Labat series Diagnostic Ultrasound System

K210154 · Shenzhen Wisonic Medical Technology Co. , Ltd. · IYN · Jul 30, 2021 · Radiology

Device Facts

Record IDK210154
Device NameLabat series Diagnostic Ultrasound System
ApplicantShenzhen Wisonic Medical Technology Co. , Ltd.
Product CodeIYN · Radiology
Decision DateJul 30, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

The Labat series Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in gynecology, obstetric, abdominal, pediatric, small organ (breast, thyroid, testes,etc.), neonatal cephalic, adult cephalic, trans-vaginal, trans-rectal, musculo-skeletal (conventional and superficial), cardiac (adult and pediatric), trans-esoph.(cardiac), peripheral vessel and urology exams. The operator for Labat series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics.

Device Story

General-purpose diagnostic ultrasound system; touch-screen controlled; acquires/displays ultrasound data via various modes (B, M, Color, Power, PW, CW, Holo PW, Tissue Harmonic, Tissue Doppler, Anatomic M). Used in hospitals/clinics by trained clinical staff. System includes main unit, probes, ECG cables, gel warmer, foot switch, barcode reader, laser marker, TEE arm. Provides digital acquisition, processing, and display of ultrasound images; performs anatomical measurements and software analysis to assist clinicians in diagnosis. Includes 'wiGuide' needle guide system for tracking magnetized needles relative to ultrasound image. Benefits include non-invasive diagnostic imaging and procedural guidance.

Clinical Evidence

No clinical data required; substantial equivalence supported by bench testing only. Testing included biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and acoustic output (NEMA UD 2-2004).

Technological Characteristics

Touch-screen controlled diagnostic ultrasound system. Modes: B, M, Color, Power, PW, CW, Holo PW, Tissue Harmonic, Tissue Doppler, Anatomic M. Connectivity includes ECG cables, barcode reader, laser marker. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-2-37, IEC 60601-1-2, NEMA UD 2-2004. Biocompatibility per ISO 10993.

