Affinity

K202882 · Hermes Medical Solutions AB · KPS · Dec 23, 2020 · Radiology

Device Facts

Record IDK202882
Device NameAffinity
ApplicantHermes Medical Solutions AB
Product CodeKPS · Radiology
Decision DateDec 23, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1200
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

AFFINITY is a software application used to process, display, analyse and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease. This application is not intended to replace visual assessment of tumors. The primary purpose of the application is to provide pre-selection of lesions for visual confirmation and to provide consistency and reproducibility when assessing tumor response to treatment.

Device Story

Affinity is a software application for processing, displaying, and managing medical imaging data (PET/CT, MR, tomographic SPECT/CT) transferred from external workstations or PACS. It enables clinicians to perform co-registration and alignment of multi-modal images. Key features include configurable clinical workflows, manual/local rigid registration, and a rendering engine for data visualization. The device provides automated segmentation tools (thresholding and split tool) to identify regions of interest (ROI). It calculates quantitative parameters including SUV (Peak, Mean, BW, BSA, LBM), SUVR, TLG, and MTV. Clinicians use these outputs to pre-select lesions for visual confirmation and assess tumor response to treatment. The device is intended for use by healthcare professionals in clinical settings to support diagnostic and monitoring decisions. It does not replace visual assessment of tumors but provides consistency and reproducibility in quantitative reporting.

Clinical Evidence

Bench testing only. Validation of the threshold/split tool involved 13 subjects (7 male, 6 female) undergoing F18FDG-PET/CT. Intra- and inter-validator variability was assessed by repeating ROI generation 5 times per user; results showed no significant change in quantitative values (variability 0 except at 14-17th decimal place). 'Hottest Connected' SUV peak validation used a synthetic 8-voxel study; Affinity results matched manual calculations in all 13 tested scenarios.

Technological Characteristics

Software-based image processing application running on Microsoft Windows 10 (64-bit). Supports DICOM standard for data transfer. Features include manual/local rigid registration, threshold-based segmentation, and multi-frame secondary capture. Quantification algorithms calculate SUV metrics (Peak, Mean, BW, BSA, LBM), TLG, and MTV. Rendering engine optimized for data loading. No hardware components; standalone software.

