K090683 · Siemens Medical Solutions USA, Inc. · IYE · May 8, 2009 · Radiology
Device Facts
Record ID
K090683
Device Name
SYNGO RT THERAPIST CONNECT
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYE · Radiology
Decision Date
May 8, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. The syngo® workspaces includes a number of syngo® based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Connect Workspace contains software applications that permits patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Connect Workspace is an option when third party OIS, Treatment Planning Systems and/or PACS devices conforming to the DICOM® Standard are intended to be used in conjunction with the Siemens branded Linear Accelerator system.
Device Story
Software application for Siemens linear accelerators (LINAC); enables patient selection/setup, positioning verification, treatment delivery/verification, and recording. Unlike predicate workspaces, it excludes internal Oncology Information System (OIS) or Verify and Record (V&R) applications; instead, it interfaces with third-party OIS/TPS via DICOM. Radiation therapists use the workspace to review treatment parameters, initiate delivery, and submit reports as DICOM RT Structured Reports. Operates in clinical oncology settings; facilitates workflow integration between Siemens LINACs and third-party oncology information systems. Benefits include flexible clinical infrastructure by allowing facilities to use preferred third-party OIS/TPS while maintaining direct control over LINAC treatment delivery.
Clinical Evidence
Bench testing only. Verification and validation testing performed, including software unit-level, integration, and system-level tests to confirm performance and functional requirements.
Technological Characteristics
Software-based workstation application; interfaces with LINAC control console and third-party OIS/TPS via DICOM standard. Operates on syngo® software architecture. No changes to LINAC hardware, dosimetry, or imaging subsystems.
Indications for Use
Indicated for patients requiring X-Ray photon and electron radiation therapy for cancer treatment. Used by trained medical professionals under the supervision of a therapeutic radiologist.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
ARTISTE™ Solution (aka ARTISTE MV) with the syngo® RT Therapist workspace (K072485)
ONCOR™ Expression with COHERENCE™ RT Therapist workspace (K060226)
{0}------------------------------------------------
# SIEMENS
MAY - 8 2009
## 510(k) Summary of Safety and Effectiveness
## SIEMENS Medical Solutions USA, Inc
### syngo® RT Therapist Connect Workspace
| Submitted By | Siemens Medical Solutions USA, Inc.<br>Oncology Care Systems<br>4040 Nelson Avenue<br>Concord, CA 94520 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Christine Dunbar<br>Senior Regulatory Affairs Specialist<br>Tel: (925) 602-8157<br>Fax: (925) 602-8008<br>Christine.dunbar.ext@siemens.com |
| Date Prepared | March 11, 2009 |
| Proprietary Name | syngo® RT Therapist Connect |
| Common Name(s) | (Accessory To) Medical Charged-Particle Radiation Therapy System |
| Classification Name(s) | 21 CFR, Part 892.5050 (90IYE) |
| Predicate Device | The syngo® RT Therapist Connect is substantially equivalent to the<br>following devices: |
| Product | 510(k) | Clearance Date |
|------------------------------------------------------------------------------------|---------|----------------------|
| ARTISTE™ Solution (aka<br>ARTISTE MV) with the<br>syngo® RT Therapist<br>workspace | K072485 | December 27,<br>2007 |
| ONCOR™ Expression with<br>COHERENCE™ RT Therapist<br>workspace | K060226 | March 15,<br>2006 |
| PreScision™ Option | K082775 | February 19,<br>2009 |
Siemens Medical Solutions USA, Inc. Oncology Care Systems Group syngo® AT Therapist Connect 510(k) Submission
CONFIDENTIAL
001172
{1}------------------------------------------------
# SIEMENS
### Description Summary
Within the submission the following internal naming conventions are used:
| Market Name | Internal naming convention |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ARTISTE™ | ARTISTE™ linear accelerator and ACCEL release 4+ |
| Bay Bridge | Internal project name for the syngo® Therapist Connect<br>Workspace. |
| ONCOR™ Expression | ONCOR™ linear accelerator and ACCEL release 2+ |
| PRIMUS™ | PRIMUS™ linear accelerator and ACCEL release 2+ |
| PRIMEVIEW™ | Siemens proprietary verify and record system. The syngo® based<br>PRIMEVIEW is hosted on the COHERENCE™ and the re-<br>branded syngo® Therapist Workspace.<br>The syngo base version is marketed as PRIMEVIEW3i and is used<br>on the PRIMUS™ linear accelerator systems. |
| COHERENCE™ Therapist Workspace | RTT Workspace contains the SIEMENS proprietary verify and<br>record system as well as access to the Oncology Information<br>System and directly connects to the LINAC control console. |
| syngo® RT Therapist Workspace and<br>syngo® RT Therapist Express™ | RTT Workspace contains the SIEMENS proprietary verify and<br>record system as well as access to the Oncology Information<br>System and directly connects to the LINAC control console on the<br>ARTISTE™ linear accelerator system. |
| syngo® Therapist Connect Workspace | The syngo® Therapist Connect Workspace does NOT contain the<br>PRIMEVIEW verify and record application. The Third Party OIS<br>system containing a verify and record application connects to the<br>syngo® Therapist Connect Workspace via a validated DICOM<br>interface.<br>The syngo® Therapist Connect Workspace directly connects to the<br>LINAC control console on the ARTISTE™ linear accelerator<br>system. |
| syngo® Suite for Oncology<br>Workspaces | Syngo based workstation, re-branded COHERENCE workspaces. |
| syngo® | Siemens proprietary software architecture and hosting SIEMENS<br>software applications organized by task cards on a dedicated<br>workstation |
For further definitions of the terms used in this submission, refer to the Glossary.
