uEXPLORER
K231762 · Shanghai United Imaging Healthcare Co., Ltd. · KPS · Jan 18, 2024 · Radiology
Device Facts
| Record ID | K231762 |
| Device Name | uEXPLORER |
| Applicant | Shanghai United Imaging Healthcare Co., Ltd. |
| Product Code | KPS · Radiology |
| Decision Date | Jan 18, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1200 |
| Device Class | Class 2 |
| Attributes | AI/ML |
Intended Use
The uEXPLORER is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images. This system is intended to be operated by qualified healthcare professionals to assist in the detection, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging. This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Device Story
uEXPLORER is a combined PET/CT diagnostic imaging system; 194 cm axial field of view PET gantry; multi-slice CT gantry; patient table; console; reconstruction system. PET inputs: radiopharmaceutical distribution; CT inputs: X-ray attenuation data. System transforms inputs into fused PET/CT images for anatomical reference. Operated by qualified healthcare professionals in clinical settings. Features include HYPER Iterative reconstruction (noise penalty term), HYPER DLR (deep learning for SNR improvement), and Digital Gating (motion correction). CT system supports advanced functions: low-dose lung screening, Auto ALARA kVp/mA, EasyRange, Injector Linkage, Shuttle Perfusion, Online MPR, and Dual Energy analysis. Output provides clinicians with functional and anatomical data to assist in detection, staging, and treatment planning. Benefits include faster imaging, improved patient experience, and enhanced diagnostic accuracy through fused imaging.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including algorithm performance testing, image quality assessment, and conformance to electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and PET performance (NEMA NU 2-2018) standards.
Technological Characteristics
PET/CT system; 760 mm bore; LYSO scintillator; 672 detector rings; 194 cm axial FOV; 250 kg table load. Software includes HYPER Iterative (regularized iterative reconstruction), HYPER DLR (deep learning), and Digital Gating. CT system supports low-dose screening, Auto ALARA, and Dual Energy analysis. Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, IEC 60825-1, IEC 62304, and NEMA NU 2-2018. Biocompatibility per ISO 10993.
Indications for Use
Indicated for qualified healthcare professionals to assist in detection, diagnosis, staging, restaging, treatment planning, and treatment response evaluation for diseases, inflammation, infection, and disorders (e.g., oncology, cardiology, neurology). Includes low-dose CT lung cancer screening for early detection of lung nodules in patients meeting established screening program criteria.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Reference Devices
- uMI 550 (K193241)
- HYPER DLR (K193210)
- uCT 780 with uWS-CT-Dual Energy Analysis (K230162)
Related Devices
- K182938 — Emission Computed Tomography System · Shanghai United Imaging Healthcare Co., Ltd. · Dec 18, 2018
- K170490 — Celesteion, PCA-9000A/3 · Toshibamedical Systems Corporation · Jun 1, 2017
- K211846 — Discovery MI Gen2 · Ge Medical Systems, LLC · Aug 20, 2021
- K232712 — uMI Panorama · Shanghai United Imaging Healthcare Co., Ltd. · Oct 3, 2023
- K223325 — uMI Panorama · Shanghai United Imaging Healthcare Co., Ltd. · Mar 10, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 18, 2024
Shanghai United Imaging Healthcare Co., Ltd. Xin Gao Regulatory Affairs Manager No. 2258 Chengbei Rd., Jiading Industrial District Shanghai, Shanghai 201807 China
Re: K231762
Trade/Device Name: uEXPLORER Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, JAK Dated: December 18, 2023 Received: December 18, 2023
Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Reg
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. Ray K
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K231762
Device Name uEXPLORER
### Indications for Use (Describe)
The uEXPLORER is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.
This system is intended to be operated by qualified healthcare professionals to assist in the detection, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging.
This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use *(Select one or both, as applicable)*
| <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is in bold, sans-serif font and is stacked on top of each other. To the right of the text is a stylized "U" shape that is dark gray with a white horizontal line through the middle.
## 510(k) SUMMARY
K231762
# 1. Date of Preparation:
June 16, 2023
#### 2. Sponsor Identification
### Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Contact Person: Xin GAO Position: Regulatory Affair Manager Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
#### Identification of Proposed Device 3.
Trade Name: uEXPLORER Common Name: Positron Emission Tomography and Computed Tomography Systems Model(s): uEXPLORER
Regulatory Information Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK, Review Panel: Radiology
#### Identification of Primary/ Reference Device(s) 4.
Predicate Device
510(k) Number: K182938 Device Name: uEXPLORER Regulatory Information Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK, Review Panel: Radiology
Reference Device#1
510(k) Number: K193241
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape that is divided into three sections by two vertical lines. The logo is simple and modern, and the use of bold fonts and geometric shapes gives it a strong and recognizable presence.