Indications for Use

Indicated for all patients undergoing molecular imaging investigations for diagnosis and monitoring of disease, including tumor response assessment.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 23, 2020 Hermes Medical Solutions AB % Joakim Arwidson VP Quality and Regulatory Strandbergsgatan 16 Stockholm, 11251 SWEDEN Re: K202882 Trade/Device Name: Affinity Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS Dated: September 9, 2020 Received: September 28, 2020 Dear Joakim Arwidson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202882 Device Name Affinity #### Indications for Use (Describe) AFFINITY is a software application used to process, display, and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease. Warning! This application is not intended to replace visual assessment of tumors. The application is to provide pre-selection of lesions for visual confirmation and to provide consistency and reproducibility when assessing tumor response to treatment Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> | |----------------------------------------------|-----------------------------------------| | Over-The-Counter Use (21 CFR | <span style="font-size: 20px;">☐</span> | |X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5.0 510 (k) SUMMARY #### A. Submitted by: - . Submitters name and address: Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden #### . Submitters telephone number Phone: +46 8 19 03 25 E-mail: joakim.arwidson@hermesmedical.com #### ● Contact person Joakim Arwidson VP Quality and Requlatory Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden - Registration number 9710645 B. Preparation date: 2020-09-09 #### C. Proprietary/Trade name, Common name, Classification name: - Proprietary/Trade name ● Affinity - . Common name Image processing systems - . Classification name Emission Computer Tomography System, Class II, 21CFR892.1200 #### D. Legally marketed device (predicate device): The following legally marketed device have been used for comparison. - Hybrid3D (K181468) primary ● #### E. Description of the device that is subject of this premarket notification: Affinity can process data as whole body or constrained field of view (e.g. abdomen, brain) where scanning was performed with any of the following modalities PET/CT, MR or tomographic reconstructed SPECT/CT. The studies are read in DICOM format and if there are studies with different modalities, on the same patient, a co-registration and alignment is performed. Modifications since version 1.0 is that SUVpeak definition has changed to correlate with the results from the predicate device, together with the option to configure the SUV peak definition in order to accommodate the users need of having quantitative values that help the users in its reporting #### K202882 {4}------------------------------------------------ process. Implementation of manual co-registration of images and local rigid registration, to give the user the possibility to improve the fit of two images and around specific areas of the medical image that may be of extra importance in the reporting process. Adjustments to make it possible for the user to in an easy manor create clinical workflows and layout that are configurable down to the smallest detail in order to simplify and personalize the workflow of each clinicians for maximum efficiency and accuracy in their reporting process. Optimization on the rendering engine has been performed to enable the data loading to be as low a reasonably achievable, hence reducing waiting time. Improvement have been made to the general Ul to make it more intuitive for the user to work in the Affinity environment. Possibility to Create, Save and reload Region Of Interest (ROI), from Affinity itself. But it has also been updated so it can load regions created in other medical imaging software's. Possibility to create, save and reload Multi-frame secondary captures. Adjustments to make it possible for the user to in an easy manor create clinical important ROI's that are configurable to best meet the user's need to obtain relevant quantitative values that can be a guidance in the reporting process. Single click segmentation is implemented that uses the feature of threshold and split into disconnected islands, which already have verification and validation test established. Software output - After the studies have been processed, these are presented visually in 2D or 3D for the user who can use tools to quantify the significant parameters SUV, SUVR, SUVbsa, SUVIbm, SUVbw, SUV Peak, SUV Mean, TLG and MTV. #### F. Intended use AFFINITY is a software application used to process, display, analyse and manage nuclear medicine and other medical imaging data transferred from other workstations, PACS or acquisition stations. The information acquired from viewing the images is used, in conjunction with other patient related data, for diagnosis and monitoring of disease. This application is not intended to replace visual assessment of tumors. The primary purpose of the application is to provide pre-selection of lesions for visual confirmation and to provide consistency and reproducibility when assessing tumor response to treatment. {5}------------------------------------------------ #### G. Technological characteristics Comparison of the proposed device Affinity and the primary predicate device Hybrid3D (K181468). | Trade Name | Affinity | Hybrid3D | Comparison | |--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510k # | This application | K181468 | N/A | | Operating System | Microsoft® Windows 10, 64 bit | Microsoft® Windows 7 and 10, 64 bit | Affinity supports Windows 10 OS | | Indications for<br>use | AFFINITY is a software application<br>used to process, display, analyse<br>and manage nuclear medicine and<br>other medical imaging data<br>transferred from other workstations,<br>PACS or acquisition stations. The<br>information acquired from viewing<br>the images is used, in conjunction<br>with other patient related data, for<br>diagnosis and monitoring of disease. | Hybrid3D is a software application that can<br>be used to process, display, analyze and<br>manage nuclear medicine and other<br>medical imaging data transferred from<br>other workstations or acquisition stations. | Equivalent Intended Use. Affinity also<br>includes the specification that "The<br>information acquired from viewing the<br>images is used, in conjunction with other<br>patient related data, for diagnosis and<br>monitoring of disease." | | Patient population | All patients undergoing molecular<br>imaging investigations. | All patients undergoing molecular imaging<br>investigations. | Equivalent patient population | | Software Input /<br>Modalities<br>PET/CT | Yes | Yes | Equivalent to Hybrid3D | | SPECT/CT | Yes (tomographic reconstructed) | Yes (tomographic reconstructed) | Equivalent to Hybrid3D | | MR | Yes | Yes | Equivalent to Hybrid3D | | Anatomical sites | Whole body or constrained field of<br>view (e.g. abdomen, brain) | Whole body or constrained field of view<br>(e.g. abdomen, brain) | Equivalent anatomical sites | | Software Output<br>Parameters | Methods used for quantification | Methods used for quantification | Affinity uses equivalent algorithms for<br>quantification. | | Trade Name | Affinity | Hybrid3D | Comparison | | SUV | Yes | Yes | Equivalent to Hybrid3D | | SUVR | Yes | Yes (in percist criteria) | Equivalent to Hybrid3D | | SUVbsa | Yes | Yes | Equivalent to Hybrid3D | | SUVIbm | Yes | Yes | Equivalent to Hybrid3D | | SUVbw | Yes | Yes | Equivalent to Hybrid3D | | SUV Peak | Yes | Yes | Two options<br>- Spherical kernel, option equivalent to Hybrid3D<br>- Hottest