← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K103400

# FLUOROMAP SOFTWARE, 9 FLUORODISC (BELT), 12 FLUORO DISC (BELT), CLOSE TUBE CLIP (K103400)

_Stryker Corporation · OLO · Mar 7, 2011 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-e%E2%80%94neurological-surgical-devices/OLO/K103400

## Device Facts

- **Applicant:** Stryker Corporation
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Mar 7, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

FluoroMap is a computer controlled stereotaxic image guided surgery system. It provides static localization information derived from image processing of intra-operatively acquired Xray images, by superposition of virtual implants and instruments onto those X-ray images.

## Device Story

FluoroMap is a computer-assisted surgical planning system for proximal femur fracture surgery. It utilizes intra-operative 2D X-ray images as input. The system includes the FluoroMap software, a FluoroDisc (attached to a C-arm), and a Closed Tube Clip (attached to the Gamma3 Nail implant). The software recognizes metal marker spheres on the FluoroDisc and Closed Tube Clip to register X-ray images and localize the implant system. It overlays virtual implants and instruments onto the X-ray images, providing visual and numerical guidance to the surgeon. Used in the surgical suite by surgeons and clinical staff, the system assists in selecting and precisely positioning implants. This guidance aims to improve surgical accuracy and implant placement, potentially benefiting patients by optimizing fracture fixation.

## Clinical Evidence

No clinical tests were conducted. Evidence is based on bench testing, including cadaveric studies, saw bone tests, and computer simulations. The system demonstrated a 2 mm accuracy (maximal deviation between displayed and actual implant position) with statistical significance. Usability was validated against IEC 62366:2007 standards in simulated use scenarios.

## Technological Characteristics

Computer-assisted stereotaxic image-guided surgery system. Components: FluoroMap software, FluoroDisc (radiolucent plate with metal marker spheres), and Closed Tube Clip (radiolucent body with metal marker spheres). Connectivity: Integrates with standard C-arm imaging. Sterilization: Closed Tube Clip is sterile, single-use. Compliance: ISO 14971 (risk management), IEC 62366 (usability), and various AAMI/ANSI/ISO standards for biocompatibility and sterility (e.g., 10993 series, 11137 series).

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- VectorVision trauma ([K062358](/device/K062358.md))
- TenZing System ([K102130](/device/K102130.md))
- Gamma3 Lag Screw ([K020677](/device/K020677.md))
- Gamma3 Trochanteric Nail ([K032244](/device/K032244.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103400

