ROTATING GAMMA SYSTEM INFINI (INFINI)

K102533 · Masep Medical Science & Tech.Development (Shenzhen · IWB · Mar 22, 2011 · Radiology

Device Facts

Record IDK102533
Device NameROTATING GAMMA SYSTEM INFINI (INFINI)
ApplicantMasep Medical Science & Tech.Development (Shenzhen
Product CodeIWB · Radiology
Decision DateMar 22, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Rotating Gamma System Infini™ (Infini™) is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures.

Device Story

Rotating Gamma System Infini™ is a teletherapy device for stereotactic irradiation of intracranial structures. System components include 60Co radiation sources, main unit with mechanical/electric systems, stereotactic system, and treatment planning system. Device operates by delivering radiation to target intracranial structures; modifications from predicate include 3D treatment bed movement (X, Y, Z axis), updated collimator sizes (4, 8, 14, 18 mm), increased number of 60Co sources (30 total), and 6-valve radiation switching design. Used in clinical settings by trained professionals. Treatment planning system assists clinicians in defining target areas and dose delivery. Output is focused ionizing radiation for therapeutic purposes, intended to treat intracranial disease.

Clinical Evidence

Bench testing only. Performance testing, including software validation to a major level of concern, verified that system modifications and electrical control systems function as intended. Compliance with IEC60601-1, IEC60601-2, ISO14971, and ISO 2919 standards was demonstrated.

Technological Characteristics

Radionuclide radiation therapy system (60Co). 30 radiation sources. 3D treatment bed (X, Y, Z). 4 collimator sizes (4, 8, 14, 18 mm). 6-valve radiation switching mechanism. Software-controlled electrical system. Standards: IEC60601-1, IEC60601-2, ISO14971, ISO 2919.

Indications for Use

Indicated for stereotactic irradiation of intracranial structures in patients requiring teletherapy.

