K110772 · Carbon Medical Technologies, Inc. · IYE · Apr 28, 2011 · Radiology
Device Facts
Record ID
K110772
Device Name
BIOMARC FIDUCIAL MARKER
Applicant
Carbon Medical Technologies, Inc.
Product Code
IYE · Radiology
Decision Date
Apr 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The BiomarC Fiducial Markers are intended to be implanted into the body to accurately visualize and constitute the reference frame for stereotactic radiosurgery and radiotherapy target localization.
Device Story
BiomarC Fiducial Marker is a sterile, pyrolytic carbon-coated zirconium oxide implant. Supplied pre-loaded in a single-use deployment device consisting of a cannula, handle, push rod, plunger, and bone wax retainer. Implanted into soft tissue to serve as a radiopaque reference frame for stereotactic radiosurgery and radiotherapy target localization. Visible on standard radiographs (x-ray, mammography, fluoroscopy, kV, CT), ultrasound, and MRI (up to 4.0 Tesla). Used by clinicians to improve target accuracy during radiation therapy planning and delivery. Benefits include precise anatomical marking for tumor localization, facilitating accurate radiation dose delivery to target tissues while sparing surrounding structures.
Clinical Evidence
Bench testing only. Biocompatibility testing performed in accordance with ISO 10993 series, including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity, implantation (muscle and bone), and haemocompatibility. Results confirmed the device is non-toxic, non-sensitizing, non-irritating, non-mutagenic, and non-hemolytic. Visualization performance verified across standard imaging modalities (x-ray, CT, ultrasound, MRI).
Technological Characteristics
Material: Pyrolytic carbon-coated zirconium oxide. Form factor: Discrete marker pre-loaded in a cannula-based deployment device. Energy: Passive radiopaque marker. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 1x10^-6. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for patients requiring target localization for stereotactic radiosurgery and radiotherapy. Used to mark soft tissue for visualization during surgical or radiation procedures.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K132064 — BIOMARC FIDUCIAL MARKER · Carbon Medical Technologies, Inc. · Feb 28, 2014
K130678 — BIOMARC GOLD FIDUCIAL MARKER · Carbon Medical Technologies, Inc. · Apr 12, 2013
K211590 — Fiducial Marker · Carbon Medical Technologies, Inc. · Jun 10, 2021
K063193 — MODIFICATION TO BIOMARC TISSUE MARKER · Carbon Medical Technologies, Inc. · Nov 21, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary 5
## 5. 510(k) Summary
.APR 2 8 2011
Andrew Adams Director - Quality Assurance Carbon Medical Technologies, Inc. 1290 Hammond Road Saint Paul, MN 55110
Phone: 651-653-8512 Fax: 651-407-1975
Submitted: March 17, 2011
#### Device Name
Trade Name: BiomarC Fiducial Marker Accelerator, Linear, Medical, 21 CFR 892.5050 Classification Name: Common/Usual Name: Tissue Marker
#### Predicate Devices
Fiducial Markers (K071614) BiomarC Tissue Marker (K063193)
#### Indication for Use
The BiomarC Fiducial Markers are intended to be implanted into the body to accurately visualize and constitute the reference frame for stereotactic radiosurgery and radiotherapy target localization.
#### Device Description
BiomarC Fiducial Marker is a sterile, pyrogen free, single patient use, pyrolytic carbon coated zirconium oxide discrete marker that is visible on standard radiographs (x-ray, mammography, fluoroscopy, kV, and CT) as well as ultrasound and Magnetic Resonance Imaging (MRI) at up to 4.0 Tesla field strength.
BiomarC Fiducial Marker is supplied pre-loaded in a sterile, pyrogen free, single patient use deployment device. The deployment device consists of a cannula with handle, a push rod with plunger, and bone wax as a marker retaining mechanism.
{1}------------------------------------------------
# Technological Characteristics and Performance
BiomarC Fiducial Marker is identical to BiomarC Tissue Marker (K063193) except the indications for use have been expanded in accordance with Fiducial Markers (K071614). Biocompatibility and performance testing confirms that the BiomarC Fiducial Marker is substantially equivalent to the Fiducial Markers for the expanded indications for use (Table 5).