Device Name: uMI 550 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK Review Panel: Radiology
### Reference Device#2
510(k) Number: K193210 Device Name: HYPER DLR Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS Review Panel: Radiology
Reference Device#3
510(k) Number: K230162 Device Name: uCT 780 with uWS-CT-Dual Energy Analysis Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Review Panel: Radiology
### 5. Device Description
The proposed device uEXPLORER combines a 194 cm axial field of view (AFOV) PET and multi-slice CT system to provide high quality functional and anatomical images, fast PET/CT imaging and better patient experience. The system includes PET gantry, CT gantry, patient table, power supply cabinet, console and reconstruction system, chiller, vital signal module.
The uEXPLORER has been previously cleared by FDA via K182938. The mainly modifications performed on the uEXPLORER (K182938) in this submission are due to the addition of HYPER Iterative, HYPER DLR, Digital gating, remote assistance and CT system modification.
Details about the modifications are listed as below:
- HYPER Iterative (has been cleared in K193241), uses a regularized iterative reconstruction algorithm, which allows for more iterations while keeping the image noise at an acceptable level by incorporating a noise penalty term into the objective function.
- HYPER DLR (has been cleared in K193210), uses a deep learning technique to produce better SNR (signal-to-noise-ratio).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED" is stacked on top of the text "IMAGING" in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is likely a graphical element of the logo. The overall design is clean and modern.
- Digital Gating (has been cleared in K193241), uses motion correction method to ● provide better alternatives to reduce motion effects without sacrificing image statistics.
- Remote assistance.
- PET recon matrix: 1024×1024.
- TG-66 compliant flat tabletop.
- Update the performance according to the NEMA NU 2-2018 standard.
- Update the operation system.
- CT system modification: add Low Dose CT Lung Cancer Screening, Auto ALARA kVp, Organ-Based Auto ALARA mA, EasyRange, Injector Linkage, Shuttle Perfusion, Online MPR and Dual Energy analysis function. All functions have been cleared via K230162.
# 6. Indications for Use
The uEXPLORER is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.
This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging.
This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.*
* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
#### 7. Comparison of Technological Characteristics with the Predicate Devices
The uEXPLORER employs the same basic operating principles and fundamental technologies, and has the similar indications for use as the predicate device. A comparison between the technological characteristics of proposed and predicate devices is provided as below.
| ITEM | Proposed Device | Predicate Device (K182938) |
|------|-----------------|----------------------------|
|------|-----------------|----------------------------|
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.
| Gantry | 760 mm bore | 760 mm bore |
|--------------------------|-------------------------------|-------------------------------|
| PET system | Scintillator material: LYSO | Scintillator material: LYSO |
| | Number of detector rings: 672 | Number of detector rings: 672 |
| | Axial FOV: 194 cm | Axial FOV: 194 cm |
| | Gantry bore: 76 cm | Gantry bore: 76 cm |
| CT System | Cleared via K230162 | Cleared via K172135 |
| Maximum table load | 250 kg | 250 kg |
| Post-processing software | | |
| HYPER Iterative | Yes | No |
| HYPER DLR | Yes | No |
| Digital gating | Yes | No |
uEXPLORER's technological characteristics do not raise new safety and effectiveness concerns.
#### Performance Data 8.
The following performance data were provided in support of the substantial equivalence determination.
### Non-Clinical Testing
Non-clinical testing including Algorithm and Image performance tests were conducted for the uEXPLORER during the product development.
UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:
## Electrical Safety and Electromagnetic Compatibility (EMC)
- A ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005. MOD)
- A IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- A IEC 60601-2-44 Edition 3.2: 2016 Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography
- > IEC 60601-1-3 Edition 2.1 2013-04 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- A IEC 60825-1 Edition 3.0 2014 Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)]
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol. The symbol is a dark teal color and has a white vertical line running through the center of the "U".
- A IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- A IEC 62304 Edition 1.1 2015-06 Consolidated version, Medical device software -Software life cycle processes
- A NEMA NU 2-2018. Performance Measurements of Positron Emission Tomographs
### Software
- A NEMA PS 3.1-3.20(2022d): Digital Imaging and Communications in Medicine (ICOM)
- A Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
## Biocompatibility
- ISO 10993-1: 2018, Edition 5.0, Biological evaluation of medical devices Part 1: A Evaluation and testing within a risk management process.
- A ISO 10993-5: 2009, Edition 3.0, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10: 2010, Edition 3.0, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
## Other Standards and Guidance
- ISO 14971: 2019, Edition 3.0, Medical Devices Application of risk management A to medical devices
- Code of Federal Regulations, Title 21, Part 820 Quality System Regulation A
- A Code of Federal Regulations, Title 21, Subchapter J - Radiological Health
# Clinical Testing
No Clinical Study is included in this submission.
## Summarv
The features described in this premarket submission are supported with the results of the testing mentioned above, the uEXPLORER was found to have a safety and effectiveness profile that is similar to the predicate device.
## 9. Conclusions
Based on the comparison and analysis above, the proposed device has similar intended use, performance, equivalence safety and effeteness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.