Regulatory Classification

Identification

A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MASEP This summary of Special 510(k) safety and Effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR807.92. | The assigned 510(k) number is: | K102533 | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Identification: | MASEP Medical Science & Technology Development (Shenzhen) Co., Ltd. | | Submitter's Addresses: | 601, 14th Building, Software Yard, Tech Mid Road 2nd, High-Tech Zone,<br>518057 Shenzhen, P.R. China | | Date Summary Prepared: | January 11, 2011 | | Contact Person: | Mr. Danny Qiu<br>CEO<br>Masep Infini Medical Science Technology Development Co. LTD.<br>17870 Castleton Street, Suite 395, City of Industry, CA 91748 USA<br>Tel: (626) 965-3200 Fax: (626) 965-3222<br>Http://www.masepinfini.com | | Name of Device: | Rotating Gamma System Infini™ | | | Classification: Radionuclide Radiation Therapy System | | | The Radiology Panel of DRAED has classified this radiology therapeutic<br>device as Class II, Performance Standards, 21 CFR Part 892.5760,<br>Radionuclide Radiation Therapy System, Product Code 90IWB. | | Predicate Device Information: | The Rotating Gamma System Infini™ is substantially equivalent to the<br>Gamma Therapy System (K#041125), manufactured and marketed by<br>MASEP Medical Science & Technology Development (Shenzhen) Co.,<br>Ltd. and the Leksell Gamma Knife ® Perfexion™ (K#061941),<br>manufactured and marketed by Elekta Instruments AB for treating | intracranial disease. The equivalence is based on the intended use, performance specifications, materials and technology characteristics. .. " … . IM {1}------------------------------------------------ MASEP MASEP Medical Science & Technology Development (Shenzhen) Co., Ltd. 510(k) Documents | | Exhibit #1 | | 510(k) Summary | |--|------------|--------------|----------------| | | | Page: 2 of 3 | | | <b>Device Description:</b> | The Rotating Gamma System Infini™ is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures. The system is composed of four parts: | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <ul><li><sup>60</sup>Co Radiation sources</li><li>Main unit (including mechanical system and electric system)</li><li>Stereotactic system</li><li>Treatment planning system</li></ul> | | <b>Intended Use:</b> | The Rotating Gamma System Infini™ is a teletherapy device indicated for | |----------------------|-------------------------------------------------------------------------| |----------------------|-------------------------------------------------------------------------| ## Technical Characteristics Comparison with MASEP Predicate Device: The Rotating Gamma System Infini TM is an improvement of the MASEP Gamma Therapy System for which MASEP was cleared under K041125. All safety and effectiveness testing (including risk analysis and performance testing) were conducted and completed with passing results. use in stereotactic irradiation of intracranial structures Infini TM has the following modifications: - · Change in treatment bed movement. · · · · · · · · · · · · · · · · · · · · · · · MASEP Gamma Therapy System: Bed only moves in Z axis; InfiniTM: Bed moves in 3 dimensions (X, Y and Z axis). - · Change in collimator diameter sizes MASEP Gamma Therapy System: Φ4 mm, Φ1 mm, Φ18 mm, Φ22 mm, total of 5 sizes Infini™: Φ4 mm, Φ8 mm, Φ18 mm, Φ18 mm, total of 4 sizes eliminating Φ22 mm collimator. - · Change in number of Co radioactive sources and total initial loading activity MASEP Gamma Therapy System: number of 00 radioactive sources = 25, total initial loading ≥ 6500 Ci. Infini "M: number of 00 radioactive sources = 30, total initial loading ≥ 7800 Ci. - Change in radioactive source switching design � MASEP Gamma Therapy System: Through rotation of the switching body, collimator passage is opened thus permitting passage of radiation. Likewise through reverse rotation of switching body, collimator passage is blocked thus closing radiation passage. Infini ™: Through rotation of the 6 switching valves, collimator passage is opened thus permitting passage of radiation. Likewise through reverse rotation of 6 switching valves, collimator passage is blocked thus closing radiation passage. For all design changes, a risk analysis and safety and effectiveness testing were performed. All testing that was conducted showed passing results. {2}------------------------------------------------ MASEP Medical Science & Technology Development (Shenzhen) Co., Ltd. 510(k) Documents Exhibit #1 # Comparison to Predicate Device: Both the Infinit™ and the Gamma Therapy System devices (K041125) are intended for stereotactic irradiation of intra-cranial structures. Infini™ can perform the same function as realized by the Gamma Therapy System (K041125). The performance of the subject Rotating Gamma System Infini™ is similar in both geometric accuracy and dosimetric characteristics to the predicate device, with other similar technological characteristics, Equivalencies include: same principle, intended use, energy used or transmitted, materials, similar design and operations. Similarities are reflected in safety, effectiveness, labeling, compatibility, standards, and application characteristics. #### Discussion of Non-Clinical Testing Performed for Determination of Substantial Equivalence: Non-clinical testing was conducted to validate and verify that the Rotating Gamma System Infini" met all design specifications and is substantially equivalent to the predicate device. Rotating Gamma System Infinit has also been tested to assure compliance to the requirements of various published standards, including IEC60601-1, IEC60601-2, ISO14971 and ISO 2919. Performance testing, including software validation testing to a major level of software concern for the Rotating Gamma System Infini " device, was conducted to verify that the operation of the entire Infini™ system along with its electrical control system has not been adversely affected by the device modifications. #### Conclusion: The Rotating Gamma System Infini™ has the same intended use and similar characteristics as the predicate devices. Testing demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. The Rotating Gamma System Infini™ is substantially equivalent to the predicate devices. inLini {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 MASEP Medical Science & Technology Development (Shenzhen) Co., Ltd. % Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Boulevard, Suite 200 MAR 2 2 25jj GREAT NECK NY 11021 Re: K102533 Trade/Device Name: Rotating Gamma System Infini™ (Infini™) Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB. MUJ Dated: March 14, 2011 Received: March 16, 2011 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of {4}------------------------------------------------ Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours. Mary Pastel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ of 1 SEP Medical Science & Technology Development (Shenzhen) Co., Ltd. 0(k) Documents Exhibit B Page: 1 Rotating Gamma System Infini™ # Indications for Use Statement 510(k) Number (if known): צום 25333 Device Name: Rotating Gamma System Infini™ (Infini™) Indications for Use: The Rotating Gamma System Infini™ (Infini™) is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures. Prescription Use _ X_ AND/OR (Part 21 CFR 801 Subpart D) 810x Over-The Counter Use (21 CFT 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Mary S. Patel (Division Sign Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety infilini™
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...