INSIGHT - FD MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM

K120388 · Hologic, Inc. · OXO · Apr 6, 2012 · Radiology

Device Facts

Record IDK120388
Device NameINSIGHT - FD MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM
ApplicantHologic, Inc.
Product CodeOXO · Radiology
Decision DateApr 6, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1650
Device ClassClass 2

Intended Use

The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide the physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.

Device Story

Portable Mini C-Arm fluoroscopic imaging system; utilizes CMOS flat detector technology for rectangular imaging; replaces conventional circular image intensifier. Operated by medical personnel in diagnostic, surgical, and post-operative settings for extremity imaging (hand, wrist, forearm, shoulder, foot, ankle, knee). Hardware includes X-ray source assembly with manual collimator rotation (+/-90 degrees). Software synchronizes collimator rotation with display orientation to maintain rectangular view. Provides real-time fluoroscopic visualization to physicians; assists in surgical guidance and diagnostic assessment; enables improved visualization of anatomy compared to circular intensifiers.

Clinical Evidence

Bench testing only. Performance characterized via non-clinical imaging studies comparing the new flat panel detector system to the prior image-intensified system. Metrics included resolution, contrast detail, uniformity, and motion. Compliance with 21 CFR 1020.30 and 1020.32 radiation protection standards verified.

Technological Characteristics

Portable Mini C-Arm; CMOS flat panel detector; X-ray source with manual collimation; +/-90 degree collimator rotation synchronization; rectangular field of view; complies with 21 CFR 1020.30 and 1020.32.

Indications for Use

Indicated for patients requiring general fluoroscopic visualization during surgical orthopedic, podiatry, critical care, emergency care, and light anatomy imaging procedures.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section 5 -510(k) Summary K120388 APR - 6 2012 ### Manufacturing Establishment and Contact Information 1 - 1.1 Manufacturer Name and Address: Hologic, Inc. 35 Crosby Drive Bedford, MA 01730 #### Establishment Registration Number: 1.2 1221300 #### Name, Title, and Telephone Number of Contact: 1.3 Name: Eileen M. Boyle Regulatory Affairs Specialist II Title: (781) 999-7781 Phone: (866) 652-8674 Fax : Email: eileen.boyle@hologic.com #### 2 Device Identification #### 2.1 Device Trade Name: Hologic® InSight - FD Mini C-Arm Fluoroscopic Imaging System #### Common/Usual Name 2.2 Fluoroscopic Imaging System #### 2.3 Proposed Intended Use: The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide the physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations. {1}------------------------------------------------ #### 3 Device Classification 3.1 Classification: Class II #### Classification Name and Rule 3.2 Image-Intensified Fluoroscopic X-Ray System, 21 CFR 892.1650 - Classification Panel 3.3 Radiology #### 3.4 Product Code охо #### 3.5 Predicate Devices - 510(k) No: K051754 ● Trade Name: Orthoscan, OrthoScan HD (which includes OrthoScan HD with Flat Detector) SE Date: 8/9/2005 Manufacturer: Orthoscan, Inc. - 510(k) No: K113708 . Trade Name: Orthoscan Mobile DI SE Date: 1/5/2012 Manufacturer: Orthoscan, Inc. - 510(k) No: K051025 . Trade Name: Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System SE Date: 5/13/2005 Manufacturer: Hologic, Inc. Page 2 of 4 {2}------------------------------------------------ #### 4.0 Device Description The Hologic® InSight-FD System , is a portable Mini C-Arm fluoroscopic imaging system with Flat Detector technology used by medical personnel in a variety of diagnostic, surgical and post-operative procedures for imaging extremities, including the hand, wrist, forearm, shoulder, foot, ankle and knee. The hardware enhancements for the InSight-FD System exist primarily within the detector assembly and collimation. The InSight-FD System offers CMOS Flat Detector technology which provides a rectangular viewing area in comparison t the circular view with the conventional image intensifier. For user convenience, additional buttons were added to the X-ray source assembly control panel from the available functions. The software enhancements for the InSight-FD System synchronize the rotation of the X-ray collimator with a manual rotation of +/-90 degrees. The synchronization enables the user to rotate the display image in a rectangular image as compared to the circular view of the image intensifier. #### 5 Performance Testing Software verification and validation test was performed following the recommendations in the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The purpose of the testing was to verify the incorporation and functionality of the flat panel detector on the Insight-FD System, that the accompanying software changes did not negatively impact the functionality, and that the functions and features tested meet their specific acceptance criteria. Nonclinical and clinical information was provided following the recommendations in the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices." This included information to characterize the solid state detector. Pairs of images were acquired to test the radiographic and fluoroscopic parameters {e.g., resolution, contrast detail, uniformity and motion) of the new Insight-FD with flat panel detector in comparison with the prior image intensified system. Testing was also performed to demonstrate that the new Insight-FD with flat panel detector continues to comply with the applicable regulatory performance standards for radiation protection, including: 21 CFR 1020.30 (Diagnostic x-ray systems and their major components) and 21 CFR 1020.32 (Fluoroscopic equipment). Testing was successfully conducted and demonstrated that the Hologic InSight-FD System meets all of its functional requirements and performance specifications. {3}------------------------------------------------ ### Substantial Equivalence б 7 The InSight-FD System is substantially equivalent to commercially available devices used for fluoroscopic imaging. The predicate devices selected for comparison are the OrthoScan- HD and the OrthoScan–HD with Flat Detector (K051754), OrthoScan Mobile DI (K113708) and the Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System (K051025). The predicate devices provide substantially equivalent or the same intended use, features and functions as the proposed device. ### Conclusion The features and functions on the InSight-FD System are substantially equivalent to those of the indicated commercially distributed devices: OrthoScan-HD with Flat Detector (K051754), OrthoScan Mobile DI (K113708) and Insight (K051025) with regard to intended use, performance, safety and effectiveness. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Eileen M. Boyle Regulatory Affairs Specialist II Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730 2012 Re: K120388 Trade/Device Name: InSight-FD Mini C-Arm Fluoroscopic Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO Dated: March 12, 2012 Received: March 22, 2012 ### Dear Ms. Boyle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {5}------------------------------------------------ ### Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device · Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## INDICATIONS FOR USE FORM 510(k) Number (if known): K120388 ### Device Name: InSight - FD Mini C-Arm Fluoroscopic Imaging System Indications for Use: The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and safety 510(k) K120388 Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...