← Product Code [IXI](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IXI) · K040804

# .DECIMAL TISSUE COMPENSATOR/INTENSITY MODULATOR (K040804)

_Southeastern Radiation Products, Inc. · IXI · Apr 28, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IXI/K040804

## Device Facts

- **Applicant:** Southeastern Radiation Products, Inc.
- **Product Code:** [IXI](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IXI.md)
- **Decision Date:** Apr 28, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5710
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

.decimal tissue compensator/intensity modulator manufacturing service manufactures the solid filters for missing tissue and intensity modulation for external beam radiation therapy.

## Device Story

Manufacturing service for custom solid filters used in external beam radiation therapy; produces tissue compensators and intensity modulators. Service utilizes patient-specific data to fabricate physical filters from 6061-T6 aluminum or 360 brass. Used in clinical radiation oncology settings; operated by medical physicists or radiation therapists. Output consists of custom-machined physical hardware; used by clinicians to shape radiation beams for precise dose delivery to target volumes while sparing healthy tissue. Benefits include improved dose distribution and treatment accuracy.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Custom-manufactured solid filters. Materials: 6061-T6 Aluminum, 360 Brass. Function: Radiation therapy beam-shaping/intensity modulation. Form factor: Patient-specific custom geometry.

## Regulatory Identification

A radiation therapy beam-shaping block is a device made of a highly attenuating material (such as lead) intended for medical purposes to modify the shape of a beam from a radiation therapy source.

## Predicate Devices

- CMS (K0000137)
- NOMOS ([K940412](/device/K940412.md))
- Varian ([K943224](/device/K943224.md))
- PAR Scientific ([K021987](/device/K021987.md))
- Huestis ([K880047](/device/K880047.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR 2 8 2004

K040804
Page 1 of 2

Section 807.87 (h) A 510(k) Summary as described in Section 807.92 or a 510(k) statement as described in 807.93

# Premarket Notification [510(k)] Summary as required by 21 CFR 807.92

#### Date summary was prepared:

March 24, 2004

#### Submitter's Name:

Southeastern Radiation Products, Inc. 2651 North Design Court Sanford, Florida 32773

#### Contact Person:

Linda Moon Quality Manager Phone: 407-330-3300 407-322-7546 Fax: Email:quality(a)seradiation.com

#### Device Name:

.decimal tissue compensator / intensity modulator manufacturing service

#### Classification Name:

Class II

## Predicate Device(s):

| CMS            | K0000137 |                                            |
|----------------|----------|--------------------------------------------|
| NOMOS          | K940412  | (as referred to in CMS' 510(k) submission) |
| Varian         | K943224  | (as referred to in CMS' 510(k) submission) |
| PAR Scientific | K021987  |                                            |
| Huestis        | K880047  |                                            |

{1}------------------------------------------------

Ko40804
page 2 of 2

#### Intended Use:

.decimal tissue compensator/intensity modulator manufacturing service manufactures the solid filters for missing tissue and intensity modulation for external beam radiation therapy.

## Technological Characteristics:

The technological characteristics of .decimal are discussed in Southeastern Radiation Product's attached documents:

Device Master Record - . decimal Tissue Compensators DMR-03 PD-03 Processing a .decimal tissue Compensator Work Order MSDS Sheet for 6061-T6 Aluminum and 360 Brass

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# APR 2 8 2004

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Mr. Richard Sweat President Southeastern Radiation Products, Inc. 2651 N. Design Court SANFORD FL 32773

Re: K040804 Trade/Device Name: .decimal Tissue Compensator/Intensity Modulator Regulation Number: 21 CFR 892.5710 Regulation Name: Radiation therapy beam-shaping block Regulatory Class: II Product Code: 90 IXI Dated: March 24, 2004 Received: March 29, 2004

Dear Mr. Sweat:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in your finding of substantial equivalence of your device to a legally promaticated predicated in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you dosh bpooms at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brygdon

Nancv C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known) _ K.O 40 804

Device Name: .decimal Tissue Compensator/Intensity Modulator

Indication for Use:

.decimal tissue compensator/intensity modulator manufacturing service manufactures the solid filters for missing tissue and intensity modulation for external beam radiation therapy.

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONFINUE ON A NOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use (Pcr 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device:

510(k) Number K040804

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IXI/K040804](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/IXI/K040804)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
