K122616 · Prowess, Inc. · IYE · Dec 21, 2012 · Radiology
Device Facts
Record ID
K122616
Device Name
PANTHER OIS/R & V MODEL VERSION 2.0
Applicant
Prowess, Inc.
Product Code
IYE · Radiology
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Panther OIS|R&V is an information management system used to manage medical data and clinical workflow in a hospital or clinic. To support radiation oncology users, it allows the user to: - Enter or import, modify, store and archive treatment plans and images from . treatment planning systems. - Import, view, manipulate, enhance, annotate, store, and archive radiological . images. - Select and provide radiation treatment plans to a radiation treatment delivery . system for patient treatment. - Store and view treatment records provided by the radiation treatment delivery . system. - Monitor and track treatment progress. - When Panther OIS|R&V is acting as an R&V system, verify the delivery device's . settings against the planned parameters, record the treatment parameters settings and prevent treatment if the parameters are out of tolerance.
Device Story
Panther OIS|R&V is a Windows-based oncology information and record-and-verify (R&V) system. It ingests treatment plans (DICOM/RTP) and radiological images from treatment planning systems. The software manages patient schedules, resource scheduling, and central storage of treatment records. In R&V mode, it interfaces with linear accelerators to download treatment beams, verify delivery parameters against the plan, and record delivered treatments. It prevents treatment delivery if parameters exceed defined tolerances. Operated by radiation oncology staff in clinical settings, the system provides a centralized database for administrative and clinical tasks. It enhances clinical decision-making by ensuring treatment accuracy through automated verification and data locking to prevent unauthorized or simultaneous access. Benefits include optimized workflow, improved treatment quality, and safeguarded patient health information.
Clinical Evidence
No clinical data required for substantial equivalence. Manufacturer conducted in-house verification and validation testing against predetermined specifications. Beta/external testing was performed at a clinical user site (Community Memorial Hospital) to evaluate device performance in a real-world clinical setting.
Technological Characteristics
Windows-based software system; utilizes a centralized database for multi-client access. Complies with ISO 13485, ISO 14971:2007 (risk management), IEC 62304 (software life cycle), and IEC 62274 (radiotherapy R&V systems). Features include role-based access control, patient ID photo verification, and data locking. Connectivity via DICOM/RTP standards.
Indications for Use
Indicated for use in hospital or clinical settings to manage medical data and clinical workflow for radiation oncology patients. Supports management of treatment plans, radiological images, and radiation treatment delivery records; verifies delivery device settings against planned parameters to prevent treatment if out of tolerance.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
IMPAC Medical Systems, Inc's (Elekta) MOSAIQ (K120067)
Varian Medical System Inc.'s ARIA Radiation Oncology (K093527)
Varian Medical System's 4D Integrated Treatment Console (4DITC) (K091132)
K133572 — ARIA RADIATION THERAPY MANAGEMENT · Varian Medical Systems, Inc. · Apr 4, 2014
K230699 — ARIA Radiation Therapy Management (v18.0) · Varian Medical Systems, Inc. · Apr 10, 2023
K132935 — ARTISTE SOLUTION WITH SYS_VC10C AND SYNGO RT THERAPIST WORKSPACE, V4.3.1 UPGRADE KIT, ARIA CONNECTIVITY OPTION FOR SIEME · Siemens Medical Solutions USA, Inc. · Feb 21, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K122616
Contact Person:
## 510(k) Summary
## DEC 2 1 2012
くし
### As required by 21 CFR Part 807.92
August 23, 2012
1. Submitter: Prowess, Inc. 1844 Clayton Road Concord, CA. 94520
> Rachel Scarano Regulatory Affairs Manager Prowess, Inc. 1844 Clayton Road Concord, CA. 94520 Phone: (925) 356-0360
> > Fax: (925) 356-0363
E-mail: rachel.scarano@prowess.com
Device Manufacturer: Prowess, Inc. 1844 Clayton Road Concord, CA. 94520
2. Device Trade Name: Panther OIS|R&V
Classification Name:
Medical charged-particle radiation therapy system (21 CFR § 892.5050), Product Code MUJ, Class II
Establishment Reg. No.: 2939248
Common Name: Oncology Information System Radiation Therapy Record and Verify System
#### 3. Predicate Devices:
- IMPAC Medical Systems, Inc's (Elekta) MOSAIQ K120067 a)
- Varian Medical System Inc.'s ARIA Radiation Oncology K093527 b)
- Varian Medical System's 4D Integrated Treatment Console (4DITC) K091132 c)
- d) Prowess Inc's Puma (version 1.0) K100801 (*as previously cleared)