connected, new feature in Affinity 2.0. | | SUV Mean | Yes | Yes | Equivalent to Hybrid3D | | TLG | Yes | Yes | Equivalent to Hybrid3D | | MTV | Yes | Yes | Equivalent to Hybrid3D | | Clinical modules | | | | | Viewing 2D | Yes | Yes | Equivalent to Hybrid3D | | Viewing 3D | Yes | Yes | Equivalent to Hybrid3D | | SIRT (Selective Internal Radiation Therapy) - post treatment | No | Yes | No support of SIRT in Affinity | | Lung Lobar Quantification | No | Yes | No support of Lung Lobar Quantification in Affinity | | Trade Name | Affinity | Hybrid3D | Comparison | | SPECT Recon-<br>struction | No | No | Affinity can load and present tomographic<br>reconstructed SPECT data, but not do the<br>reconstruction itself. | | Communication<br>DICOM | Yes | Yes | Equivalent to Hybrid3D | {6}------------------------------------------------ {7}------------------------------------------------ Affinity is equivalent to predicate device in terms of IFU, patient population, modalities, anatomical sites and communication (DICOM). The algorithms used in the quantification of parameters are equivalent for Affinity and Hybrid3D. {8}------------------------------------------------ #### H. Changes to the Software Following modifications have been implemented since version 1.0: - SUVpeak, the definition has changed to correlate with the results from the predicate . device. together with the option to configure the SUV peak definition in order to accommodate the users need of having quantitative values that help the users in its reporting process. - . Hottest SUV peaks, as the uptake in the patient is non-homogeneous, low uptake voxels may occur right next to high uptake voxels, which will affect the SUVpeak value and mav reduce its value. This has led to new methods being developed to get around these fluctuations, using a small kernel of any shape that sweeps over a region looking for the highest possible occupancy in order to find the SUV peak. This is particularly important when imaging for Hogkins Lymphoma in children, as the tumors may be very small. It is this method which has been implemented in this version. Hasenclever D. Kurch L. gPET - a quantitative extension of the Deauville scale to assess response in interim FDG-PET scans in lymphoma. Eur J Nucl Med Mol Imaging 2014 Jul;41(7):1301-8. doi: 10.1007/s00259-014-2715-9. Epub 2014 Mar 7. - . Implementation of manual co-registration of images and local rigid registration, to give the user the possibility to improve the fit of two images and around specific areas of the medical image that may be of extra importance in the reporting process. Adjustments to make it possible for the user to in an easy manor create clinical workflows and layout that are configurable down to the smallest detail in order to simplify and personalize the workflow of each clinicians for maximum efficiency and accuracy in their reporting process. Optimization on the rendering engine has been performed to enable the data loading to be as low a reasonably achievable, hence reducing waiting time. Improvement have been made to the general UI to make it more intuitive for the user to work in the Affinity environment. Possibility to Create, Save and reload Region Of Interest (ROI), from Affinity itself. But it has also been updated so it can load regions created in other medical imaging software's. Possibility to create, save and reload Multi-frame secondary captures. Adjustments to make it possible for the user to in an easy manor create clinical important ROI's that are configurable to best meet the user's need to obtain relevant quantitative values that can be a guidance in the reporting process. - . Single click seamentation is implemented, which uses the threshold region feature to enable the user to select one disconnected region of interest at a time with the uptake higher than the threshold value set by the user. - . The split tool function is used to separate individual regions which meet the selected criteria as it is information about these individual regions which is required clinically in order to assess tumor size/metabolic activity. Split tool has verification and validation tests established. ### I. Testing The tests for verification and validation followed Hermes Medical Solutions AB design-controlled procedures. The Risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results confirm that all the software specifications have met the acceptance criteria. ### J. Substantially Equivalent/Conclusions The proposed device Affinity and the predicate devices Hybrid 3D (K181468) have equivalent indications for use. {9}------------------------------------------------ The proposed device uses similar technology and the fundamental concepts and operation are also the same, as described in the 510(k) submission. Validation of the features Split tool and the SUV peak tool in Affinity v2.0 has been done as described below Summary of performance validation of intra or inter variability The validation of Affinity v2.0 included performance testing data that demonstrates the subject device can provide reproducible tumor response assessment. The validation included 13 subjects (7 males, 6 female) that received a F18FDG-PET/CT study. Regions of interest were drawn by using the threshold tool in Affinity, with the predefined threshold value of 4 SUVbw applied to the PET study to include all voxels higher than 4 SUVbw. The regions created were split based on 1ml criteria, i.e. all disconnected regions larger than 1 ml are generated as a separate region. Brain, heart, and urinary bladder regions were deleted, based on the fact that they have no clinical importance in the pre-selection of lesions in order to perform visual confirmation and that they have normal uptake. In cases for which pathological areas where connected to non-pathological areas such as brain, heart, and urine bladder, the pathological areas were deleted and thus not included in this analysis. For every region of clinical importance, MTVs of 41% where created. This process was repeated for each patient 5 times and with 2 different validators to establish the intra and inter variability of the threshold/split tool. For each user, the mean of all quantitative value generated for every region and patient and corresponding standard deviation were calculated, where the Standard deviation is a clear indication of the variability of the quantitative values in each region. None of the quantitative values showed a significant change between the repeat processing of each validator and between the two validators. The variability was in all cases 0, except for in the 14-17th decimal place. The Split tool provides quantitative parameters for tumor response assessment which from this validation displays no intra or inter variability that can have any clinical impact. #### Summary of validation of the Hottest Connected SUV peak value The introduction of the Hottest Connected region SUV Peak value provides an additional SUV peak option. The validation was performed using a synthetic software generated study containing 8 voxels, each with a different SUV value. As there is no predicate device available for validating the SUV Hottest Connected region SUV peak values, the validation was performed by comparing the results obtained from manual calculation to those provided by Affinity, using a region placed in turn over different voxels in the synthetic study and using different values of SUV Hottest Connected Size. In all 13 scenarios tested the results were identical. In summary, the Affinity v2.0 described in this submission is in our opinion substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...