VERITON CT whole body SPECT/CT system

K182484 · Spectrum Dynamics Medical, Ltd. · KPS · Nov 9, 2018 · Radiology

Device Facts

Record IDK182484
Device NameVERITON CT whole body SPECT/CT system
ApplicantSpectrum Dynamics Medical, Ltd.
Product CodeKPS · Radiology
Decision DateNov 9, 2018
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1200
Device ClassClass 2
AttributesPediatric

Intended Use

The Spectrum Dynamics Medical's VERITON™ CT is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and trauma. The system output can be used for planning, guiding, and monitoring therapy. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; whole body and tomographic imaging. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles. The CT part is indicated for pediatric and adult patients. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.

Device Story

VERITON™ CT integrates SPECT scanners with X-ray CT; captures physiological data via radionuclide distribution and anatomical data via X-ray transmission. System performs whole-body and tomographic imaging; enables independent or combined SPECT/CT operation. CT data used for attenuation correction and image registration with SPECT data. Operated by trained healthcare professionals in clinical settings. Output provides cross-sectional images for lesion detection, localization, staging, and therapy planning/monitoring. Benefits include precise anatomical reference for metabolic activity and improved diagnostic accuracy through multimodal fusion.

Clinical Evidence

No formal clinical trial data provided. Evidence consists of clinical image evaluation by a board-certified radiologist to confirm diagnostic quality of SPECT, CT, and fused SPECT/CT images.

Technological Characteristics

System integrates SPECT (CZT detectors) and CT (X-ray transmission). Connectivity includes integrated SPECT/CT modality or standalone operation. Manufactured under ISO 13485 and 21 CFR 820. Complies with 21 CFR 1020.30 and 1020.33 radiation safety standards.