# MAR - 7 2011

# 510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 .

| Submitter Information                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Stryker Leibinger GmbH & Co. KG - Navigation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Address                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Boetzinger Strasse 41, D-79111 Freiburg, Germany                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Phone number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | +49-761-4512-117                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Fax number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | +49-761-4512-49117                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Establishment Registration Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 3007582679                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Name of contact person                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Lilian Eckert                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date prepared                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 17 November 2010                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Name of device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Trade or proprietary name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | FluoroMap Computer Assisted Surgery System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Common or usual name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | FluoroMap System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Orthopedic stereotaxic instrument and Picture archiving and communications system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthopedic and Radiology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | §882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Code(s)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | OLO and LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Legally marketed device(s) to which equivalence is claimed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | VectorVision trauma from BrainLAB AG, K062358<br>TenZing System from Mazor Surgical Technologies Ltd., K102130<br>Gamma3 Lag Screw from Stryker Trauma GmbH, K020677<br>Gamma3 Trochanteric Nail from Stryker Trauma GmbH, K032244                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Reason for 510(k) submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Traditional 510(k) Notification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device description                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | The FluoroMap System is designed as a computer assisted surgical planning device to assist the surgeon in proximal femur fracture surgery.<br>The FluoroMap System provides static localization information derived through image processing of intra-operatively acquired X-ray images, by superposition of virtual implants and instruments onto those X-ray images.<br>The FluoroMap System consists of the FluoroMap Software, the FluoroDisc and the Closed Tube Clip. The Closed Tube Clip is a device which contains a radiolucent body with integrated metal marker spheres. The Closed Tube Clip is attached to the implant system of the Gamma3 Nail. The FluoroDisc |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| is a device which contains a radiolucent plate with integrated metal marker<br>spheres. It is attached to a standard C-arm, an intra-operative mobile 2D X-<br>ray imaging device. The FluoroMap Software is a surgery planning system<br>which recognizes the metal marker spheres of the FluoroDisc and the Close<br>Tube Clip and performs a registration of the X-ray images to localize the<br>position of the implant system, e.g. the Gamma3 Nail system. The<br>FluoroMap Software overlays virtual implants and instruments onto the X-<br>ray images and provides visual and numerical information to assist the<br>surgeon in selecting the appropriate implant and to position it precisely. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Intended use of the<br>device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | FluoroMap is a computer controlled stereotaxic image guided surgery<br>system. It provides static localization information derived from image<br>processing of intra-operatively acquired X-ray images, by superposition of<br>virtual implants and instruments onto those X-ray images.                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | FluoroMap assists the surgeon to determine the needed size and position<br>of orthopedic implants during proximal femur fracture surgery using the<br>Stryker Gamma3 Nail system.<br>The system should be operated only by trained personnel such as<br>surgeons and clinic staff.                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Summary of the technological characteristics of the device compared to the predicate device<br>1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Subject Device<br>FluoroMap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device 1<br>VectorVision trauma, BrainLAB<br>AG, K062358                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended Use /<br>Indication for Use<br>Statement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Intended Use:<br>FluoroMap is a computer controlled<br>stereotaxic image guided surgery<br>system. It provides static localization<br>information derived from image<br>processing of intra-operatively acquired<br>X-ray images, by superposition of virtual<br>implants and instruments onto those X-<br>ray images.<br>Indications:<br>FluoroMap assists the surgeon to<br>determine the needed size and position<br>of orthopedic implants during proximal<br>femur fracture surgery using the Stryker<br>Gamma3 Nail system.<br>The system should be operated only by<br>trained personnel such as surgeons and<br>clinic staff.                                          | BrainLAB VectorVision trauma is<br>intended to be a pre- and intraoperative<br>image guided localization system to<br>enable minimally invasive surgery. It<br>links a freehand probe, tracked by a<br>passive marker sensor system to virtual<br>computer image space on a patient's<br>pre- or intraoperative image data being<br>processed by a VectorVision<br>workstation. The system is indicated for<br>any medical condition in which the use<br>of stereotactic surgery may be<br>appropriate and where a reference to a<br>rigid anatomical structure, such as the<br>skull, a bone structure like tubular<br>bones, pelvic, calcaneus and talus, or<br>vertebra, can be identified relative to a<br>CT, fluoroscopic, X-ray or MR based<br>model of the anatomy.                                                                                                           |
| Intended Patient<br>Population                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Patients undergoing orthopedic surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Patients undergoing orthopedic surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Surgical Approach                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Orthopedic surgery, specifically at the<br>proximal femur                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Orthopedic surgery, specifically surgery<br>at tubular bones, pelvic, calcaneus and<br>talus, or vertebra                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Operational<br>Environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Surgical suite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Surgical suite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Primary Device<br>Function                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Assistance in planning and positioning of<br>orthopedic implants using intra-operative<br>X-ray images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Assistance in planning and positioning<br>of orthopedic implants using intra-<br>operative X-ray images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Main System<br>Components                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Software Reference device for C-arm Reference device for instrumentation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Software Reference device for C-arm Reference device for instrumentation Reference system for patient                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| User Interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Computer and monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Computer and monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Bench Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Accuracy testing of implant positioning in<br>simulated use scenarios using cadavers,<br>saw bones and computer simulation;<br>system accuracy of 2 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Unknown, expected to be similar<br>accuracy tests with similar results                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Conformance to<br>Recognized<br>Consensus<br>Standards                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Compliance with standards in regard to<br>risk management, usability, software,<br>sterilization, biocompatibility, packaging,<br>development process and quality<br>system, see section 1.11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Body Contact and<br>Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Reference device for instrumentation<br>(Closed Tube Clip) has patient contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference device for instruments and<br>reference device for patient have body<br>contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterile, Single-use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference device for instrumentation<br>(Closed Tube Clip) is delivered sterile<br>and as a single use product                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Parts of the reference device (infra-red<br>reflectors) are delivered sterile and are<br>single use products                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Summary of the technological characteristics of the device compared to the predicate device<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Characteristic…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-e%E2%80%94neurological-surgical-devices/OLO/K103400](https://fda.innolitics.com/submissions/RA/subpart-e%E2%80%94neurological-surgical-devices/OLO/K103400)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