## Technological Characteristics:
The syngo® Therapist Connect Workspace is an optional feature to the existing SIEMENS branded ARTISTE™ and ONCOR™ family of medical linear accelerator devices [LINAC]. The basic design, safety features and function of the LINAC control console and the linear accelerator treatment delivery, dosimetry, treatment recording and portal imaging subsystems remain unchanged.
Siemens Medical Solutions USA, Inc. CONFIDENTIAL Oncology Care Systems Group syngo® RT Therapist Connect 510(k) Submission
Page 92 of 100
001173
{2}------------------------------------------------
The syngo® Therapist Connect Workspace differs from the predicate COHERENCE and syngo® Therapist Workspaces in that it does not contain a Oncology Information System (OIS) or Verify and Record (V&R) application such as PRIMEVIEW. These functions are located on the Third Party OIS system. The syngo® Therapist Connect Workspace receives the treatment plan from a Third Party OIS system or Third Party Treatment Planning System (TPS), supports the final review of the treatment parameters and enables the Radiation Therapist to initiate the delivery of the treatment prescription.
Upon completion of the treatment prescription, a report is submitted to the Third Party OIS or TPS system in the form of a DICOM RT Structured Report. All other features such as patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording that are supported by the predicate COHERENCE and syngo® Therapist Workspaces remain unchanged.
2x Required: The Linear Accelerator device and its accessories are intended to be used by trained medical professionals under the supervision and direction of a therapeutic radiologist.
#### General Safety and Effectiveness:
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software development. verification of requirements and validation testing. To minimize electrical, mechanical and radiation hazards, SIEMENS adheres to recognized and established industry practice and relevant international standards.
#### Intended Use:
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
#### The syngo® Suite for Oncology Workspaces:
The syngo® workspaces includes a number of syngo® based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.
The syngo® RT Therapist Connect Workspace is an additional option when third party OIS, Treatment Planning Systems and/or PACS devices are intended to be used in conjunction CONFIDENTIAL Siemens Medical Solutions USA, Inc. Page 93 of 100 Oncology Care Systems Group syngo@ RT Therapist Connect 510(k) Submission
001174
{3}------------------------------------------------
# SIEMENS
with the Siemens branded Linear Accelerator system. The syngo® RT Therapist Connect Workspace is a software application that permits patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Connect Workspace can be interfaced with third party devices conforming to the DICOM® Standard.
The addition of the syngo® RT Therapist Connect Workspace does not change the intended use of the Siemens branded Linear Accelerator System.
### Verification and Validation Testing
Verification and Validation testing was performed to evaluate the performance and functionality of the syngo® RT Therapist Connect workspace software. This testing included software unit-level test, integration and system-level tests. The results demonstrate that the device satisfies all performance and functional requirements as stated in the Product Reguirement Specifications.
#### Summary:
In summary, it is SIEMENS' belief that the syngo® RT Therapist Connect workspace option does not introduce any new potential safety risks and is substantially equivalent to. and performs as well as, the predicate devices.
#### CONFIDENTIAL
001175
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in black. The overall design is simple and recognizable, representing the department's role in public health and welfare.
#### Public Health Service
MAY - 8 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Christine Dunbar Senior Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 4040 Nelson Avenue CONCORD CA 94520
Re: K090683
Trade/Device Name: syngo® RT Therapist Connect Workplace Regulation Number: 21 CFR 892.5050
Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: April 9, 2009
Received: April 10, 2009
Dear Ms. Dunbar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050;
{5}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 884.xxx (Obstetrics) Gynecology) (240) 276-0115 21 CFR 892.xxx (Radiology) (240) 276-0120 Other (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Samme M. Marre
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## SIEME
## Indications for Use Statement
510(k) Number:
K0.90683
Device Name:
syngo® RT Therapist Connect Workspace
#### Indications for Use:
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
### The syngo® Suite for Oncology Workspaces:
The syngo® workspaces includes a number of syngo® based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.
### The syngo® RT Therapist Connect Workspace:
The syngo® RT Therapist Connect Workspace contains software applications that permits patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Connect Workspace is an option when third party OIS, Treatment Planning Systems and/or PACS devices conforming to the DICOM® Standard are intended to be used in conjunction with the Siemens branded Linear Accelerator system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Colin Pollard
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Prescription Use
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 807 Subpart C)
Siemens Medical Solutions USA, Inc. Oncology Care Systems Group syngo® RT Therapist Connect 510(k) Submission
CONFIDENTIAL
Over-The-Counter Use
Page _1_ of _1 : Page 86 of 100
001165
AND/ OR