| Characteristic | Proposed Device | Predicate Device | Predicate Device |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | BiomarC Fiducial<br>Marker | Fiducial Markers | BiomarC Tissue Marker |
| 510(k) number | K110772 | K071614 | K063193 |
| 510(k) holder | Carbon Medical<br>Technologies, Inc. | Civco Medical Solutions | Carbon Medical<br>Technologies, Inc |
| Indications for use | The BiomarC Fiducial<br>Markers are intended to<br>be implanted into the<br>body to accurately<br>visualize and constitute<br>the reference frame for<br>stereotactic radiosurgery<br>and radiotherapy target<br>localization. | The Fiducial Markers<br>are intended to be<br>implanted into the body<br>to accurately visualize<br>and constitute the<br>reference frame for<br>stereotactic radiosurgery<br>and radiotherapy target<br>localization.<br><br>Specifically, they can be<br>used in intracranial<br>diseases as gliomas,<br>neuromas,<br>meningiomas,<br>astroctyomas,<br>arteriovenous<br>malformations, and<br>metastatic carinomas.<br><br>Additionally, they can<br>be used in the body for<br>treating hepatic,<br>pancreatic,<br>retroperitoneal,<br>paraspinal, skeletal,<br>prostatic and breast<br>tumors. | The BiomarC Tissue<br>Marker is indicated for<br>use to radiographically<br>mark soft tissue during a<br>surgical procedure of for<br>future surgical<br>procedures. |
| Use | Single use | Single use | Single use |
| Sterility | Sterilized by EO with an<br>SAL of 1x106 | Sterilized by EO with an<br>SAL of 1x106 | Sterilized by EO with an<br>SAL of 1x106 |
| Pyrogens | Pyrogen free | Unknown | Pyrogen free |
| Marker material | Pyrolytic carbon coated<br>zirconium oxide | Gold | Pyrolytic carbon coated<br>zirconium oxide |
| Device body contact<br>category | Implant device,<br>tissue/bone, permanent | Implant device,<br>tissue/bone, permanent | Implant device, tissue,<br>permanent |
BiomarC Fiducial Marker Premarket Notification 510(k) 5-2
{2}------------------------------------------------
| (Table 5) (continued) | | | | |
|-----------------------|--|--|--|--|
|-----------------------|--|--|--|--|
| Safety | Biocompatibility testing for: | Biocompatibility testing for: | Biocompatibility testing for: |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | cytotoxicity,<br>sensitization,<br>intracutaneous<br>reactivity,<br>systemic toxicity,<br>genotoxicity,<br>implantation (muscle<br>and bone), and<br>haemocompatibility | cytotoxicity,<br>sensitization,<br>irritation,<br>systemic toxicity,<br>implantation, and<br>hemolysis | cytotoxicity,<br>sensitization,<br>intracutaneous<br>reactivity,<br>systemic toxicity,<br>genotoxicity,<br>implantation (muscle),<br>and haemocompatibility |
| | has demonstrated that<br>the BiomarC Fiducial<br>Marker is: | has demonstrated that<br>the Fiducial Marker is: | has demonstrated that<br>the BiomarC Tissue<br>Marker is: |
| | non-toxic (cyto and<br>systemic),<br>non-sensitizing,<br>non-irritating<br>(intracutaneous and<br>implantation),<br>non-mutagenic, and<br>non-hemolytic | non-cytotoxic,<br>non-sensitizing,<br>non-irritating,<br>non-reacting implanted,<br>and non-hemolytic | non-toxic (cyto and<br>systemic),<br>non-sensitizing,<br>non-irritating<br>(intracutaneous and<br>implantation),<br>non-mutagenic, and<br>non-hemolytic |
| | per testing in accordance<br>with the ISO 10993<br>series. | per testing in accordance<br>with ISO 10993. | per testing in accordance<br>with the ISO 10993<br>series. |
| Visualization | Visible on standard<br>radiographs (x-ray,<br>mammography,<br>fluoroscopy, kV, and<br>CT), ultrasound, and<br>magnetic resonance<br>imaging (MRI) | Visible on EPID, film,<br>kV and CR. | Visible on standard<br>radiographs (x-ray,<br>mammography,<br>fluoroscopy, kV, and<br>CT), ultrasound, and<br>magnetic resonance<br>imaging (MRI) |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Andrew J. Adams Director-Quality Assurance Carbon Medical Technologies, Inc. 1290 Hammond Road SAINT PAUL MN 55110-5867
APR 2 8 2011
Re: K110772
Trade/Device Name: BiomarC Fiducial Marker Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: March 17, 2011 Received: March 21, 2011
Dear Mr. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll me (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 4. Indications for Use Statement
K110772 510(k) Number (if known): __
Device Name: BiomarC Fiducial Marker
Indications for Use:
The BiomarC Fiducial Markers are intended to be implanted into the body to accurately visualize and constitute the reference frame for stereotactic radiosurgery and radiotherapy target localization.
Prescription Use X (Part 21 CFR 801 Subpart D)
.
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S Patel
---
(Division Sign-Off)
SIOK
adiningica Office 31 10
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
BiomarC Fiducial Marker Premarket Notification 510(k) 4-1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.