KODAK DIGITAL SCIENCE REMOTE CASSETTE
Device Facts
| Record ID | K962896 |
|---|---|
| Device Name | KODAK DIGITAL SCIENCE REMOTE CASSETTE |
| Applicant | Eastman Kodak Company |
| Product Code | KPR · Radiology |
| Decision Date | Sep 20, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1680 |
| Device Class | Class 2 |
Intended Use
The user will enter Exam Information and/or display Patient information from HIS/RIS at one of the RCIS terminal(s). This information will be transmitted via a network to the primary Quality Control Workstation of the CR-400 System. The primary QCW will update other QCW's and their respective KESPR Readers. After the phosphor cassette containing the Latent Image is scanned by a KESPR Reader the images are transmitted to a workstation which allows for the proper display of the images for diagnostic purposes. The information stored in the Readers and the will be updated by removing Exam and Patient Information. The phosphor readers are not for supporting or sustaining human life, nor do they present a potential for unreasonable risk for illness or injury.
Device Story
KODAK Digital Science Remote Cassette Identification Station (RCIS) functions as accessory to KESPR and QCW components of KODAK Ektascan Image Link Model 400 System. Hardware includes touch screen monitor, CPU, and bar-code reader. Used by clinical staff to input exam/patient data or retrieve information from HIS/RIS. Data transmitted via network to primary Quality Control Workstation (QCW) to update KESPR readers. Facilitates association of patient/exam data with latent images on phosphor cassettes prior to scanning. Enables diagnostic image display workflow. Does not support/sustain life.
Clinical Evidence
Bench testing only. Compliance with safety and EMC standards including UL 1950, CSA C22.2 950-M89, IEC 950, IEC 60601-1, and various EN/IEC EMC standards.
Technological Characteristics
Hardware: touch screen monitor, CPU, bar-code reader. Networked connectivity for data transmission to QCW. Safety standards: UL 1950, CSA C22.2 950-M89, IEC 950, IEC 60601-1. EMC compliance: EN 55022, EN 50082-1, IEC 801-2/3/4, FCC Part 15. Seismic compliance: California Admin Code Title 22/24.
Indications for Use
Indicated for clinical environments requiring entry and display of patient and exam information for integration with KODAK Ektascan Image Link Model 400 System CR-400 phosphor cassette readers.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Reference Devices
- KESPR
- QCW
- KODAK Ektascan Image Link Model 400 System
Related Devices
- K103703 — REGUIS SIGMA · Konica Minolta Medical & Graphic, Inc. · Jun 2, 2011
- K061538 — ALARA CRYSTALVIEW COMPUTED RADIOGRAPHY SYSTEM, MODEL R200 · Alara, Inc. · Jun 30, 2006
- K041990 — FUJI MEDICAL CR CONSOLE, MODEL FLASH IIP · Fujifilm Medical System U.S.A., Inc. · Aug 6, 2004
- K152094 — GR40CW · Samsung Electronics Co., Ltd. · Oct 16, 2015
- K062390 — CRYSTALVIEW R200 · Alara, Inc. · Sep 18, 2006