Indications for Use

Indicated for adults, pregnant women, pediatric patients, and neonates for diagnostic ultrasound imaging, measurement, and analysis in gynecology, obstetric, abdominal, pediatric, small organ, cephalic, trans-vaginal, trans-rectal, musculo-skeletal, cardiac, peripheral vessel, and urology exams.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". Shenzhen Wisonic Medical Technology Co.,Ltd. % Jiang Xiaosan Regulatory Manager 1st, 2nd, 5th & 6th Floor, NO.6 Building, Pingshan Technology Park, Taoyuan Street, Nanshan District Shenzhen, Guangdong 518055 CHINA July 30, 2021 ### Re: K210154 Trade/Device Name: Labat series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: June 8, 2021 Received: June 21, 2021 ### Dear Jiang Xiaosan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210154 Device Name Labat series Diagnostic Ultrasound System ### Indications for Use (Describe) The Labat series Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in gynecology, obstetric, abdominal, pediatric, small organ (breast, thyroid, testes,etc.), neonatal cephalic, adult cephalic, trans-vaginal, trans-rectal, musculo-skeletal (conventional and superficial), cardiac (adult and pediatric), trans-esoph.(cardiac), peripheral vessel and urology exams. The operator for Labat series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics. Modes of operation include: B-Mode, Color Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging, Anatomic M and combined mode. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary K210154 - 1. Submitter | Manufacturer: | Shenzhen Wisonic Medical Technology Co., Ltd. | | |--------------------|---------------------------------------------------------------|------------------| | Address: | 1st, 2nd, 5th & 6th Floor, NO.6 Building, Pingshan Technology | | | | Park, Taoyuan Street, Nanshan District, Shenzhen. Guangdong, | | | | 518055, P.R. CHINA | | | Contact person: | Name: | Jiang Xiaosan | | | Phone: | +86-755-86007788 | | | Fax: | +86-755-86007799 | | Date prepared: | November 12, 2020 | | | 2. Device | | | | Name of Device: | Labat series Diagnostic Ultrasound System | | | Models: | Labat SE, Labat SP, Labat SG, Labat PE, Labat PG, Labat | | | | TE, Labat TG, Labat IE and Labat IG. | | | Common/Usual Name: | Diagnostic Ultrasound System | | | Regulatory Class | II | | | Product Code: | IYN, IYO, ITX | | ### 3. Device Description The proposed Labat series Diagnostic Ultrasound System is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement and analysis of the human body and fluid that provides digital acquisition, processing and display capabilities. The Labat series Diagnostic Ultrasound System is a touch screen controlled ultrasonic system. Nine models for the main units are included in this submission, that is Labat SE, Labat SP, Labat SG, Labat PE, Labat PG, Labat TE, Labat TG, Labat IE and Labat IG. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, Color-Mode, Power (Dirpower)-Mode, PW-Mode, CW-mode, Holo PW, Anatomic M (AMM), Tissue Doppler Imaging (TDI) and the combined modes. The system can also measure anatomical structures and offer software analysis packages performance to provide information based on which the competent health care professionals can make the diagnosis. The Labat series Diagnostic Ultrasound System consists of the main unit, ultrasound probes, ECG cables, gel warmer, magnetic cup, foot switch, barcode reader, laser {4}------------------------------------------------ marker. TEE( transesophageal echocardiography) architecture arm, etc. Ten different models of probes are available for the Labat series, that is C6-1B-H, C8-3-H, L15-4WB-H, L15-4NB-H, L22-10-H, L15-6-H, LH15-6-H, SP5-1-H, EV11-3-H, P7-3T-H. 4. Indications for Use The Labat series Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in gynecology, obstetric, abdominal, pediatric, small organ (breast, thyroid, testes,etc.), neonatal cephalic, adult cephalic, trans-vaginal, trans-rectal, musculo-skeletal (conventional and superficial), cardiac (adult and pediatric), trans-esoph.(cardiac), peripheral vessel and urology exams. The operator for Labat series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics. include: B-Mode, Modes of operation M-Mode, Color Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging, Anatomic M and combined mode. | 510(k)<br>Clearance | Trade Name | Manufacturer | Predicate Device | |---------------------|------------------------------------------------------------------|---------------------------------------------------------|------------------| | K180912 | TE7/TE5 Diagnostic<br>Ultrasound System | Shenzhen Mindray<br>Bio-Medical<br>Electronics Co., Ltd | Predicate Device | | K163712 | Clover 50/Clover 60/Clover 70<br>Diagnostic Ultrasound<br>System | Shenzhen Wisonic<br>Medical<br>Technology Co.,<br>Ltd. | Reference device | | K180461 | Navi e/ Navi s/ Navi X<br>Diagnostic Ultrasound<br>System | Shenzhen Wisonic<br>Medical<br>Technology Co.,<br>Ltd. | Reference device | | K201693 | DC-90/ DC-90S/<br>DC-90Q/ DC-95/<br>DC-95S/ DC-88/ | Shenzhen Mindray<br>Bio-Medical<br>Electronics Co.,Ltd | Reference device | 5. Comparison of Technological Characteristics between Proposed Subject Device and Predicate Device {5}------------------------------------------------ | | DC-88S/ DC-80A/ | |--|------------------------------| | | DC-80AExp/ DC-80A | | | Pro/ DC-8X/ DC-8Q/ | | | DC-81/ DC-82 | | | Diagnostic Ultrasound System | Labat series Diagnostic Ultrasound System has the same technology as the predicate device(K180912). The proposed device has the same intended uses and similar operation modes as the predicate device. All systems have the same capability in term of measurements and calculation functions . The proposed device has the same intended uses and probe type as predicate device TE7/TE5(K180912); The acoustic power levels of the proposed device are below the limits of FDA, which is the same as the predicate device (K180912); The proposed device has similar operation modes as the predicate device TE7/TE5(K180912), except for Holo PW, AMM and TDI. For Holo PW and TDI, it has been cleared in reference device (K163712). For AMM mode, it can be considered same as reference device (K201693); The proposed device has the same function, except for Needle Guide System(wiGuide), which detects the position and orientation of magnetized needles in the presence of the probe and displays this information relative to the ultrasound image. This function has similar operating principals and specifications as the reference device (K180461); This difference in technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device. 6. Performance Data The following performance data were provided in support of the substantial equivalence determination. # Biocompatibility testing The biocompatibility evaluation for the Labat series Diagnostic Ultrasound System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk {6}------------------------------------------------ Management Process," as recognized by FDA. The battery of testing included the following tests: - Cytotoxicity - Sensitization - Irritation or intracutaneous reactivity # Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Labat series Diagnostic Ultrasound System. The system complies with the IEC 60601-1, IEC 60601-2-18 and IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC. ## Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "Moderate" level of concern. ## Acoustic testing Acoustic testing is conducted on the Labat series Diagnostic Ultrasound System in accordance with the NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic ultrasound equipment revision 3. ## Clinical study The subject of this premarket submission, Labat series Diagnostic Ultrasound System, did not require clinical studies to support substantial equivalence. ## 7. Conclusion The differences between the Labat series Diagnostic Ultrasound System and its predicate device do not raise different questions of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the Labat series Diagnostic Ultrasound System should perform as intended in the specified use conditions. From the results of performance data described, Shenzhen Wisonic concludes that the Labat series Diagnostic Ultrasound System is as safe and as effective as the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...