← Product Code [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE) · K072485

# ARTISTE MV SA (K072485)

_Siemens Medical Solutions USA, Inc. · IYE · Dec 27, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K072485

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [IYE](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE.md)
- **Decision Date:** Dec 27, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the ARTÍSTE MV SA™ linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. The ARTÍSTE MV SA™ includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngo™ RT Therapist Express Workspace with MVision™. The syngo™ RT Therapist Express Workspace is a software application that uses syngo™ based applications.

## Device Story

ARTÍSTE MV SA is a linear accelerator for cancer radiation therapy. System components include an Electronic Portal Imaging Device (EPID), 160-leaf multi-leaf collimator, and syngo RT Therapist Express Workspace. Device delivers x-ray radiation to target tumors. Integrated software manages patient data, treatment delivery, and verification. MVision system performs Megavoltage Cone Beam acquisition for 3D reconstruction and patient positioning verification via OPTIVUE imaging. Operated by clinical staff in oncology settings. Output includes radiation beams for treatment and imaging data for positioning verification. Benefits include precise radiation delivery and improved patient setup accuracy.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Linear accelerator system; includes 160-leaf multi-leaf collimator and EPID. Energy source: Megavoltage x-ray. Software: syngo RT Therapist Express Workspace with MVision. Features include patient data management, treatment verification, and Megavoltage Cone Beam CT (CBCT) acquisition/reconstruction. Connectivity: local on-board database.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- ONCOR EXPRESSION with COHERENCE Workspaces ([K060226](/device/K060226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SIEMENS

K072485

2007

#### Section 5 510(k) Summary and Statement

| Submitter:                          | Siemens Medical Solutions USA, Inc., Oncology Care Systems |         |
|-------------------------------------|------------------------------------------------------------|---------|
| Address:                            | 4040 Nelson Ave.<br>Concord, CA 94520<br>USA               |         |
| Phone number:                       | 925-246-8378                                               |         |
| Fax number:                         | 925-602-8008                                               |         |
| Contact person:                     | Bill Collins, Manager, Regulatory Affairs                  |         |
| Date prepared:                      | 8/31/2007                                                  | DEC 2 7 |
| Trade name:                         | ARTÍSTE MV SATM                                            |         |
| Device classification:              | 892.5050/ IYE/Radiology                                    |         |
| Substantial equivalence claimed to: | ONCOR EXPRESSION with COHERENCE Workspaces (K060226)       |         |

# Device Description:

The ACCEL system is the internal name for the Siemens program which includes the ONCOR Linear Accelerator product lines. The newest member of the ACCEL systems is the ARTISTE MV SATM Linear Accelerator product. The ARTÍSTE MV SA™ is composed of upgraded hardware, firmware and software as standard features.

### Intended use:

The intended use of the ARTÍSTE MV SA™ linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

The ARTÍSTE MV SA™ includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngo™ RT Therapist Express Workspace with MVision™.

The syngo™ RT Therapist Express Workspace is a software application that uses syngo™ based applications.

# Summary of technological characteristics:

The ARTISTE MV SA™ linear accelerator does not have significant changes in materials, energy source or technological characteristics as compared to the predicate device. The intended use, indication for use and the fundamental scientific technology are the same as the predicate device and therefore we believe it is substantially equivalent to the predicate device.

### Substantial equivalence:

The ARTÍSTE MV SA™ linear accelerator is substantially equivalent to the ONCOR EXPRESSION with COHERENCE Workspaces (K060226) in that it provides x-ray radiation for therapeutic treatment of cancer.

{1}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized lines, possibly representing an eagle or a similar bird.

#### Public Health Service

DEC 2 7 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions USA, Inc. % Mr. Erik Rodriguez Senior Software Quality Engineer Certified Compliance Solutions, Inc. (CCS, Inc) 16787 Bernardo Center Drive SAN DIEGO CA 92128

Re: K072485

Trade/Device Name: ARTISTE MV SATM Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 28, 2007 Received: November 28, 2007

# Dear Mr. Rodriguez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdri/industry/support/index.html.

Sincerely yours.

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

# SIEMENS

Indications for Use Statement Section 4 510(k) Number: K077403

Device Name: ARTÍSTE MV SATM

# Indications for Use:

The intended use of the ARTÍSTE MV SA™ linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

The ARTISTE MV SA™ includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngo™ RT Therapist Express Workspace with MVision™.

The syngo™ RT Therapist Express Workspace is a software application that uses syngo™ based applications with these embedded features:

- patient data management, selection and setup by use of a local, on-board database, .
- treatment delivery/verification and treatment recording by use of on-board applications, .
- Megavoltage Cone Beam acquisition method for acquiring projection data for 3D . reconstruction.
- Patient positioning verification by use of the OPTIVUE imaging system, including . MVision™(Cone Beam acquisition and reconstruction and the Adaptive Targeting application).

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aruith Wh

luctive. Abdomin 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K072485](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IYE/K072485)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
