K203146 · Akesis, Inc. · IWB · Mar 5, 2021 · Radiology
Device Facts
Record ID
K203146
Device Name
Akesis Galaxy RTi
Applicant
Akesis, Inc.
Product Code
IWB · Radiology
Decision Date
Mar 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device intended for the stereotactic irradiation of human head structures in adults
Device Story
Akesis Galaxy RTi is a teletherapy device for stereotactic irradiation of adult head structures. It utilizes 30 Cobalt-60 sources housed in a hemispheric shield with a primary collimator and an independently rotating secondary collimator hemisphere (4 sizes plus blocking). The system integrates a y-ray treatment unit, a stereotactic localization interface, and a kV x-ray positioning/verification system. Operated by clinical staff in a medical facility, the device delivers radiation therapy based on pre-planned stereotactic coordinates. The kV x-ray system provides image-based positioning and verification to ensure accurate delivery. The device benefits patients by enabling precise, targeted radiation treatment for intracranial conditions.
Clinical Evidence
No clinical data. Bench testing only. System verified for electrical safety and EMC per IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-68, 60601-2-11, 61217, 62274, and 62366. Hardware and software verification and validation conducted per 21 CFR 820, ISO 13485, and ISO 14971.
Technological Characteristics
Radionuclide radiation therapy system using 30 Cobalt-60 sources. Features hemispheric shield, primary collimator, and rotating secondary collimator. Includes kV x-ray imaging system for positioning. Complies with IEC 60601 series standards. No direct tissue-contacting components; biocompatibility testing not required.
Indications for Use
Indicated for stereotactic irradiation of human head structures in adults.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
K200050 — Akesis Galaxy RTx · Akesis, Inc. · Feb 4, 2020
K133565 — LEKSELL GAMMA KNIFE PERFEXION · Elekta Instrument AB · May 9, 2014
K203250 — TaiChiC · Our United Corporation · Mar 12, 2021
K151159 — Leksell Gamma Knife Perfexion · Elekta Instrument AB · Oct 29, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
March 5, 2021
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Akesis Inc % Mr. Christopher Brown Director Quality Assurance and Regulatory Affairs 5129 Commercial Circle CONCORD CA 94520
Re: K203146
Trade/Device Name: Akesis Galaxy RTi Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: Class II Product Code: IWB Dated: January 28, 2021 Received: January 29, 2021
Dear Mr. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203146
Device Name Akesis Galaxy RTi
#### Indications for Use (Describe)
The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device intended for the stereotactic irradiation of human head structures for adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification [510(k)] Summary Akesis Galaxy Rotating Gamma System
The following information is provided following the format of 21 CFR 807.92.
# I. GENERAL INFORMATION
K203146
| Submitter's Name: | Akesis Inc.<br>5129 Commercial Circle<br>Concord, CA 94520 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name: | Christopher Brown<br>Phone: +1 925 408 1381<br>Fax: +1 925 326 2646 |
| Date Prepared: | October 19, 2020 |
| II. DEVICE INFORMATION | |
| Proprietary Name: | Akesis Galaxy RTi |
| Classification Name: | Radionuclide radiation therapy system |
| Device Classification: | 21 CFR 892.5750 |
| Device Class: | Class II |
| Product Code: | IWB |
| Common/Usual Name: | Radionuclide radiation therapy system |
| III. PREDICATE DEVICE | Akesis Galaxy RTx Rotating Gamma System (K200050) |
| IV. DEVICE DESCRIPTION | The Akesis Galaxy RTi, is a medical teletherapy device, which contains<br>(30) Cobalt-60 sources, distributed in a single, compact source carrier or<br>source drawer, which is inside a hemispheric shield. Immediately adjacent<br>to the source unit is the "built-in" primary collimator. An independently<br>rotating secondary collimator hemisphere contains 4 sizes of collimators<br>and a blocking position. |
| | consists of a) The y-ray treatment unit, and b) interface to a stereotactic<br>localization system, and c) kV x-ray positioning and verification system |
| V. INTENDED USE | The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device<br>intended for the stereotactic irradiation of human head structures in adults |
| VI. INDICATIONS FOR USE | The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device<br>intended for the stereotactic irradiation of human head structures in adults |
{4}------------------------------------------------
# VII. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Akesis Galaxy RTi has the same intended use, uses the same radioisotope, applies the same operating principle, has the same technical characteristics and meets the same performance specifications. Both the predicate and the Akesis Galaxy meet the same set of regulations and standards.
The following technological differences exists between the subject and the predicate device:
- -Addition of a kV x-ray imaging system
- Provides an interface to compatible immobilization devices -
# VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence:
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and electromagnetic compatibility (EMC) testing was conducted on this medical device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC. Successful testing was performed in accordance with IEC standards 60601-1-3, 60601-1-6, 60601-2-68, 60601-2-11, 60601-1-2, 61217, 62274 and 62366.
## Hardware and Software Verification and Validation Testing
Hardware and software verification and validation testing was conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards. Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software in Medical Devices." The software for this device was considered as a "major" level of concern.
## Biocompatibility Testing
Biocompatibility testing was not performed since this device does not have any direct tissue contacting components / parts.
This is in accordance with the Guidance for Industry and FDA Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", September 2020.
## Animal and Clinical Studies
Animal and clinical testing was not required.
{5}------------------------------------------------
## Summarv
These tests have demonstrated that the Akesis Galaxy RTi has met its specifications, demonstrated substantially equivalent performance to the predicate device and is suitable for its intended use.
# IX. CONCLUSIONS
The Akesis Galaxy RTi has the same intended use, fundamental scientific technology and principles of operation as the predicate device.
Use of the Akesis Galaxy RTi does not raise any new or different issues of safety or effectiveness when compared with the predicate device.
Performance tests have demonstrated that the Akesis Galaxy RTi has met its specifications, demonstrated substantially equivalent performance to the predicate device and is suitable for its intended use.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.