STEP OPTION TO PRISM 2000
Device Facts
| Record ID | K960865 |
|---|---|
| Device Name | STEP OPTION TO PRISM 2000 |
| Applicant | Philips Medical Systems (Cleveland), Inc. |
| Product Code | KPS · Radiology |
| Decision Date | Sep 20, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1200 |
| Device Class | Class 2 |
Intended Use
The STEP Option to the Prism Dual Head system is a modification of a gamma camera system. This device is intended to be used for diagnostic imaging of organs and lesions.
Device Story
Modification to Prism Dual Head gamma camera system; adds hardware and software to perform attenuation correction. Operated by trained nuclear medicine professionals in clinical settings. Input: raw gamma camera imaging data. Processing: applies STEP correction algorithm to minimize body attenuation effects. Output: corrected diagnostic images. Benefit: improved image quality by reducing attenuation artifacts; assists clinicians in diagnostic interpretation of organs and lesions.
Clinical Evidence
Bench testing only. Laboratory tests demonstrated that the STEP option minimizes degrading effects of body attenuation compared to standard gamma cameras without correction. No clinical data provided.
Technological Characteristics
Gamma camera system modification; includes hardware and software for attenuation correction. Electrical safety certified to IEC-601 or UL-544 standards.
Indications for Use
Indicated for diagnostic imaging of organs and lesions in patients undergoing nuclear medicine examinations.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Related Devices
- K961577 — PRISM 3000XPV SYSTEM (210714) · Philips Medical Systems (Cleveland), Inc. · Jul 23, 1996
- K983948 — BEACON, MODEL 211060 · Philips Medical Systems (Cleveland), Inc. · Apr 16, 1999
- K971980 — MCD-AC · Adac Laboratories · Oct 23, 1997
- K983516 — TRANSMISSION KIT, TRANS-CT, MODEL NSTR-050A · Toshiba America Medical Systems, In.C · Dec 23, 1998
- K040616 — CARDIOMD-AC · Danish Diagnostic Development A/S · Mar 24, 2004