← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K081935

# STEREOTACTIC HEAD AND NECK LOCALIZER, MODEL 50-100 (K081935)

_Aktina Medical Physics Corp. · IYE · Oct 7, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K081935

## Device Facts

- **Applicant:** Aktina Medical Physics Corp.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Oct 7, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This product is used for patients who require external beam stereotactic radiation therapy of the head and neck region or radiosurgery of the cranial region. The system provides cranial and head and neck fixation and stereotactic localization with automatic software fiducial localization. It is intended to be used during both Computed Tomography (acquisition of the imaging series used for the patient's treatment plan) and each of the patient radiation treatments.

## Device Story

System provides patient immobilization and stereotactic localization for radiotherapy/radiosurgery. Inputs: CT imaging series; physical patient positioning. Components: customized dental tray (vacuum-assisted), head/neck support, stereotactic fiducial frame. Operation: fiducial frame positioned over treatment area; software reads CT images to determine patient coordinate system within frame. Output: coordinate data for treatment planning/setup. Used in clinical radiotherapy settings by radiation oncology staff. Benefits: enables precise, reproducible patient positioning and target localization for radiation delivery.

## Clinical Evidence

Bench testing only. Hardware and software specification testing performed to verify safety and performance requirements. Biocompatibility demonstrated per ISO-10993-1 for surface devices in contact with skin/mucosa (<24 hours).

## Technological Characteristics

Carbon fiber base board; dental impression tray/mask; rigid support apparatus; stereotactic fiducial frame. Biocompatible materials per ISO-10993-1. Software-based coordinate determination. System is non-invasive, used in supine position.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- VBH HeadFix ([K030439](/device/K030439.md))
- X-Plan-2 and Head and Neck Localizer (HNL) ([K991237](/device/K991237.md))
- GTC Headholder ([K934523](/device/K934523.md))
- Stereotactic Mask System, CT/X-Ray Localizer, and Target Positioner ([K945903](/device/K945903.md), [K954861](/device/K954861.md), [K073018](/device/K073018.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081935

# OCT 0 7 2008

### 510(k) Summary

### In Compliance with 21 CFR Section 807.92(c)

#### 1. General Provisions

| Device Trade Name:         | Stereotactic Head and Neck Localizer                                                                                                                              |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:               | Head and Neck Fixation System                                                                                                                                     |
| Owner Name<br>and Address: | Aktina Medical Physics Corporation<br>360 North Route 9 W<br>Congers, New York, 10920<br>Phone: 845-268-0101<br>Fax: 845-268-1700<br>Registration Number: 2436865 |

#### 2. Classification

This device is classified as a class II device according to 21 CFR 892.5050, "Medical chargedparticle radiation therapy system." The product code is TYE.

#### 3. Predicate Devices

- 1) VBH HeadFix, K030439, manufactured by Medical Intelligence Medizintechnik, Schwabmunchen, Germany
- 2) X-Plan-2 and Head and Neck Localizer (HNL), K991237 and GTC Headholder, K934523, manufactured by Radionics Software Applications, Inc, Burlington, MA
- 3) Stereotactic Mask System, CT/X-Ray Localizer, and Target Positioner, K945903, K954861, and K073018, BrainLAB AG, Heimstetten, Germany

{1}------------------------------------------------

#### 4. Description

The Aktina Head and Neck Stereotactic Localizer, Part Number 50-100, is used for the localization and fixation of patients undergoing stereotactic radiotherapy and radiosurgery of the cranial area, as well as radiotherapy of the head and neck area. Fixation is accomplished via two components: a customized Dental Tray that with the aid of slight vacuum suction fixes to the roof of the patient's mouth, and a customized head and neck support. Localization is accomplished via two components: a hardware component which comprises of a Stereotactic Fiducial Frame that is positioned over the patient's treatment area while being accurately interfaced to the Dental Tray, and a software component that reads the patient's Computed Tomography (CT) imaging series and determines the coordinate system of the patient within the fiducial frame. The Fiducial Frames are used during the patient's initial CT and then as a setup target box for each treatment thereafter.

The following are optional components of the system:

- . shoulder restraints which can be employed for head and neck treatments,
- . a table adapter which allows for manual angulations to level the system within the treatment room in order to remove any angular errors induced by the treatment couch,
- . a software tool that rotates the patient's imaging series (used for treatment planning) to remove any angular errors incurred during the initial CT imaging.

#### 5. Intended Use

This product is used for patients who require external beam stereotactic radiation therapy of the head and neck region or radiosurgery of the cranial region. The system provides cranial and head and neck fixation and stereotactic localization with automatic software fiducial localization. It is intended to be used during both Computed Tomography (acquisition of the imaging series used for the patient's treatment plan) and each of the patient radiation treatments.

#### 6. Technological Characteristics

- The Aktina Medical Physics' Stereotactic Head and Neck Localizer Software has the same technological characteristics as: the VBH HeadFix manufactured by Medical Intelligence Medizintechnik, the X-Plan-2 and Head and Neck Localizer (HNL) / GTC Headholder manufactured by Radionics Software Applications, Inc., and the Stereotactic Mask System, CT/X-Ray Localizer, and Target Positioner, K945903 manufactured by BrainLAB AG. All of these systems accomplish non-invasive cranial and head and neck immobilization as well as tumor localization with the patient in a supine position by using:

{2}------------------------------------------------

- A carbon fiber base board adapted to the CT or treatment patient table, a.
- A dental impression tray or face mask attached to a rigid support apparatus for fixation b. of the patient's cranium.
- A fiducial localizing frame to provide initial CT study data and a target box for patient ். set-up during LINAC treatment,
- A computerized determination of the target position based on the localizing frame data. d.

#### 7. Performance Standards and Data

The FDA under Section 514 of the Food, Drug and Cosmetic Act has not established performance standards for this product

Hardware and software specification testing have been performed on the Stereotactic Head and Neck Localizer to show that the verification, validation and safety requirements have been met.

#### 8. Biocompatibility

The Aktina Medical Corporation Stereotactic Head and Neck Localizer and its associated components have been shown to be biocompatible per the requirements of the FDA's Blue Book Memorandum #G95-1, entitled Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" for surface devices in contact with the skin or a mucosal membrane with a contact duration of less than 24 hours.

#### 9. Summary of Substantial Equivalence

This device is similar in design and intended use, technological, physical, and performance characteristics to the predicate devices. No new issues of safety or effectiveness are introduced by using this device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" in capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 0 7 2008

Mr. Tony Spaccarotella Director, QA/RA Aktina Medical Physics Corporation 360 North Route 9W CONGERS NY 10920

Re: K081935

Trade/Device Name: Stereotactic Head and Neck Localizer Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 15, 2008 Received: September 16, 2008

### Dear Mr. Spaccarotella:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, or vo and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device , can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must complive of all the Act's requirements, including, but not limited to: registration and listing (21 CFR Dan) 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html.

Sincerely yours.

forque M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 4

## Indications for Use

510(k) Number (if known): K081935

Device Name: _ Stereotactic Head and Neck Localizer

Indications for Use:

This product is used for patients who require external beam stereotactic radiation therapy of the head and neck region or radiosurgery of the cranial region. The system providen creatial and head and neck fixation and stereotactic localization with automatic software fiducial location. It is intended to be used during both Computed Tomography (acquisition of the imagiong series used for the patient's treatment plan) and each of the patient radiation free on ants.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hulda Kerner

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devi 510(k) Number

Page of of

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K081935](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K081935)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