#### Device Description 4.
Panther OIS|R&V is medical device software which allows the user to deliver radiation treatment on linear accelerator to a patient. The software has been developed to be a record and verify system that also manages oncology medical information in order to simplify the
{1}------------------------------------------------
oncologist administrative tasks and to optimize the quality of patient treatment. It is a Windows® based system, which uses a single database that can be accessed by any Panther OIS|R&V client station in the treatment facility. Additionally, Panther OIS|R&V is capable of performing such functions as:
- lmporting treatment plans in DICOM or RTP format -
- Downloading of beams contained in the treatment plans to the linear accelerator ﺖ
- Verifying that the beams from the treatment plan are correctly set up on the linear accelerator
- Recording treatments delivered by the linear accelerator -
- Managing patient schedules ・
- Managing schedules for resources such as treatment machines and personnel "
- Maintaining a central repository of treatment plans and records -
- Documenting, monitoring and updating treatments and treatment plan information -
- Performing administrative tasks such as patient management. -
- Data locking or warning to prevent simultaneous access to patient (Tx) data -
- -Reports of patients and treatment history
- Manual Recording of treatment beams ﺖ
- Late Resumption of incomplete treatments -
#### ട. Intended Use
Panther OIS|R&V is an information management system used to manage medical data and clinical workflow in a hospital or clinic. To support radiation oncology users, it allows the user to:
- Enter or import, modify, store and archive treatment plans and images from . treatment planning systems.
- Import, view, manipulate, enhance, annotate, store, and archive radiological . images.
- Select and provide radiation treatment plans to a radiation treatment delivery . system for patient treatment.
- Store and view treatment records provided by the radiation treatment delivery . system.
- Monitor and track treatment progress. �
- When Panther OIS|R&V is acting as an R&V system, verify the delivery device's . settings against the planned parameters, record the treatment parameters settings and prevent treatment if the parameters are out of tolerance.
#### 6. Summary of Comparisons to Predicate Devices
Panther OIS|R&V is substantially equivalent to the predicate devices, IMPAC Medical Systems, Inc's (Elekta) MOSAIQ (K120067), Varian Medical System Inc.'s ARIA Radiation Oncology (K093527) and 4D Integrated Treatment Console (K091132), and Prowess Inc.'s Puma (K100801), for purposes of FDA clearance for commercial distribution, as demonstrated and documented in this premarket notification submission. It has the same intended use and indications for use. In addition, the rationalization for substantial equivalence is further evidenced through discussion
{2}------------------------------------------------
of similar technological characteristics between Panther OIS|R&V and the predicates, as well as test results, which prove that Panther OIS|R&V is as safe and effective as the predicate devices.
#### 7. Summary of Technological Considerations
Panther OIS | R&V has many of the same technological characteristics as the predicate devices. There is a limited amount of distinguishing factors when comparing Panther OIS|R&V to the predicates, and those features that are different do not affect safety or effectiveness. Panther OIS|R&V provides a set of similar functionalities in such a manner that it is substantially equivalent to the MOSAIQ system and the combined ARIA+4DITC.
#### Summary of Non-clinical Tests 8.
A hazard analysis was completed for Panther OIS/R&V. Methods for preventing such hazards follow Prowess Inc. and ISO 14971 Risk Management process. Verification and validation of the software was conducted in house according to the Verification and Validation (V&V) Protocol, Panther OIS/R&V Testing Protocol. Test results have been included in this submission, showing documented evidence of successful fulfillment of pass/fail criteria. These tests have demonstrated that Panther OIS|R&V has met its predetermined specifications, is substantially equivalent in performance as compared to the predicate devices, functions as intended, and is safe and effective for its specified use.
#### 9. Summary of Clinical Tests
Although clinical testing is not required to demonstrate substantial equivalence in safety and effectiveness, we elected to conduct testing at Community Memorial Hospital. We feel that no matter how carefully a product is tested at the manufacturer's facility, such testing cannot replace actual use of the device in a clinical setting. As such, we consider both in-house testing and beta/external testing at a user site during device development.
#### 10. Labeling
The CD media labeling, Instructions for Use and User Manual are provided this submission. The entire User Manual, in digital format, is also included in the software media and can be viewed as part of the on-line help.