Indications for Use

Indicated for pediatric and adult patients requiring diagnostic imaging for cardiovascular disease, neurological disorders, trauma, and lesion staging/restaging. Used by trained healthcare professionals for detecting, localizing, and monitoring disease and organ function.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 9, 2018 Spectrum Dynamics Medical Ltd % Igor Naroditsky Director QA/RA 22 Bareket St. North Industrial Park CAESAREA 3088900 ISRAEL Re: K182484 Trade/Device Name: VERITON CT whole body SPECT/CT system Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, JAK Dated: August 15, 2018 Received: September 11, 2018 Dear Mr. Naroditsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Hol 2. Nils for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k182484 Device Name VERITON™ CT whole body SPECT/CT system Indications for Use (Describe) The Spectrum Dynamics Medical's VERITON™ CT is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and trauma. The system output can be used for planning, guiding, and monitoring therapy. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; whole body and tomographic imaging. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles. The CT part is indicated for pediatric and adult patients. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="text-decoration: overline;">☒</span> For use in connection with a PPE-related Settlement Program | |---------------------------------------------------------------------------------------------------------------| | <span style="text-decoration: overline;">☐</span> For use in connection with a PPE-related Settlement Claim | 2 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Spectrum Dynamics Medical. The logo consists of a blue triangular prism on the left and the words "SPECTRUM" and "DYNAMICS MEDICAL" in blue on the right. The word "SPECTRUM" is in a larger font than "DYNAMICS MEDICAL". The logo is set against a white background with a curved blue line at the bottom. #### PREMARKET NOTIFICATION 510(K) SUMMARY 7. The Company's 510(k) summary as required by 21 CFR Part 807.87(h) and 21 CFR Part 807.92(c) is provided on the following pages. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Spectrum Dynamics Medical. The logo consists of a blue triangular shape on the left and the words "SPECTRUM" and "DYNAMICS MEDICAL" in blue on the right. The word "SPECTRUM" is in a larger font than "DYNAMICS MEDICAL". The logo is clean and professional, and the colors are calming. # 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS 510(k) Number K182484 | Date of submission: | July 31, 2018 | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Spectrum Dynamics Medical Ltd.<br>22 Bareket St. North Industrial Park<br>Caesarea, Israel 3088900. | | Submitter Contact: | Mr. Igor Naroditsky<br>Director QA/RA<br>Tel: + (972) 73-737-4548<br>Fax: + (972) 73-737-4502<br>Email: igorn@spectrum-dynamics.com | | Device Trade Name: | VERITON™ CT | | Common<br>Name/Classification: | Single Photon Emission Computed Tomography<br>(SPECT)/Computed Tomography (CT) system | | Class: | II | | Product Code: | KPS and JAK | | Classification Panel: | Radiology | | Regulation No: | Emission Computed Tomography per 21 CFR<br>892.1200 and Computed Tomography X-Ray<br>System per 21 CFR 892.1750 | ## Marketed Devices: The Spectrum Dynamics VERITON™ CT system is a modification to the VERITON™ NM system (K180514) with the addition of the CT component to extend system's intended use. All system parts which are not related to the CT component remain the same as in the VERITON™ NM system (K180514) and have the same functionality and performance. #### Predicate device: Spectrum Dynamics Medical's VERITON™ NM (K180514) GE Healthcare Discovery NM_CT 670 CZT (K153402) {6}------------------------------------------------ #### Referenced device: Analogic Corporation CT6485 (K172058) Computed Tomography (CT) Scanner to support CT part used. Siemens Symbia 6.5 system (K162337) to support SPECT/CT combination and intended use. ## Device Description: VERITON™ CT consists of Single Photon Emission Computed Tomography (SPECT) scanners and integrated X-Ray Computed Tomography (CT). The SPECT subsystem images and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and integrates the CT's anatomical details for precise reference of the location of the metabolic activity. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The system can be used as an integrated SPECT and CT modality while also enabling independent functionality of SPECT and CT as stand- alone diagnostic imaging devices. ## Intended Use: The Spectrum Dynamics Medical's VERITON™ CT is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and trauma. The system output can be used for planning, guiding, and monitoring therapy. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; whole body and tomographic imaqing. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles. The CT part is indicated for pediatric and adult patients. {7}------------------------------------------------ SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) imaqes. ## Technological characteristic: Spectrum Dynamics' VERITON™ CT employs the same fundamental scientific technology as it's predicate device VERITON™ NM (K180514). The CZT detectors, system architecture and SPECT performance specifications have not changed between the commercially available VERITON™ NM (K180514) and proposed VERITON™ CT system. The CT part is similar to the Analogic Corporation CT6485 (K172058) Computed Tomography (CT) Scanner. ## Determination of substantial equivalence: #### Performance testing Verification & Validation includes testing for Biocompatibility, Software, Electrical Safety / EMC, Consensus Standard testing applicable for SPECT/CT scanners, mitigation measures set forth in device specific regulations (21CFR §1020.30, §1020.33), and usability. These tests were conducted to characterize the performance of the proposed device functionality against that of the predicate device. The supplementary non-clinical performance evaluations used a variety of test methods and phantoms appropriate for the performance metric/claim that was to be tested and evaluated. Mathematical and physics analysis were performed to demonstrate that each performance metric/claim was successfully verified and substantiated. The areas additionally evaluated included energy resolution, count rate linearity, uniformity, system resolution and lesion detectability. All testing has met the acceptance criteria for the proposed device. #### Summary of Non-Clinical Testing: The device has successfully completed all design control testing as per our quality system. No new hazards were identified, and no unexpected test results were obtained. The VERITON™ CT system was designed and will be manufactured under the Quality System Requlations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system: - . Risk Analysis {8}------------------------------------------------ - Design Reviews ● - . Software Development Lifecycle - Testing on unit level (Module verification) ● - . Integration testing (System verification) - . Performance testing (Verification) - . Safety testing (Verification) Spectrum Dynamics believes that the VERITON™ CT system is substantially equivalent to the VERITON™ NM system (K180514). The substantial equivalence was also based on software documentation for a "Moderate" level of concern device. #### Summary of Clinical Testing: Sample clinical images of the SPECT, CT and SPECT/CT modalities were evaluated by a board-certified radiologist to confirm that the images were of diagnostic quality. ## Conclusion: Based on the conformance to standards, development under Spectrum Dynamics quality system, the successful verification testing, additional engineering testing, and the clinical evaluation, Spectrum Dynamics Medical believes that the VERITON™ CT is substantially equivalent to the predicate device, VERITON™ NM (K180514) and hence is safe and effective for its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...