------------------------------

K131807
Page 4 of 4

.

.

·

 .

·

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with text around the perimeter and a stylized symbol in the center. The symbol appears to be an abstract representation of an eagle or bird-like figure.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Cuntrol Center - WO66-G609 Silver Spring, MI) 20993-0002

October 11. 2013

Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Global Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304

Re: K131807

Trade/Device Name: Varian High Energy Linear Accelerator Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle.radiation therapy system Regulatory Class: 11 Product Code: IYE Dated: October 2, 2013 Received: October 8, 2013

Dear Mr. Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practive requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Peter Coronado

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known): K131807

Device Name: Varian High Energy Linear Accelerator

Indications for Use:

The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of In Vitro Diagnostics and Radiological Health (OIR)

Michael D. O'Hara

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

> K131807 510(k)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K131807](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K131807)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