Commercial and marketing materials have been included in this submission.
Product labels comply with 21 CFR 1040.10 and 1040.11 as applicable. In addition, labeling complies with applicable requirements of 21 CFR 801, including the requirement that the device be provided with adequate directions for use.
#### Summary of Safety and Effectiveness Information 11.
- Prowess, Inc. is a registered medical device establishment, whose quality system a. meets the requirements of ISO 13485, the Medical Device Directive 93/42/EEC Annex II and FDA's Quality System Regulation, 21 CFR 820.
- Panther OIS|R&V was designed and implemented according to the established b. design and development as well as quality management procedures of Prowess Inc. In addition, design and development of the medical device software complies with internationally recognized standards including ISO 14971:2007Medical devices –
{3}------------------------------------------------
Application of risk management to medical devices, IEC 62304 Medical device software - Software life cycle processes, and IEC 62274 Medical electrical equipment - Safety of radiotherapy record and verify systems.
- The management of the company is committed to the highest standards of quality C. management. The Quality Management System is subject to regular, planned and documented audits by external consultants and by the FDA.
- A comprehensive risk analysis has been conducted. Detailed methods of mitigating d. these potential risks have been identified by the development team, and verified by clinical physicists contracted by Prowess and determined to be adequate.
- e. The software has been verified and validated based on established testing plans. The functionalities have been tested by in-house test engineers. In addition to in-house testing, the system was also tested by our beta-site using clinical cases.
- Key safety features of Panther OIS|R&V include: f.
- Defined privileges to access specified areas of the functionality for different . roles and users
- Patient identification by an ID photo associated to the patient during registration
- -Data locking and messaging to prevent simultaneous access to patient treatment data
- . Verifying that the beams contained in the treatment plan are correctly set up on the linear accelerator
- Panther OIS|R&V includes features specifically designed to protect confidential g. patient health information in accordance with the HIPAA regulations, including such security measures as password protection and access privilege set up and modification. By restricting access and other rights, patient health information can be safeguarded against deletion and/or alteration.
- h. Steps and precautions for the safe and effective use are included in the Instructions for Use and User Manual. Training by a Prowess specialist may also be provided as at the time of product distribution, when requested by the customer.
#### Level of Concern 12.
As medical device software, the submission for a follow Panther OIS|R&V FDA's Guidonce for the Content of Premarket Submissions for Software Contained in Medical Devices. Since prior to mitigation of hazards, a failure of the software device could result in death or serious injury to a patient, it has been determined that the software correlates to a Major Level of Concern, and as such, the associated documentation is included in this submission.
#### 13. Conclusions
Panther OIS|R&V is substantially equivalent to the predicate devices for the purposes of FDA clearance for commercial distribution. It has the same intended use and similar technical characteristics. The software has been found to perform as intended and the benefits to patient and user outweigh any inherent risks, which has been demonstrated via in-house testing as well
{4}------------------------------------------------
.
.
. .
as in our field tests. Its use does not raise any new or different safety and effectiveness concerns when compared to the predicates.
Comments of the country of the country of
1999 - 1999 - 1999
:
.
.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 21, 2012
Ms. Rachel Scarano Regulatory Affairs Manager Prowess Inc. 1844 Clayton Road CONCORD, CA 94520
Re: K122616
Trade/Device Name: Panther OIS|R&V Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 16, 2012 Received: November 21, 2012
Dear Ms. Scarano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Rachel Scarano
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
## Janine M. Morris -S
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
### 510(k) Number (if known): K122616
Device Name: Panther OIS|R&V
### Indications for Use:
Panther QIS|R&V is an information management system used to manage medical data and clinical workflow in a hospital or clinic. To support radiation oncology users, it allows the user to:
- . Enter or import, modify, store and archive treatment plans and images from treatment planning systems.
- Import, view, manipulate, enhance, annotate, store, and archive radiological . images.
- Select and provide radiation treatment plans to a radiation treatment delivery . system for patient treatment.
- Store and view treatment records provided by the radiation treatment delivery . system.
- Monitor and track treatment progress. +
- When Panther OIS|R&V is acting as an R&V system, verify the delivery device's . settings against the planned parameters, record the treatment parameters settings and prevent treatment if the parameters are out of tolerance.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Janine M. Morris -S 2012.12.21 16:13:44 -05'00'
> (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
k122616 510(k)
Page 1 of
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.